Cianna Medical SCOUT Check Quick start guide

  • Hello! I am an AI chatbot trained to assist you with the Cianna Medical SCOUT Check Quick start guide. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
SCOUT
®
Check
2
Overview
SCOUT Check SAVI SCOUT System
Console housing Same Same
Console front face Gray Black
Sound Constant cadence no
change with proximity
Increasing cadence with
proximity
Visual LED Lights for
modes/messages
Display with numbers,
modes/messages
Volume control None Up/down
Handpiece Reusable, gray cable Single use, white cable
3
System Setup: Components
Console
Handpiece
Reusable
4
Console Overview
Start/Stop
Button
Handpiece
receptacle
connector
SYSTEM READY
Indicator
DETECT MODE
Indicator
HOLD STEADY
Indicator
5
System Setup: Console
Place console on top of cart or other equipment
Position console to accommodate 5 ft. handpiece cable
(i.e. side of patient table vs. foot of table)
Attach power cord to console and plug into outlet
Power on/off
switch
Power cord
receptacle
Fuse Drawer
6
System Setup: Handpiece
Attach handpiece to console
Align arrow on handpiece connector with notch on console port
Press until secure connection made between handpiece and console
Handpiece
receptacle
connector
Alignment
arrow
Alignment
notch
7
System Setup: Power up
Turn on ‘Power’ switch on back of
console
START/STOP button on front of
console will activate
Wait for ‘SYSTEM READY’ indicator
light: green
Blinking ‘Ready’ light = error
Check handpiece connections
Turn console off, then on again
If blinking persists, attach new
handpiece
Power on/off
switch
8
Press ‘START/STOP’ button to activate
handpiece
‘DETECT MODE’ indicator: Green
Place handpiece on Handpiece Test Card as
indicated
Positive test audible clicking
Negative test no sound
Replace with new handpiece
System Setup: Handpiece Test
System is now ready for reflector verification
9
Post-Reflector Verification
System shut down
Press START/STOP button
Turn off power switch
Disconnect handpiece and store in safe place
Unplug power cord if console will be moved or stored
Cleaning
Console: wipe all console surfaces, including power cord with a
mild cleaning solution or disinfectant and a damp cloth
Handpiece (see Handpiece IFU for full cleaning instructions)
Wipe off obvious contaminants
Use moistened wipe (CaviWipe) to remove any remaining contaminants
or gross debris
For use as a disinfectant, use second CaviWipe to thoroughly wet
surface
/