USER MANUAL
MODEL 465L
OZONE MONITOR
Teledyne API (TAPI)
9970 Carroll Canyon Road
San Diego, CA 92131-1106
Phone:
800-324-5190 (toll free)
+1 858-657-9800 (direct)
FAX:
+1 858-657-9816
Email:
Website:
http://www.teledyne-api.com/
© 2007-2018 TAPI 05509K DCN8017
Teledyne API 31 October2018
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual i
Notice of Copyright
© 2007-2018 Teledyne API Inc. All rights reserved.
Trademarks
All trademarks, registered trademarks, brand names or product names appearing in this
document are the property of their respective owners and are used herein for identification
purposes only.
ii Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
WARRANTY
WARRANTY POLICY (02024J)
Teledyne API (TAPI), a business unit of Teledyne Instruments, Inc., provides that:
Prior to shipment, TAPI equipment is thoroughly inspected and tested. Should equipment failure
occur, TAPI assures its customers that prompt service and support will be available. (For the
instrument-specific warranty period, please refer to the “Limited Warranty” section in the Terms
and Conditions of Sale on our website at the following link: http://www.teledyne-
api.com/terms_and_conditions.asp).
COVERAGE
After the warranty period and throughout the equipment lifetime, TAPI stands ready to provide
on-site or in-plant service at reasonable rates similar to those of other manufacturers in the
industry. All maintenance and the first level of field troubleshooting are to be performed by the
customer.
NON-TAPI MANUFACTURED EQUIPMENT
Equipment provided but not manufactured by TAPI is warranted and will be repaired to the extent
and according to the current terms and conditions of the respective equipment manufacturer’s
warranty.
Product Return
All units or components returned to Teledyne API should be properly packed for handling and
returned freight prepaid to the nearest designated Service Center. After the repair, the equipment
will be returned, freight prepaid.
The complete Terms and Conditions of Sale can be reviewed at http://www.teledyne-
api.com/terms_and_conditions.asp
CAUTION – Avoid Warranty Invalidation
Failure to comply with proper anti-Electro-Static Discharge (ESD) handling
and packing instructions and Return Merchandise Authorization (RMA)
procedures wh
en returning parts for repair or calibration may void your
warranty. For anti-ESD handling and packing instructions please refer to the
manual, Fundamentals of ESD, PN 04786, in its “Packing Components for
Return to Teledyne API’s Customer Service” section
. The manual can be
downloaded from our website at http://www.teledyne-
api.com. RMA
procedures can also be found on our website.
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual iii
IMPORTANT SAFETY INFORMATION
Important safety messages are provided throughout this manual for the purpose of
avoiding personal injury or instrument damage. Please read these messages
carefully. Each safety message is associated with a safety alert symbol and placed
throughout this manual and inside the instrument. The symbols with messages are
defined as follows:
WARNING: Electrical Shock Hazard
HAZARD: Strong oxidizer
GENERAL WARNING/CAUTION: Read the accompanying message
for specific information.
CAUTION: Hot Surface Warning
Do Not Touch: Touching some parts of the instrument without
protection or proper tools could result in damage to the part(s) and/or
the instrument.
Technician Symbol: All operations marked with this symbol are to be
performed by qualified maintenance personnel only.
Electrical Ground: This symbol inside the instrument marks the central
safety grounding point for the instrument.
CAUTION
GENERAL SAFETY HAZARD
This instrument should only be used for the purpose and in the manner
described in this manual. If you use it in a manner other than that for
which it was intended, unpredictable behavior could ensue with possible
hazardous consequences.
NEVER use any gas analyzer to sample combustible gas(es).
Teledyne API Model 465L O
3
Monitor Operation Manual
iv Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
CONSIGNES DE SÉCURITÉ
Des consignes de sécurité importantes sont fournies tout au long du présent manuel
dans le but d’éviter des blessures corporelles ou d’endommager les instruments.
Veuillez lire attentivement ces consignes. Chaque consigne de sécurité est
représentée par un pictogramme d’alerte de sécurité; ces pictogrammes se retrouvent
dans ce manuel et à l’intérieur des instruments. Les symboles correspondent aux
consignes suivantes :
AVERTISSEMENT : Risque de choc électrique
DANGER : Oxydant puissant
AVERTISSEMENT GÉNÉRAL / MISE EN GARDE : Lire la consigne
complémentaire pour des renseignements spécifiques
MISE EN GARDE : Surface chaude
Ne pas toucher : Toucher à certaines parties de l’instrument sans
protection ou sans les outils appropriés pourrait entraîner des dommages
aux pièces ou à l’instrument.
Pictogramme « technicien » : Toutes les opérations portant ce symbole
doivent être effectuées uniquement par du personnel de maintenance
qualifié.
Mise à la terre : Ce symbole à l’intérieur de l’instrument détermine le
point central de la mise à la terre sécuritaire de l’instrument.
MISE EN GARDE
Cet instrument doit être utilisé aux fins décrites et de la manière décrite
dans ce manuel. Si vous utilisez cet instrument d’une autre manière que
celle pour laquelle il a été prévu, l’instrument pourrait se comporter de
façon imprévisible et entraîner des conséquences dangereuses.
NE JAMAIS utiliser un analyseur de gaz pour échantillonner des gaz
combustibles!
Teledyne API Model 465L O
3
Monitor Operation Manual
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual v
ABOUT THIS MANUAL
This manual is comprised of multiple documents as listed below.
Part
Name/Description
465L Operation Manual (this manual)
Rack Mount Spare Parts List (integrated into this manual as Section 12.1)
NEMA Spare Parts List (integrated into this manual as Section 12.2)
NOTE
We recommend that this manual be read in its entirety before any attempt is made to operate the
instrument.
Teledyne API Model 465L O
3
Monitor Operation Manual
vi Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
Intentionally left blank.
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual vii
TABLE OF CONTENTS
SAFETY MESSAGES .................................................................................................... iii
ABOUT THIS MANUAL .................................................................................................. v
1. INTRODUCTION .................................................................................................... 11
2. SPECIFICATIONS AND AGENCY APPROVALS .................................................. 13
2.1. SPECIFICATIONS ................................................................................................ 13
2.2. APPROVALS ...................................................................................................... 14
2.2.1. Safety ..........................................................................................................................14
2.2.2. EMC ...........................................................................................................................14
2.2.3. Other Type Certifications ..........................................................................................15
3. PRINCIPLE OF OPERATION ................................................................................ 17
4. GETTING STARTED .............................................................................................. 19
4.1. WIRING REQUIREMENTS FOR NEMA ENCLOSURES .............................................. 20
4.2. WIRING INSTRUCTIONS ...................................................................................... 21
4.3. PNEUMATIC CONNECTIONS ................................................................................. 22
4.3.1. Sample Gas Connection .............................................................................................24
4.3.2. Exhaust Connection ...................................................................................................28
4.4. ELECTRICAL I/O CONNECTIONS .......................................................................... 29
4.4.1. Location of I/O Connectors in the NEMA Configuration ..........................................29
4.4.2. Analog Output ............................................................................................................29
4.4.3. Status Outputs ............................................................................................................29
4.4.4. Hi-Current Relay Outputs ..........................................................................................31
4.4.5. Multi-Stream Relay Outputs (Optional) ....................................................................32
4.4.6. Digital Serial Interfaces ............................................................................................33
5. FREQUENTLY ASKED QUESTIONS .................................................................... 35
6. OPTIONS ................................................................................................................ 37
6.1. STREAM SELECTOR (3 OR 6 STREAM) ................................................................. 37
6.2. DISPOSABLE SAMPLE INLET FILTERS ................................................................... 37
6.3. RACK MOUNT SLIDES ........................................................................................ 37
6.4. ELECTRONIC FLOWMETER .................................................................................. 37
6.5. SAMPLE CONDITIONER ....................................................................................... 37
6.6. PERMEATION GAS DRYER .................................................................................. 37
7. OPERATION........................................................................................................... 39
7.1. FRONT PANEL OVERVIEW .................................................................................. 39
7.1.1. Ozone Concentration Display ....................................................................................39
7.1.2. Menu Navigation Keypad ..........................................................................................39
7.1.3. Stream Selector Status/Control (Optional) ................................................................40
Teledyne API Model 465L O
3
Monitor Operation Manual
viii Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
7.1.4. Concentration Alarm Status/Reset .............................................................................40
7.1.5. Status LEDs ...............................................................................................................41
7.2. FRONT PANEL MENUS ....................................................................................... 41
7.2.1. Menu Navigation Basics ............................................................................................41
7.2.2. Editing Numeric Values .............................................................................................42
7.2.3. DIAG Menu ................................................................................................................45
7.2.4. VIEW Menu ................................................................................................................50
7.2.5. CONFIG Menu ..........................................................................................................50
7.2.6. SETUP-VARS Menu ...................................................................................................51
7.2.7. SETUP-STREAM Menu .............................................................................................52
7.3. MULTI-STREAM OPERATION ............................................................................... 53
8. DIGITAL COMMUNICATIONS ............................................................................... 55
8.1. DIGITAL COMMUNICATIONS OVERVIEW ................................................................ 55
8.2. CHANGING COM PORT HARDWARE CONFIGURATION ............................................ 55
8.3. RS-232 CONFIGURATION ................................................................................... 57
8.4. RS-485 CONFIGURATION ................................................................................... 58
8.5. ETHERNET CONFIGURATION ............................................................................... 59
8.5.1. Hardware Setup for Configuring the Ethernet Module .............................................59
8.5.2. Verifying Network Hardware Connection .................................................................59
8.5.3. Determining a Dynamically Assigned IP Address (DHCP Mode) ............................60
8.5.4. Manually Assigning a Static IP Address ...................................................................60
8.6. MODBUS REGISTER MAPS ............................................................................... 61
8.6.1. MODBUS Commands Supported ..............................................................................61
8.6.2. Notes on MODBUS registers .....................................................................................61
8.6.3. Real-Time Concentration and Status Registers (Read Only) ....................................62
8.6.4. Instrument Setup and Configuration Registers (Read / Write) ..................................63
9. MAINTENANCE AND ADJUSTMENTS ................................................................. 65
9.1. MAINTENANCE SCHEDULE .................................................................................. 65
9.2. INSTRUMENT LAYOUT ......................................................................................... 66
9.3. REPLACING INTERNAL PARTICULATE FILTER ........................................................ 67
9.4. UV LAMP ADJUSTMENT ...................................................................................... 68
9.5. UV LAMP REPLACEMENT ................................................................................... 69
9.6. SAMPLE PUMP REPLACEMENT ............................................................................ 70
9.7. SENSOR VALVE REPLACEMENT .......................................................................... 71
9.8. WATER DROP-OUT FILTER (OPTION) MAINTENANCE ........................................... 72
9.9. PARTICULATE FILTER REPLACEMENT .................................................................. 72
9.10. CONFIGURING THE ANALOG OUTPUT ............................................................... 72
9.11. CLEANING EXTERIOR SURFACES OF THE 465L ................................................. 73
9.12. BOOT LOADER OPERATION ............................................................................. 73
9.12.1. Boot Loader – Hardware Setup ..............................................................................74
9.12.2. Boot Loader – Software Setup.................................................................................74
Teledyne API Model 465L O
3
Monitor Operation Manual
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual ix
10. TROUBLESHOOTING ....................................................................................... 77
10.1. REFERENCE DRAWINGS ................................................................................. 77
10.1.1. Pneumatic Diagram ................................................................................................77
10.1.2. Interconnect Diagram .............................................................................................78
10.2. TROUBLESHOOTING WITH FRONT PANEL STATUS LED’S OR STATUS OUTPUTS ... 79
10.2.1. Sensor OK ...............................................................................................................80
10.2.2. Invalid Reading .......................................................................................................80
10.2.3. Check Lamp .............................................................................................................80
10.2.4. Pneumatic Error ......................................................................................................80
11. CALIBRATION ................................................................................................... 83
11.1. FACTORY CALIBRATION .................................................................................. 83
11.2. ZERO CALIBRATION ........................................................................................ 83
12. SPARE PARTS .................................................................................................. 85
12.1. 465L RACK MOUNT SPARE PARTS LIST ........................................................... 85
12.2. 465L NEMA SPARE PARTS LIST .................................................................... 86
FIGURES
Figure 4-1: Mounting Hole Locations and Dimensions .......................................... 20
Figure 4-2: NEMA AC Power Connection to Monitor .............................................. 21
Figure 4-3: Pneumatic Connections, Rack Mount Configuration .......................... 22
Figure 4-4: Pneumatic Connections, NEMA Configuration, Single Stream .......... 23
Figure 4-5: Pneumatic Connections, NEMA Configuration, Multi-Stream ............. 23
Figure 4-6: 465L Pneumatic Diagram with Sample Conditioner and Permeation
Gas Dryer Options (in single-stream configuration) ................................. 25
Figure 4-7: Rear Panel Electrical I/O Connections .................................................. 28
Figure 4-8: NEMA Electrical I/O Connections .......................................................... 28
Figure 4-9: Status Output Schematic ....................................................................... 30
Figure 7-1: Front Panel Display ................................................................................ 39
Figure 7-2: Front Panel Menu Diagram (Page 1)...................................................... 43
Figure 7-3: Front Panel Menu Diagram (Page 2)...................................................... 44
Figure 7-4: Span Cal Menu ........................................................................................ 49
Figure 7-5: Flow Cal Menu ......................................................................................... 49
Figure 8-1: Com Configuration Jumper Settings (SW2) ......................................... 56
Figure 9-1: Instrument Layout, Rack Mount Configuration .................................... 66
Figure 9-2: Instrument Layout, NEMA Configuration .............................................. 67
Figure 9-3: Sensor Detail ........................................................................................... 71
Teledyne API Model 465L O
3
Monitor Operation Manual
x Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
Figure 9-4: Mainboard Analog Output Configuration ........................................... 73
Figure 10-1: Pneumatic Diagram, 3 Stream Configuration ..................................... 77
Figure 10-2: Interconnect Diagram ........................................................................... 78
TABLES
Table 4-1: Status Output Definitions ..............................................................30
Table 4-2: Relay Functions for Single Stream Config ...................................31
Table 4-3: Relay Functions for Multi-Stream Configs ...................................31
Table 4-4: Multi-Stream Relays .......................................................................32
Table 7-1: Alarm LED’s .....................................................................................40
Table 7-2: Status LED’s ...................................................................................41
Table 7-3: Analog Step Test Values ................................................................45
Table 7-4: Signal I/O List .................................................................................46
Table 7-5: VIEW Menu Parameters .................................................................50
Table 7-6: CONFIG Menu Parameters .............................................................50
Table 7-7: VARS List ........................................................................................51
Table 7-8: STREAM Properties ........................................................................52
Table 8-1: RS-232 Connector Details ..............................................................57
Table 8-2: RS-232 Port Setup ..........................................................................57
Table 8-3: RS-485 Connector Details ..............................................................58
Table 8-4: RS-485 Port Setup ..........................................................................58
Table 8-5: Ethernet LED Definitions ...............................................................59
Table 8-6: Supported MODBUS Commands ..................................................61
Table 9-1: Maintenance Schedule ...................................................................65
Table 9-2: Required Software ..........................................................................74
Table 10-1: Status LED/Output Definitions ....................................................79
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual 11
1. INTRODUCTION
Teledyne API’s Model 465L (or the 465L) is a UV photometric ozone monitor designed and
manufactured for single/multi-channel industrial hygiene ozone analysis. It is a very stable
analyzer, ideal for critical low-level PPM safety and process applications. Whether you have
selected the rack mount or wall mount unit in NEMA 4X enclosure, the 465L is designed to be
highly reliable and simple to operate with easy setup and user-friendly menudriven software that
can also be set and monitored remotely.
As the 465L was designed for stability, reliability and simplicity, it includes built-in tests and
diagnostics to facilitate maximum uptime. Teledyne API is pleased to provide you with any
support required so that you may utilize our equipment to the fullest extent. Our full time
technical support department is available to answer your questions.
Teledyne API, Technical Support,
9970 Carroll Canyon Road
San Diego, California 92131-1106 USA
Toll-free Phone: 800-324-5190
Phone: +1 858-657-9800
Fax: +1 858-657-9816
Email: api-techsuppo[email protected]
Website: http://www.teledyne-api.com
Teledyne API Model 465L O
3
Monitor Operation Manual
12 Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
This page intentionally left blank.
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual 13
2. SPECIFICATIONS AND AGENCY
APPROVALS
2.1. Specifications
PARAMETER
DESCRIPTION
Ranges
User selectable to any full scale range from:
1 PPM to 500 PPM
1000 PPB to 10,000 PPB
2000 µg/m
3
to 10,000 µg/m
3
100 PPHM to 10,000 PPHM (optional configuration)
Low conc scrubber configuration: 0 to 20 PPM
High conc scrubber configuration: >20 PPM
Measurement Units
PPM, PPB, µg/m
3
, PPHM (optional configuration)
Zero Noise
< .0015 PPM (rms)
Span Noise
< .5% of reading (rms) (above 0.1 PPM)
Lower Detectable Limit
< .003 PPM (rms)
Accuracy
+/- 1% of full scale range
Response Time (95%)
<30 sec
(<60 sec with the Sample Conditioner option)
Sample Flow Rate
0.8 L/min
Temperature Range
5-45
o
C
Humidity Range
10-90% RH, Non-Condensing
Dimensions (H x W x D)
5.22” x 19.0" x 15.3" (133 mm x 483 mm x 388 mm)
– Rack Mount Version
16.85" x 15.60 (±0.25)" x 6.65" (428mm x 396
(±6.35)mm x 169mm) – with NEMA 4X enclosure,
typical product envelope
Weight
12.6 lb. (5.73 kg) - Rack Mount Version
15.2 lb. (6.91 kg) – with NEMA 4X enclosure
Power
100-240 V~, 47-63 Hz, 74 W
Environmental Conditions
Installation Category (Overvoltage Category) II
Pollution Degree 2
For Indoor Use Only
Maximum Operating Altitude 2000 meters
Analog Output, Voltage Mode
0 - 5V
Teledyne API Model 465L O
3
Monitor Operation Manual
14 Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
PARAMETER
DESCRIPTION
Isolated Analog Output,
4-20mA Mode
Maximum voltage between outputs and ground 60V
peak
System OK Status Relay
SPDT (Form C) Dry Contact, 250 VAC, 5A
Global Alarm Relays
SPDT (Form C) Dry Contact, 250 VAC, 5A (x2, HI
and HI-HI)
Instrument Status Outputs
6 - SPST Dry Contact, 50VDC, 250mA Resistive
Load
Multi-Stream Specifications:
PARAMETER
DESCRIPTION
Minimum Stream Duration
(Cycle Time)
1.0 min
Stream Alarm Relays (HI and
HI-HI)
SPST Dry Contact, 50VDC, 250mA Resistive
Load (2 per stream)
Stream ID Relays
SPST Dry Contact, 50VDC, 250mA Resistive
Load (1 per stream)
2.2. Approvals
This section presents Safety and Electromagnetic Compatibility (EMC) compliance
approvals for the Model 465L monitor.
2.2.1. Safety
IEC 61010-1:2001 (3
rd
Edition), Safety requirements for electrical
equipment for measurement, control, and laboratory use.
CE: 2006/95/EC, Low-Voltage Directive
2.2.2. EMC
EN 61326-1 (IEC 61326-1), Class A Emissions/Industrial Immunity
EN 55011 (CISPR 11), Group 1, Class A Emissions
FCC 47 CFR Part 15B, Class A Emissions
CE: 2004/108/EC, Electromagnetic Compatibility Directive
Teledyne API Model 465L O
3
Monitor Operation Manual
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual 15
2.2.3. Other Type Certifications
For additional certifications, please contact Technical Support:
Toll-free: 800-324-5190
Phone: +1 858-657-9800
Fax: +1 858-657-9816
Email: api-techsuppo[email protected]
Teledyne API Model 465L O
3
Monitor Operation Manual
16 Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
This page intentionally left blank.
05509K DCN8017 Teledyne API Model 465L O
3
Monitor Manual 17
3. PRINCIPLE OF OPERATION
The detection of ozone molecules is based on absorption of 254 nm UV light due to
an internal electronic resonance of the O
3
molecule. The Model 465L uses a
mercury lamp constructed so that a large majority of the light emitted is at the
254nm wavelength. Light from the lamp shines down a hollow quartz tube that is
alternately filled with sample gas, then filled with gas scrubbed to remove ozone.
The ratio of the intensity of light passing through the scrubbed gas to that of the
sample forms a ratio I/I
o
. This ratio forms the basis for the calculation of the ozone
concentration.
The Beer-Lambert equation, shown below, calculates the concentration of ozone
from the ratio of light intensities.
o
o
O
inHg
C
Ι
Ι
×
Ρ
×
Κ
Τ
×
×
= ln
92
.29
273
10
9
3
α
Where:
I = Intensity of light passed through the sample
I
o
= Intensity of light through sample free of ozone
α
= absorption coefficient
= path length
3
O
C
= concentration of ozone in ppb
T = sample temperature in degrees Kelvin
P = pressure in inches of mercury
As can be seen the concentration of ozone depends on more than the intensity ratio.
Temperature and pressure influence the density of the sample. The density changes
the number of ozone molecules in the absorption tube which impacts the amount of
light removed from the light beam. These effects are addressed by directly
measuring temperature and pressure and including their actual values in the
calculation. The absorption coefficient is a number that reflects the inherent ability
of ozone to absorb 254 nm light. Most current measurements place this value at 308
cm
-1
atm
-1
at STP. The value of this number reflects the fact that ozone is a very
efficient absorber of UV radiation which is why stratospheric ozone protects the life
forms lower in the atmosphere from the harmful effects from solar UV radiation.
Lastly, the absorption path length determines how many molecules are present in the
column of gas in the absorption tube.
Teledyne API Model 465L O
3
Monitor Operation Manual
18 Teledyne API Model 465L O
3
Monitor Manual 05509K DCN8017
The intensity of light is converted into a voltage by a high resolution A/D (analog-
to-digital) converter. The digitized signal and other variables are used by the CPU to
compute the concentration using the above formula.
About every 2.5 seconds the 465L completes a measurement cycle consisting of a 1
second wait period for the sample tube to flush, followed by a 150 ms measurement
of the UV light intensity to obtain I. The sample valve is switched to admit scrubbed
sample gas for 1 second, followed by a 150 ms measurement of the UV light
intensity to obtain I
o
. Measurement of the I
o
every 2.5 seconds eliminates
instrument drift due to changing intensity of the lamp caused by aging and dirt.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91

Teledyne API 465L User manual

Category
Measuring, testing & control
Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI