Seitron Chemist 600/600 Be Green Quick start guide

Type
Quick start guide

Seitron Chemist 600/600 Be Green is a multi-functional combustion analyzer that can be used to measure various parameters related to combustion processes, including flue gas analysis, calculating stack heat loss and efficiency, CO environment measurement, gas pipes tightness test, store Smoke value, calculating mean value, measuring differential pressure, Draught measurement, and pressure measurement in the gas supply pipe. The device is compliant with EN 50379-1, EN 50379-2 UNI 10389-1 standards and features a user-friendly interface with context keys for easy navigation.

Seitron Chemist 600/600 Be Green is a multi-functional combustion analyzer that can be used to measure various parameters related to combustion processes, including flue gas analysis, calculating stack heat loss and efficiency, CO environment measurement, gas pipes tightness test, store Smoke value, calculating mean value, measuring differential pressure, Draught measurement, and pressure measurement in the gas supply pipe. The device is compliant with EN 50379-1, EN 50379-2 UNI 10389-1 standards and features a user-friendly interface with context keys for easy navigation.

OK
Compliant with: EN 50379-1, EN 50379-2 UNI 10389-1
Features
Flue gas analysis
Calculating of stack heat loss and efficiency
CO environment measurement
Gas pipes tightness test according to standards UNI 7129 and UNI 11137
Store Smoke value, calculating mean value
Measuring differential pressure
Draught measurement
Pressure measurement in the gas supply pipe.
Date: 15/01/14
Time: 10.10
Fuel: Natural gas
Altitude: 0 m
R.H. air: 50 %
O2 4.2 ٪
CO2 9.3 ٪
λ,n 1.25
T flue 190.2 °C
T air 15.4 °C
ΔT 174.8 °C
QS 8.6 ٪
ηs 91.4 ٪
ET 4.9 ٪
ηt 91.4 ٪
CO 148 ppm
NO 40 ppm
NOX/NO: 1.03
NOX 41 ppm
CO amb 0 ppm
Draft: 0.05 hPa
T out: 20 °C
Smoke: 3 1 2
Aver n. : 2
15/01/14
10:00
ηs
%
91.4
Qs
%
8.6
ΔT
°C
74.7
T air
°C
15.4
T flue
°C
190.1
λ,n
1.25
CO
2
%
9.3
O
2
%
4.2
Memory
Average analysis
15/01/14
10:00
Print
Report
Model partial
Analysis average
OK
Memory 1
15/01/14
10:00
Print
Report
Model partial
Analysis In average
F1
Memory 1
In case the automatic print mode is set then the average analysis printing start automatically.
Conversely, after the sampling of the third analysis, the average analysis is shown, so that it can be sent to the
printer as explained in the following.
WARNING! Before proceeding with the combustion analysis remember to set the proper fuel.
OK
Memory
Save
15/01/14
10:00
OK
Memory
1
Mode
UNI 10389
Samples
3
Interval
s
30
Automatically
saves the first
sample when the
set interval
expires, and so on
until the third
sample.
15/01/14
10:00
ηs
%
91.4
Qs
%
8.6
ΔT
°C
74.7
T air
°C
15.4
T flue
°C
190.1
λ,n
1.25
CO
2
%
9.3
O
2
%
4.2
Combustion analysis
UNI 10389
15/01/14
10:00
ηs
%
91.4
Qs
%
8.6
ΔT
°C
74.7
T air
°C
15.4
T flue
°C
190.1
λ,n
1.25
CO
2
%
9.3
O
2
%
4.2
Combustion analysis
UNI 10389
o
1
120
KEYBOARD FUNCTIONS
KEYS FUNCTION
Activate the context keys shown on
the display
Enters the Configuration menu
Starts the flue gas analysis
Enters the measurements menu
Enters the Memory menu
Enters the Print menu
Turns the instrument on/off
Quits the current screen
Selects and/or modifies
Confirms data
CONTEXT KEYS
CONTEXT KEY FUNCTION
Makes the pressure zero
Updates the measure
Saves the current measurement or
analysis in the memory location selected
from the ‘Select Memory’ menu
Starts the report ticket
Enters the Print menu
Stops the flue gas analysis
Enters the modify mode for the selected
parameter
Confirms the settings
Cancels the change/operation and
returns to the previous screen
Zooms the screen
Shows the details for the selected
parameter
‘Find’ function; starts a quick search for
the analysis to recall
OK
O
Esc
o
K600000000SE 034136 190220
OK
Configuration
Fuel
15/01/14
10:00
Biogas
Propane-Air
Fuel oil
Diesel oil
Butane
L.P.G.
Propane
Natural gas
OK
15/01/14
10:00
Analysis
Operator
Information
Language
Instrument
Alarms
Diagnostic
Configuration
Restore
WARNING
F1: stop
Printing.
Please wait...
The magnets in the back of the instrument can damage credit
cards, hard driver, mechanical watches, pacemakers,
defibrillators and other devices proven sensitive to magnetic
fields.
It is recommended to keep the instrument at a distance of at
least 25cm away from these devices.
15/01/14
10:00
Fuel
O
2
reference
Measure units
Measures list
Condensation
NO
x
/NO ratio
Autozero
Configuration
Analysis
Air temp.
Via del Commercio, 9/11 36065 - MUSSOLENTE (VI) - ITALY
0424 567842 - 0424 567849
[email protected] - www.seitron.com
Seitron respects the nature and the environment, therefore provides this quick user
guide to avoid any unwanted waste of paper.
You can download the PC software “CHEMIST SMART ANALYSIS” and the complete
user and maintenance manual from our website www.seitron.com.
Respect your environment: think before printing the full manual on paper.
15/01/14
10:00
Measurements
Ambient CO
CO
ppm
0
CO max
ppm
0
15/01/14
10:00
Measurements
Ambient CO
CO
ppm
0
CO max
ppm
0
Connect the gas probe to the instrument and perform
the measurement.
Wait 5 minutes.
15/01/14
10:00
Measurements
Draft
P diff
hPa
0.000
T outdoor
°C
20
O
15/01/14
10:00
Measurements
Draft
P diff
hPa
0.001
T outdoor
°C
20
O
Once the
pressure zeroing
is completed
insert the probe in
the chimney and
measure the
draught.
15/01/14
10:00
ηs
%
91.4
Qs
%
8.6
ΔT
°C
74.7
T air
°C
15.4
T flue
°C
190.1
λ,n
1.25
CO
2
%
9.3
O
2
%
4.2
Combustion analysis
UNI 10389
AAC DP02
AA SF--A
AAC TA03A
AAC KP01
AA AL05
AA SA08
KEEP
PRESSED
FOR 2
SECONDS
UNTIL IT
BEEPS
WARNING!
During autozero DO NOT
insert the gas probe in the
chimney.
Memory
Select
15/01/14
10:00
Customer
Seitron
Memory 1
Address
Via M. Prosdocimo
n° 30
Phone
0424 567842
Bassano del Grappa
Boiler
xxxx
Date 15/01/14
In "Select" can be
inserted all the
customer data.
Memory
Data logger
15/01/14
10:00
Samples 3
Mode UNI 10389
Interval
ppm
120
Print auto
Memory auto
In "Data logger" the
user can define the
analysis, memory
selection and printing
modes.
Suggested settings:
Mode: UNI 10389
Print: auto
Memory: auto (assigns
a new memory at each
instrument turn-on).
15/01/14
10:00
Save
Select
Delete
Average
Data logger
Usage %
Memory
15/01/14
10:00
Measurements
Draft
P diff
hPa
0.001
T outdoor
°C
20
O
Before zeroing the
pressure it is
mandatory to
extract the probe
from the chimney.
Save the
measurement.
Zeroing
instrument
O
Save the
measurement
OK OK
Data logger
Select
Draft
Ambient CO
It is mandatory to perform the
instrument autozero in clean air.
  • Page 1 1
  • Page 2 2

Seitron Chemist 600/600 Be Green Quick start guide

Type
Quick start guide

Seitron Chemist 600/600 Be Green is a multi-functional combustion analyzer that can be used to measure various parameters related to combustion processes, including flue gas analysis, calculating stack heat loss and efficiency, CO environment measurement, gas pipes tightness test, store Smoke value, calculating mean value, measuring differential pressure, Draught measurement, and pressure measurement in the gas supply pipe. The device is compliant with EN 50379-1, EN 50379-2 UNI 10389-1 standards and features a user-friendly interface with context keys for easy navigation.