Symbols appearing in the numeric
fi eld:
Result lower than 10 mg/dL
(0.6 mmol/L)
Result higher than 600 mg/dL
(33.3 mmol/L)
+ ashing : code chip not
correctly inserted into the meter
or missing
+ : average cannot be
calculated
Error messages
Test strip not inserted correctly /
completely; test strip removed
and not re-inserted
Code chip from a different
system
Code chip cannot be read; code
chip from a different system
Measuring window dirty, test
strip is bent / not positioned
straight and at
Used test strip inserted,
measuring window dirty, test
strip not inserted correctly /
completely
Blood or control solution applied
too early, test strip bent / moved,
strong electromagnetic fi eld
Code chip removed while testing
was in progress
Test strip bent / moved, test strip
not inserted correctly /
completely
Meter error
Too little blood or control
solution applied
Temperature too low or too high
for testing