ver.20221202E
Resonac Corporation (https://www.shodex.com/)
5
7.3 Solvent Exchange
To replace the solvent, set the system at less than 0.3 mL/min. Recommended solvent volume to introduce at
each step is 3 to 5 times of the column volume.
(1) Check miscibility/solubility of the desired new solvent and the solvent currently filled in the column.
(2) When replacing the current solvent with a miscible solvent, first introduce 1:1 mixture of the current solvent
and the new solvent, and then replace it with 100 % new solvent.
e.g. When replacing highly concentrated buffer solution (or aqueous salt solution) to methanol, first run 1:1
mixture of buffer (or aqueous salt solution) and methanol, then replace it with methanol.
(3) When replacing the current solvent with a solvent with low miscibility/solubility to the current solvent, first
use a solvent that is miscible/soluble to both solvents, and then replace it with the new solvent.
e.g. When replacing highly concentrated buffer solution or salt solution to water/acetonitrile, first run water
and then replace it with the eluent.
Attention! · Frequent solvent replacement deteriorates the column, and thus not recommended.
7.4 Column Cleaning
Problems in peak shapes and elution time changes or elevated column pressure are often caused by the
deposition of insoluble or adsorbing components from the sample/flow-line inside the column. These problems
may be resolved by cleaning the column.
If a guard column is used with an analytical column(s), first remove the guard column and check the
performance of the analytical column alone. If the problem is solved, most likely the cause was from the guard
column. In this case, clean the guard column.
If the problem remains even after removing the guard column, clean both guard and analytical columns. Make
sure to clean the guard and the analytical columns separately. In case multiple number of analytical columns
are used together, wash them separately. During the column cleaning, disconnect the detector and collect the
washing solution directly from the column outlet into a waste container (i.e., do not let the solution go through
the detector).
If the column performance does not improve (recover) after performing the column cleaning, please
replace the column with a new one.
<Cleaning method>
(1) Insoluble components that block the column inlet may be removed by reversing the flow direction, i.e.,
introducing the eluent from the column outlet, with flow rate at less than half of the recommended flow rate.
(2) Follow below cleaning steps for adsorbing components. For an efficient cleaning, reverse the flow direction.
Set the flow rate at less than 0.5 mL/min. Recommended solvent volume to introduce is 5 to 10 times of the
column volume.
Method 1: Adsorption of hydrophobic compounds (when using aqueous eluent)
Introduce the eluent with a high polar organic solvent (acetonitrile or methanol).
Method 2: Adsorption of ionic compounds
Introduce the eluent with a higher salt concentration.
Attention! · Keep the organic solvent and salt concentrations within the concentrations stated in
4.2 List of Applicable Solvents.
· Keeping the washing solution in the column for a long time will lead to column deterioration.
Please replace the washing solution with the eluent immediately after cleaning.
8. Column Storage
Remove the column from HPLC system after replacing the in-column solvent with the initial shipping solvent.
Securely tighten the end caps and store the column at a location with stable temperature (a cool and dark space
is recommended). Refer to section 7.3 Solvent Exchange for how to replace the eluent.
Attention! · Never allow inside the column to dry. It can damage the column.