Anritsu MS2691A Operating instructions

Category
Measuring, testing & control
Type
Operating instructions

This manual is also suitable for

MS2690A/MS2691A/MS2692A
and MS2830A
Signal Analyzer
Operation Manual
Phase Noise Measurement Function
R
e
m
ote
Co
n
t
r
o
l
Third Edition
For safety and warning information, please read this
manual before attempting to use the equipment.
Additional safety and warning information is provided
within the “MS2690A/MS2691A/MS2692A and
MS2830A Signal Analyzer Operation Manual
(Mainframe Operation)”. Please also refer to this
document before using the equipment.
Keep this manual with the equipment.
ANRITSU CORPORATION
Do
cument No.: M-W3118AE-3.0
Safety Symbols
To prevent the risk of personal injury or loss related to equipment malfunction, Anritsu Corporation uses the
following safety symbols to indicate safety-related information. Ensure that you clearly understand the meanings of
the symbols BEFORE using the equipment. Some or all of the following symbols may be used on all Anritsu
equipment. In addition, there may be other labels attached to products that are not shown in the diagrams in this
manual.
Symbols used in manual
DANGER
This indicates a very dangerous procedure that could result in serious injury or
death if not performed properly.
WARNING
T
his indicates a hazardous procedure that could result in serious injury or death if
not performed properly.
CAUTION
T
his indicates a hazardous procedure or danger that could result in light-to-severe
injury, or loss related to equipment malfunction, if proper precautions are not taken.
Safety Symbols Used on Equipment and in Manual
The following safety symbols are used inside or on the equipment near operation locations to provide information
about safety items and operation precautions. Ensure that you clearly understand the meanings of the symbols and
take the necessary precautions BEFORE using the equipment.
This indicates a prohibited operation. The prohibited operation is indicated
symbolically in or near the barred circle.
ii
This indicates an obligatory safety precaution. The obligatory operation is
indicated symbolically in or near the circle.
This indicates a warning or caution. The contents are indicated symbolically in or
near the triangle.
This indicates a note. The contents are described in the box.
These indicate that the marked part should be recycled.
MS2690A/MS2691A/MS2692A and MS2830A
Signal Analyzer
Operation Manual Phase Noise Measurement Function Remote Control
25 April 2008 (First Edition)
24 June 2011 (Third Edition)
Copy
right © 2008-2011, ANRITSU CORPORATION.
All rights reserved. No part of this manual may be reproduced without the prior written permission of the
publisher.
The contents of this manual may be changed without prior notice.
Printed in Japan
Notes On Export Management
This product and its manuals may require an Export License/Approval by
the Government of the product's country of origin for re-export from you
r
country.
Before re-exporting the product or manuals, please contact us to confirm
whether they are export-controlled items or not.
When you dispose of export-controlled items, the products/manuals need
to be broken/shredded so as not to be unlawfully used for military purpose.
iii
iv
About This Manual
Associated Documents
The operation manual configuration of the MS2690A/MS2691A/MS2692A
and MS2830A Signal Analyzer is shown below.
MS2690A/MS2691A/MS2692A
Signal Analyzer Operation Manual
(Main Frame Operation)
MS2690A/MS2691A/MS2692A and MS2830A Signal Analyzer Operation Manual
(Phase Noise Measurement Function Operation)
MS2690A/MS2691A/MS2692A and MS2830A Signal Analyzer Operation Manual
(Phase Noise Measurement Function Remote Control)
MS2690A/MS2691A/MS2692A and MS2830A
Signal Analyzer Operation Manual (Main Frame Remote Control)
MS2830A Signal Analyzer
Operation Manual
(Main Frame Operation)
Or
MS2690A/MS2691A/MS2692A
Signal Analyzer Operation Manual
(Signal Analyzer Function
Operation)
MS2690A/MS2691A/MS2692A and MS2830A Signal Analyzer Operation Manual
(Signal Analyzer Function Remote Control)
MS2830A
Signal Analyzer Operation Manual
(Signal Analyzer Function
Operation)
Or
MS2690A/MS2691A/MS2692A
Spectrum Analyzer Operation Manual
(Spectrum Analyzer Function
Operation)
MS2690A/MS2691A/MS2692A and MS2830A
Signal Analyzer Operation Manual (Spectrum Analyzer Function Remote Control)
MS2830A
Spectrum Analyzer Operation Manual
(Spectrum Analyzer Function
Operation)
Or
I
II
Signal Analyzer Operation Manual (Mainframe Operation)
Signal Analyzer Operation Manual (Mainframe Remote Control)
Description of basic operations, maintenance procedures, common
functions and common remote functions of the mainframe
Signal Analyzer Operation Manual (Signal Analyzer Function)
Signal Analyzer Operation Manual (Signal Analyzer Function Remote
Control)
Description of basic operations, functions and remote functions of the
signal analyzer
Signal Analyzer Operation Manual (Spectrum Analyzer Function)
Signal Analyzer Operation Manual (Spectrum Analyzer Function Remote
Control)
Description of basic operations, functions and remote functions of the
spectrum analyzer
Signal Analyzer Operation Manual (Phase Noise Measurement Function)
Signal Analyzer Operation Manual (Phase Noise Measurement Function
Remote Control) <This document>
Description of basic operations, functions and remote functions of the Phase
Noise
III
3
2
1
Table of Contents
About This Manual ........................................ I
Chapter 1 Overview.................................... 1-1
1.1 Overview ....................................................................... 1-2
1.2 Native Mode.................................................................. 1-4
1.3 Setting Numeric Program Data..................................... 1-7
Chapter 2 SCPI Device Message Details.. 2-1
2.1 Selecting Application..................................................... 2-4
2.2 Setting Basic Parameter ............................................... 2-11
2.3 Utility Function .............................................................. 2-26
2.4 Common Measurement Function ................................. 2-30
2.5 Log Plot Measurement.................................................. 2-35
Chapter 3 SCPI Status Register................ 3-1
3.1 Querying Measurement Status ..................................... 3-2
3.2 STATus:QUEStionable Register................................... 3-3
3.3 STATus:OPERation Register ....................................... 3-13
IV.
Chapter 1 Overview
1-1
1
Overview
This chapter provides an overview of the remote control of the Phase
Noise Measurement Function (hereinafter, referred to as “this
application”).
1.1
Overview ....................................................................... 1-2
1.1.1
Interface............................................................ 1-2
1.1.2
Controlled Application....................................... 1-3
1.2
Native Mode.................................................................. 1-4
1.3
Setting Numeric Program Data..................................... 1-7
Chapter 1 Overview
1-2
1.1 Overview
This application can be controlled from an external controller (PC) by
remote control commands using the MS2690/MS2691/MS2692A or
MS2830A Signal Analyzer (hereinafter, referred to as “this instrument”).
The remote control commands are defined by the SCPI format.
1.1.1 Interface
This instrument has GPIB, Ethernet, and USB interfaces for remote
control. Those interfaces cannot be used at the same time.
The interface is automatically determined when a command is received
at the start of communication. This instrument enters the remote state
after the interface has been determined. The front panel
lamp
lights during remote interface operation, and goes off during
local-interface operation.
Refer to the “MS2690A/MS2691A/MS2692A and MS2830A Signal
Analyzer Operation Manual (Mainframe Remote Control)” for more
details of the basic operations for remote control, such as interface
setting.
1.1 Overview
1-3
1
Overview
1.1.2 Controlled Application
Two types of the remote control commands can be used with this
instrument: commands that are commonly applied to this instrument
itself or all the applications (hereinafter, referred to as “Common
commands”), and the other commands unique to the applications.
Common commands can be executed at any time and do not depend on
the currently controlled application. However, commands unique to an
application can be executed at the controlled application. If it is executed
at another application, the command is not executed or an error occurs.
In this instrument, multiple applications can be activated at the same
time. Only one application resource can be executed at one time per piece
of hardware. This application measures an input signal by using the
resource of RF Input. Thus, this application cannot be executed at the
same time with another application using the same resource such as the
signal analyzer function. In order to execute a function unique to the
application by using the remote control, you need to select this
application while it has been running. Furthermore, this application can
be executed at the same time with another application that uses by itself
a resource not used by this application, such as the Vector Signal
Generator option.
Chapter 1 Overview
1-4
1.2 Native Mode
In this instrument, the syntax/format types of the remote control
commands are defined as “Language mode.” The language mode has two
modes: SCPI and Native.
(1) SCPI mode
The SCPI mode processes commands conforming to the
syntax/format defined in SCPI (ver1999.0). For programming, you
can use character strings in long/short form and can omit definition
character strings within square brackets ([ ]).
On the Configuration screen, the SCPI mode is automatically set
after transmitting command
SYST:LANG SCPI
.
(2) Native mode
The Native mode processes commands that are in this instrument's
own definition type. Unless otherwise specified, the character string
of a command header is fixed. If application commands are defined
only by the SCPI mode, character strings converted according to the
conversion rules are the commands in the Native mode. For
programming, you cannot use the grammar of the SCPI mode, such
as character strings in long/short form and cannot omit any
definition character strings within square brackets ([ ]).
On the Configuration screen, the Native mode is automatically set
after transmitting
SYST:LANG NAT
.
Note:
The
STATus:QUEStionable
and
STATus:OPERation
registers
cannot be used in the Native mode, even if the corresponding
commands are converted to Native-mode commands according to the
conversion rules.
1.2 Native Mode
1-5
SCPI Mode
Command definition
AAAAaa:BBBBbb[:CCCCcc]:D|E <n>
Programming examples:
AAAAaa:BBBBbb:CCCCcc:D 0
AAAA:BBBB:CCCC:D 0
AAAA:BBBB:D 0
AAAA:BBBB:CCCC:E 0
Native Mode (Initial setting)
Command definition
Original
VWXYZ1 <n>
Programming examples:
VWXYZ 0
Command Definition (Converte
d
from SCPI)
AAAA:BBBB:D <n>
Programming examples:
AAAA:BBBB:D 0
SYST:LANG SCPI
SYST:LANG NAT
AAAA:BBBB:E 0
1
Overview
Figure 1.2-1
SCPI and Native modes
This application is only defined as the SCPI mode commands. You need
to follow the conversion rules below in order to control this application by
using the Native mode.
Conversion Rules
[1] Move the numeric parameter in the program header of an SCPI
command before the argument. Delete a numeric parameter which
only has one value and can be deleted. Describe the argument if it
cannot be deleted.
[2] Use the first one if multiple nodes can be selected.
[3] Delete those layers which can be deleted.
[4] Alter all long forms into short forms.
[5] Delete the colon mark (“:”) at the head.
Example
To convert
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real>
into a Native mode command.
[1]
Delete a numeric parameter in the program header which only has
one value and can be deleted.
:DISPlay:WINDow
[1]
:TRACe:Y[:SCALe]:RLEVel <real>
:DISPlay:WINDow:TRACe:Y:[SCALe]:RLEVel <real>
Chapter 1 Overview
1-6
[2]
Delete those layers which can be deleted.
:DISPlay:WINDow:TRACe:Y
[:SCALe]
:RLEVel <real>
:DISPlay:WINDow:TRACe:Y:RLEVel <real>
[3]
Alter all long forms into short ones.
:DISPlay:WINDow:TRACe:Y:RLEVel
<real>
:DISP:WIND:TRAC:Y:RLEV <real>
[4]
Delete the colon mark (“
:
”) at the head.
:
DISP:WIND:TRAC:Y:RLEV <real>
DISP:WIND:TRAC:Y:RLEV <real>
1.3 Setting Numeric Program Data
1-7
1
Overview
1.3 Setting Numeric Program Data
The following character programs can be used for setting numeric
program data (numeric parameter).
(1)
DEFault
After
DEFault
is set to numeric program data, the target
parameter is set to the initial value.
(2)
MINimum
After
MINimum
is
set to numeric program data, the target
parameter is set to the minimum value.
(3)
MAXimum
After
MAXimum
is
set to numeric program data, the target
parameter is set to the maximum value.
In this application,
DEFault
,
MINimum
, and
MAXimum
can be used for the
following parameters.
<freq>
<real>
<rel_ampl>
<rel_power>
<integer>
<rel_freq>
Chapter 1 Overview
1-8.
Chapter 2 SCPI Device Message Details
2-1
2
SCPI Device Message Details
This chapter describes the detailed specifications of SCPI remote control
commands for executing the functions of this application. The device
messages are listed according to each function. Refer to the
“MS2690A/MS2691A/MS2692A and MS2830A Signal Analyzer Operation
Manual (Mainframe Remote Control)” for the detailed specifications of
the IEEE488.2 common device messages and application common device
messages.
2.1
Selecting Application ............................................................................................................. 2-4
2.1.1
Load Application........................................................................................................ 2-5
:SYSTem:APPLication:LOAD PNOISE..................................................................... 2-5
:SYSTem:APPLication:UNLoad PNOISE ................................................................. 2-5
2.1.2
Selecting Application................................................................................................. 2-6
:INSTrument[:SELect] PNOISE|SIGANA|SPECT|CONFIG...................................... 2-6
:INSTrument[:SELect]? ............................................................................................. 2-7
:INSTrument:SYSTem PNOISE,[ACTive]|INACtive|MINimum ................................. 2-8
:INSTrument:SYSTem? PNOISE.............................................................................. 2-9
2.1.3
Initialization ............................................................................................................... 2-10
:INSTrument:DEFault................................................................................................ 2-10
:SYSTem:PRESet ..................................................................................................... 2-10
2.2
Setting Basic Parameter........................................................................................................ 2-11
2.2.1
Carrier Frequency ..................................................................................................... 2-12
[:SENSe]:FREQuency:CENTer <freq> ..................................................................... 2-12
[:SENSe]:FREQuency:CENTer?............................................................................... 2-13
2.2.2
Start Offset ................................................................................................................ 2-14
[:SENSe]:FREQuency:OFFSet:STARt <rel_freq>.................................................... 2-14
[:SENSe]:FREQuency: OFFSet:STARt? .................................................................. 2-14
2.2.3
Stop Offset ................................................................................................................ 2-15
[:SENSe]:FREQuency:OFFSet:STOP <rel_freq>..................................................... 2-15
[:SENSe]:FREQuency: OFFSet:STOP? ................................................................... 2-15
2.2.4
Reference Level ........................................................................................................ 2-16
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel <real>.......................................... 2-16
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel?................................................... 2-17
2.2.5
RF Attenuator............................................................................................................ 2-18
[:SENSe]:POWer[:RF]:ATTenuation <rel_ampl>...................................................... 2-18
[:SENSe]:POWer[:RF]:ATTenuation? ....................................................................... 2-19
2.2.6
RF Attenuator Auto/Manual ...................................................................................... 2-20
[:SENSe]:POWer[:RF]:ATTenuation:AUTO OFF|ON|0|1 ......................................... 2-20
[:SENSe]:POWer[:RF]:ATTenuation:AUTO?............................................................ 2-20
2.2.7
Level Offset ............................................................................................................... 2-21
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet <rel_power> .................. 2-21
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet? ..................................... 2-21
2.2.8
Level Offset State...................................................................................................... 2-22
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet:STATe OFF|ON|0|1 ...... 2-22
Chapter 2 SCPI Device Message Details
2-2
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RLEVel:OFFSet:STATe?......................... 2-22
2.2.9
Pre Amp .................................................................................................................... 2-23
[:SENSe]:POWer[:RF]:GAIN[:STATe] OFF|ON|0|1 .................................................. 2-23
[:SENSe]:POWer[:RF]:GAIN[:STATe]?..................................................................... 2-23
2.2.10
Log Scale Line .......................................................................................................... 2-24
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LINE <line> .............................................. 2-24
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:LINE? ....................................................... 2-24
2.2.11
Reference Value ....................................................................................................... 2-25
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]:RVALue <real> ........................................ 2-25
:DISPlay:WINDow[1]:TRACe:Y[:SCALe]: RVALue?................................................. 2-25
2.3
Utility Function....................................................................................................................... 2-26
2.3.1
Erase Warm Up Message......................................................................................... 2-27
:DISPlay:ANNotation:WUP:ERASe .......................................................................... 2-27
2.3.2
Display Title............................................................................................................... 2-27
:DISPlay:ANNotation:TITLe[:STATe] OFF|ON|0|1 ................................................... 2-27
:DISPlay:ANNotation:TITLe[:STATe]? ...................................................................... 2-28
2.3.3
Title Entry .................................................................................................................. 2-28
:DISPlay:ANNotation:TITLe:DATA <string>.............................................................. 2-28
:DISPlay:ANNotation:TITLe:DATA?.......................................................................... 2-29
2.4
Common Measurement Function .......................................................................................... 2-30
2.4.1
Measurement and Control ........................................................................................ 2-31
:INITiate:CONTinuous OFF|ON|0|1 .......................................................................... 2-31
:INITiate:CONTinuous?............................................................................................. 2-31
:INITiate:MODE:CONTinuous................................................................................... 2-32
:INITiate:MODE:SINGle ............................................................................................ 2-32
:INITiate[:IMMediate]................................................................................................. 2-33
:CONFigure? ............................................................................................................. 2-33
2.4.2
Save Result Data ...................................................................................................... 2-34
:MMEMory:STORe:TRACe [<filename>[,<device>]] ................................................ 2-34
2.5
Log Plot Measurement .......................................................................................................... 2-35
2.5.1
Measure .................................................................................................................... 2-37
:CONFigure:LPLot..................................................................................................... 2-37
:INITiate:LPLot .......................................................................................................... 2-37
:FETCh:LPLot[n]? ..................................................................................................... 2-38
:READ:LPLot[n]?....................................................................................................... 2-39
:MEASure:LPLot[n]? ................................................................................................. 2-39
2.5.2
Average Count .......................................................................................................... 2-40
[:SENSe]:LPLot:AVERage:COUNt <integer> ........................................................... 2-40
[:SENSe]:LPLot:AVERage:COUNt? ......................................................................... 2-40
2.5.3
Marker Mode ............................................................................................................. 2-41
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:MODE
NORMal|INTEgralnoise|RMSNoise|JITTer|RESIdualfm|OFF .................................. 2-41
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:MODE? ................................................ 2-42
2.5.4
Analysis Width........................................................................................................... 2-43
2.1 Selecting Application
2-3
2
SCPI Device Message Details
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:WIDTh:STARt <rel_freq>..................... 2-43
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:WIDTh:STARt? .................................... 2-44
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:WIDTh:STOP <rel_freq> ..................... 2-45
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:WIDTh:STOP?..................................... 2-46
2.5.5
Marker Frequency Offset .......................................................................................... 2-47
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:X <rel_freq> ......................................... 2-47
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:X?......................................................... 2-48
2.5.6
Marker Result............................................................................................................ 2-49
:CALCulate:LPLot:MARKer[1]|2|3|4|5|6|7|8:Y?......................................................... 2-49
Chapter 2 SCPI Device Message Details
2-4
2.1 Selecting Application
Table 2.1-1 lists the device messages for setup operations such as
starting/selecting/initializing an application.
Table 2.1-1 Selecting Application
Parameter Device Message
Load Application
:SYSTem:APPLication:LOAD PNOISE
Unload Application
:SYSTem:APPLication:UNLoad PNOISE
:INSTrument[:SELect] PNOISE
Application Switch
:INSTrument[:SELect]?
:INSTrument:SYSTem PNOISE,[ACTive]|INACtive|MINimum
Application Status
:INSTrument:SYSTem? PNOISE
:INSTrument:DEFault
Initialization
:SYSTem:PRESet
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84