1) Turn pan machine OFF.
2) Remove the cover.
3) Check Voltage at connector XC201.
a) Turn machine OFF
b) set DMM to DC Volts (ensure DMM is capable of measuring >300
VDC).
c) Connect probes to XC201.
d) Power machine ON.
e) Enter Service Mode and go to Service Test 10 (see Chapter 8 of the
CDR PanX Installation and Service Guide PN: B1051402).
f) Stand clear of X-ray path and energized components.
g) Press and hold the exposure switch momentarily.
h) Check DMM for proper reading of about 300 VDC.
If voltage is significantly low, the cable between the Power Supply Board
and Power Switching Unit may be bad.
If the voltage is correct then
3) If the voltage is correct then replace the Power Switching Unit board
(Schick PN: A4504000). Be aware of proper rewiring and cable replacement.
4) Turn machine ON and retest as above.
5) If the error still appears then replace the Tubehead.
Tubehead Replacement
1) Turn machine OFF
2) Remove Tubehead covers.
3) Loosen locknut on Tubehead shaft.
4) Back out alignment pin from alignment disk.
5) Remove disk screw.
6) Remove disk by gently prying it up.
7) Remove Tubehead wiring and shield clamps. Make a note of wiring
for proper replacement.
Loosen locknut and
back out alignment
pin. These fasteners
are metric.
Remove disk
screw and gently
pry the disk off
the Tubehead