Corning Ascent® FBR PD System Peristaltic Pump Owner's manual

Type
Owner's manual
Corning® Ascent® FBR PD System
Peristaltic Pump
Guidelines for Use
The Corning Ascent Fixed Bed Reactor (FBR) PD System consumables (Cat. Nos. 6971 - 6976) use high quality tubing designed for
extended-use operations with peristaltic pumps for the recirculation pump section. As with all peristaltic pumps, in rare cases of
prolonged use or improper loading, tubing can wear abnormally, fracture, and eventually leak. The following are best practices to load and
monitor tubing to prevent leaks from occurring on the Ascent FBR PD System.
During initial loading of the tubing loop through the recirculation pump or when opening the pump head for tubing inspection, tubing
should be taut and centered in the pump channel prior to closing the pump head. A slight tension can be placed on the tubing by
manually sliding the consumable sensor tray away from the pump head while closing the pump head (Figure 1). The tray may retract
slightly following release.
Following the Media Maintenance Phase, the tubing should be inspected for proper centering in the channel and re-tensioned, as the
tubing may stretch slightly during initial use (Figures 2 and 3).
Discoloration and attening are normal during peristaltic pumping operation (Figure 4), and external particulate generation on the
tubing and pump head during pumping is common. Particulate should be removed from the pump head regularly as it is observed.
Particulates can be removed using a lint-free wipe. Excessive particulate generation can be a sign of abnormal or excessive tubing wear.
Additional inspection, centering and tensioning may be recommended in these cases.
For operations with consistent high RPM (>250) and extended time periods (>3 days), the tubing should be inspected daily for abnormal
wear per the pump and tubing manufacturers’ recommendations and repositioned in the center of the rollers. If the presence of deep
ssures, gouges, or cuts through the thickness of tubing are observed, the risk of the tube leaking is elevated. The user should consider
stopping the run (Figure 5).
Figure 1. Tubing should be taut and centered in the channel while the pump head is closed. Gently sliding the
consumable sensor tray away from the pump head (white arrow) can ensure the tubing is taut during loading and
reloading.
2
Figure 2. Examples where the tubing is positioned too high (A) or low (B) in the channel of the pump.
The tubing should be centered vertically with respect to the roller.
A B
Figure 3. Properly loaded, the tube is centered in the channel without excess tubing through the pump head (A).
Approximately 3 mm of roller should be visible when the tube is centered (B).
A B
Figure 4. Normal tubing wear. The tubing lacks ridges, gouges, dents, or material loss. Discoloration and attening
are normal during peristaltic pumping operation.
Figure 5. In the rare occurrence that the tubing fails in a peristaltic pump, the failure is generally near the end of
the pumping channel out the out-ow side (A). If the presence of ssures, gouges, or cuts are observed (B), the risk
of leaking is elevated, and the user should consider stopping the run as a leak may occur.
BA
Flow
Elevated
wear region
For a listing of trademarks, visit www.corning.com/clstrademarks. All other trademarks are the property of their respective owners.
© 2022 Corning Incorporated. All rights reserved. 11/22 CLS-AN-719DOC REV1
Corning Incorporated
Life Sciences
www.corning.com/lifesciences
NORTH AMERICA
t 800.492.1110
t 978.442.2200
ASIA/PACIFIC
Australia/New Zealand
t 61 427286832
Chinese Mainland
t 86 21 3338 4338
India
t 91 124 4604000
Japan
t 81 3-3586 1996
Korea
t 82 2-796-9500
Singapore
t 65 6572-9740
Taiwan
t 886 2-2716-0338
EUROPE
France
t 0800 916 882
Germany
t 0800 101 1153
The Netherlands
t 020 655 79 28
United Kingdom
t 0800 376 8660
All Other European Countries
t +31 (0) 206 59 60 51
LATIN AMERICA
Brazil
t 55 (11) 3089-7400
Mexico
t (52-81) 8158-8400
Warranty/Disclaimer: Unless otherwise specied, all products are for research use or general laboratory use only.* Not intended for use in diagnostic
or therapeutic procedures. Not for use in humans. These products are not intended to mitigate the presence of microorganisms on surfaces or in the
environment, where such organisms can be deleterious to humans or the environment. Corning Life Sciences makes no claims regarding the performance
of these products for clinical or diagnostic applications. *For a listing of US medical devices, regulatory classications or specic information on claims,
visit www.corning.com/resources.
Corning’s products are not specically designed and tested for diagnostic testing. Many Corning products, though not specic for diagnostic testing, can be
used in the workow and preparation of the test at the customers discretion. Customers may use these products to support their claims. We cannot make any
claims or statements that our products are approved for diagnostic testing either directly or indirectly. The customer is responsible for any testing, validation,
and/or regulatory submissions that may be required to support the safety and efcacy of their intended application.
  • Page 1 1
  • Page 2 2
  • Page 3 3

Corning Ascent® FBR PD System Peristaltic Pump Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI