Thermo Fisher Scientific Array Plate Scanner Site Preparation Owner's manual

Type
Owner's manual
P/N 08-0257 Rev.B
February 2007
Site Preparation Guide—HT Array Plate
Scanner
For Research Use Only. Not for Use in Diagnostic Procedures.
Trademarks
Affymetrix®, , GeneChip®, HuSNP®, GenFlex®, Flying Objective™,
CustomExpress®, CustomSeq®, NetAffx™, Tools To Take You As Far As Your Vision®,
The Way Ahead™, Powered by Affymetrix™, and GeneChip-compatible™, are
trademarks of Affymetrix, Inc.
All other trademarks are the property of their respective owners.
Limited License
Limited License. Subject to the Affymetrix terms and conditions that govern your use of
Affymetrix products, Affymetrix grants you a non-exclusive, non-transferable, non-sublicensable
license to use this Affymetrix product only in accordance with the manual and written
instructions provided by Affymetrix. You understand and agree that except as expressly set forth
in the Affymetrix terms and conditions, that no right or license to any patent or other intellectual
property owned or licensable by Affymetrix is conveyed or implied by this Affymetrix product. In
particular, no right or license is conveyed or implied to use this Affymetrix product in
combination with a product not provided, licensed or specifically recommended by Affymetrix
for such use.
Copyright ©2006 Affymetrix Inc. All rights reserved.
Compliant with directive 2002/96/EC (WEEE)
Contents
HT Array Plate Scanner Site Preparation Guide Overview . . .3
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Preparation Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Site Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Hazards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Transportation Details and Requirements . . . . . . . . . . . . . . . . . . . . . . . 6
Space . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Electrical Outlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
HT Array Plate Scanner Specifications . . . . . . . . . . . . . . . . . . . . . . . . . 7
Networking Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Server Hardware and Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Pre-Installation Requirements and Fax . . . . . . . . . . . . . . . . .11
HT Array Plate Scanner Confirmation of Site Preparation FAX . . . . . . 11
Facilities . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Space Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
Electrical Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
Networking Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
SQL Server License . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
GCOS Sever Pre-Installation Checklist . . . . . . . . . . . . . . . . . .17
Location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Information Needed Prior to Configuration . . . . . . . . . . . . . . . . . . . . . 17
SQL Server Configuration (if using SQL Server) . . . . . . . . . . . . . . . . . 19
Oracle® Configuration (if using Oracle) . . . . . . . . . . . . . . . . . . . . . . . . 19
Information Needed from the Oracle® DBA . . . . . . . . . . . . . . . . . . . . . 20
Client Stations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Staff Requirement During Installation . . . . . . . . . . . . . . . . . . . . . . . . . 21
Training . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Other Comments/Questions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
ii GeneChip® HT Array Plate Scanner Site Preparation Guide
This page intentionally left blank.
HT Array Plate Scanner Site Preparation Guide Overview
Affymetrix, Inc. developed this document to help you through the process of preparing your
laboratory, to ensure a prompt and successful installation of the GeneChip® HT Array Plate
Scanner.
In preparing your site for the installation of the HT Array Plate Scanner, please note the
following points:
Suitable preparation of your laboratory is essential to a successful installation.
You may wish to work with your facilities personnel to ensure that your laboratory
meets the minimum electrical and other environmental requirements as outlined
in this document.
Once your site location meets all of the requirements outlined, your laboratory is suitable
for installation of your new HT Array Plate Scanner. At that time:
1. Complete the site verification checklist in Pre-Installation Requirements
and Fax on page 11 and fax the completed checklist to Affymetrix
Instrument Support at (408) 481-9670.
2. Upon receipt of this written confirmation, Affymetrix will contact you
by telephone to schedule the installation of your HT Array Plate
Scanner.
COMPONENTS
The GeneChip HT Array Plate Scanner is one of several components of the HT System (See
Table 1.1 on page 4). This site prep guide lists only the requirements for the scanner and
does not include the other components. Please refer to the GeneChip® Array Station Site
Prep Guide (P/N 702020) for additional information.
Please do not unpack the HT Array Plate Scanner shipping container
prior to the arrival of an Affymetrix Representative. The following
information is provided for reference purposes only.
4GeneChip® HT Array Plate Scanner Site Preparation Guide
PREPARATION CHECKLIST
Site preparation is essential for a successful installation of your instruments. It is important,
therefore, to ensure that all site requirements are met prior to the installation date. Please
use this check list for preparing your laboratory. Detailed information is provided in Pre-
Installation Requirements and Fax on page 11.
Appropriate power connections for the HT Array Plate Scanner are available.
Network requirements are met.
Adequate space for the dimensions of the HT Array Plate Scanner and benchtop
support to sustain the total weight are allocated with available service access.
Once the HT Array Plate Scanner is on-site, ensure that shipping skid and
associated boxes are close to the laboratory installation location. This equipment
must be unpacked only by an Affymetrix Representative.
Two people are available with proper access to the loading dock and pallet jacks to
move the HT Array Plate Scanner and associated boxes to a location close to the lab
for unpacking by an Affymetrix Representative.
Table 1.1
HT System components depending on use
Target Prep HT Array Plate
Processing
1 Array Station ✔✔
2 Spectrophotometer (user-supplied) ✔✔
3 HT Array Plate Scanner -
4 GCOS server hardware (user supplied) -
5 HT Software Suite (includes GCOS Server
software) -
| Site Requirements 5
SITE REQUIREMENTS
HAZARDS
Table 1.2 summarizes possible hazards.
A qualified Affymetrix Representative must uncrate and install the
HT Array Plate Scanner. Facilities personnel should be assigned to
assist the Affymetrix Representative.
Table 1.2
Hazards
Hazard Present? Description
Chemical No
Control No Control software
Electrical Yes 100-240V power
Ergonomic Yes User interface
Gas No
Mechanical Yes, scanning stage, scanner weight (heavy instrument)
Laser Yes, if you remove the scanner enclosure.
Noise No
Pneumatic No
Radiation No
Temperature No
Ultrasonic No
Vibration No
Heat No Total power generated is less than 100 watts.
E-Fields No
H-Fields No
6GeneChip® HT Array Plate Scanner Site Preparation Guide
SPACE REQUIREMENTS
At the installation site, verify that there is sufficient space and air circulation to
accommodate the GeneChip HT Array Plate Scanner.
Refer to the following diagram for space dimensions.
TRANSPORTATION DETAILS AND REQUIREMENTS
1. Loading dock or a delivery truck with a lift gate
2. Transportation to final location: elevator with 1200 lb. (545 kg)
capacity.
3. Doorways: >40” (1.016 m) wide
4. Cushioned rolling cart: 400 lb. (181.4 kg) capacity (Uncrated
instrument)
5. Lift truck: to remove instrument from crate and place on cart: 600 lb.
(272.1 kg) capacity
HT Array
Plate
Scanner
Array Plate Access
Clearance
29"
(76 cm)
16"
(41 cm)
12"
(30 cm)
3"
(7.6 cm)
Clearance 2"
(5 cm)
| Space 7
SPACE
1. The instrument footprint is 29” (76 cm) depth x 26” (66 cm) height x
16” (41 cm) width.
2. Additional clearance: 12” (30 cm) on left side, 3” (7.5 cm) right side,
and 2” (5 cm) in the rear.
3. Computer workstation: Bench space for monitors: 36” (0.91 m) wide
by 23” (0.58 m) deep; CPU tower needs to be within 80” (2.03 m) of
the right side of the instrument
ELECTRICAL OUTLETS
1. Two separate, dedicated grounded outlets within 4 ft. (1.22 m) of the
instrument (one for the scanner and one for the workstation).
2. Outlet requirements: 100-120 VAC, 50/60 Hz, 20 Amps or more, or
220-240 VAC, 50/60 Hz, 10 Amps or more.
HT ARRAY PLATE SCANNER SPECIFICATIONS
Table 1.3 lists the important instrument specifications.
Table 1.3
The Specifications of the GeneChip® HT Array Plate Scanner
Item Parameter Value
Weight Shipping approx 216 pounds (98
Kg)
Free-standing approx 145 lbs (66 Kg)
Dimensions Width ~16 in.
Depth ~29in.
Height ~26 in.
Power
Voltage
Current
Line Frequency.
100 - 240 V ~
2.8 - 1.2 A
50 - 60 Hz
Working Environment Temperature 59°F-85°F (15°C-30°C)
Humidity 10-90% Non-condensing
Clearance
2 in. (5 cm) in rear
12 in. (30.5 cm) on left
side
8GeneChip® HT Array Plate Scanner Site Preparation Guide
Pollution Degree 2
Installation Category II
Altitude <2000m
Electrical Supply Provide voltage, frequency or power rating per unit
label. Circuit breaker.
Main Supply Voltage
Fluctuations
Are not to exceed ±10% of the nominal supply
voltage
Table 1.3 (Continued)
The Specifications of the GeneChip® HT Array Plate Scanner
Item Parameter Value
Networking Requirements
The GeneChip HT Array Plate Scanner will be installed and tested in local mode. If you
later want to connect the workstation to your network, you may need to arrange for your
network services department to install anti-virus software and to prepare for attachment in
other ways. If your network policies require that the name of the workstation be changed,
you will need to inform Affymetrix so that the workstation software can be reconfigured
accordingly.
SERVER HARDWARE AND SOFTWARE
The GCOS Server software is provided with the HT Array Plate Scanner while the server
hardware component is user-supplied. Due to the large amount of data generated, a GCOS
server (or a server that meets the minimum specifications) is recommended. The
configuration of the server is dependent on the amount of data generated by the HT System.
As such, additional memory, processor speed, and disk drive capacity should be considered
depending on the number of plates processed.
GCOS Server Hardware minimum configuration:
Dell PowerEdge 6800
Tower Configuration
Quad 3.0 GHz Xeon Processors
•8MB Cache
•4GB RAM
4x300 GB Hard Drives
RAID Configuration
• Microsoft® Windows® 2000
•3.5" Floppy Drive
24X IDE CD-RW/DVDROM
17" Monitor
10 GeneChip® HT Array Plate Scanner Site Preparation Guide
For a more detailed recommendation of hardware specifications tailored to your institution,
please refer to on page 16.
Affymetrix does NOT supply licenses for Microsoft® Windows® 2000
Server, Microsoft® Windows® 2000 Professional or your chosen
RDBMS software (SQL/ORACLE®).
If you plan to purchase a server, additional requirements and site
information are required prior to installation. Please refer to on
page 16 for detailed information.
The USB and Firewire (IEEE 1394 High Speed Serial Bus) cables are
not locking connectors. Insure that the USB and Firewire cables are
plugged securely and in the correct orientation in both the scanner
and the workstation. Avoid snagging or pulling these cables during
the scanner operation.
Pre-Installation Requirements and Fax
To ensure that the GeneChip® HT Array Plate Scanner is installed successfully, please fill
out the information requested for shipping and receiving, space requirements and the site
preparation checklist and fax this information (on page 11 through page 15) to Affymetrix
Instrument Support (408) 481-9670.
HT ARRAY PLATE SCANNER CONFIRMATION OF SITE PREPARATION FAX
Order Reference Number ____________
Contact ___________________(printed)
_________________________(signed)
Institution/Company Name ______________________________________________
Shipping Address
_____________________________________________________
_____________________________________________________
_____________________________________________________
State ___________ZIP code ___________ Country_______
Phone _______________________Fax ____________________
Contact Email_________________________________________
If you plan to purchase a server, please provide a list of contacts for the server installation.
Refer to on page 16 for additional server requirements.
Fax page 1 of 5
12 GeneChip® HT Array Plate Scanner Site Preparation Guide
FACILITIES
Receiving Facilities
Check with your Receiving Department about the shipping requirements we need to
consider when shipping your unit. Please provide Affymetrix with detailed shipping and
receiving requirements deemed necessary by your Receiving Department.
Do you have a loading dock?
Do you require a delivery truck with a lift gate?
I, ____________________________________, have verified the statement checked
above.
Transporting the Instrument
Determine the path that the GeneChip HT Array Plate Scanner system will follow from the
loading dock to your installation site. Note that the scanner is shipped as two units: a large
crate containing the main instrument and a pallet with several boxes. If the instrument will
be transported in the crate, all doorways and passages along that path must be at least 40”
(1.016 m) inches wide. Please verify passageway dimensions with a measuring device.
Should it be necessary to transport the crated scanner on an elevator, make sure that the
elevator can carry at least 1200 lb. (545 kg), the combined weight of the crated instrument
(about 360 lb., or 164 kg) and the transport mechanism and personnel. Consult with on-site
personnel.
I, ____________________________________, have verified our facility meets the
transport requirements above.
Uncrating the GeneChip® HT Array Plate Scanner
Should it not be possible to move the crated instrument through the facility, the instrument
may be transferred to a cushioned rolling cart. This cart should be able to accommodate the
full weight of the instrument (145 lbs., 66 kg, when uncrated), and it should be sufficiently
stable to prevent tipping.
Due to the nature of the instrument, a manual, winch-operated lift truck is recommended
for removing the instrument from the crate and transporting it to the transfer cart or to its
place of installation. The lift truck must be rated in excess of 600 lbs. (272.15 kg). A lift
truck is not recommended for transport of the instrument for significant distances. The lack
of cushioning poses a risk to the instrument, and the high center of gravity of the lift truck
and instrument poses a risk of tipping.
SPACE REQUIREMENTS
At the installation site, verify that there is sufficient space and air circulation to
accommodate the GeneChip HT Array Plate Scanner. The instrument footprint is 29” (76
Fax page 2 of 5
| Space Requirements 13
cm) depth x 26” (66 cm) height x 16” (41 cm) width. In addition to the footprint, when
looking at the instrument from the front, allow for 2” (5 cm) in the back and on the right
side for proper airflow. During installation, access to the front, back, and right-hand side of
the instrument will be required.
The table or bench supporting the instrument must be rated for 500 pounds (227 kg) and
support must be sufficient to prevent tipping.
In addition to space for the instrument, there must be room for the workstation. The
workstation includes a CPU tower and two LCD monitors as well as a keyboard and mouse.
Bench space at least 36” (0.91 m) wide by 23” (0.58 m) deep is recommended. The CPU
tower will need to be within 80” (2.03 m) of the right side of the instrument.
I, ____________________________________, have verified our facility meets the space
requirements above.
HT Array
Plate
Scanner
Array Plate Access
Clearance
29"
(76 cm)
16"
(41 cm)
12"
(30 cm)
3"
(7.6 cm)
Clearance 2"
(5 cm)
Fax page 3 of 5
14 GeneChip® HT Array Plate Scanner Site Preparation Guide
ELECTRICAL REQUIREMENTS
Verify that there are at least two separate, dedicated, properly grounded electrical outlets
that are within about 4 feet (1.22 m) of the GeneChip HT Array Plate Scanner location. The
scanner outlets should be 100-120 VAC, 50/60 Hz, 20 Amps or more, or 220-240 VAC,
50/60 Hz, 10 Amps or more. One outlet will be used for the scanner, and the other for the
workstation and its peripherals.
I, ____________________________________, have verified our facility meets the
electrical requirements above.
NETWORKING REQUIREMENTS
The GeneChip HT Array Plate Scanner will be installed and tested in local mode. If you
later want to connect the workstation to your network, you may need to arrange for your
network services department to install anti-virus software and to prepare for attachment in
other ways. If your network policies require that the name of the workstation be changed,
you will need to inform Affymetrix so that the workstation software can be reconfigured
accordingly.
The workstation will ultimately be attached to a network.
The workstation will not be attached to a network.
The workstation name will need to be changed during attachment.
The workstation name will not need to be changed during attachment.
I, ____________________________________, have verified the statement(s) checked
above.
Fax page 4 of 5
| SQL Server License 15
SQL SERVER LICENSE
The GeneChip HT Array Plate Scanner workstation comes with a license to Microsoft® SQL
Server 2000 Standard Edition that can be transferred to an end user in the United States. All
other end users will need to purchase a license on their own.
I, ____________________________________, certify that I am an end user in the
United States, or
I, ____________________________________, certify that I will purchase a license to
Microsoft SQL Server.
Special Installation Requirements
If there are any special requirements on your part that we need to know, please list them
here:
________________________________________________________
________________________________________________________
________________________________________________________
________________________________________________________
________________________________________________________
________________________________________________________
________________________________________________________
Fax page 5 of 5
16 GeneChip® HT Array Plate Scanner Site Preparation Guide
This page intentionally left blank.
GCOS Sever Pre-Installation Checklist
Please provide the following information.
LOCATION
Please provide your
Site Name:
City/ State:
• Date:
INFORMATION NEEDED PRIOR TO CONFIGURATION
Contacts
Name of contacts who will be involved with the GCOS System:
1. IT personnel:
a. Name
b. Phone number:
c. Email address:
2. DBA personnel:
a. Name
b. Phone number:
c. Email address:
3. Bioinfomatics Specialist personnel:
a. Name:
b. Phone number:
c. Email address:
4. Scientist or Primary End user personnel:
18 GeneChip® HT Array Plate Scanner Site Preparation Guide
a. Name:
b. Phone number:
c. Email Address:
Installation Details
1. Are you purchasing a pre-configured server from Affymetrix?
Yes/No
2. If Yes, are you purchasing the Dell or Compaq? Dell/Compaq
3. Is this a one server or two server configuration? One/Two
4. Which of the following databases will be used:
MS SQL Server 2000? Yes /No
Oracle® 8.1.7 for Microsoft® Windows® 2000? Yes/No
5. The site license information: License Key #__________
Non-Configured Server Information
1. Configured for the network/domain
2. Microsoft® Windows® Network Domain name that server resides on:
___________________________
3. Operating System running on the Primary Domain Controller:
______________________
4. Name of local server: ______________________
5. Name of remote AADM server (if applicable): _______________
If purchasing a pre-configured server, please skip to Information
Needed from the Oracle® DBA on page 20.
It is important to make sure that an underscore is not used in the
computer name of the GCOS Server. Otherwise, the GCOS Server
1.0 software will not function properly.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19