RTD TRANSMITTER
MODEL
RB/RBS
INSTRUCTION MANUAL
BEFORE USE ....
Thank you for choosing M-System. Before use, check the
contents of the package you received as outlined below.
If you have any problems or questions with the product,
please contact M-System’s Sales Office or representatives.
■ PACKAGE INCLUDES:
Signal conditioner (body + base socket) .........................(1)
■ MODEL NO.
Confirm that the model number described on the product is
exactly what you ordered.
■ INSTRUCTION MANUAL
This manual describes necessary points of caution when
you use this product, including installation, connection and
basic maintenance procedures.
POINTS OF CAUTION
WARNING
•RiskofElectricalShock:TheFrontCoveristobeOpened
Only by Qualified Service Personnel.
•BeforeYouRemovetheUnitfromitsBaseSocketorMount
it, Turn Off the Power Supply and Input Signal for Safety.
■ CONFORMITY WITH EU LVD DIRECTIVE OR UL
(CE/ULisunavailableforcertainmodelsufxoptions.)
•ThisequipmentissuitableforuseinaPollutionDegree2
environment and in Installation Category II.
Reinforcedinsulation(signalinputoroutputtopowerin-
put: 300V) andbasic insulation (signal input to output:
300V)aremaintained.(NoI/OisolationforRB).
Prior to installation, check that the insulation class of this
unit satisfies the system requirements.
•Altitudeupto2000meters.
•Theequipmentmustbemountedinsideapanel.
•The equipment mustbe installed such that appropriate
clearance and creepage distances are maintained to con-
formtoCE/ULrequirements.Failuretoobservethesere-
quirementsmayinvalidatetheCE/ULconformance.
■ CONFORMITY WITH EU EMC DIRECTIVE
•InsertnoiseltersforthesignalI/Oandthepowersource
connected to the unit. TDK Model ZCAT 3035-1330 or
equivalent is recommended.
•The actualinstallation environments such as panelcon-
figurations, connected devices, connected wires, may affect
the protection level of this unit when it is integrated in
apanelsystem.TheusermayhavetoreviewtheCEre-
quirements in regard to the whole system and employ ad-
ditionalprotectivemeasurestoensuretheCEconformity.
■ POWER INPUT RATING & OPERATIONAL RANGE
•Operationalrange&powerconsumption:Checkthepow-
er rating for the unit with the specifications.
ACratings:Rating±10%,50/60±2Hz,
approx.2VA(approx.3VAwithOption/E2)
DCratingsexcept110V:Rating±10%,approx.2W
(approx.3WwithOption/E2)
110VDCrating:85–150V,approx.2W
■ UNPLUGGING THE UNIT
•Beforeyouremovetheunitfromitsbasesocketormount
it, turn off the power supply and input signal for safety.
■ ENVIRONMENT
•Indooruse.
•Whenheavydustormetalparticlesarepresentintheair,
install the unit inside proper housing with sufficient ven-
tilation.
•Donotinstalltheunitwhereitissubjectedtocontinuous
vibration.Donotapplyphysicalimpacttotheunit.
•Environmentaltemperaturemust bewithin -5 to+60°C
(23to140°F)withrelativehumiditywithin30to90%RH
in order to ensure adequate life span and operation.
■ WIRING
•Do not install cables close to noise sources (relay drive
cable, high frequency line, etc.).
•Do not bind these cables together with those in which
noisesarepresent.Donotinstalltheminthesameduct.
■ AND ....
•Theunitisdesignedtofunctionassoonaspowerissup-
plied,however,awarmupfor10minutesisrequiredfor
satisfying complete performance described in the data
sheet.
COMPONENT IDENTIFICATION