Dwyer Series PFG2 User manual

Type
User manual
2-5/16 [58.74]
2-3/4 [69.85]
111
O-RING
PORT & MTG
BOLT 3/8-24
or M10x1.5 THD
2-1/8
[53.98]
2-1/2
[63.50]
1-3/4 [44.45]
7/8
[22.23]
3/8 [9.53]
HIGH
PRESSURE
INLET
LOW
PRESSURE
OUTLET
FLOW
Series PFG2 Process Filter Gage
Specications - Installation and Operating Instructions
Bulletin P-PFG2
The Series PFG2 Process Filter Gage is designed for determining the state of an
in-line lter. The differential pressure indicator determines the pressure drop on either
side of a lter through its 1/8˝ female NPT pressure connections, and relates the
value to one of three zones: clean (green), change (yellow), or dirty (red). The Series
PFG2 is perfectly suited for lter applications, line loss, valve drop, and many other
differential pressure applications where a simple indicator is needed.
MOUNTING
The Series PFG2 features two #10-32 UNF threaded holes in its mounting block for
easy mounting. The holes are 3/4˝ apart and located at the same level as the end
process connections. The direction of process ow is indicated on the dial, with the
arrow pointing to the low pressure port. In order to change the high and low pressure
connections, simply remove the indicator from the mounting base and rotate 180°.
Retighten the plastic bolts to 20 to 25˝ pounds of torque. By removing the mounting
block, the PFG2 can be directly mounted through the outlet/inlet connections on the
bottom of the unit.
MAINTENANCE
Upon nal installation of the Series PFG2 Process Filter Gage, no routine maintenance
is required. A periodic check of system calibration is recommended. The Series PFG2
is not eld serviceable and should be returned if repair is needed (eld repair should
not be attempted and may void warranty). Be sure to include a brief description of the
problem plus any relevant application notes. Contact customer service to receive a
return goods authorization number before shipping.
PFG2 WITHOUT MOUNTING BASE
SPECIFICATIONS
Service: Liquids/gases compatible with SS, GFN, and uoropolymer.
Wetted Materials: Aluminum, SS, glass lled nylon, and uoropolymer.
Accuracy: ±5% F.S.
Maximum Temperature: 200°F (93°C).
Maximum Pressure: 300 psig (20.7 bar).
Materials: Body: Glass lled nylon; Mounting Block: Aluminum; Lens: Polyester;
Elastomers: Fluoroelastomer.
Process Connection: 1/8˝ female NPT.
Mounting Orientation: Any orientation with 10-32 threaded holes 3/4˝ apart.
Dimensions: 3.25˝ H x 3˝ W x 2.31˝ D.
Weight: 9.6 oz (272.2 g).
Printed in U.S.A. 10/18 FR# 443837-00 Rev. 1©Copyright 2018 Dwyer Instruments, Inc.
HIGH
PORT
1-1/8
[28.58]
3
[76.20]
3/4
[19.05]
2X #10-32 UNF
MOUNTING HOLES
LOW
PORT
3/4
[19.08]
3/8
[9.37]
2-1/2
[63.50]
2-3/4
[69.85]
2-5/16
[58.74]
1-1/2
[38.10]
2X 1/8 NPT
CONNECTION
®
DWYER INSTRUMENTS, INC.
P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.
Phone: 219/879-8000
Fax: 219/872-9057
www.dwyer-inst.com
  • Page 1 1