Roche ACCU-CHEK Inform II User manual

Category
Measuring, testing & control
Type
User manual
Accu-Chek
®
Inform II
BLOOD GLUCOSE MONITORING SYSTEM
Operator’s Manual
Revision History
Manual content version Revision date Changes
Version 1.0 2008-01 New document, international launch with document
0 4807839001 (01) EN 2008-01.
Version 2.0 2009-08 Update, new SW version 02.00; creation document
versions 0 5956200001 (01) 2009-08 EN-CAN/
0 5956218001 (01) 2009-08 FR-CAN for Canada
Version 3.0 2010-09 Update, new SW version 03.00. This manual version
not available in Canada.
Version 4.0 2012-11 Update, new SW version 03.04 (OTS, OTE), revision
Cleaning/Disinfection section; misc. editorial revi-
sions; international launch with document
0 4807839001 (04) 2012-11 EN. This document not
available in Canada.
Version 4.0 2013-11 Creation documents 0 5956200001 (02) 2013-11
EN-CAN/ 0 5956218001 (02) 2013-11 FR-CAN
containing Canada-specific changes. Additionally,
implemented information on barcode symbologies;
misc. corrections/amendments to v4.0 international.
Version 5.0 2013-09 Transition to new meter hardware: 2D barcode scan-
ner, modifications to wireless LAN option (separate
insertable RF card replaced by integrated WLAN com-
ponent); new battery pack. Update to SW 04.00; misc.
editorial revisions; international launch with document
0 4807839001 (05) 2013-09 EN. This document not
available in Canada.
Version 5.0 2014-09 Creation documents 0 5956200001 (03) 2014-09
EN-CAN/ 0 5956218001 (03) 2014-09 FR-CAN
containing Canada-specific changes to international
master document 0 4807839001 (05) 2013-09 EN.
Additionally, implemented Li-ion battery safety
information.
Version 6.0 2017-03 Creation documents 0 5956200001 (04) 2017-03
EN-CAN/ 0 5956218001 (04) 2017-03 FR-CAN
containing Canada-specific changes to international
master document 0 4807839001 (06) 2017-03 EN.
SW 04.01 and low power icons implemented; update
to SW 04.02; new Base Unit in “System overview”; new
section “What is new…”; new Appendix E for legacy
Base Unit; misc. editorial revisions.
Version 6.1 2020-04 Update of C&D information, minor editorial revisions.
0 5956200001 (05) 2020-04 EN-CAN
Operators Manual
Version 6.1
A
A
CCU-
CCU-C
HEK
HEK
®
Inform II System
© 2008-2020, Roche Diagnostics. All rights reserved.
The contents of this document, including all graphics, are the property of Roche Diagnostics. No part of this
document may be reproduced or transmitted in any form or by any means, electronic or mechanical, for any
purpose, without the express written permission of Roche Diagnostics. Roche Diagnostics has made every
reasonable effort to ensure that all the information contained in this manual is correct at the time of printing.
However, Roche Diagnostics reserves the right to make any changes necessary without notice as part of
ongoing product development.
Please send questions or comments about this manual to your local Roche representative.
ACCU-CHEK, ACCU-CHEK INFORM, ACCU-CHEK PERFORMA and COBAS are trademarks of Roche.
All other trademarks are the property of their respective owners.
The Wi-Fi CERTIFIED Logo is a certification mark of the Wi-Fi Alliance.
On the packaging, on the identification plate of the meter, the battery pack, the base unit, or the code key
reader you may encounter the following symbols, shown here with their meaning:
On meters with WLAN capability:
For other WLAN certifications, see label on bottom of battery compartment.
Caution, consult accompanying documents. Refer to safety-related notes in the instructions
for use accompanying this product.
Temperature limitation (Store at)
Manufacturer
Date of manufacture
Catalog number
In vitro diagnostic medical device
Global Trade Item Number
Serial Number
This product fulfills the requirements of the European Directives 98/79/EC on in vitro
diagnostic medical devices and 1999/5/EC on radio and telecommunications terminal
equipment (R&TTE).
The system fulfills the Canadian and U.S. safety requirements (UL LISTED, in accordance
with UL 61010-1 and CAN/CSA-C22.2 No. 61010-1).
Power supply connection (Base Unit Light)
Power supply connection (Base Unit)
This device complies with Part 15 of the FCC Rules and with RSS-210 of Industry Canada.
The compliance mark indicates that the product complies with the applicable standard and
establishes a traceable link between the equipment and the manufacturer, importer or their
agent responsible for compliance and for placing it on the Australian and New Zealand
market.
IVD
SN
7.5V 1.
7
12V 1,25
A
What is new in publication version 6.0?
This section provides an overview of all major changes
from Operator’s Manual version 5 to version 6. Deletions
or minor corrections are not listed.
SW 04.01 Description of new features SW 04.01 and amendments
from the supplement (English master document
0 7466161001 (01) 2015-03 EN and all derived transla-
tions) have been implemented.
New icon (bottle with flashing blue drop) used to
prompt sample application for control, proficiency,
and linearity tests. For example, see page 78,
Applying the control solution”.
Configurable delay between activation of the scan-
ner and reading a barcode (PreAimer Delay). See
section “Barcode Configuration” on page 178.
More barcode types supported. See section
“Barcode Configuration” on page 178.
Additional information regarding lithium-ion
(Li-ion) batteries (Avoidance of electrical shock,
fire, and explosions; Possible hazards posed by the
battery pack). See section 1.2.
Information regarding shipping. See page 132.
Low power icons If the battery pack is running low or not charging prop-
erly, the meter will display a series of low power icons.
See section “Low power icons” on page 153.
New layout safety messages The presentation of safety messages in this manual has
been updated throughout. See section 1.2 “Important
safety instructions and additional information”.
Test principle Description of the test principle used by the meter. See
page 17.
SW 04.02 Based on customer feedback, new features have been
added and existing ones improved.
Support of isolation room workflows. Scan the test
strip vial before scanning the patient ID. See
page 51.
In case of an automatic power-off, results will be
automatically flagged with a standard comment.
See page 22.
Shut down meter. See page 23.
When wireless connectivity is enabled, signal
strength bars show the quality of the connection.
See page 45.
The date of birth (DOB) can now be displayed on
patient glucose test screens. See page 59.
Patient ID validation. See page 174.
Configurable QC lockout. See page 57.
There are separate patient and QC result comment
settings. See page 80.
Delete Patient Information function has been added
to a new Service Menu. See page 147.
OTS workflow: the assessment buttons on the dis-
play are bigger and easier to use. See page 215.
Workflow for replacing the battery pack has been
modified. See page 122.
Accu-Chek Inform II Base Unit
The Accu-Chek Inform II Base Unit with new hard-
ware and software (REF 07671717190) has
replaced the legacy version (REF 05060290001) in
this manual.
An overview of the legacy Accu-Chek Inform II
Base Unit and technical information have been
moved to Appendix E.
This page intentionally left blank.
9
What is new in publication version 6.0? 6
SW 04.01................................................................................................................................................ 6
Low power icons ................................................................................................................................ 6
New layout safety messages.......................................................................................................... 6
Test principle ....................................................................................................................................... 6
SW 04.02................................................................................................................................................ 7
Accu-Chek Inform II Base Unit..................................................................................................... 7
1 Introduction 15
1.1
Before you start......................................................................................................................................... 15
Intended use ...................................................................................................................................... 15
Important information regarding use ....................................................................................... 15
If you need help................................................................................................................................ 16
Note on the use of “base unit” in this manual...................................................................... 16
Note on illustrations in this manual .......................................................................................... 16
What can the system do for you? .............................................................................................. 17
Test principle ..................................................................................................................................... 17
1.2 Important safety instructions and additional information......................................................... 18
Important information regarding safety................................................................................... 19
Disposal of the system ................................................................................................................... 21
Product safety.................................................................................................................................... 21
General care....................................................................................................................................... 21
Accessory box ................................................................................................................................... 22
Meter .................................................................................................................................................... 22
Power off meter ................................................................................................................................ 22
Automatic power-off....................................................................................................................... 22
Shut down meter.............................................................................................................................. 23
Automatic shutdown....................................................................................................................... 23
Battery Pack ....................................................................................................................................... 23
Touchscreen....................................................................................................................................... 26
Electromagnetic compatibility (EMC)....................................................................................... 26
Electrostatic discharge (ESD)...................................................................................................... 27
Wireless connectivity ...................................................................................................................... 28
Radiofrequency radiation exposure information.................................................................. 28
Local Area Network: protection from unauthorized access ............................................ 30
Wired network connection ........................................................................................................... 30
1.3 System components................................................................................................................................ 32
1.4 Overview of the meter ............................................................................................................................ 34
1.5 Overview of the code key reader ....................................................................................................... 36
1.6 Overview of the Accu-Chek Inform II Base Unit .......................................................................... 37
1.7 Overview of the Accu-Chek Inform II Base Unit Hub................................................................. 39
1.8 Overview of the accessory box ........................................................................................................... 40
1.9 Reagents and consumables................................................................................................................. 40
1.10 Instructions for initial setup.................................................................................................................. 41
10
2 Powering Up and Entering an Operator ID 43
2.1 Powering up the meter........................................................................................................................... 43
Adjusting the display ...................................................................................................................... 44
Enabling/disabling wireless connectivity ................................................................................ 44
Closing startup .................................................................................................................................. 46
2.2 Entering the operator ID........................................................................................................................ 46
Entering an operator ID with barcode scanner .................................................................... 47
Entering the operator ID manually............................................................................................. 48
Entering a password ....................................................................................................................... 48
3 Patient Glucose Testing 49
3.1
Information regarding blood glucose testing ................................................................................ 49
Preparing to test ............................................................................................................................... 49
3.2 Performing a patient glucose test...................................................................................................... 51
Overview of test procedure .......................................................................................................... 51
Isolation Room Workflow .............................................................................................................. 51
Entering or selecting the patient ID .......................................................................................... 52
Entering the patient ID manually................................................................................................ 54
Selecting the patient ID from a list............................................................................................ 55
Entering a patient ID with barcode scanner .......................................................................... 56
Confirming or selecting the test strip lot................................................................................. 57
Patient identification information............................................................................................... 59
Inserting test strips .......................................................................................................................... 60
Obtaining a blood sample............................................................................................................. 61
Applying a blood sample............................................................................................................... 62
Results screen ................................................................................................................................... 63
Adding comments............................................................................................................................ 66
Additional Patient Test ................................................................................................................... 68
4 Glucose Control Testing 69
4.1
Information regarding glucose control tests.................................................................................. 69
Glucose control testing intervals................................................................................................ 70
Information stored during glucose control testing.............................................................. 71
Control solutions............................................................................................................................... 72
Preparing to run a glucose control test ................................................................................... 72
4.2 Performing glucose control tests ....................................................................................................... 73
Overview of test procedure .......................................................................................................... 73
Starting a glucose control test .................................................................................................... 74
Confirming or selecting the lot number for control solutions ......................................... 75
Confirming or selecting the test strip lot................................................................................. 76
Inserting test strips .......................................................................................................................... 77
Applying the control solution....................................................................................................... 78
Results screen ................................................................................................................................... 79
Performing a STAT test .................................................................................................................. 81
11
5 Review Results 83
5.1 Displaying test results from the memory ........................................................................................ 83
Information stored in data records for test results.............................................................. 83
List of results stored in the memory ......................................................................................... 84
6 Storing Test Strip, Control Solution, and Linearity
Solution Information in the Meter 87
6.1
Storing information about test strips................................................................................................ 87
Transferring code key information to the meter .................................................................. 88
Editing test strip data...................................................................................................................... 91
6.2 Storing control solution information ................................................................................................. 94
Entering the lot number of the control solution ................................................................... 94
Selecting a stored lot number as the current lot number ................................................ 97
6.3 Storing linearity test information........................................................................................................ 99
Entering the lot number of the linearity test.......................................................................... 99
Selecting a stored lot number as the current lot number ............................................. 101
7 Linearity Testing 103
7.1
Information regarding linearity tests.............................................................................................. 103
Linearity testing intervals ........................................................................................................... 104
Information stored during linearity testing.......................................................................... 104
Linearity test kit ............................................................................................................................. 105
Preparing to run a linearity test ............................................................................................... 105
7.2 Performing a linearity test.................................................................................................................. 106
Overview of test procedure ....................................................................................................... 106
Starting a linearity test ................................................................................................................ 106
Confirming or selecting the lot number for linearity test kits....................................... 107
Confirming or selecting the test strip lot.............................................................................. 107
Inserting test strips....................................................................................................................... 108
Applying a linearity test sample .............................................................................................. 109
Results screen ................................................................................................................................ 110
8 Proficiency Testing 111
8.1
Information regarding proficiency tests ....................................................................................... 111
Information stored during proficiency testing.................................................................... 112
Preparing a proficiency test ...................................................................................................... 112
8.2 Performing a proficiency test............................................................................................................ 113
Overview of test procedure ....................................................................................................... 113
Starting a proficiency test.......................................................................................................... 113
Entering the proficiency sample ID........................................................................................ 114
Confirming or selecting the test strip lot.............................................................................. 114
Inserting test strips....................................................................................................................... 115
Applying a proficiency sample................................................................................................. 116
Results screen ................................................................................................................................ 117
12
9 Initial Startup 119
9.1 Connecting the base unit .................................................................................................................. 119
9.2 Installing or replacing the battery pack ....................................................................................... 121
Removing the battery pack........................................................................................................ 122
Installing the battery pack ......................................................................................................... 124
9.3 Docking the meter ................................................................................................................................ 126
9.4 Setting the date and time................................................................................................................... 128
9.5 Beeper options ....................................................................................................................................... 129
10 Maintenance and Care 131
10.1
Conditions for storage and shipping ............................................................................................. 131
General operating conditions ................................................................................................... 131
Shipping............................................................................................................................................ 132
Storage .............................................................................................................................................. 133
10.2 Cleaning/disinfecting the Accu-Chek Inform II system.......................................................... 133
Guide to cleaning and disinfecting the Accu-Chek Inform II system ...................... 134
Difference between cleaning and disinfecting .................................................................. 134
When to clean/disinfect.............................................................................................................. 134
Approved cleaning and disinfecting products................................................................... 135
Technical Assistance ................................................................................................................... 136
What to clean/disinfect............................................................................................................... 137
How to clean/disinfect ................................................................................................................ 139
Cleaning the barcode scanner window................................................................................ 141
Cleaning/disinfecting the base unit ....................................................................................... 141
Cleaning/disinfecting the accessory box ............................................................................. 142
Cleaning the code key reader .................................................................................................. 143
Cleaning the Accu-Chek Inform II Base Unit Hub............................................................ 143
10.3 Logging maintenance activities....................................................................................................... 144
10.4 Diagnostics view.................................................................................................................................... 145
10.5 Service menu .......................................................................................................................................... 147
Deleting patient information ..................................................................................................... 147
11 Troubleshooting 149
Errors and unusual behavior without error messages
.................................................... 149
Low power icons ........................................................................................................................... 153
Pop-up messages.......................................................................................................................... 154
Meter reset ...................................................................................................................................... 156
12 General Product Information 159
12.1
Technical data ....................................................................................................................................... 159
12.2 Further Information............................................................................................................................... 163
Ordering ........................................................................................................................................... 163
Accu-Chek Inform II Operator’s Manual and Quick Reference Guide..................... 165
Reagents and Solutions.............................................................................................................. 165
Product limitations........................................................................................................................ 165
Warranty ........................................................................................................................................... 165
Information about software licenses...................................................................................... 166
Contact Roche................................................................................................................................ 167
13
A Appendix 169
A.1 Table of configuration options ......................................................................................................... 169
A.2 Example of barcode symbologies ................................................................................................... 181
B Appendix 185
B.1
Option: Wireless network (WLAN) ................................................................................................. 185
Preliminary note............................................................................................................................. 185
Background..................................................................................................................................... 185
Technical implementation.......................................................................................................... 186
RF specific functionalities and effective performance claims...................................... 188
C Supplement for Other Test Entry 191
C.1
Before you start...................................................................................................................................... 191
Description ...................................................................................................................................... 191
C.2 Overview of Other Test Entry (OTE) ............................................................................................... 192
Introduction..................................................................................................................................... 192
C.3 Recording Other Patient Tests ......................................................................................................... 195
C.4 Recording Other Control Tests ........................................................................................................ 202
Introduction..................................................................................................................................... 202
Other Control Testing Intervals................................................................................................ 202
Stored Control Information........................................................................................................ 202
Warning messages ....................................................................................................................... 208
C.5 Reviewing Other Test Results........................................................................................................... 209
C.6 Other Test Entry Configuration Options ....................................................................................... 212
D Supplement for Observed Test Sequence 213
Observed Test Sequence (OTS)
............................................................................................... 213
Using the OTS function............................................................................................................... 214
E Appendix for Accu-Chek Inform II Base Unit (legacy version) 217
E.1
Overview of the base unit with older hardware......................................................................... 217
E.2 Connecting the base unit (legacy version) ................................................................................ 219
E.3 Cleaning/disinfecting the base unit (legacy version).............................................................. 219
E.4 Technical data ....................................................................................................................................... 220
Index 221
14
This page intentionally left blank.
Introduction • 1
15
1 Introduction
1.1 Before you start
Intended use The Accu-Chek
®
Inform II system is intended for in vitro
diagnostic use in the quantitative determination of blood
glucose levels in venous, capillary, arterial, and neonatal
whole blood samples for monitoring blood glucose.
This product is for monitoring hypoglycemia in neonates
diagnosed with laboratory glucose methods. The system
is indicated for use by healthcare professionals. For
healthcare professionals, the system is a bedside unit that
can help you provide quality patient care by measuring
blood glucose and by automating the record keeping
associated with blood glucose and glucose control tests.
Important information regarding
use
Read this operator's manual, as well as the package
inserts for all relevant consumables, before using the
system for the first test.
You must configure the Accu-Chek Inform II system
according to your needs before initial use. You can con-
figure the system by using a suitable data management
system. Refer to Appendix A for configuration via a data
management system. The device may not be used in
screening for neonatal hypoglycemia.
Be sure to read the “Important safety instructions and
additional information” section in this chapter before
operating the system.
16
If you need help Information about using the system, the screen menus and
performing a test can be found in this operator's manual.
Error messages that appear on screen include informa-
tion or instructions on how to correct the error.
For all questions about the Accu-Chek Inform II system
that are not answered in this manual, contact your Roche
representative (see Chapter 12). In order to expedite
troubleshooting, please have ready your Accu-Chek
Inform II meter, its serial number, this manual, and all
related consumables when you call. If you suspect a com-
munication error beyond the meter, also have your
Accu-Chek Inform II Base Unit serial number ready to
help assist our customer care group in troubleshooting.
Note on the use of “base unit” in
this manual
Note on illustrations in this
manual
Unless otherwise specified, the term “base unit” refers
to both the Accu-Chek Inform II Base Unit and the
Accu-Chek Inform II Base Unit Light.
Illustrations in this manual show two different kinds of
hands:
Hand without glove Hand with glove
A dashed arrow between screen illustrations indicates
that some screens have been skipped in these
illustrations.
Introduction • 1
17
What can the system do for you? The Accu-Chek Inform II system has the following
features and properties:
Perform patient blood glucose tests and glucose
control tests with control solution.
Automatically record all relevant data for the appli-
cation, which includes:
Time and date of test
IDs for operator, patient, and samples
Information about control solutions, test strips, and
linearity
Test results and comments
Record patient test results, quality control test
results and reagent information for certain off-
meter manual tests.
For purposes of quality assurance, information on
the following areas can be collected, stored, and
transferred:
Meters
Test strips
Glucose control solutions
Linearity solutions
Test results
Test principle An enzyme on the test strip converts the glucose in the
blood sample to gluconolactone. This reaction creates a
harmless DC electrical current that the meter interprets
and converts into a blood glucose result. The sample and
environmental conditions are also evaluated using AC
and DC signals.
For more details, refer to the test strip package insert.
18
1.2 Important safety instructions and additional information
This section explains how safety-related messages and
information related to the proper handling of the system
are presented in the Accu-Chek Inform II manual. Please
read these passages carefully.
These symbols and signal words are used for specific
hazards:
The safety alert symbol alone (without a signal word)
promotes awareness to hazards which are generic or
directs the reader to related safety information.
WARNING
Indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
CAUTION
Indicates a hazardous situation which, if not avoided,
could result in minor or moderate injury.
NOTICE
Indicates a hazardous situation which, if not avoided, may
result in damage to the system.
Important information that is not safety relevant is
presented against a colored background (without a
symbol). Here you will find additional information on
correct use of the meter or useful tips.
Introduction • 1
19
Important information regarding
safety
Operator qualification
Only trained healthcare professionals may operate the
Accu-Chek Inform II system. Operators must also have
received comprehensive instruction in the operation,
quality control, and care of the Accu-Chek Inform II
system.
WARNING
Protection against infection and blood-borne
pathogens
Healthcare professionals using the Accu-Chek Inform II
system to perform tests must be aware that any object
coming into contact with human blood is a potential
source of infection. Operators need to adhere to Standard
Precautions when handling or using the Accu-Chek
Inform II system. All parts of this system should be
considered potentially infectious and are capable of
transmitting blood-borne pathogens between patients
and between patients and healthcare professionals.
Use gloves. Wear a new pair of clean gloves for test-
ing each patient.
Wash hands thoroughly with soap and water before
putting on a new pair of gloves and performing the
next patient test.
Use an auto-disabling single-use lancing device for
each patient.
Dispose of used lancets in a sturdy sharps container
with lid.
Dispose of used test strips from patient and
proficiency tests according to your institution’s infec-
tion control policy.
Follow all health and safety regulations in force
locally.
20
CAUTION
Allergy or injury caused by reagents and other
working solutions
Direct contact with reagents, detergents, cleaning/disin-
fection solutions, or other working solutions may cause
skin irritation or inflammation.
Always use protective gloves.
Observe the cautions given in the package inserts of
the reagents and cleaning/disinfection solutions.
If a reagent, control, linearity, or cleaning/disinfection
solution comes into contact with your skin, wash it off
immediately with water.
Follow all health and safety regulations in force
locally.
WARNING
Avoidance of electrical shock, fire, and explosions
Only use Roche original accessories (cables, power
supply units, battery packs, and spare parts). Third-
party cables, power supply units, and battery packs
can cause the battery pack to explode or the meter to
become damaged.
Do not use loose power sockets or damaged power
supply units, cables, plugs, or battery packs.
Do not short circuit the power supply unit, the base
unit charging contacts, or the battery pack.
Do not drop the Accu-Chek Inform II meter, the power
supply unit, or the battery pack and protect these
against shaking and vibrations.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208
  • Page 209 209
  • Page 210 210
  • Page 211 211
  • Page 212 212
  • Page 213 213
  • Page 214 214
  • Page 215 215
  • Page 216 216
  • Page 217 217
  • Page 218 218
  • Page 219 219
  • Page 220 220
  • Page 221 221
  • Page 222 222
  • Page 223 223
  • Page 224 224
  • Page 225 225
  • Page 226 226
  • Page 227 227
  • Page 228 228