Mettler Toledo Test Procedure for Moisture Analyzer Routine Testing Operating instructions

Type
Operating instructions
Routine Testing
SOP for Periodic SmartCal
Test
of Moisture Analyzers
Title: SOP for Periodic SmartCal test with Halogen
Moisture Analyzers
Document No.:
Author: METTLER TOLEDO Laboratory & Weighing Technologies
Number of Pages: 4
Controls
Replacement Document: N/A
Reason for Revision: N/A
Release Date: 21. Aug 2013
Release for Routine Operation
Reviewed by:
Date:
Signature:
Approved by:
Date:
Signature:
2 of 4
SOP SmartCal
METTLER TOLEDO
2 of 4
Routine Testing of Moisture Analyzers
General
The SmartCal test is a rapid test which checks the moisture analyzer’s overall functionality. SmartCal is a
granular substance which releases a defined amount of moisture depending on the selected drying tempera-
ture. Intended use is the verification of METTLER TOLEDO Halogen Moisture Analyzers.
Basic Rules for Handling Moisture Analyzers
•Beforeusingamoistureanalyzer,makesuretheinstrumentwasleftonpowerforasufficientperiodoftime
(mentioned in the user manual).
•Makesurethemoistureanalyzerisleveled.
•Minimizeenvironmentalinfluences,e.g.draftordirectsunlight.
•Distributesamplesgentlyandevenlyonthesamplepan.
Basic Rules for Handling SmartCal
Important to know
•Thegranulartestsubstanceisdouble-packedinastickinsideablisterpack.
•Thestickshouldonlyberemovedfromtheblisterpackimmediatelybeforeuse.
•Donotuseanydamagedorexpiredsticks.
•AftertheSmartCaltest,theusedtestsubstancecanbedisposedofasnormalwaste.
How to Store SmartCal
•SmartCalshouldbestoredinthesameroomasthemoistureanalyzertheyareusedwith,sincetemperature
differences between the test substance and their surroundings can lead to measurement errors.
How to Handle SmartCal
• TheSmartCaltestresultrequiresanormalizationtoambientconditions(%RHandroomtemperature).
• WhenperformingaSmartCaltest,usethesamplepanhandler.
• Usethecompletecontentsofthestick(approx.weight8.5g).
3 of 4
SOP SmartCal
METTLER TOLEDO
3 of 4
1)
– Values within the control limit: no action is necessary.
–Valuesbeyondthecontrollimitshowthatthemoistureanalyzerisnolongerundercontrolandimmediateactionisthereforerequired.
SmartCal Routine Test
Preparation
•Beforethetestisperformed,SmartCaltestsubstanceandthethermohygrometermustbeacclimatizedtothe
ambient temperature of the moisture analyzer.
•Whenthetestisperformed,theheatingunitofthemoistureanalyzershouldbeatroomtemperature.
•Ifasensitivitytestneedstobedoneaswell,pleaseperformthesensitivitytestfirst(toavoidwaitingtimefor
cooling down).
•UseinstrumentpredefinedtestproceduresorsetinstrumenttotheSmartCaltestparameters:
–Switch-offtime:10minutes
– Drying program: standard drying
–Display:%MC
–Dryingtemperature:70,100,130or160°C
(Select the temperature closest to the drying temperature normally used)
Test Procedure
•Ifrequiredbyinstrumentmenuenterambienttemperatureandambienthumidity.
•Placethesamplepanhandlerwiththealuminumsampledishintheinstrumentandtare.
•RemoveaSmartCalstickfromtheblisterpack,openitanddistributetheentirecontentsevenlyoverthe
sample pan (whole pan covered with granules).
•Startthetestmeasurementimmediatelyafterdistributingthesample.
•Readthedisplayedresult(%MC,%MC
N
) from the instrument.
•IftheresultisshownasMC%recordambienttemperatureandambienthumidityaswell.
Evaluation
•Iftheunitofthedisplayedresultis%MC
N
,nonormalizationisneeded.Iftheunitofthedisplayedresultis
MC%normalizethedisplayedresult–usingtheambientconditionsrecordedabove–eithermanuallywith
normalizationtableorwiththeExcelmeasurementreport.
•Evaluatewhetherthenormalizedresultexceedsthedefinedcontrollimits.
1)
Deviation
Control Limit
1)
•Ifthecontrollimitisexceeded,reduceerrorsources(seeSmartCalusermanual)andrepeatthetest
(after cooling down of the moisture analyzer).
•Ifthecontrollimitisexceededagain,reportsupervisorthatthecontrollimitwasnotmet.
•Letthemoistureanalyzercooldown,performweightandthentemperatureadjustment,letthemoisture
analyzer cool down and repeat the SmartCal test.
•Ifthecontrollimitisstillexceeded,reporttheproblemtothesupervisor.Optionally,contactMETTLER
TOLEDO’s service organization for advice.
•Markthemoistureanalyzeras“outofcontrollimits”.
4 of 4
SOP SmartCal
METTLER TOLEDO
Recommended Control Limits for SmartCal Test
Control limits are valid for METTLER TOLEDO Halogen Moisture Analyzers.
Drying Temperature cSmartCal SmartCal
70°C 3.3-4.3%MC
N
3.2-4.4%MC
N
100°C 5.3-6.3%MC
N
5.2-6.4%MC
N
130°C 7.5-8.7%MC
N
7.4-8.8%MC
N
160°C 10.0-11.6%MC
N
9.9-11.7%MC
N
Mettler-Toledo AG
Laboratory & Weighing Technologies
CH-8606Greifensee
Tel. +41 44 944 22 11
Fax.+41449443170
Subjecttotechnicalchanges
©11/2013Mettler-ToledoAG
PrintedinSwitzerland30031480
GlobalMarComSwitzerland
Routine Testing of Moisture Analyzers
For more information
www.mt.com/moisture
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4

Mettler Toledo Test Procedure for Moisture Analyzer Routine Testing Operating instructions

Type
Operating instructions

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI