Teledyne 2000XTC User manual

Category
Oxygen Equipment
Type
User manual
OPERATING INSTRUCTIONS FOR
2000 XTC
Thermal Conductivity Transmitter
P/N M74838
ECO: 04-0203
REV 1
DANGER
Toxic gases and/or flammable liquids may be present in this monitoring system.
Personal protective equipment may be required when servicing this instrument.
Hazardous voltages exist on certain components internally which may persist
for a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing.
Teledyne Analytical Instruments
2000 XTC
Before conducting any maintenance or servicing, consult with authorized
supervisor/manager.
Copyright © 2002 Teledyne Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed,
stored in a retrieval system, or translated into any other language or computer language in
whole or in part, in any form or by any means, whether it be electronic, mechanical, magnetic,
optical, manual, or otherwise, without the prior written consent of Teledyne Analytical
Instruments, 16830 Chestnut Street, City of Industry, CA 91749-1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from
defects of material and of construction, and that our liability shall be limited to replacing or
repairing at our factory (without charge, except for transportation), or at customer plant at our
option, any material or construction in which defects become apparent within one year from the
date of shipment, except in cases where quotations or acknowledgements provide for a shorter
period. Components manufactured by others bear the warranty of their manufacturer. This
warranty does not cover defects caused by wear, accident, misuse, neglect or repairs other than
those performed by Teledyne or an authorized service center. We assume no liability for direct
or indirect damages of any kind and the purchaser by the acceptance of the equipment will
assume all liability for any damage which may result from its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus, and
to make any alterations in the dimensions, shape or weight of any parts, in so far as such
alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which valuable
data can be gathered. The information provided by the instrument may assist the user in
eliminating potential hazards caused by his process; however, it is essential that all personnel
involved in the use of the instrument or its interface, with the process being measured, be properly
trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process conditions.
While this instrument may be able to provide early warning of imminent danger, it has no
control over process conditions, and it can be misused. In particular, any alarm or control
systems installed must be tested and understood, both as to how they operate and as to how
they can be defeated. Any safeguards required such as locks, labels, or redundancy, must be
provided by the user or specifically requested of Teledyne at the time the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser is
responsible for the training of personnel, for providing hazard warning methods and
instrumentation per the appropriate standards, and for ensuring that hazard warning devices and
instrumentation are maintained and operated properly.
Teledyne Analytical Instruments, the manufacturer of this instrument, cannot accept
responsibility for conditions beyond its knowledge and control. No statement expressed or
implied by this document or any information disseminated by the manufacturer or its agents, is
to be construed as a warranty of adequate safety control under the user’s process conditions.
Teledyne Analytical Instruments ii
Thermal Conductivity Transmitter
Specific Model Information
It is not recommended that this instrument be used for analysis on any
other gas or gas mixture than that specified at the time of purchase.
Thermal conductivity analyzers are calibrated at the factory for a specific
application using a known gas mixture that is representative of the
customers’ process. Using this instrument to analyze any other gas mixture
may result in serious error. Consult the factory for additional information
for gas analysis not specified at the time of purchase.
Instrument Serial Number: _______________________
Instrument Range: _______________
Calibrated for: _______________
Background Gas: _______________
Zero Gas: _______________
Span Gas: _______________
Teledyne Analytical Instruments iii
2000 XTC
Safety Messages
Your safety and the safety of others is very important. We have
provided many important safety messages in this manual. Please read these
messages carefully.
A safety message alerts you to potential hazards that could hurt you or
others. Each safety message is associated with a safety alert symbol. These
symbols are found in the manual and inside the instrument. The definition
of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions
for details on the specific danger. These cautions warn of
specific procedures which if not followed could cause bodily
Injury and/or damage the instrument.
CAUTION: HOT SURFACE WARNING: This warning is specific
to heated components within the instrument. Failure to heed the
warning could result in serious burns to skin and underlying
tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous voltages
appear within this instrument. This warning is specific to an
electrical hazard existing at or nearby the component or
procedure under discussion. Failure to heed this warning could
result in injury and/or death from electrocution.
Technician Symbol: All operations marked with this symbol
are to be performed by qualified maintenance personnel only.
No
Symbol
NOTE: Additional information and comments regarding a
specific component or procedure are highlighted in the form of a
note.
CAUTION:
THE ANALYZER SHOULD ONLY BE USED FOR THE
PURPOSE AND IN THE MANNER DESCRIBED IN THIS
MANUAL.
Teledyne Analytical Instruments iv
Thermal Conductivity Transmitter
IF YOU USE THE ANALYZER IN A MANNER OTHER
THAN THAT FOR WHICH IT WAS INTENDED,
UNPREDICTABLE BEHAVIOR COULD RESULT
POSSIBLY ACCOMPANIED WITH HAZARDOUS
CONSEQUENCES.
This manual provides information designed to guide you through the
installation, calibration operation and maintenance of your new analyzer.
Please read this manual and keep it available.
Occasionally, some instruments are customized for a particular
application or features and/or options added per customer requests. Please
check the front of this manual for any additional information in the form of
an Addendum which discusses specific information, procedures, cautions
and warnings that may be peculiar to your instrument.
Manuals do get lost. Additional manuals can be obtained from
Teledyne at the address given in the Appendix. Some of our manuals are
available in electronic form via the internet. Please visit our website at:
www.teledyne-ai.com.
Teledyne Analytical Instruments v
2000 XTC
Table of Contents
List of Figures viii
List of Tables ix
Introduction………………………………………………………………1
1.1 Overview 1
1.2 Typical Applications 1
1.3 Main Features of the Transmitter 2
1.4 Operator Interface 2
Operational Theory……………………………………………………...5
2.1 Introduction 5
2.2 The Thermal Conductivity Sensor 5
2.3 Sample System 7
2.4 Electronics and Signal Processing 9
2.4.1 Transmitter 9
2.4.2 Galvanic Isolator 10
2.5 Connection Cable 11
Installation………………………………………………………………12
3.1 Unpacking the Transmitter 12
3.2 Mounting the Transmitter 13
3.3 Mounting the Interface Unit 13
3.4 Gas Connections 14
3.5 Electrical Connections 15
3.6 Calibration 17
3.6.1 Adjusting Internal Temperature 18
3.6.2 Zero Calibration 19
3.6.3 Span Calibration 20
Teledyne Analytical Instruments vi
Thermal Conductivity Transmitter
Maintenance…………………………………………………………….22
4.1 Routine Maintenance 22
4.2 Troubleshooting 22
4.3 Fuse Replacement 24
Appendix………………………………………………………………...26
Specifications 26
Recommended Spare Parts List 28
Reference Drawings 28
Index……………………………………………………………………...29
Teledyne Analytical Instruments vii
2000 XTC
List of Figures
Figure 1-1: XTC Transmitter Modules…………………………………..3
Figure 2-1: Thermal Conductivity Sensor ………………………………6
Figure 2-2: Recommended External Sample System…………………8
Figure 2-3: Internal Piping………………………………………………..8
Figure 2-4: Block Diagram of 2000 XTC………………………………..9
Figure 2-5: Model 2000 XTC External Wiring Diagram……………..11
Figure 3-1: Mounting Dimensions for Sensor Unit…………………...13
Figure 3-2: Mounting Dimensions for Interface Unit…………………14
Figure 3-3: Power and Output Connections to Galvanic Isolator…..16
Figure 3-4: Cable Connection to Galvanic Isolator Module…………16
Figure 3-5: Temperature and Calibration Adjustment Location…….18
Figure 4-1: Cable Identification in Transmitter………………………..24
Figure 4-2 Fuse Location………………………………………………25
Teledyne Analytical Instruments viii
Thermal Conductivity Transmitter
List of Tables
Table 3-1: Heater Temperature…………………………………..19
Table 3-2: Concentration from Output Current………………….21
Table 4-1 Troubleshooting………………………………………...222
Table 4-2: Fuses……………………………………………………244
Teledyne Analytical Instruments ix
2000 XTC
DANGER
COMBUSTIBLE GAS USAGE
WARNING
This instrument is approved as an intrinsically safe gas
analyzer for usage in a category (ib) Group IIC hazardous area.
This approval applies only to the equipment specified and
installed in accordance with the information contained within
this manual. It is the customer's responsibility to ensure
safety, especially when combustible gases are being analyzed
since the potential of gas leaks always exists.
The customer should ensure that the principles of operating of
this equipment are well understood by the user and that the
instrument as well as any approved support equipment is
properly installed. Misuse of this product in any manner,
tampering with its components, or unauthorized substitution of
any component may adversely affect the certification and
safety of this instrument.
Since the use of this instrument is beyond the control of
Teledyne, no responsibility by Teledyne, its affiliates, and
agents for damage or injury from misuse or neglect of this
equipment is implied or assumed.
Teledyne Analytical Instruments x
Thermal Conductivity Transmitter Introduction
Introduction
1.1 Overview
The Teledyne Analytical Instruments 2000 XTC Thermal Conductivity
Transmitter is an instrument designed to analyze the gas concentration of a
binary gas mixture. This manual covers the CENELEC/ATEX approved
model 2000 XTC Thermal Conductivity Transmitter. These units are rated as
intrinsically safe and may be used in Class I, Group A, B, C, D, Div. 1 (North
America) and EEx ib IIC T3 (IEC/Europe) hazardous environments.
1.2 Typical Applications
The 2000 XTC Thermal Conductivity Transmitter is versatile. It is
capable of measuring the gas concentration of a species in a binary or
pseudo-binary mixture of gases over a concentration range. The specific
species and gas mixture as well as the analysis range employed is fixed
at the factory for your specific application. Typical applications for this
transmitter include:
Determining the gas concentration of a binary gas, for
instance, Ar or N
2
in air
Analyzing the methane (CH
4
) concentration in natural gas
(80-100% CH
4
) sample
Analysis of H
2
in various backgrounds
CO
2
safety monitoring
Specific gas mixer and process control applications
Semiconductor manufacturing
Petrochemical process control
Quality assurance
Typical instrument ranges for this instrument include:
H
2
in N
2
(or air)
0-5%, 0-8%, 0-10%, 0-20%, 0-100%
Teledyne Analytical Instruments 1
Introduction 2000 XTC
He in N2 (or air)
0-5%, 0-8% 0-10%, 0-30%, 0-100%
Natural Gas in Methane
0-20%, 0-50%, 0-100%
Nitrogen (or air) in CO2
0-10%, 0-20%, 0-50%, 0-100%
1.3 Main Features of the Transmitter
The 2000 XTC Thermal Conductivity Transmitter is sophisticated
yet simple to use. It provides a 4-20 mA output signal proportional to
the concentration of one component of a gas mixture. The signal is
based on the difference between the thermal conductivity of the sampled
gas compared to calibration points. The main features of the analyzer
include:
Integral isolated safety barrier classified as intrinsically safe
for hazardous area use.
Nema 4X, IEC 60529 and IP 66 rated sensor housing
ATEX approved for EEx ib IIC T3 (Europe)
4-20 mA signal output proportional to the concentration of
one of the gases in the mixture.
Long-lifetime solid state thermal conductivity sensor
Temperature controlled cell block provides excellent zero
and span stability over a wide temperature range
Versatile analog signal processing IC.
Easy installation
1.4 Operator Interface
The user does not interact with the transmitter directly. The
standard 2000 XTC Thermal Conductivity Transmitter provides a 4-20
mA output signal proportional to the concentration of the gas being
analyzed. This signal can be used to drive devices such as displays,
concentration alarms, user-supplied devices, or it can be monitored
using a chart recorder.
Teledyne Analytical Instruments 2
Thermal Conductivity Transmitter Introduction
Figure 1-1 shows the 2000 XTC Thermal Conductivity Transmitter.
The instrument is comprised of two modules: the transmitter unit and the
galvanic isolator unit. The transmitter is located at the sample point
(which can be a hazardous area) and is where the sample gas input is
converted into an electrical output signal.
The galvanic isolator module contains the 24V power supply, the
power isolation safety barrier and signal output connection. This module
is used to provide safe power to the transmitter eliminating the danger of
ignition by sparks or heating due to excessive power.
Figure 1-1: XTC-Transmitter Modules
Teledyne Analytical Instruments 3
Operational Theory 2000 XTC
Teledyne Analytical Instruments 4
Thermal Conductivity Transmitter Operational Theory
Operational Theory
2.1 Introduction
The 2000 XTC is comprised of two subsystems:
1. The thermal conductivity sensor and transmitter
2. Galvanic isolator and power supply
The system directs a sample gas into the transmitter. Internally, the
sample is passed to the sensor and heated without contaminating or
altering the sample in any way. The thermal conductivity sensor is a
solid-state device that translates the amount of gas present in the sample
into a change in resistance (expressed in ohms). The electronic signal
processing and control circuits provide a linear output signal
compensated for ambient temperature fluctuations over a wide range.
Increased stability is achieved using integral heaters that control the
temperature of the sensor cell.
2.2 The Thermal Conductivity Sensor
The Thermal Conductivity Transmitter is designed to measure the
concentration of a component in a binary stream of gas. It can be used to
determine the concentration of a sample stream containing a composite
mixture of impurities by comparing the difference in thermal conductivity of
the sample stream with calibrated reference points.
Thermal conductivity measurements are non-specific by nature.
The thermal conductivity of a gas mixture depends on the conductivities
of the individual components of the mixture and their relative
concentration. The transmitter is calibrated at the factory to be linear
over the range and mixture of interest specified by the customer at the
time of purchase. In order to accurately detect and quantify a component
in a sample stream, the sample must be composed of the particular
component and background gas specified at the time of purchase. The
transmitter can be used with other binary gas mixtures, but the output
signal may or may not be linear across the sensor output range. Consult
the factory for additional information regarding the use of this
Teledyne Analytical Instruments 5
Operational Theory 2000 XTC
instrument for analyzing other gas mixtures. Generally, the user must be
able to calibrate the instrument with a gas containing the host gas and a
known concentration of the impurity or gas to be analyzed. At a
minimum, a zero gas and a mixture containing approximately 70-100%
of the species to be analyzed on the range of interest should be provided
in order to properly calibrate this instrument.
Figure 2-1 shows a cross section of the sensor used in the 2000 XTC. A
resistor is mounted on a heated membrane to which the sample gas is
exposed. This resistor is used to both heat and measure the temperature of the
membrane. A second resistor is fabricated directly on the silicon and is used
to compensate for changes in the ambient temperature. Depending on the
sample gas density, exposing the sensor to the gas will either increase the
membrane temperature or decrease or the measuring resistor temperature.
Implementing each of the two resistors in the analog signal conditioning
stages enables the null point of the circuit to be directly related to the thermal
conductivity of the sample gas.
Figure 2-1: Thermal Conductivity Sensor
During calibration, the analog signal is balanced for the zero gas at
one end of the measurement range, and span gas at the other end. The
resulting electrical signal is fed to a microprocessor which linearizes the
output based on data programmed at the factory. An E-to-I converter
produces an isolated 4–20 mA DC current output spanning the analysis
range.
Teledyne Analytical Instruments 6
Thermal Conductivity Transmitter Operational Theory
Control circuitry precisely regulates the temperature of the
measuring cell. These additional controlled heaters are installed to heat
the sample gas and compensate for ambient temperature fluctuations,
enhancing the stability and accuracy of the measuring process. The
heater temperature is adjustable using a set-screw adjustment
potentiometer inside the transmitter unit.
Control of the sample and supporting gases is not provided in the
basic design. TAI offers a variety of supporting gas control panels as
companion accessories to the analyzer to fill this need. In any case,
means must be provided for controlling the flow rates of sample and
calibration gas through the analyzer. A control manifold will be required
for the introduction of zero and span gas, as well as sample gas, into the
sample path. Appropriate pressure reducing regulators will have to be
installed at all gas supply sources; for those customers wishing to
incorporate their own sample controls, a recommended system piping
drawing is shown in Figure 2.2 and a reference drawing is included
among the drawings at the rear of the manual.
2.3 Sample System
A suitable external sample system must be provided by the
customer. The external sample system delivers calibration or sample gas
to the transmitter at suitable pressure and flow rate. Internally, the 2000
XTC uses a brass cell block with stainless steel tubing and connectors.
CAUTION: IT IS THE RESPONSIBILITY OF THE END-USER TO
ENSURE THAT THE GASES USED IN THIS
INSTRUMENT ARE COMPATIBLE WITH THE
MATERIALS USED IN THE SAMPLE SYSTEM. FAILURE
TO DO SO COULD RESULT IN PERSONAL INJURY AS
WELL AS DAMAGE TO THE INSTRUMENT.
The external sample system should be capable of supplying clean,
particulate-free and moisture-free sample gas to the transmitter. The
sample gas must be regulated between 2-20 psig. A flow control device
such as a flow meter with a needle valve should be incorporated to
maintain gas flow within 0.4-2 SCFH. It is convenient to install valves
in the sample system to allow switching between sample and calibration
gas. Figure 2-2 is a typical sample system for delivering sample and
calibration gases to the transmitter. It uses two (2) 3-way valves to
deliver either sample or calibration (zero or span) gas to the transmitter.
Teledyne Analytical Instruments 7
Operational Theory 2000 XTC
Figure 2-2: Recommended External Sample System
The sample path within the transmitter is a straight-through path.
Gas encounters no dead space. This minimizes residual gas pockets that
can interfere with accurate analysis.
Stainless steel 1/4” tube fittings are installed (6mm are available)
for connecting the external sample system to the transmitter. For metric
system installations, 6 mm adapters can be supplied if needed. The
sample gas path through the transmitter is shown in Figure 2-3.
Figure 2-3: Internal Piping
Teledyne Analytical Instruments 8
Thermal Conductivity Transmitter Operational Theory
2.4 Electronics and Signal Processing
For safety reasons, the electronics and signal processing section are
located remotely from the transmitter module. A galvanic isolator is
employed to limit the current that can be passed through the interconnect
cable to levels that are incapable of providing an ignition source. The
galvanic isolator also provides an optically-isolated 4-20 mA output
signal reconstructed from the 0-4 mA output signal sent from the
transmitter. All electronics in the transmitter are contained on a single
PC board accessible by removing the front cover. The transmitter is
powered from two (2) separate 15VDC lines from the galvanic isolator
module. The galvanic isolator incorporates a 24VDC power supply.
The processing electronics for the sensor are located inside the
transmitter case in addition to the sample cell heaters and E-to-I
converter. The control circuitry for the sample cell heaters, the 24 VDC
power supply, and electronics used to condition the output signal are
located in the galvanic isolator case. Figure 2-4 is a block diagram of the
instrument electronics.
Figure 2-4: Block Diagram of 2000 XTC
2.4.1 Transmitter
The differences in thermal conductivities of the sample gas provide
a small but detectable change in temperature of the resistive sensor
Teledyne Analytical Instruments 9
Operational Theory 2000 XTC
membrane. The amount of voltage needed to offset the analog output
(i.e. to null the circuit when the resistance changes due to temperature
effects) is measured and applied. A preprogrammed microcontroller
with EPROM uses the offset voltage from the analog stages to generate
a linearized output voltage signal, which is then converted to a 0-4 mA
current. This signal is fed back to the galvanic isolator for conversion to
a standard 4-20 mA current, to be used as interface with the user’s
readout device or alarm circuit. The linearized output is based on
calibrations performed and adjustments made at the factory for the
specific application at hand.
Certain factors outside of the gas concentrations bear on the
accuracy of the thermal conductivity sensor. The major contributor
involves temperature deviations or transients from a baseline
temperature. For this reason, a pair of controlled heaters is used to fix
the operating temperature of the sample cell. The control circuitry is
capable of maintaining a temperature setting from 18°C to 67°C within
the operating range of the instrument (0 to 50°C). Electronics for the
temperature control are housed in the transmitter module. Power is
supplied to the heaters by pins 2-3 for heater #1 and 1-4 for heater #2 on
the 5-wire interconnection cable.
2.4.2 Galvanic Isolator
The purpose of the safety barrier is to limit power to and from the
transmitter to levels that are incapable of generating enough energy to
cause ignition in a flammable or explosive gas mixture. The galvanic
isolator accepts a 24 volt DC input and provides two separate and
isolated 15 volt DC power outputs, plus receiving a 0-4mA current input
from the transmitter. Both power outputs are mutually isolated and
comply with the CENELEC requirement intrinsically safe to 60V. The
output from the isolator supplies power to the two (2) heaters in the
transmitter. Additionally, it supplies power to the sensor and PC board
for the compensation circuitry within the transmitter.
Each output side of the galvanic isolator is protected with a series
200mA fuse. A pair of zener diodes limits the open circuit voltage to
below 18V and short circuit current to 503mA or less. Power
fluctuations are reduced at the input side of the 24V input transformer
using a crowbar network of diodes.
Teledyne Analytical Instruments 10
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40

Teledyne 2000XTC User manual

Category
Oxygen Equipment
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI