  • Hello! I am an AI chatbot trained to assist you with the Lovibond 56B006101 Owner's manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
www.lovibond.com
Legionella pneumophila serogroup 1 in Water
Instruction Manual
Part Number: 56B006101
Systems Test Kit
Lovibond
®
Water Testing
Tintometer
®
Group
Overview
The Legionella Systems Test Kit is for detecting Legionella in a water sample -- it is ideal for on-line and a variety
of industrial applications.
The test operates via a Lateral Flow Immuno-Chromatographic Assay (LFICA).
Each Systems Test Kit contains the following:
•5Individuallyfoil-wrappedtests--eachcontainsanexactvolumepipette
•5HollowFiberFilters
•5SyringeswithRecoveryBuffer
•1SampleCollectionBottle
•1Samplepointpipeadapter
•Instructionsforuse
This test is intended for analysis of water samples only. This product is not intended for clinical or medical
diagnostic use.
Theproductisintendedtobeusedaspartofanoverallwatertreatment,management,andriskreductionap-
proachandshouldNOTbeusedasthesolemethodforassessingrisksassociatedwithLegionellabacteria.
Test Operating Limits
Thetesthasbeenevaluatedforoperationbetween10-40ºC(50-104ºF).
Thetesthasbeenvalidatedforsamplesthatlterinlessthan10minutes.Samplesrequiringmorethan10min-
utestoltermaygiveerroneousresults.Samplesthatrequirelongtimestosamplemaybeindicativeofpoor
systemmaintenance.Awiderangeofnon-oxidizingbiocidesandbiodispersantshavebeencheckedforcross
reaction and inference with the test. This test should not be used on systems treated with biguanide or
THPS based biocides!
PerformanceofthetestwithsamplesotherthanthosetakenfromCoolingTowers,DomesticHot&Coldwater
systemsandWhirlpoolspashasnotbeenestablished.
Disposal
Thetest,lter,syringeandcapscannotbereusedorrecycled.Thepackagingmaterialsandthisinstruction
manualcanberecycled.
Storage
Thetestisintendedforstorageatroomtemperature.DoNotFreeze.Whenstoredcorrectly,thetestwill
continuetooperatewithindesignspecications,untilthespeciedexpirationdate.Do not use the test after
the stated expiration date specied on the packaging, Do not use any test where the foil packaging
is damaged.
Technical Information
Limit of Detection
Laboratoryanalysishasdemonstratedthat98-100%(91%Cl)oftestsarepositiveforcleanwatersamples
containing100CFU/LiterLegionella pneumophilaserogroup1.Thetheoreticalmathematicallimitofdetec-
tionofthetestisequivalent100CFU/Lwhena250mlsampleisltered.Ifsmallervolumesareprocessed,the
detectionlimitwillbealteredaccordingly:
Suspendedsolidcontentinthewatersamplesimpactsltrationandtestperformance,includinganalyticalsen-
sitivity.Acutualresultswillvary.Watersampleswithhighlevelsofsuspenedsolidsmayblockltrationentirely.
Legionella pneumophilaserogroup1bacteriarecoveryfromwatersamplescanrangefrom<10to100%,
dependingonwatersamplecomposition.Thisissimilartoltrationconcentrationtechniquestusedinother
microbiologicalanalysismethods.
limit of detection =
100CFU
Sample Volume (L)
Fit Sample Point
Identifythewatersystemtobesampledandanappropriatelocationfromwhichtoobtainarepresentative
sample.Largesystemsmayneedtobesampledandtestedatmultiplelocations.Attachthesamplepoint
adaptertothepointinthesystemtobetested.Itisrecommendedthatasamplepointadapterbettedand
retained to each sample point to avoid cross contamination. The sample line pressure should not exceed 6
bar (87 psi).
Resuspend the Bacteria
Disconnectthelterfromthesamplepointadapterbytwistingthelockringandpullingthelterfromthet-
ting.Removethewhitescrewcapfromthelterandreplaceitontheoppositeendofthelter.
•Pullthesyringebacktothe0.5mlmarktosuspendtherecoverybuffer,thenpushthesyringeall
thewaytothe0mlmark.
•Repeatthisprocesstwo(2)additionaltimes.
•Drawthesyringebacktocollect0.1mlofsample,thendisconnectfromthelter.Thesyringe
nowconainstherecoveredbacteriareadyfortesting.
Add Sample to the Test Strip
Removethetestfromitsfoilwrapping,andplacethetestonaatsurface.
Thefoilwrappingshouldnotbeopeneduntilimmediatelypriortorunningthetest.Ifthefoilisopenedand
thenthetestisnotperformedwithin60minutes,discardthetest.
Beforeuse,thetestshouldhave2palebluelinesacrosstheresultwindow.Ifthesearenotpresent,notifyyour
suppliertoreplacethetestkit.Takethepipettefromthefoilwrapping.
Placethesyringeoverthesmallsamplewindowattheendoftheteststrip,andpresstheplungersothat0.1ml
of sample is dispensed onto the test strip. Record the time.
Read the Test Strip
Leavetheteststripsittingonaatsurfaceduringincubation.Foroptimumresultsthetestresultsshouldbe
performed at room temperature. After waiting 25 minutes, examine the test strip in good lighting. If the
testisnotreadwithin60minutesofaddingthewatersample,itshouldbediscardedandanewtestshouldbe
run. The test should show one of the following results on the large result window on the test strip:
Negative Result:One(1)REDlineacrosstheresultwindowattheendfurthestfromthesamplewindow.
Positive Result:Two(2)REDlinesacrosstheresultwindow.Theredlineclosesttothesamplewindowmay
beveryfaint(palepink).Anydistinctline,nomatterhowfaint,shouldbecountedaspositive.
Invalid Result: If the test does not show any red lines; or if it only shows a line at the end closest to the sample
window;orifthelinefurthestfromthesamplewindowisveryfaint,thenthetestisinvalid.Repeatthetestand
notify your supplier for technical support.
Test Procedure
Take the Sample
Beforetakingthesample,ushthesampleportfor15seconds.Attachtheltertothesamplepointadapter;
pushtheopenendofthelterintotheplasticluerttingandtwistthelockingringtosecure.Holdthesample
collectionbottleundertheoutletportofthelter.
Openthesystemvalvetoallowtheltrationtostart,lter250mlofthewaterandcloseoffthevalve.
Ifthesampletakeslongerthan10minutestolter,thenstoptheltrationprocessandmeasuretheamountof
liquidinthecollectionbottlebeforeproceedingtothenextstep(thedetectionlimitwillbeaffected,seLimitof
Detectionsection).
Tintometer Inc.
6456ParklandDrive
Sarasota,FL34243
Phone:941.756.6410
Fax:941.727.9654
Email: sales@tintomter.us
Web:www.lovibond.com
Interpreting the Results
Positive Results:
A positive test result indicates that Legionella pneumophilaserogroup1bacteriawerepresentinthesample
abovethedetectionlimit.
Thetestdoesnotdifferentiatebetweenviable(living)andnon-viable(dead)organisms.Thetestwilldetectvi-
ablebutnon-cultruablebacteriawhicharenotdetectablebytraditionallaboratorytechniques.Apositiveresult
doesnotnecessarilymeanthatviablebacteriaarepresent.
Whenapositiveresultisobserved,seekadvicefromyourriskmanagementplan,orwatertreatmentspecialist.
Negative Results:
A negative test result indicates that Legionella pneumophilaserogroup1bacteriawerenotdetectedorthe
numberofbacteriainthesamplewerebelowthedetectionlimit.
A negative result does not necessarily mean that Legionella bacteriaareabsent.
AnegativeresultdoesnotmeanthatthesystemiscompletelyfreefromrisksassociatedwithLegionella bacte-
ria. The test only detects Legionella pneumophilaserogroup1.Thetestdoesnotdetectthepresenceofother
Legionella species or serogroups.
Test Procedure, continued.
Specicity
The test has been shown to be non-reactive with the following bacteria ( at 1 x 10
8
organisms per sample):
Staphylococcus aureus and Legionella pneumophilaserogroups4&7inconcentrationshigherthan1x10
8
organisms per sample may interfere with test results in negative samples. These concentrations are higher than
woudlbeexpectedtobepresentinnormalwatersamples.
Acinetobacter calcoaceticus,
Aeromonas hydrophila subsp.
hydrophila, Bacillus subtilis,
Burkholderia cepacia, Citrobacter
freudii, Citrobacter koseri,
Escherichia coli, Enterobacter
cloacae, Klebsiella oxytoca, Pseu-
domonas
aeruginosa, Pseudomonas
uorescens, Pseudomonas putida,
Pseudomonas stutzeri,
Ralstonia pickettii, Raoultella
terrigena, Streptococcus pyro-
genes, Yersinia ruckeri
HYDROSENSEisatrademarkofAlbagaiaLimited,usedworldwideandregisteredinvariousterritories.
/