Swann AMI Codes-II User manual

Category
Measuring, testing & control
Type
User manual
AMI Codes-II
Version 6.20 and higher
A-96.250.571 / 050517
Operator’
s Manual
© 2016, SWAN ANALYTISCHE INSTRUMENTE AG, Switzerland, all rights reserved
subject to change without notice
Customer Support
SWAN and its representatives maintain a fully trained staff of technical specialists
around the world. For any technical question, contact your nearest
SWAN representative, or the manufacturer:
SWAN ANALYTISCHE INSTRUMENTE AG
Studbachstrasse 13
8340 Hinwil
Switzerland
Internet: www.swan.ch
Document Status
Title:
Monitor AMI Codes-II Operator’s Manual
ID:
A-96.250.571
Revision Issue
00 April 2010
01 June 2010
02 Aug. 2013 Update to Rev. 5.30, main board V2.4
03 June 2016 Update to Rev. 6.00, main board V2.5
AMI Codes-II
A-96.250.571 / 050517 1
Table of Contents
1. Safety Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1. Warning Notices . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.2. General Safety Regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3. Restrictions for use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2. Product Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1. Instrument Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
2.2. Instrument Overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
3. Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.1. Installation Check List. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3.2. Mounting of Instrument Panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3. Connecting Sample and Waste . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3.1 FEP Tube at Sample Inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
3.3.2 FEP Tube at Sample Outlet . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4. Installation of Flow Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
3.5. Install the Option pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.1 pH as Option ex works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.2 pH Option as Retrofit Kit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.6. Electrical Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6.1 Connection Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
3.6.2 Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
3.7. Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.8. Relay Contacts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.8.1 Alarm Relay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
3.8.2 Relay 1 and 2 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
3.9. Signal Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.9.1 Signal Output 1 and 2 (current outputs) . . . . . . . . . . . . . . . . . . 31
3.10 Interface Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
3.10.1 Signal Output 3 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.10.2 Profibus, Modbus Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
3.10.3 HART Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
3.10.4 USB Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
2 A-96.250.571 / 050517
AMI Codes-II
4. Instrument Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.1. Prepare Reagents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.2. Peristaltic Pump . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
4.3. Establish Sample Flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
4.4. Fill or Flush Reagent System . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.5. Programming . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
4.6. Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5. Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.1. Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.2. Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.3. Software Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.4. Changing Parameters and values. . . . . . . . . . . . . . . . . . . . . . . . . 41
6. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.1. Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
6.2. Stop of Operation for Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 43
6.3. Refill or replace Reagents. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
6.3.1 Reagents for Free Chlorine, Chlorine Dioxide, Bromine and Iodine 46
6.3.2 Reagents for Measuring Monochloramine and Ozone . . . . . . . 48
6.4. Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
6.5. Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
6.6. Cleaning the protective Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
6.7. Cleaning the Photometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
6.8. Cleaning the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.8.1 Disassemble the Flow Cell . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
6.8.2 Assemble the Flow Cell. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
6.9. Maintenance of pH sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
6.10. Tube Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
6.10.1 Replace the Pump Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
6.10.2 Replace the Reagent Tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . 61
6.11. Replacing Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
6.12. Longer Stop of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
7. Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
7.1. General Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
7.2. Calibration Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
7.2.1 Process calibration DIS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
7.2.2 Process pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
7.2.3 Standard pH . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
7.3. Error List . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
AMI Codes-II
A-96.250.571 / 050517 3
8. Program Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
8.1. Messages (Main Menu 1) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
8.2. Diagnostics (Main Menu 2). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 70
8.3. Maintenance (Main Menu 3). . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
8.4. Operation (Main Menu 4) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
8.5. Installation (Main Menu 5) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
9. Program List and Explanations. . . . . . . . . . . . . . . . . . . . . . . . . 75
1 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
2 Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 75
3 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
4 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
5 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
10. Material Safety Data sheets . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
10.1. Reagents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
11. Default Values. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
12. Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
13. Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
4 A-96.250.571 / 050517
AMI Codes-II
Safety Instructions
AMI Codes-II - Operator’s Manual
This document describes the main steps for instrument setup, oper-
ation and maintenance.
1. Safety Instructions
General The instructions included in this section explain the potential risks
associated with instrument operation and provide important safety
practices designed to minimize these risks.
If you carefully follow the information contained in this section, you
can protect yourself from hazards and create a safer work environ-
ment.
More safety instructions are given throughout this manual, at the
respective locations where observation is most important.
Strictly follow all safety instructions in this publication.
Target
audience
Operator: Qualified person who uses the equipment
for its intended purpose.
Instrument operation requires thorough knowledge of applications,
instrument functions and software program as well as all applicable
safety rules and regulations.
OM Location The AMI Operator’s Manual shall be kept in proximity of the instru-
ment.
Qualification,
Training
To be qualified for instrument installation and operation, you must:
read and understand the instructions in this manual as well as
the Material Safety Data Sheets.
know the relevant safety rules and regulations.
AMI Codes-II
Safety Instructions
A-96.250.571 / 050517 5
1.1. Warning Notices
The symbols used for safety-related notices have the following sig-
nificance:
DANGER
Your life or physical wellbeing are in serious danger if such
warnings are ignored.
Follow the prevention instructions carefully.
WARNING
Severe injuries or damage to the equipment can occur if such
warnings are ignored.
Follow the prevention instructions carefully.
CAUTION
Damage to the equipment, minor injury, malfunctions or incor-
rect process can be the consequence if such warnings are ig-
nored.
Follow the prevention instructions carefully.
Mandatory
Signs
The importance of the mandatory signs in this manual.
Safety goggles
Safety gloves
6 A-96.250.571 / 050517
AMI Codes-II
Safety Instructions
Warning Signs The importance of the warning signs in this manual.
Electrical shock hazard
Corrosive
Harmful to health
Flammable
Warning general
Attention general
AMI Codes-II
Safety Instructions
A-96.250.571 / 050517 7
1.2. General Safety Regulations
Legal
Requirements
The user is responsible for proper system operation.
All precautions must be followed to ensure safe operation
of the instrument.
Spare Parts
and
Disposables
Use only official SWAN spare parts and disposables. If other parts
are used during the normal warranty period, the manufacturer’s
warranty is voided.
Modifications Modifications and instrument upgrades shall only be carried out by
an authorized Service Technician. SWAN will not accept responsi-
bility for any claim resulting from unauthorized modification or alter-
ation.
WARNING
Risk of Electrical Shock
If proper operation is no longer possible, the instrument must be
disconnected from all power lines, and measures must be taken
to prevent inadvertent operation.
To prevent from electrical shock, always make sure that the
ground wire is connected.
Service shall be performed by authorized personnel only.
Whenever electronic service is required, disconnect instru-
ment power and power of devices connected to.
relay 1,
relay 2,
alarm relay
WARNING
For safe instrument installation and operation you must read
and understand the instructions in this manual.
WARNING
Only SWAN trained and authorized personnel shall perform the
tasks described in this document.
8 A-96.250.571 / 050517
AMI Codes-II
Safety Instructions
1.3. Restrictions for use
The sample must not contain any particles, which may block the
flow cell. Sufficient sample flow is coercive for the correct function
of the instrument.
If the sample contains only little disinfectant concentrations, or
there is the danger of biological growth, we recommend to use the
optional Cleaning module from Swan.
WARNING
Health hazard
Some reagents are etching and can cause severe burns or eye
damage.
For safe handling of the reagents you must read and under-
stand the instructions in this manual, as well as the Material
Safety Data Sheets (MSDS)
Download
MSDS
The current Material Safety Data Sheets (MSDS) for the below list-
ed Reagents are available for downloading at www.swan.ch.
OXYCON ON-LINE DPD
OXYCON ON-LINE Buffer
OXYCON ON-LINE KI
Buffer solution pH 4
Buffer solution pH 7
Buffer solution pH 9
AMI Codes-II
Product Description
A-96.250.571 / 050517 9
2. Product Description
Application
Range
The AMI Codes-II analyzer is a complete monitoring system for the
automatic, continuous measurement and dosing control of chlorine
and other disinfectants based on the DPD colorimetric method
AWWA 4500 Cl-G and on EN ISO 7393-2.
It can be used for measuring disinfectants in:
Potable water
Swimming pools
Cooling water
Effluent
Seawater
It is also applicable for water containing additives like corrosion in-
hibitors, cyanuric acids and antiscalants.
For the determination of free chlorine, chlorine dioxide, bromine
and iodine the reagents:
Oxycon on-line DPD
Oxycon on-line Buffer
are needed.
The measurement of monochloramine or ozone requires an addi-
tional reagent:
Oxycon on-line KI; which is added to Oxycon on-line Buffer
Disinfectant
Measurement
Signal
Outputs
Two signal outputs programmable for measured values (freely
scaleable, linear or bilinear) or as continuous control output (control
parameters programmable).
Current loop: 0/420 mA
Maximal burden: 510
Third signal output available as an option. The third signal output
can be operated as a current source or as a current sink (selectable
via switch).
Disinfectant Measuring range Accuracy
Ozone 0.0001.000 ppm ± 0.005 ppm
HOCl / free chlorine /
Monochloramine
0.001.00 ppm
1.003.00 ppm
3.005.00 ppm
± 0.01 ppm
± 0.06 ppm
± 0.20 ppm
Chlorine dioxide /
Iodine / Bromine
0.002.00 ppm
2.006.00 ppm
± 0.02 ppm
± 0.12 ppm
10 A-96.250.571 / 050517
AMI Codes-II
Product Description
Relay Two potential-free contacts programmable as limit switches for
measuring values, controllers or timer for system cleaning with au-
tomatic hold function. Both contacts can be set as normally open or
normally closed with a jumper.
Maximum load: 1 A/250 VAC
Alarm Relay One potential free contact.
Alternatively:
Open during normal operation, closed on error and loss of
power.
Closed during normal operation, open on error and loss of
power.
Summary alarm indication for programmable alarm values and in-
strument faults.
Input For potential-free contact to freeze the measuring value or to inter-
rupt control in automated installations (hold function or remote-off)
Special
Feature
Possibility to interrupt the measurement by activating the input. See
Program List and Explanation, 5.3.4, p. 93.
Safety
Features
No data loss after power failure. All data is saved in non-volatile
memory. Over voltage protection of in- and outputs.Galvanic sepa-
ration of measuring inputs and signal outputs.
Communica-
tion Interface
(optional)
USB Interface for logger download
Third signal output (can be used in parallel to the USB interface)
RS485 with Fieldbus protocol Modbus or Profibus DP
HART interface
pH Optional pH measurement is possible (pH correction or calibration).
Relay Box The AMI Relay Box is designed for the direct power supply and ac-
tivation of dosing devices which are controlled with an AMI trans-
mitter, e. g. to connect two solenoid valves or one motor valve for
disinfectant addition.
Cleaning
Module
An optional Cleaning module is available.
AMI Codes-II
Product Description
A-96.250.571 / 050517 11
Fluidics The sample flows through the sample inlet [F] and the filter vessel
[G] into the constant head [A]. Adjust the flow regulating valve [D]
so that always a small part of the sample flows through the overflow
tube [B] into the drain [H]. A part of the sample flows through the
photometer inlet [C] into the mixing chamber [E], where the re-
agents [J] and [K] are added by the peristaltic pump [O] and mixed
with the sample. The mixed sample flows through the photometer
[N] and disinfectant is measured. If KI and Buffer are mixed in the
reagent canister [K] monochloramine is measured.
After the measurement the sample flows through the outlet of the
photometer where it will be aerated through air inlet [P] to generate
bubbles. Then the sample flows through the bubble detector [M]
into the photometer drain [I].
A
B
C
D
E
F
G
H
Constant head
Overflow tube
Photometer inlet
Flow regulating valve
Mixing chamber
Sample inlet
Inlet Filter
Constant head drain
I
J
K
L
M
N
O
P
Photometer drain
Reagent Oxycon on-line DPD
Reagent Oxycon on-line Buffer
Reagent level detector
Air bubble detector
Photometer
Peristaltic pump
Photometer air inlet
Level
D
B
A
C
F
G
H
I
J
K
L
M
N
O
P
E
12 A-96.250.571 / 050517
AMI Codes-II
Product Description
Time
interval of a
measurement
The measuring interval can be set between 1 and 12 minutes. The
time sequence of a measurement with a measuring interval of
5 min is shown in the diagram below.
The blue bar represents the sample which flows continuously
through the photometer. A short time before the measurement
starts, a zero point measurement is performed. Then the peristaltic
pump starts and a small portion of the reagents [J] and [K] is
pumped into the mixing chamber. Shortly after, if the mixture is in
the photometer, the sample is measured.
J
K
OXYCON ON-LINE DPD
OXYCON ON-LINE Buffer Solution
Zero point measurement
Sample measurement
1
0
2345678910111213141516171819 20
m
J
K
AMI Codes-II
Product Description
A-96.250.571 / 050517 13
2.1. Instrument Specification
Power Supply Voltage:
Power consumption:
100240 VAC (± 10%)
50/60 Hz (± 5%)
or 24 VDC (± 10%)
max. 30 VA
Electronics Aluminium with a protection degree of IP 66 / NEMA 4X
housing Ambient temperature:
Limit range of operation:
Storage and transport:
Humidity:
Display:
-10 to +50 °C
-25 to +65 °C
-30 to +85 °C
1090% rel., non condensing
backlit LCD, 75 x 45 mm
Sample
requirements
Flow rate:
Temperature:
Inlet pressure:
Outlet pressure:
min. 10 l/h
550 °C (41122 °F)
0.152 bar (228 PSI)
pressure free
NOTICE: No oil, no grease, no sand.
On-site The analyzer site must permit connections to:
requirements Sample inlet:
Sample outlet:
Tube 6 x 8 mm
1/2” hose nozzle for flexible tube
diam. 20x15 mm
Measuring
Range
Ozone
0.0001.000 ppm
Accuracy
± 0.005 ppm
HOCL, free chlorine, monochloramine
0.001.00 ppm
1.003.00 ppm
3.005.00 ppm
± 0.01 ppm
± 0.06 ppm
± 0.2 ppm
Chlorine dioxide, iodine, bromine
0.002.00 ppm
2.006.00 ppm
± 0.02 ppm
± 0.12 ppm
14 A-96.250.571 / 050517
AMI Codes-II
Product Description
Dimensions Panel:
Mounting hole distance
Screws:
Weight:
400x850 mm
374x824
5 mm or 6 mm diameter
12.0 kg / 26.5 lbs without reagents and
sample water
17.0 kg / 37.5 lbs with reagents and sam-
ple water
AMI Codes II
824 mm / 32
7
/
16
412 mm / 16 ½ ”
850 mm / 33½”
13 mm / ½”
400 mm / 15¾”
374 mm / 14¾”
13 mm / ½”
6 x dia. 6.5 mm / ¼”
30 mm / 1 ”
3
16
/
AMI Codes-II
Product Description
A-96.250.571 / 050517 15
2.2. Instrument Overview
A
B
C
D
E
F
G
H
Panel
Transmitter
Peristaltic pump
Reagent Oxycon on-line DPD
Reagent Oxycon on-line Buffer
pH sensor
Temperature sensor
Constant head
I
J
K
L
M
N
O
Flow regulating valve
Sample inlet
Inlet filter
Photometer
Air bubble detector
Constant head drain
Photometer drain
A
B
C
D
E
F
G
H
I
J
K
L
M
N
O
16 A-96.250.571 / 050517
AMI Codes-II
Installation
3. Installation
3.1. Installation Check List
Check Instrument’s specification must conform to the National Electrical
Code, all state and local codes, and all plant codes and standards
for electrical equipment.
On site require-
ments
100240 VAC ( 10%), 50/60 Hz ( 5%) or 24 VDC (±10%),
isolated power outlet with ground connection and 30 VA.
Sample line with min. 10 l/h and 0.152 bar.
Waste line with pressure free drain.
Installation Mount the instrument in vertical position.
Display should be at eye-level.
Mount the filter, filter vessel, and constant head cover. Connect the
sample and waste line. See Connecting Sample and Waste, p. 17
Electrical Wiring Do not switch on the Instrument until all electrical connections
have been completed.
Connect all external devices like limit switches, current loops and
pumps.
Connect power cord.
See Electrical Connections, p. 24
If ordered:
Option pH
See Install the Option pH, p. 20
Reagents Prepare reagents. See Refill or replace Reagents, p. 43.
Insert the suction lances.
Power-up Perform exactly in this order:
Lock pump tubes.
Turn on the sample flow and wait until the flow cell is completely
filled.
Switch on power.
Start <Fill system>. See 4.4., 2 36.
Instrument
Setup
Program all parameters for external devices (interface, recorders,
etc.). Program all parameters for instrument operation (limits,
alarms, measuring interval).
Run-in period Let the instrument run continuously for 1 h.
pH sensor
calibration
If ordered: Calibrate pH sensor. See Process pH, p. 51.
AMI Codes-II
Installation
A-96.250.571 / 050517 17
3.2. Mounting of Instrument Panel
The first part of this chapter describes the preparing and placing of
the system for use.
The instrument must only be installed by trained personnel.
Mount the instrument in vertical position.
For ease of operation mount it so that the display is at eye
level.
For the installation a kit containing the following installation
material is available:
6 Screws 6x60 mm
–6 Dowels
6 Washers 6.4/12 mm
Mounting re-
quirements
The instrument is only intended for indoor installation.
For dimensions see Dimensions, p. 14.
3.3. Connecting Sample and Waste
3.3.1 FEP Tube at Sample Inlet
Use plastic tube (FEP, PA, or PE 6 x 8 mm) to connect the sample
line.
Mounting of
SERTO fitting
Process
calibration
Make 3 manual measurements. Use a high quality photometer,
e.g. Chematest from Swan. Calculate average value and compare
this value to the value indicated by the AMI. If necessary, correct
the value. The zero point is done automatically before each mea-
surement.
A
B
C
D
Screw connection
Compression ferrule
Knurled nut
Flexible tube
B
C
D
A
18 A-96.250.571 / 050517
AMI Codes-II
Installation
3.3.2 FEP Tube at Sample Outlet
WARNING
Risk of water pollution
The drain of the photometer outlet contains DPD.
At no means recirculate it into the water system.
Connect the 1/2” tubes [F] to the hose nozzles [E] and place it into
a pressure free drain with sufficient capacity.
A
B
C
D
E
F
Tube from photometer
Drain Photometer
Tube from constant head
Drain constant head
Hose nozzles
1/2” tubes
B
D
E
F
A
C
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106