15
Function (8)
Selects the parameter to adjust.
•The FUNCTION switch is active only in Manual or Local
mode. The enunciators above the FUNCTION switch will be
blank when in Local Restricted or Remote mode.
Voltage
For an MPS-3000 CE FC power supply, voltage is adjustable for the
Beam, Accelerator, or Discharge power supply. The word VOLTAGE
appear above the FUNCTION switch.
For an MPS-3000 CE HC power supply, voltage is adjustable for the
Beam and Accelerator power supplies only. The word VOLTAGE
appears above the FUNCTION switch.
Current
MPS-3000 CE FC: The current is adjustable only when the Neutralizer
or Cathode power supply has been selected.
MPS-3000 CE HC: The current is adjustable only when the Hollow
Cathode Neutralizer or Discharge power supply has been selected.
MPS-3000 CE PBN: The current is adjustable only when the Cathode
or PBN power supply has been selected.
The beam current may be selected when operating in Local mode. The
word CURRENT appears above the FUNCTION switch indicating that
beam current has been selected.
A/B Ratio
The A/B Ratio may be adjusted only when the Accelerator power supply
has been selected when operating in Manual or Local mode.
• The letters A/B appear above the FUNCTION switch indicat-
ing that A/B Ratio has been selected. The A/B Ratio may also
be adjusted through a computer.
Limit
MPS-3000 CE FC, HC, or PBN: The limit is adjustable, when the Cath-
ode, Neutralizer or Discharge power supply has been selected.
• The word LIMIT appears above the FUNCTION switch. Limits
may be adjusted in Manual mode or through a computer.
MPS-3000 CE PBN: The PBN filament current may be viewed when in
Local mode. Select PBN and LIMIT with the MODULE and FUNCTION
switches.
Heater
The letters HCN appear above the module switch indicating that Hol-
low Cathode Neutralizer has been selected. The word HEATER appears
above the FUNCTION switch.