EKO MS-80 Albedometer Quick start guide

  • Hello! I'm your chat assistant. I've reviewed the Quick Start Guide for the EKO pyranometers MS-80, MS-60, MS-40, MS-805, MS-605, and MS-405. This document outlines the installation, wiring, and output checking procedures for these devices. It also provides key information on calculating solar irradiance and periodic maintenance, such as recalibration. I'm ready to answer any questions you might have about these pyranometers.
  • What is the recommended recalibration period for MS-80/80S?
    What is the recommended recalibration period for MS-60/60S and MS-40/40S?
    What should I do if any part is missing or damaged?
    What is the purpose of the shunt resistor?
SDI 12
5-36V
SDI-12
V
V
Voltage
V
M
Voltage
Branch line
4-20mA 0-10mA (0-1V*)
DC 5 - 36V (+) DC 5 - 36V (+)
4-20mA(-)/GND
4-20mA(+)
- - - - -
0-10mA(-)/GND/0-1V*(-)
0-10mA(+)/0-1V*(+)
- - - - -
- - - - -
FG
- - - - -
FG
Modbus RTU SDI-12
DC 5 - 36V (+)
GND
RS485/-/A
- - - - -
DC12V (+)
GND
SDI-12 Data
RS485/+/B
- - - - -
- - - - -
FG FG
* When connect the precision resistor 100
S-2016-2019 QSG-MS80_60_40_19-07E_Ver.3
6
https://eko-usa.comhttps://eko-eu.comhttps://eko-asia.com
/
/ /
9
9
between the line of White and Gray.
*
*0-10mA: 0-1V through 100Ω external shunt resistor
When using the 0-1V, please prepare the precision resister 100
0-10mA/0-1V output line is disabled and can be enabled through the EKO congurator.
When using a shunt resistor to measure voltage, please use a resistor of <150
Voltage
V
Connect to a PC through the optional RS-485/USB converter cable
(Download free configurator software through EKO website)
[default settings]
/
/
Voltage
Add the shunt resistor 100 between
Gray(+)and White(-) terminal.
100R
0. 01%
Ω
1
Pin numbers
Pin numbers
(Please prepare separately)
In case of the lower 2digits of S/N is “00",
Defalut settings of Modbus(MS-80S/60S/40S)
Baud rate 19200bps
Stop bit 1bit
Parity check None
Data length 8bit
Address Lower 2digits of product S/N
**
**
the address is set to “100.
/