MODULAR ANALYTICS Core

Roche MODULAR ANALYTICS Core User manual

  • Hello! I am an AI chatbot trained to assist you with the Roche MODULAR ANALYTICS Core User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Replacement Page
MODULAR ANALYTICS EVO
SWA
Operator’s Manual
Software Version 08-08
MODULAR ANALYTICS EVO SWA
Roche Diagnostics Replacement Page
2 Operator’s Manual · Version 4.1.3
Document information
Revision history
Manual version Software version Revision date Changes
1.0 07-01 30.05.2002
1.1 07-01 30.09.2002 Data alarms
2.0 07-04 05.05.2004 Structural and content
revision
3.0 08-01 01.03.2006 Supplementation of new
software features
3.1 08-02 10.03.2008 Supplementation of
corrections
4.0 08-07 09.2013 Supplementation of
corrections
4.1 08-07 09.2014 Maintenance E module
Clean
ProCell M/CleanCell M
aspiration tube filters
deleted.
Caution message to
Replacing ProCell M and
CleanCell M procedure
added.
4.1.1 08-07 03.2015 Replacement pages with
minor additions:
Equation for Sans data
alarm revised.
4.1.2 08-07 11.2016 Replacement pages with
minor additions:
Explanation of product
symbols inclusive RoHS
and Cont.
Safety message for
incomplete display of test
results in case the column
is too narrow
Fluorinated greenhouse
gas
4.1.3 08-08 10.2017 Replacement page for the
front cover, showing
updated software version.
Publication version 4.1 of
the Operator’s Manual and
Online Help also applies to
software version 08-08.
Replacement page for
chapter System
description > Core unit >
Sample racks: Safety
message level changed to
WARNING.
Roche Diagnostics Replacement Page
Operator’s Manual · Version 4.1.3 A-73
MODULAR ANALYTICS EVO SWA 4 Core unit
Specifications
Specifications for maximum
sample volumes
The maximum filling level is 10 mm below the upper rim of the sample container.
This applies to all sample tubes, standard cups, and micro cups. For sample cups the
maximum filling level results in the following maximum sample volumes:
Do not exceed maximum sample volumes. Otherwise sample may get spilled on rack
conveyor and cause instrument problems.
Certified standard cups Only the following original standard cups are certified by Roche Diagnostics:
o Hitachi standard cup
o cobas sample cup (catalog no. 10394246001)
Although non-Roche Diagnostics sample cups are visually similar, they may have a
different inner geometry and may have been produced using non-specified materials.
The liquid level detection and the level tracking are influenced by the inner geometry
of the cup. Thus, air may be pipetted when non-Roche Diagnostics cups are used.
This could not be detected under certain circumstances.
Standard cup
1.5 mL
Micro cup
500 L
CAUTION
Incorrect results when using non-certified cups
Sample cups from other vendors may cause inaccurate sample pipetting which may lead
to incorrect results.
o Use only cups certified by Roche Diagnostics.
Roche Diagnostics Replacement Page
A-74 Operator’s Manual · Version 4.1.3
4 Core unit MODULAR ANALYTICS EVO SWA
Host interface
Host interface
The instrument can be bidirectionally interfaced with a Host computer. Details
concerning the interfacing of the system are available in the MODULAR
ANALYTICS EVO SWA System Host Interface Document.
An interface port for a Host computer is available for use, if desired. This port is
located at the rear of the loader.
Sample racks
Samples are placed in 5-position sample racks, the racks are placed on a supply tray,
and the tray is placed on the loader.
Sample cups or tubes are placed in sample racks. Each sample rack holds a maximum
of 5 samples. Each tube slot is separated by a rack identification strip that is read by
the system.
Figure A-29 Sample racks
WARNING
Incorrect results when sample tubes are not aligned vertically
Incorrect placement of sample tubes in racks may cause incorrect sample pipetting, which
may lead to false low results, especially when testing immunoassays.
o Always use the Roche cup adapter for sample tubes with an outer diameter of 13 mm
or less.
o Only use sample tubes specified for the use with the MODULAR ANALYTICS EVO
SWA instrument.
/