081854-P Rosemount Analytical January 2001
P3
Model 340 Trace Moisture Analyzer
SPECIFICATIONS - GENERAL
NON-PORTABLE ANALYZERS
CATALOG NUMBER 193000
AC Power
1
: 107 to 127 VAC 50/60 Hz ±0.5 Hz, 80 Watts maximum
Sample System Materials: 316 stainless steel, Viton, Teflon
Case: General purpose, panel mounting, purgeable
2
CATALOG NUMBER 193001
AC Power
1
: 107 to 127 VAC 50/60 Hz ±0.5 Hz, 80 Watts maximum
Sample System Materials: Brass, aluminum, neoprene, Buna-N, stainless
steel.
Case: General purpose, panel mounting, purgeable
2
C
ATALOG
N
UMBER
193004
AC Power
1
: 107 to 127 VAC 50/60 Hz ±0.5 Hz, 80 Watts maximum
Sample System Materials: 316 stainless steel, Viton, Teflon
Case: Detector module: Class I, Group D, Division 1.
3
Control Module: General purpose, purgeable
2
PORTABLE ANALYZERS
C
ATALOG
N
UMBER
193005
AC Power: 107 to 127 VAC 50/60 Hz ±0.5 Hz, 20 Watts maximum
Sample System Materials: Brass, aluminum, neoprene, Buna-N, stainless
steel.
Case: Portable with carrying handle.
C
ATALOG
N
UMBER
193006
DC Power: ±15 VDC from self-contained battery pack.
Sample System Materials: Brass, aluminum, neoprene, Buna-N, stainless
steel.
Case: Portable with carrying handle.
1
When requested by customer as a "special", the Model 340 Trace Moisture Analyzer has been modified for operation on
230 VAC power.
To verify AC power requirements for an analyzer, see the instrument name-rating plate.
If the instrument has the 230 VAC power modification, the following drawings are applicable:
780213 Schematic Diagram, 230 VAC Operation
780796 Pictorial Wiring Diagram, 230 VAC Operation
2
Air purge. When installed with user supplied components, meets requirements of Type Z purge per ANSI/NFPA 496-
1986, Chapter 2, for installation in Class I, Division 2 locations per National Electrical Code (ANSI/NFPA-70) for
analyzers sampling non-flammable gases. Analyzer sample flammable gases must be contained in a suitable explosion
proof enclosure or protected by a continuous dilution purge.
3
Air purge. When installed with user supplied components, meets requirements of Type Z purge per ANSI/NFPA 496-
1986, Chapter 2, for installation in Class I, Division 2 locations per National Electrical Code (ANSI/NFPA-70).