General 1-9
kina40e 01.fm technical handbook(1007)
1.6 Installation
1.6.1 Unpacking
Unpack the HLD5000 immediately after delivery, even if it is to be installed later on.
Examine the shipping container for any external damage. Completely remove the
packaging materials.
Notice Retain the shipping container and the packaging materials in the event of
complaints about damage.
Check that the HLD5000 is complete and carefully examine the HLD5000 visually.
If any damage is discovered, report it immediately to the forwarding agent and
insurer. If the damaged part has to be replaced, please contact the orders
department.
Remove the protection foil from the display.
After switching on the HLD5000 the programmed type of gas has to appear in the
upper left corner of the display and has to match with the sticker on the probe.
1.6.2 Mechanical Connections
In order to operate the HLD5000, a sniffer line must be connected. The connection
(Fig. 2-1/2) is located on the left side on the front of the main unit. Insert the plug into
the opening until it engages. For this the red mark on the plug must correspond with
the mark on the main unit.
To disconnect the plug, retract the coupling and remove the probe’s line.
1.6.3 Electrical Connection
The HLD5000 is equipped with a wide voltage range power supply unit covering 100
V to 230 V (± 10 %, 50/60 Hz). The mains cord is connected to the mains socket (Fig.
2-1/7) on the rear of the main unit. A fuse (Fig. 2-1/8) for each conductor of the mains
cord has been integrated in the mains socket (Fig. 2-1/7) of the main unit.
1.6.4 Wall Mounting
A fixture at the bottom of the HLD5000 (Fig. 2-2/7) allows mounting the leak detector
on walls, (for example due to space restrictions). It is recommended to place the
HLD5000 high enough (at least 2 m) so that the display points downwards.
Only mains cords having three conductors with a protective ground conductor may
be used. The HLD5000 must never be operated with a disconnected protective
ground conductor.