Peak Scientific Infinity 1051 User manual

Type
User manual
Infinity 1051
User Manual
2014 © Peak Scientific - Rev. 0 - 21/01/14
Infinity 1051 User Manual
Page 2 2014 © Peak Scientific - Rev. 0 - 21/01/14
Contents
Change History 3
How to use this manual 3
Warranties and Liabilities 4
Safety Notices 5
Symbols 5
Safety Notice to Users 5
Technical Specifications 6
Environment 6
Inlet Conditions 6
Outlet Gas 6
General 6
Introduction 7
Unpacking and Installation 7
Fittings Kit 8
General Dimensions 8
Air Connection 9
Commissioning 10
Pressure & Flow Settings 10
Cleaning 10
Maintenance Schedule 11
Service Schedule 11
Safely Isolated Condition 11
IMPORTANT DOCUMENT
S
12
Infinity 1051 User Manual
Page 3 2014 © Peak Scientific - Rev. 0 - 21/01/14
Change History
Rev. Comment Name Date
1
2
3
4
5
How to use this manual
This manual is intended for end users and has been written so that it can either
be read as a step by step guide to installation and usage; or as a reference
document where you can skip to the relevant information.
Please review each of the following sections carefully.
Thank you for selecting Peak Scientific to meet your Gas Generation needs,
should you require any further assistance or support please do not hesitate to
contact Peak Scientific or the Peak Partner from which you purchased your
Generator.
Infinity 1051 User Manual
Page 4 2014 © Peak Scientific - Rev. 0 - 21/01/14
Warranties and Liabilities
1. The Company warrants that it has title to the Goods.
2. Subject to the provisions of this clause the Company warrants that the
Goods shall comply in all material respects with any specification referred
to in the Order Confirmation (as the same may be amended) and shall,
subject thereto, be free from defects in material and workmanship for the
lesser of a period of twelve months from the date of delivery or thirteen
months from the date of dispatch from the factory.
3. Save as provided in this clause and except where the Goods are sold to a
person dealing as a consumer (within the meaning of the Unfair Contract
Terms Act 1977) all warranties, conditions or other terms implied by statute
or common law are hereby expressly excluded save to the extent they may
not be lawfully excluded. When the Goods are sold to a consumer within
the meaning of the Unfair Contract Terms Act 1977 their statutory rights are
not affected by the provisions of this clause.
4. In the event of the Customer making a claim in respect of any defect in
terms of clause 2 hereof the Customer must.
4.1. Reasonably satisfy the Company that the Goods have been properly
installed, commissioned, stored, serviced and used and without
prejudice to the generality of the foregoing that any defect is not the
direct or indirect result of lack of repair and/or servicing, incorrect
repair and/or servicing, use of wrong materials and/or incorrect spare
parts
4.2. Allow the company to inspect the Goods and/or any installation and
any relevant packaging as and when reasonably required by the
Company.
5. Subject to the Company being notified of any defect as is referred to in
sub-clause 2 hereof within a reasonable time of it becoming apparent and
subject always to the terms of sub-clause 4 hereof, the Company shall, in its
option, replace or repair the defective Goods or refund a proportionate part
of the Price. The Company shall have no further liability to the Customer
(save as mentioned in sub-clause 6 hereof).
6. The Company shall be liable to indemnify the Customer in respect of any
claim for death or personal injury to any person in so far as such is
attributable to the negligence or breach of duty of the Company or any
failure by the Company to comply with the provisions of sub-clause 2
hereof.
7. Save as provided in sub-clause 2 hereof the Company shall not be liable in
respect of any claim by the Customer for costs, damages, loss or expenses
(whether direct, indirect, consequential or otherwise) or indemnity in any
respect howsoever arising including, but not by way of limitation, liability
arising in negligence (other than pursuant to clause 6 above) that may be
suffered by the Customer or any third party.
Infinity 1051 User Manual
Page 5 2014 © Peak Scientific - Rev. 0 - 21/01/14
Safety Notices
Symbols
This manual uses the following symbols to highlight specific areas important to
the safe and proper use of the Generator.
Table 1 - Safety Symbols
Safety Notice to Users
These instructions must be read thoroughly and understood before installation
and operation of your Peak Scientific Infinity 1051. Use of the Generator in a
manner not specified by Peak Scientific MAY impair the SAFETY provided by
the equipment.
When handling, operating or carrying out any maintenance, personnel must
employ safe engineering practices and observe all relevant local health and
safety requirements and regulations. The attention of UK users is drawn to the
Health and Safety at Work Act 1974, and the Institute of Electrical Engineers
regulations.
Nitrogen is not a poisonous gas, but if the concentration in
the inhaled air becomes too high there will be a risk of
asphyxiation.
A WARNING notice denotes a hazard. It calls attention to an operating procedure, process
or similar, which if not correctly performed or adhered to, could cause personal injury or in
the worst case death. Do not proceed beyond a WARNING notice until the indicated
conditions are fully understood or met.
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, process
or similar, which if not correctly performed or adhered to, could cause damage to the
Generator or the Application. Do not proceed beyond a CAUTION notice until the indicated
conditions are fully understood or met.
Infinity 1051 User Manual
Page 6 2014 © Peak Scientific - Rev. 0 - 21/01/14
Technical Specifications
Environment
1051
Min/Max Air Operating Temperature
5°C - 30°C / 41°F - 86°F
Maximum Relative Humidity
70%
Min/Max Storage Temperature
-20°C - 60°C / -4°F - 140°F
Inlet Conditions
Min/Max Air Inlet Pressure
8.3-10 bar / 120-145 psi / 830-1000 kPa
Minimum Air inlet Flow
110 L/min
Outlet Gas
Maximum Gas output Pressure
110 psig / 760 kPa
Maximum Pressure Drop (Outlet-Inlet)
8 psig / 56 kPa
Max Nitrogen gas output flow
25 L/min (0.88 cfm)*
Max Air output flow
25 L/min (0.88 cfm)*
Start up time for purity
60 minutes
Particles
0.01µm
* Refer to Gas Flow Combinations section.
General
Dimensions (cm) WxDxH
31x25x73
Dimensions (ins) WxDxH
12x10x29
Generator Weight (Kg/lbs)
21.3 / 47
Shipping Weight (Kg/lbs)
29.3 / 64.5
Infinity 1051 User Manual
Page 7 2014 © Peak Scientific - Rev. 0 - 21/01/14
Introduction
The Peak Scientific Infinity 1051 Generator is designed specifically for use with
Mass Spectrometer applications. The Generator has been designed to produce
Nitrogen from an existing source of dry, oil free, compressed air in the
laboratory.
Unpacking and Installation
Although Peak Scientific take every precaution with safe transit and packaging,
it is advisable to fully inspect the unit for any sign of transit damage.
Check ‘SHOCKWATCH’ label for signs of rough handling prior to un-packing
Any damage should be reported immediately to the carrier and Peak Scientific
or the Peak Partner from where the unit was purchased.
Follow the unpacking instructions posted on the side of the crate. It will require
two people to remove the unit from the shipping crate and to manoeuvre the
Generator onto the floor.
Please save the product packaging for storage or future shipment of the
Generator.
After unpacking and a visual inspection, the unit should be
placed in a ventilated area away from direct sunlight. Care
should be taken not to obstruct the ventilation holes on the
side of the unit. The Generator should be placed on a steady
and level base. The Infinity 1051 has been designed to fit
under most workbenches. Alternatively, the unit may be wall
mounted using the fixing slots provided.
Performance of the Generator (like all sophisticated
equipment) is affected by ambient temperatures. Prolonged
operation in temperatures exceeding 30°C will shorten the
life of the unit.
Infinity 1051 User Manual
Page 8 2014 © Peak Scientific - Rev. 0 - 21/01/14
Fittings Kit
Supplied in the Fittings Kit are all the fittings required to connect the Infinity
1051 Generator to the application. The contents of the Fittings Kit are as
follows:
1. ¼” Compression Fittings × 2
2. ¼” Teflon Tubing × 8m
3. Screw N8 x 1 ½ Supa Counter Sunk x 3
4. Raw-Plug for N8 Screw (521-3087) x 3
5. Installation GuideInfinity 1051 × 1
6. User Manual CDInfinity 1051 × 1
General Dimensions
Infinity 1051 in mm
Infinity 1051 User Manual
Page 9 2014 © Peak Scientific - Rev. 0 - 21/01/14
Air Connection
Infinity 1051 Generator should be connected to a clean, dry, OIL - FREE source
of compressed air. A minimum inlet pressure of 120 psi / 830 kPa / 8.3 bar is
required. Any doubts as to the suitability of your compressed air supply should
be referred to Peak Scientific or any of their authorised partners.
The generator has a Breathing Air Filter with ¼” BSPT connection to the left
side of the unit. The Compressed Air supply should be connected here. This
filter will drain moisture and is equipped with an automatic drain. The drain
should be led to a convenient place. There are 2-off ¼” BSPT Female bulkhead
connections to the right of the unit. Fittings are provided to connect these
ports to the gas connections of the Mass Spectrometer.
LC/MS with 2-off Connections (e.g. LC-TQ)
Connect the 2-off gas ports at the right side of the generator to their
respective ports on the Mass Spectrometer using the 2-off ¼” BSPT x ¼”
Swagelok fittings. Use ¼” tubing throughout.
Infinity 1051 User Manual
Page 10 2014 © Peak Scientific - Rev. 0 - 21/01/14
Commissioning
With the Generator installed (as described earlier), connect compressed air to
the unit and open the air supply to the Unit. Disconnect the Outlet connections
to allow the generator to vent to atmosphere until the unit is stabilised. The
Generator has been pre-set in the factory to give the specified output flow-rate
and pressure in line with known requirements. Once the Membranes reach the
design pressures the Generator will stabilise and produce Nitrogen. Maximum
purity will be achieved after approximately 1 hour. The generator can then be
re-connected to the application.
The design of the generator is such that it can deliver up to the supplies
specified in the
Pressure & Flow Settings
table below. Should the demand for
gasses be less than the rated output flow at any time, the demand will be
determined by the consuming equipment. In circumstances of no demand, the
Generator remains operational and will provide Nitrogen immediately as
demand resumes.
Pressure & Flow Settings
The above settings should allow the Infinity 1051 to be operated with all
standard configurations for the Shimadzu LCMS 8050. Should the above
settings not provide sufficient flow or pressure for your application please
contact Peak Scientific for assistance.
Cleaning
Clean the outside of the Generator only using warm soapy water and a clean
damp cloth. Ensure the cloth is thoroughly rung out to remove excess fluid
prior to use.
Under no circumstances should any solvents or abrasive
cleaning solutions be used as these can contain fumes that
could be harmful to the Generator.
Flow/Pressure (l/min @ psi)
Port
Usage/Connection
1051
Nitrogen
Curtain Gas
25 @ 100 /690 kPa
Air
Ion Source/Nebuliser
25 @ 100 / 690 kPa
Infinity 1051 User Manual
Page 11 2014 © Peak Scientific - Rev. 0 - 21/01/14
Maintenance Schedule
Servicing and/or repair of the Generator should only be
undertaken by a TECHNICALLY COMPETENT PERSON with
the Generator in a safely isolated condition.
Due to the simplicity of the design and the small number of moving parts, the
Infinity 1051 Nitrogen Generator will have a long and trouble free life. However
as with all scientific and technical equipment it should be regularly inspected
and serviced as below.
Service Schedule
As an alternative to purchasing the 12 month service items individually an
Annual Service Kit is available as one part number. This contains all the filters
required for this Generator.
Safely Isolated Condition
The unit is in a safely isolated condition when it is disconnected from its
application and fully de-pressurised. Directions for isolating the Generator are
shown below.
Failure to place the Generator in a safely isolated condition
when instructed to do so may lead to personal injury or injury
to others and even death.
a) Disconnect from air supply.
b) Ensure the output pressure gauge read zero. (If gauge does not fall to
zero, loosen the outlet fitting slightly to allow trapped gas to escape).
Disconnect from the application.
Service interval
Component
Part No.
Qty.
12 months
Element Active Carbon (Black)
00-0046
1
Element 0.01 micron (Blue)
00-0045
1
RAC Filter Element
00-0039
2
Purchase Interval
Component
Part No.
Qty.
12 months
Annual Service Kit
08-0317
1
Infinity 1051 User Manual
Page 12 2014 © Peak Scientific - Rev. 0 - 21/01/14
IMPORTANT
DOCUMENT
S
Warranty Entitlement
To register your generator for your warranty entitlement, send the completed
form to Peak Scientific by:
Email
warranty@peakscientific.com
Online
http://www.peakscientific.com/service-and-support/warranty_registration
Phone
F
ax
+44 (0)141 530 4185
+44 (0)141 812 8200
PRODUCT WARRANTY REGISTRATION
COMPANY:
CONTACT NAME:
ADDRESS:
EMAIL ADDRESS:
CITY/TOWN:
GENERATOR SERIAL NUMBER:
POSTCODE:
COUNTRY:
MODEL TYPE:
TELEPHONE:
INSTALLATION DATE (DD/MM/YYYY):
Important Please Note:
You have 1 month to register your Peak Scientific product from the date of shipment.
If you wish to defer installation of your generator you must notify Peak Scientific within
1 month of the shipment date. This can be done by emailing
warranty@peakscientific.com Once registered the warranty will be honoured for a
period of 12 months after the installation date.
For any generators that remain unregistered the warranty will begin from date of
shipment.
Thank you on behalf of Peak Scientific.
2014 © Peak Scientific - Rev. 0 - 21/01/14
Peak Scientific UK
Fountain Crescent
Inchinnan Business Park
Inchinnan
PA4 9RE
Scotland, UK
Peak Scientific North
America
19 Sterling Road
Suite #1
Billerica, MA 01862
USA
Peak Scientific India
202, Amsri Shamira
Old Lancer Line
Opp. St. Mary’s Degree
College
S.D. Road Secunderabad
500 025, India
Peak Scientific Africa
PO Box 478
Somerset Mall
7137
South Africa
Tel: +44 (0)141 812 8100
Fax:
+44 (0)141 812 8200
Tel: +1 866 647 1649
Fax:
+1 978 608 9503
Tel: +91 40 2780 0663
Fax:
+91 40 2780 0663
Tel: +27 (0)2185 16542
Fax:
+27 (0)8654 64473
Peak Scientific Germany
Herriotstrasse 1
60528 Frankfurt
Germany
Peak Scientific China
Room 606 Building 1
Lane 2277 Zuchongzhi
Road
Pudong New Area
Shanghai
201203, China
Peak Scientific Mexico
Solon 352
Col. Los Morales Polanco
11530 Mexico, D.F.
Mexico
Peak Scientific Australia
PO Box 65
Belmont
Victoria 3216
Australia
Tel: +49 (0)69 677 33 490
Fax:
+49 (0)69 677 33 200
Tel: +86 21 5079 1190
Fax:
+86 21 5079 1191
Tel: +1 866 647 1649
Fax:
+1 978 608 9503
Tel: +61 1300 965 352
Peak Scientific Taiwan
4F.-6, No.736
Zhongzheng Rd.
Zhonghe Dist.
New Taipei City
Tel:
+886 2 8226 2383
Fax:
+886 2 8226 9499
Peak Scientific Brazil
Peak Scientific Brasil
Av. Quieroz Filho
1700 torre Sunny-Conj.
602
São Paulo SP Brasil
Tel:
+1 866 647 1649
Peak Scientific Japan K.K.
2-7-56, 2F Fuji Building 28
Kita Aoyama, Minato-Ku
Tokyo, Japan
107-0061
Tel:
+81 3-6864-0468
Peak Scientific Singapore
3 Science Park Drive
#03-14 The Franklin
Science Park Drive 1
Singapore, 118223
Tel:
+65-6777 1966
Fax:
+65-6777 1978
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13

Peak Scientific Infinity 1051 User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI