National Instruments NI 4050 User manual

Category
Multimeters
Type
User manual
Computer-Based
Instruments
NI 4050 User Manual
Digital Multimeter Card for PCMCIA
NI 4050 User Manual
March 2000 Edition
Part Number 321427C-01
Worldwide Technical Support and Product Information
www.ni.com
National Instruments Corporate Headquarters
11500 North Mopac Expressway Austin, Texas 78759-3504 USA Tel: 512 794 0100
Worldwide Offices
Australia 03 9879 5166, Austria 0662 45 79 90 0, Belgium 02 757 00 20, Brazil 011 284 5011,
Canada (Calgary) 403 274 9391, Canada (Ontario) 905 785 0085, Canada (Québec) 514 694 8521,
China 0755 3904939, Denmark 45 76 26 00, Finland 09 725 725 11, France 01 48 14 24 24,
Germany 089 741 31 30, Greece 30 1 42 96 427, Hong Kong 2645 3186, India 91805275406,
Israel 03 6120092, Italy 02 413091, Japan 03 5472 2970, Korea 02 596 7456, Mexico (D.F.) 5 280 7625,
Mexico (Monterrey) 8 357 7695, Netherlands 0348 433466, New Zealand 09 914 0488, Norway 32 27 73 00,
Poland 0 22 528 94 06, Portugal 351 1 726 9011, Singapore 2265886, Spain 91 640 0085,
Sweden 08 587 895 00, Switzerland 056 200 51 51, Taiwan 02 2528 7227, United Kingdom 01635 523545
For further support information, see the Technical Support Resources appendix. To comment on the
documentation, send e-mail to [email protected]
© Copyright 1999, 2000 National Instruments Corporation. All rights reserved.
Important Information
Warranty
The NI 4050 is warranted against defects in materials and workmanship for a period of one year from the date of shipment, as
evidenced by receipts or other documentation. National Instruments will, at its option, repair or replace equipment that proves to
be defective during the warranty period. This warranty includes parts and labor.
The media on which you receive National Instruments software are warranted not to fail to execute programming instructions,
due to defects in materials and workmanship, for a period of 90 days from date of shipment, as evidenced by receipts or other
documentation. National Instruments will, at its option, repair or replace software media that do not execute programming
instructions if National Instruments receives notice of such defects during the warranty period. National Instruments does not
warrant that the operation of the software shall be uninterrupted or error free.
A Return Material Authorization (RMA) number must be obtained from the factory and clearly marked on the outside of
the package before any equipment will be accepted for warranty work. National Instruments will pay the shipping costs of
returning to the owner parts which are covered by warranty.
National Instruments believes that the information in this document is accurate. The document has been carefully reviewed
for technical accuracy. In the event that technical or typographical errors exist, National Instruments reserves the right to
make changes to subsequent editions of this document without prior notice to holders of this edition. The reader should consult
National Instruments if errors are suspected. In no event shall National Instruments be liable for any damages arising out of
or related to this document or the information contained in it.
E
XCEPT
AS
SPECIFIED
HEREIN
, N
ATIONAL
I
NSTRUMENTS
MAKES
NO
WARRANTIES
,
EXPRESS
OR
IMPLIED
,
AND
SPECIFICALLY
DISCLAIMS
ANY
WARRANTY
OF
MERCHANTABILITY
OR
FITNESS
FOR
A
PARTICULAR
PURPOSE
. C
USTOMER
S
RIGHT
TO
RECOVER
DAMAGES
CAUSED
BY
FAULT
OR
NEGLIGENCE
ON
THE
PART
OF
N
ATIONAL
I
NSTRUMENTS
SHALL
BE
LIMITED
TO
THE
AMOUNT
THERETOFORE
PAID
BY
THE
CUSTOMER
. N
ATIONAL
I
NSTRUMENTS
WILL
NOT
BE
LIABLE
FOR
DAMAGES
RESULTING
FROM
LOSS
OF
DATA
,
PROFITS
,
USE
OF
PRODUCTS
,
OR
INCIDENTAL
OR
CONSEQUENTIAL
DAMAGES
,
EVEN
IF
ADVISED
OF
THE
POSSIBILITY
THEREOF
. This limitation of the liability of National Instruments will
apply regardless of the form of action, whether in contract or tort, including negligence. Any action against National Instruments
must be brought within one year after the cause of action accrues. National Instruments shall not be liable for any delay in
performance due to causes beyond its reasonable control. The warranty provided herein does not cover damages, defects,
malfunctions, or service failures caused by owner’s failure to follow the National Instruments installation, operation, or
maintenance instructions; owner’s modification of the product; owner’s abuse, misuse, or negligent acts; and power failure or
surges, fire, flood, accident, actions of third parties, or other events outside reasonable control.
Copyright
Under the copyright laws, this publication may not be reproduced or transmitted in any form, electronic or mechanical, including
photocopying, recording, storing in an information retrieval system, or translating, in whole or in part, without the prior written
consent of National Instruments Corporation.
Trademarks
CVI
, ComponentWorks
, LabVIEW
, National Instruments
, ni.com
, NI-DAQ
, SCXI
, and VirtualBench
are trademarks
of National Instruments Corporation.
Product and company names mentioned herein are trademarks or trade names of their respective companies.
WARNING REGARDING USE OF NATIONAL INSTRUMENTS PRODUCTS
(1) NATIONAL INSTRUMENTS PRODUCTS ARE NOT DESIGNED WITH COMPONENTS AND TESTING FOR A LEVEL
OF RELIABILITY SUITABLE FOR USE IN OR IN CONNECTION WITH SURGICAL IMPLANTS OR AS CRITICAL
COMPONENTS IN ANY LIFE SUPPORT SYSTEMS WHOSE FAILURE TO PERFORM CAN REASONABLY BE
EXPECTED TO CAUSE SIGNIFICANT INJURY TO A HUMAN.
(2) IN ANY APPLICATION, INCLUDING THE ABOVE, RELIABILITY OF OPERATION OF THE SOFTWARE PRODUCTS
CAN BE IMPAIRED BY ADVERSE FACTORS, INCLUDING BUT NOT LIMITED TO FLUCTUATIONS IN ELECTRICAL
POWER SUPPLY, COMPUTER HARDWARE MALFUNCTIONS, COMPUTER OPERATING SYSTEM SOFTWARE
FITNESS, FITNESS OF COMPILERS AND DEVELOPMENT SOFTWARE USED TO DEVELOP AN APPLICATION,
INSTALLATION ERRORS, SOFTWARE AND HARDWARE COMPATIBILITY PROBLEMS, MALFUNCTIONS OR
FAILURES OF ELECTRONIC MONITORING OR CONTROL DEVICES, TRANSIENT FAILURES OF ELECTRONIC
SYSTEMS (HARDWARE AND/OR SOFTWARE), UNANTICIPATED USES OR MISUSES, OR ERRORS ON THE PART OF
THE USER OR APPLICATIONS DESIGNER (ADVERSE FACTORS SUCH AS THESE ARE HEREAFTER
COLLECTIVELY TERMED “SYSTEM FAILURES”). ANY APPLICATION WHERE A SYSTEM FAILURE WOULD
CREATE A RISK OF HARM TO PROPERTY OR PERSONS (INCLUDING THE RISK OF BODILY INJURY AND DEATH)
SHOULD NOT BE RELIANT SOLELY UPON ONE FORM OF ELECTRONIC SYSTEM DUE TO THE RISK OF SYSTEM
FAILURE. TO AVOID DAMAGE, INJURY, OR DEATH, THE USER OR APPLICATION DESIGNER MUST TAKE
REASONABLY PRUDENT STEPS TO PROTECT AGAINST SYSTEM FAILURES, INCLUDING BUT NOT LIMITED TO
BACK-UP OR SHUT DOWN MECHANISMS. BECAUSE EACH END-USER SYSTEM IS CUSTOMIZED AND DIFFERS
FROM NATIONAL INSTRUMENTS' TESTING PLATFORMS AND BECAUSE A USER OR APPLICATION DESIGNER
MAY USE NATIONAL INSTRUMENTS PRODUCTS IN COMBINATION WITH OTHER PRODUCTS IN A MANNER NOT
EVALUATED OR CONTEMPLATED BY NATIONAL INSTRUMENTS, THE USER OR APPLICATION DESIGNER IS
ULTIMATELY RESPONSIBLE FOR VERIFYING AND VALIDATING THE SUITABILITY OF NATIONAL
INSTRUMENTS PRODUCTS WHENEVER NATIONAL INSTRUMENTS PRODUCTS ARE INCORPORATED IN A
SYSTEM OR APPLICATION, INCLUDING, WITHOUT LIMITATION, THE APPROPRIATE DESIGN, PROCESS AND
SAFETY LEVEL OF SUCH SYSTEM OR APPLICATION.
Compliance
FCC/Canada Radio Frequency Interference Compliance*
Determining FCC Class
The Federal Communications Commission (FCC) has rules to protect wireless communications from interference.
The FCC places digital electronics into two classes. These classes are known as Class A (for use in industrial-
commercial locations only) or Class B (for use in residential or commercial locations). Depending on where it is
operated, this product could be subject to restrictions in the FCC rules. (In Canada, the Department of
Communications (DOC), of Industry Canada, regulates wireless interference in much the same way.)
Digital electronics emit weak signals during normal operation that can affect radio, television, or other wireless
products. By examining the product you purchased, you can determine the FCC Class and therefore which of the two
FCC/DOC Warnings apply in the following sections. (Some products may not be labeled at all for FCC; if so, the
reader should then assume these are Class A devices.)
FCC Class A products only display a simple warning statement of one paragraph in length regarding interference and
undesired operation. Most of our products are FCC Class A. The FCC rules have restrictions regarding the locations
where FCC Class A products can be operated.
FCC Class B products display either a FCC ID code, starting with the letters EXN,
or the FCC Class B compliance mark that appears as shown here on the right.
Consult the FCC web site
http://www.fcc.gov for more information.
FCC/DOC Warnings
This equipment generates and uses radio frequency energy and, if not installed and used in strict accordance with the
instructions in this manual and the CE Mark Declaration of Conformity**, may cause interference to radio and
television reception. Classification requirements are the same for the Federal Communications Commission (FCC)
and the Canadian Department of Communications (DOC).
Changes or modifications not expressly approved by National Instruments could void the user’s authority to operate
the equipment under the FCC Rules.
Class A
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to part 15
of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to
radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in
which case the user will be required to correct the interference at his own expense.
Canadian Department of Communications
This Class A digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe A respecte toutes les exigences du Règlement sur le matériel brouilleur du
Canada.
Class B
Federal Communications Commission
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15
of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio frequency energy and, if not installed
and used in accordance with the instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful
interference to radio or television reception, which can be determined by turning the equipment off and on, the user
is encouraged to try to correct the interference by one or more of the following measures:
Reorient or relocate the receiving antenna.
Increase the separation between the equipment and receiver.
Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
Consult the dealer or an experienced radio/TV technician for help.
Canadian Department of Communications
This Class B digital apparatus meets all requirements of the Canadian Interference-Causing Equipment Regulations.
Cet appareil numérique de la classe B respecte toutes les exigences du Règlement sur le matériel brouilleur du
Canada.
European Union - Compliance to EEC Directives
Readers in the EU/EEC/EEA must refer to the Manufacturer's Declaration of Conformity (DoC) for information**
pertaining to the CE Mark compliance scheme. The Manufacturer includes a DoC for most every hardware product
except for those bought for OEMs, if also available from an original manufacturer that also markets in the EU, or
where compliance is not required as for electrically benign apparatus or cables.
* Certain exemptions may apply in the USA, see FCC Rules §15.103 Exempted devices, and §15.105(c).
Also available in sections of CFR 47.
** The CE Mark Declaration of Conformity will contain important supplementary information and instructions
for the user or installer.
Conventions
The following conventions are used in this manual:
» The » symbol leads you through nested menu items and dialog box options
to a final action. The sequence File»Page Setup»Options directs you to
pull down the File menu, select the Page Setup item, and select Options
from the last dialog box.
This icon denotes a note, which alerts you to important information.
This icon denotes a caution, which advises you of precautions to take to
avoid injury, data loss, or a system crash.
bold Bold text denotes items that you must select or click on in the software,
such as menu items and dialog box options. Bold text also denotes
parameter names.
italic Italic text denotes variables, emphasis, a cross reference, or an introduction
to a key concept. This font also denotes text that is a placeholder for a word
or value that you must supply.
monospace Text in this font denotes text or characters that you should enter from the
keyboard, sections of code, programming examples, and syntax examples.
This font is also used for the proper names of disk drives, paths, directories,
programs, subprograms, subroutines, device names, functions, operations,
variables, filenames and extensions, and code excerpts.
© National Instruments Corporation vii NI 4050 User Manual
Contents
Chapter 1
Taking Measurements with the NI 4050
Cable and Probes ...........................................................................................................1-1
Introduction to the VirtualBench-DMM Soft Front Panel.............................................1-3
Use the Soft Front Panel................................................................................................1-5
Measure DC and AC Voltage..........................................................................1-5
Measure 2-Wire Resistance.............................................................................1-6
Measure the Voltage Drop Across a Diode.....................................................1-7
Measure Current..............................................................................................1-8
Measure Temperature......................................................................................1-9
Chapter 2
NI 4050 Operation
Safety Instructions .........................................................................................................2-1
Measurement Fundamentals ..........................................................................................2-2
Warm Up.........................................................................................................2-2
Selecting the Resolution..................................................................................2-2
Grounding........................................................................................................2-2
Voltage Measurements ..................................................................................................2-2
DC Voltage......................................................................................................2-2
Input Ranges .....................................................................................2-3
Measurement Considerations............................................................2-3
Input Impedance .................................................................2-3
Thermal EMF .....................................................................2-4
Noise Rejection...................................................................2-4
AC Voltage......................................................................................................2-6
Input Ranges .....................................................................................2-7
Measurement Considerations............................................................2-7
AC Offset Voltage..............................................................2-7
Frequency Response...........................................................2-7
Resistance Measurements..............................................................................................2-8
2-Wire Resistance Measurements ...................................................................2-8
Input Ranges .....................................................................................2-8
Continuity Measurements................................................................................2-9
Diode Measurements .....................................................................................................2-9
Contents
NI 4050 User Manual viii www.ni.com
Appendix A
Specifications
Appendix B
Technical Support Resources
Glossary
Index
Figures
Figure 1-1. Installing the NI 4050 and Cables.........................................................1-2
Figure 1-2. NI-DMM Soft Front Panel....................................................................1-3
Figure 1-3. Digits of Precision.................................................................................1-5
Figure 1-4. Connecting Probes for Voltage Measurement......................................1-6
Figure 1-5. Connections for Resistance Measurement............................................1-7
Figure 1-6. Connecting Signals for Diode Test.......................................................1-7
Figure 1-7. Connections for Current Measurement.................................................1-8
Figure 1-8. Connecting Signals for RTDs and Thermistors....................................1-9
Figure 2-1. Effect of Input Impedance on Signal Measurements............................2-3
Figure 2-2. Normal Mode Measurement Effects.....................................................2-5
Figure 2-3. Common Mode Measurement Effects ..................................................2-6
Figure 2-4. Circuit for 2-Wire Resistance Measurements.......................................2-8
Figure 2-5. Circuit for Diode Measurements...........................................................2-9
© National Instruments Corporation 1-1 NI 4050 User Manual
1
Taking Measurements with
the NI 4050
Thank you for buying a National Instruments 4050 digital multimeter card.
A system based on the NI 4050 offers the flexibility, performance, and
size that makes it ideal for service, repair, and manufacturing as well as
for use in industrial and laboratory environments. The NI 4050, used in
conjunction with your computer, is a versatile, cost-effective platform for
high-resolution measurements.
For the most current versions of manuals and example programs, visit
www.ni.com/instruments for free downloads.
Detailed specifications for the NI 4050 are in Appendix A, Specifications.
Note
Before using any measurement equipment, it is important that you thoroughly
understand the safety instructions for that product. The beginning of Chapter 2, NI 4050
Operation, covers the safety guidelines for your NI 4050.
Cable and Probes
The NI 4050 instrument kit contains the NI 4050 accessory cable that
connects the NI 4050 to a pair of test probes with shrouded banana plugs,
which are also included in the kit. Both the NI 4050 accessory cable and the
test probes meet international safety requirements including UL 3111 and
IEC 1010-1 for the full ranges of applications supported by the NI 4050.
Before using any probes or accessories not supplied by National
Instruments, ensure that they meet applicable safety requirements for the
signal levels you may encounter.
To use the NI 4050 accessory cable and probes with the NI 4050, first
connect the cable to the card as shown in Figure 1-1. The accessory cable
connector is polarized so that it cannot be plugged in incorrectly.
Chapter 1 Taking Measurements with the NI 4050
NI 4050 User Manual 1-2 www.ni.com
Figure 1-1. Installing the NI 4050 and Cables
The test probes connect to the NI 4050 accessory cable via shrouded
banana jacks. The shrouds around the banana jacks prevent you from
contacting potentially hazardous voltages connected to the test probes.
You can also connect the cable to standard, unshrouded banana jack probes
or accessories; however, use unshrouded probes or accessories only when
the voltages are less than 30 V
rms
or 42 V
pk-to-pk
.
Caution To prevent possible safety hazards, the maximum voltage between either of the
inputs and the ground of the computer should never exceed ±250 VDC or 250 V
rms
.
PCMCIA Slot
Portable
Computer
NI 4050
Probes
Accessory Cable
Chapter 1 Taking Measurements with the NI 4050
© National Instruments Corporation 1-3 NI 4050 User Manual
Introduction to the VirtualBench-DMM Soft Front Panel
The following sections explain how to make connections to your NI 4050
and take simple measurements using the VirtualBench-DMM, as shown in
Figure 1-2. To launch the soft front panel, select Start»Program
National Instruments DMM»Soft Front Panel.
Figure 1-2.
NI-DMM Soft Front Panel
The following text describes the options available on the soft front panel.
Refer to Help»Online Reference located on the soft front panel for
information on front panel menus.
The range selector determines the range of measurements
VirtualBench-DMM makes. The range differs for each measurement mode.
If the measurement exceeds the range, +OVER or –OVER appears in the
measurement display. Auto selects the range that best matches the input
signal.
The value indicator displays the value measured by your NI 4060
(The value shown is an example only.).
The unit indicator displays the measurement units of the value you are
measuring. The units are expressed as VAC, VDC, mVAC, mVDC, ,
k,M, mA, AC, or mA DC. The indicator also displays the digits of
resolution. By clicking on the indicator, you can change the DMM’s
resolution.
Chapter 1 Taking Measurements with the NI 4050
NI 4050 User Manual 1-4 www.ni.com
The Function selector allows you select a measurement mode. Select
Edit»Settings and click on the tabs for Current and Resistance or
Temperature to control the data type acquired by VirtualBench-DMM.
DC volts measures the DC component of a voltage signal.
AC volts measures the AC component of a voltage signal.
DC current measures the DC component of a current source.
AC current measures the AC component of a current source.
2-wire measures resistance using the 2-wire method.
4-wire measures resistance using the 4-wire method.
Diode measures the voltage drop across a diode. The maximum voltage
VirtualBench-DMM measures is 2 V.
Temperature measures temperature.
The run button starts and stops continuous DMM measurements.
The single button performs a single measurement.
The math buttons allow you to manipulate readings mathematically.
Null starts relative mode. VirtualBench-DMM makes all subsequent
measurements relative to the measurement it makes when you click
on Null.
Max/Min displays the maximum and minimum values that occur after you
start Relative mode.
mX+B enables the mX+B calculation on all readings.
dB compresses a large range of measurements into a much smaller range
by expressing DC or AC voltage in decibels.
dBm shows decibels above or below a 1 mW reference.
Chapter 1 Taking Measurements with the NI 4050
© National Instruments Corporation 1-5 NI 4050 User Manual
% selects the percentage calculation. VirtualBench-DMM expresses the
displayed reading as a percent deviation from the reference value entered
in the Math Settings. Refer to Help»Online Reference, Math Settings topic
for more information about dB, dBm, mX+B, and percentage calculations.
The log button enables data logging. To configure the datalog file and log
interval, select Edit»Settings. Refer to Help»Online Reference, Logging
Measurements to Disk topic for more details.
Digits of Precision—A pop-up ring control in the DMM front panel display
allows you to set measurement accuracy to 3 1/2, 4 1/2, or 5 1/2. A larger
value gives greater precision but slower measurement performance. Refer
to Figure 1-3.
Figure 1-3. Digits of Precision
Use the Soft Front Panel
The following sections describe procedures for measuring DC and AC
voltage, resistance, diode, and temperature, using the soft front panel.
Measure DC and AC Voltage
Use the following procedure to measure DC and AC voltage using the soft
front panel:
1. Connect the test probes to voltage signals as shown in Figure 1-4. For
DC voltages, the HI (red) terminal is the positive terminal, and the
LO (black) terminal is negative. For AC voltages, positive and negative
terms are irrelevant.
Chapter 1 Taking Measurements with the NI 4050
NI 4050 User Manual 1-6 www.ni.com
The NI 4050 is protected against damage from voltages within
±250 VDC or 250 V
rms
in all ranges. You should never apply voltages
above these levels to the inputs.
Figure 1-4. Connecting Probes for Voltage Measurement
2. Select the mode you will measure:
•DC Volts
•AC Volts
3. Select the range for your measurement or autoranging:
DC Volts—± 20 mV, ± 200 mV, ± 2 V, ± 25 V, and ± 250 V
AC Volts—20 mV
rms
, 200 mV
rms
, 2 V
rms
, 25 V
rms
, and 250 V
rms
The value indicator displays the voltage measured.
Measure 2-Wire Resistance
Use the following procedure to measure 2-wire resistance using the soft
front panel:
1. Connect the test probes to a resistor as shown in Figure 1-5. To
accurately measure the value of a resistor, make sure the resistor is not
connected to any other circuits. Erroneous or misleading readings may
result if the resistor you are measuring is connected to external circuits
that supply voltages or currents or to external circuits that change the
effective resistance of that resistor.
AC Voltage
Source
DC Voltage
Source
+
250 V
MAX.
LO
HI
250 V
MAX.
LO
HI
Chapter 1 Taking Measurements with the NI 4050
© National Instruments Corporation 1-7 NI 4050 User Manual
Figure 1-5. Connections for Resistance Measurement
2. Select 2-wire resistance mode.
3. Select the range for your measurement—200 , 2 k, 20 k, 200 k,
2 M, 200 M, or autorange.
The value indicator indicates the resistance measured. See the 2-Wire
Resistance Measurements section of Chapter 2, NI 4050 Operation, for
more information on 2-wire resistance measurements.
Measuring the Voltage Drop Across a Diode
The NI 4050 can excite a device under test and read the resulting voltage
drop. Diode mode is useful for testing diodes. Use the following procedure
to measure the forward drop across a diode. Voltage up to 2 V can be
measured in this mode.
1. Connect the test probes to a diode as shown in Figure 1-6. To
accurately measure the forward voltage of a diode, make sure that the
diode is not connected to any other circuits. The NI 4050 biases the
diode with a current of 100 µA and measures the resulting voltage
drop. Diode measurements are made with a fixed range of 2.0 V.
Figure 1-6. Connecting Signals for Diode Test
Resistor
250 V
MAX.
LO
HI
Diode
+
100 µA
250 V
MAX.
LO
HI
Chapter 1 Taking Measurements with the NI 4050
NI 4050 User Manual 1-8 www.ni.com
2. Select diode mode.
3. Select the range for your measurement. Only the 2 V range is available
for diode measurement.
The value indicator will indicate the voltage drop measured. If the display
indicates 2.200 VDC, the diode is either reverse biased or defective. See the
Diode Measurements section of Chapter 2, NI 4050 Operation, for more
information on diode measurements.
Measure Current
You can use the NI 4050 to measure current with an optional National
Instruments CSM series current shunt module. These accessories are
connected between the NI 4050 cable and the test probes as shown in
Figure 1-7.
Figure 1-7.
Connections for Current Measurement
Current shunt accessories contain a precision resistor that converts the
current through the shunt into a voltage that the NI 4050 can measure in
voltage mode. While the soft front panel cannot measure current directly
with the NI 4050, you can calculate the value of the current flowing through
the shunt by dividing the voltage measured by the value of the precision
resistor.
Current
Source
Current Shunt
Accessory
250 V
MAX.
LO
HI
Chapter 1 Taking Measurements with the NI 4050
© National Instruments Corporation 1-9 NI 4050 User Manual
Measure Temperature
You can measure temperature using common temperature transducers such
as resistive temperature devices (RTD) and thermistors. You can measure
transducers in the 2-wire resistance mode, as shown in Figure 1-8.
Although the soft front panel does not support temperature measurements,
you can convert and scale the transducer value to temperature
programmatically through software.
Note
The NI 4050 for PCMCIA does not support 4-wire resistance measurements. To
avoid measurement errors due to resistance offset, before doing resistance measurements,
measure the resistance to your loads.
Figure 1-8.
Connecting Signals for RTDs and Thermistors
Resistor
250 V
MAX.
LO
HI
© National Instruments Corporation 2-1 NI 4050 User Manual
2
NI 4050 Operation
This chapter contains safety instructions, measurement fundamentals and
concerns, and scanning information.
Safety Instructions
Cautions
To avoid personal injury or damage to electronic equipment, observe the
following:
Do not operate this instrument in an explosive atmosphere or where there may be
flammable gases or fumes.
Equipment described in this document must be used in an Installation Category II
environment per IEC 664. This category requires local level supply mains-connected
installation.
The NI 4050 must be used in a UL-listed laptop or personal computer.
To prevent safety hazards, the maximum voltage between either of the inputs and the
ground of the computer should never exceed ±250 VDC or 250 V
rms
.
Do not operate damaged equipment. The safety protection features built into this
instrument can become impaired if the instrument becomes damaged in any way. If the
instrument is damaged, do not use it until service-trained personnel can check its safety.
If necessary, return the instrument to National Instruments for service and repair to ensure
that its safety is not compromised.
Do not operate this instrument in a manner that contradicts the information specified in this
document. Misuse of this instrument could result in a shock hazard.
Do not substitute parts or modify equipment. Because of the danger of introducing
additional hazards, do not install unauthorized parts or modify the instrument. Return the
instrument to National Instruments for service and repair to ensure that its safety is not
compromised.
Connections that exceed any of the maximum signal ratings on the NI 4050 can create a
shock or fire hazard or can damage any or all of the devices connected to the NI 4050.
National Instruments is not liable for any damages or injuries resulting from incorrect
signal connections.
Clean the instrument and accessories by brushing off light dust with a soft, nonmetallic
brush. Remove other contaminants with a stiff nonmetallic brush. The unit must be
completely dry and free from contaminants before returning to service.
Chapter 2 NI 4050 Operation
NI 4050 User Manual 2-2 www.ni.com
Measurement Fundamentals
Warm Up
The required warm-up time for the NI 4050 is 30 minutes. This warm-up
time is important because measurements made with the NI 4050
multimeter can change with temperature. This change is called a thermal
drift and affects your accuracy. To minimize the effects of thermal drift and
ensure the specified accuracies, take all measurements after the NI 4050
has had a chance to fully warm up. Depending on your environment, the
NI 4050 can operate significantly above ambient temperature. Therefore,
measurements made immediately after powering up the system can differ
significantly from measurements made after the system has fully warmed
up. The NI 4050 temperature specifications are listed in the Accuracy
sections in Appendix A, Specifications.
Selecting the Resolution
The resolution on the NI 4050 multimeter is programmable. You can select
from three different resolutions: 5 1/2 digits, 4 1/2 digits, or 3 1/2 digits.
These settings allow you to trade off speed for resolution. The 5 1/2 digit
setting has the highest resolution and slowest reading rate, while the
3 1/2 digit setting gives you the least resolution and fastest reading rate.
Measurement mode and range affect the reading rate by requiring different
conversion times to obtain a given resolution for the different modes and
ranges.
Grounding
When measuring ground-referenced signals, connect the
ground-referenced side of your signal to the IN HI + terminal for best
performance.
Voltage Measurements
DC Voltage
Your NI 4050 multimeter uses a high-resolution delta sigma, A/D
converter (ADC) to sample signals and convert them into a digital form.
The ADC is preceded by a series of gain and attenuation circuitry that allow
both small and large signals to be measured using the same converter. The
NI 4050 uses a digital filter, which heavily rejects powerline frequencies
(50–60 Hz) and their harmonics, as well as high-frequency noise.
Chapter 2 NI 4050 Operation
© National Instruments Corporation 2-3 NI 4050 User Manual
Input Ranges
The NI 4050 has five input ranges available for measuring DC voltages.
These ranges are ±20 mV, ±200 mV, ±2.0 V, ±25V, and ±250 V. Each
range has a 10% overrange, except for the 250 V range. The 250 V and
25 V input ranges have a 1 Minput impedance; the 2 V, 200 mV, and
20 mV ranges have an input impedance greater than 1 G. Take these
values into consideration when measuring high-impedance sources. When
the NI 4050 is powered off, the 250 V and 25 V input range have a 1 M
input impedance and the 2 V, 200 mV, and 20 mV ranges have an input
impedance of 100 kΩ.
If you are taking measurements that require a high degree of accuracy, you
should consider problems associated with input impedance, noise effects,
and thermal electromotive forces (thermal EMFs). These effects are
discussed in the Measurement Considerations section.
Measurement Considerations
Input Impedance
Figure 2-1 illustrates the input impedance of an NI 4050 and its effect on
the measurement of a circuit under test. If you know the source impedance
of the circuit being tested, you can correct for the attenuation caused by the
NI 4050 in software. Since R
in
is large, at least 1 M, it will require a large
source impedance, R
s
, to cause a large change in the measured voltage, V
m
.
Figure 2-1.
Effect of Input Impedance on Signal Measurements
Source
Voltage V
s
External Source
Impedance R
s
+
+
Input
Impedance
R
in
Measured
Voltage
V
m
V
s
R
in
R
s
+ R
in
V
m
-------------- -------------=
HI
LO
Input
V
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46

National Instruments NI 4050 User manual

Category
Multimeters
Type
User manual