ZOLL Pedi-Padz User manual

Type
User manual

ZOLL Pedi-Padz are designed for use on pediatric patients less than 8 years of age or weighing less than 55lbs (25kg). They are compatible with ZOLL R Series and X Series devices and can be used for defibrillation, noninvasive pacing, cardioversion, and ECG monitoring. The electrodes feature a solid gel design that conforms to the patient's skin for optimal conductivity and patient comfort. They are also preconnected to a ZOLL Multi-Function Cable for quick and easy setup.

ZOLL Pedi-Padz are designed for use on pediatric patients less than 8 years of age or weighing less than 55lbs (25kg). They are compatible with ZOLL R Series and X Series devices and can be used for defibrillation, noninvasive pacing, cardioversion, and ECG monitoring. The electrodes feature a solid gel design that conforms to the patient's skin for optimal conductivity and patient comfort. They are also preconnected to a ZOLL Multi-Function Cable for quick and easy setup.

6. These electrodes are recommended for
patients less than 8 years of age or
weighing less than 55lbs (25kg). For
patients over 25kg, adult electrodes are
recommended. Do not delay therapy to
determine actual weight of patient during
emergency.
7. Avoid electrode placement near the
generator of an internal pacemaker, other
electrodes or metal parts in contact with
the patient.
8. Some current generated by electrosurgical
units (ESU) may concentrate in the
conductive gel of pacing / defibrillation
electrodes, especially if an ESU grounding
pad other than that recommended by the
ESU manufacturer is used. Consult the
ESU operator's manual for further details.
9. To avoid electrical shock, do not touch the
pads, patient, or bed when defibrillating.
10.Do not discharge standard paddles on or
through electrodes or place separate ECG
leads under pads. Doing so could lead to
arcing and/or skin burning.
11.Device disposal shall follow hospital
protocol.
12. Use only with ZOLL pacemaker/
defibrillator products.
1. Do not use if gel is dry. Dried out gel can lead
to skin burning. Do not open pouch until ready
to use. Do not use electrodes past the
expiration date printed on the pouch label.
2. Do not fold electrodes or packaging. Any fold
in or other damage to the conductive element
could lead to the possibility of arcing and/or
skin burns.
3. After patient movement due to muscle
contraction or patient repositioning, press pads
to skin to ensure good coupling between pads
and skin.
4. Continuous pacing of patients could cause
burns. Periodic inspection of the underlying
skin, at least every 30 minutes during pacing,
is strongly advised.
5. Electrodes should be replaced after 24 hours of
use or 1 hour of pacing.
Instructions for Use
OPERATING TEMPERATURE: 0° C to 50° C (32° F to 122° F)
SHORT TERM STORAGE TEMPERATURE: -30° C to 65° C (-22° F to 149° F)
LONG TERM STORAGE TEMPERATURE: 0° C to 35° C (32° F to 95° F)
1
2
SkinPad
ELECTRODE PLACEMENT (Anterior-Posterior)
Grasp the Back/Sternum electrode at
the bottom and peel from the plastic
liner. Apply electrode centered along
spine of patient.
Grasp the Apex/Front
electrode at the bottom and
peel from the plastic liner.
Apply over cardiac apex with
the nipple under adhesive
area.
Avoid any contact between
nipple and gel treatment
area. Skin of the nipple
area is more susceptible to
burning.
Always apply back electrode
first. If front electrode is already
in place when patient is being
maneuvered for placement of
the back, the front may become
partially lifted. This could lead
to arcing and skin burns.
BACK FRONT
R2078-01 Rev. C
Pedi-padz® Solid Gel Multi-function
INDICATIONS FOR USE
Defibrillation Noninvasive Pacing Cardioversion ECG monitoring
SKIN PREPARATION
Ensure skin is clean and dry. Remove any debris, ointments, etc. with water (and mild soap if needed).
Wipe off excess moisture/diaphoresis with dry cloth.
ELECTRODE APPLICATION
Poor adherence and/or air under the electrodes can lead to the possibility of arcing and skin burns.
1. Apply one edge of the pad securely to the patient.
2. “Roll” the pad smoothly from that edge to the other. Be careful
not to trap any pockets of air between the gel and skin.
3. Connect electrodes to machine cable.
4. Connect ECG leads for demand pacing.
Recommended for defibrillation, noninvasive pacing, cardioversion, and ECG monitoring of pediatric patients.
For use on pediatric patients with ZOLL® R Series® and X Series® by trained personnel including Physicians, Nurses, Paramedics, Emergency Medical
Technicians and Cardiovascular Laboratory Technicians. The Pedi-padz electrodes are indicated for use on pediatric patients less than 8 years of age
or weighing less than 55lbs (25kg). For patients over 25kg, adult electrodes are recommended.
Preconnecting the Electrodes
1. Do not open until ready to use.
2. Periodically inspect electrode packaging for integrity & expiration date.
3. Attach to connector of a ZOLL Multi-Function Cable.
WARNINGS
Rx ONLY
NOT MADE WITH
NATURAL RUBBER
LATEX.
32° F
0° C
50° C
122° F
ZOLL International Holding B.V.
Newtonweg 18 6662 PV ELST
The Netherlands Tel. +31 (0) 481 366410
®
ZOLL Medical Corporation
269 Mill Road
Chelmsford, MA 01824-4105 USA
800-348-9011 Made in USA
  • Page 1 1

ZOLL Pedi-Padz User manual

Type
User manual

ZOLL Pedi-Padz are designed for use on pediatric patients less than 8 years of age or weighing less than 55lbs (25kg). They are compatible with ZOLL R Series and X Series devices and can be used for defibrillation, noninvasive pacing, cardioversion, and ECG monitoring. The electrodes feature a solid gel design that conforms to the patient's skin for optimal conductivity and patient comfort. They are also preconnected to a ZOLL Multi-Function Cable for quick and easy setup.