Microsonic zws-15-CI-QS User manual

Type
User manual

The Microsonic zws-15-CI-QS ultrasonic proximity switch provides non-contact distance measurement and features one analog output. With a range of 4 cm to 35 cm, it offers a resolution of 0.20 mm and accuracy of ±0.15%. The device includes a Teach-in function for easy adjustment of the analog output and characteristic curve. It has two LEDs for operation and output status indication.

The Microsonic zws-15-CI-QS ultrasonic proximity switch provides non-contact distance measurement and features one analog output. With a range of 4 cm to 35 cm, it offers a resolution of 0.20 mm and accuracy of ±0.15%. The device includes a Teach-in function for easy adjustment of the analog output and characteristic curve. It has two LEDs for operation and output status indication.

Product Description
The zws sensor offers a non-contact
measurement of the distance to an
object which must be positioned
within the sensor’s detection zone. In
dependence of the set window limits,
a distance-proportional analogue sig-
nal is output.
Via the push-button, the window li
-
mits of the analogue output and its
characteristic can be adjusted (Teach-
in). Two LEDs indicate operation and
the state of the analogue output.
Safety Notes
Ԏ
Read the operating instructions
prior to start-up.
ԎConnection, installation and ad-
justment works may only be car-
ried out by expert personnel.
Ԏ
No safety component in ac-
cordance with the EU Machine
Directive
Use for intended purpose only
zws ultrasonic sensors are used for
non-contact detection of objects.
Installation
Î
Mount the sensor at the installation
site with the aid of the enclosed
mounting plate (see Fig. 1). Maxi-
mum torque of attachment screw:
0,5 Nm
ÎConnect a connection cable to the
M8 device plug (see Fig. 2).
ÎAvoid mechanical load on the con-
nector.
Start-Up
ÎConnect the power supply.
Î
Carry out the adjustment in ac-
cordance with Diagram 1.
Factory Setting
Ԏ
Rising analogue characteristic curve
between the blind zone and the
operating range
Fig. 1: Attachment with mounting plate
1
2
3
4
colour
1 +UBbrown
3 –UBblue
4 U/I black
2 sync white
Fig. 2: Pin assignment with view onto sensor
plug and colour coding of the microso-
nic connection cable
Synchronisation
You can synchronise as many sensors
as you like.
ÎApply a square-wave signal to the
sync-input with pulse width ti and
repetition rate tp (Fig. 3 and techni-
cal data).
A high level on the sync-input will
disable the sensor.
Sensor disabled Sensor disabled
Start
+U
B
–U
B
tp
ti
Fig. 3: External synchronisation signal
Checking operation mode
Î
In normal operating mode shortly
press the push-button. The green
LED stops shining for one second,
then it will show the current cha-
racteristic of the analogue out-
put:
Ԏ1x flashing = rising
Ԏ2x flashing = falling
Maintenance
microsonic sensors are maintenance-
free. In case of excess caked-on dirt
we recommend cleaning the white
sensor surface.
Notes
Ԏ
Every time the power supply is swit-
ched on, the sensor detects its ac-
tual operating temperature and
transmits it to the internal tempera
-
ture compensation. This results in a
slight correction of the analogue
output value after 45 seconds.
Ԏ
If the sensor was switched off for at
least 30 minutes and after power
on an object is placed in the middle
of the adjusted analogue window
for 30 minutes (the analogue out-
put value is in the range of 11 to
13 mA or 4.4 to 5.6 V) a new ad-
justment of the internal tempera-
ture compensation to the actual
mounting conditions takes place.
Ԏ
The zws sensor has a blind zone,
within which distance measure-
ments are not possible.
ԎIn the normal operating mode, an
illuminated yellow LED signals the
object is within the adjusted win-
dow limits.
Ԏ
If the push-button is not pressed for
30 seconds during the teach-in set-
ting, the settings made hitherto are
deleted.
ԎThe sensor can be reset to its facto-
ry setting (see »Further settings«,
Diagram 1).
Operating Manual
Ultrasonic proximity switch
with one analogue output
zws-15/CI/QS zws-15/CU/QS
zws-24/CI/QS zws-24/CU/QS
zws-25/CI/QS zws-25/CU/QS
zws-35/CI/QS zws-35/CU/QS
zws-70/CI/QS zws-70/CU/QS
Diagram 1: Set sensor parameters via Teach-in procedure
Reset to factory setting
Wait for 10 s.
Normal operating mode
Further settings
Switch off
operating voltage.
Switch on
operating voltage.
Switch on
operating voltage.
Switch off
operating voltage.
Enable/disable Teach-in
push-button
To enable/disable Teach-in
press push-button
for about 1 s.
Press and hold push-button. Press and hold push-button.
Keep push-button pressed
for about 3 s until both
LEDs flash simultaneously.
Keep push-button pressed
for about 13 s until both
LEDs stop flashing.
flashesgreen LED:
on:
off:
push-button
enabled
push-button
disabled
yellow LED:
Set window limits Set rising/falling output
characteristic curve
Wait for about 10 s.
off: falling
characteristc
curve
on: rising
Normal operating mode
Set analogue output
Press push-button for
about 13 s until LEDs
flash alternately.
Press push-button
for about 1 s.
To change output
characteristic press
push-button for about 1 s.
Green LED:Both LEDs: flashesflash
alternately
Yellow LED:
Place object at position .
1
Place object at position .
2
Press push-button for
about 3 s, until LEDs
flash simultanously.
1 2
microsonic GmbH / Phoenixseestraße 7 / 44263 Dortmund / Germany / T +49 231 975151-0 / F +49 231 975151-51 / E [email protected] / W microsonic.de
The content of this document is subject to technical changes. Specifications in this document are presented in a descriptive way only. They do not warrant any product features.
Technical Data
+UB
–UB
Sync
1
4
2
3
1 analogue output
UU/I
RL
zws-15... zws-24... zws-25... zws-35... zws-70...
4.2
17.6
3.9
4
24
32
21.6
M8x1
12
Teach-in button
LEDs
3.2
2.6
10.7
4.2
17.6
3.9
4
24
32
23.1
M8x1
12
Teach-in button
LEDs
3.2
2.6
10.7
44.2
18.3
24
20.1
15.8
3.9 32
18
M8x1
12
23
Teach-in button
LEDs
2.6
10.7
9.4
44.2
18.3
24
20.1
15.8
3.9 32
18
M8x1
12
18
Teach-in button
LEDs
2.6
10.7
9.4
44.2
18.3
24
20.1
15.8
3.9 32
18
M8x1
12
18
Teach-in button
LEDs
2.6
10.7
9.4
blind zone 20 mm 50 mm 30 mm 64 mm 120 mm
operating range 150 mm 240 mm 250 mm 350 mm 700 mm
maximum range 250 mm 350 mm 350 mm 600 mm 1000 mm
angle of beam spread see detection zone see detection zone see detection zone see detection zone see detection zone
transducer frequency 380 kHz 500 kHz 320 kHz 400 kHz 300 kHz
resolution 0.20 mm 0.20 mm 0.20 mm 0.20 mm 0.20 mm
reproducibility ±0.15 % ±0.15 % ±0.15 % ±0.15 % ±0.15 %
detection zones
for different objects:
The dark grey areas represent the
zone where it is easy to recognise
the normal reflector (round bar).
This indicates the typical opera-
ting range of the sensors. The
light grey areas represent the
zone where a very large reflector
– for instance a plate – can still
be recognised. The requirement
here is for an optimum alignment
to the sensor. It is not possible to
evaluate ultrasonic reflections
outside this area.
4 cm 0 cm 4 cm 8 cm8 cm
0 cm
4 cm
8 cm
12 cm
16 cm
20 cm
24 cm
Plate
Round bar ø 10 mm
10 cm 5 cm 0 cm 5 cm 10 cm
0 cm
5 cm
10 cm
15 cm
20 cm
25 cm
30 cm
35 cm
Plate
Round bar ø 10 mm
0 cm
5 cm
Plate
Round bar ø 10 mm
10 cm 0 cm 10 cm 20 cm
0 cm
10 cm
20 cm
30 cm
40 cm
50 cm
60 cm
Plate
Round bar ø 10 mm
20 cm 0 cm 20 cm 40 cm
0 cm
20 cm
40 cm
60 cm
80 cm
100 cm
Plate
Round bar ø 10 mm
accuracy ±1 %; ±1 %; ±1 %; ±1 %; ±1 %;
temperature drift internal compensated temperature drift internal compensated temperature drift internal compensated temperature drift internal compensated temperature drift internal compensated
operating voltage UB20 to 30 V DC, reverse polarity protection 20 to 30 V DC, reverse polarity protection 20 to 30 V DC, reverse polarity protection 20 to 30 V DC, reverse polarity protection 20 to 30 V DC, reverse polarity protection
voltage ripple ±10 % ±10 % ±10 % ±10 % ±10 %
no-load current consumption <25 mA <35 mA <25 mA <35 mA <35 mA
housing ABS ABS ABS ABS ABS
ultrasonic transducer: polyurethane foam, ultrasonic transducer: polyurethane foam, ultrasonic transducer: polyurethane foam, ultrasonic transducer: polyurethane foam, ultrasonic transducer: polyurethane foam,
epoxy resin with glass content epoxy resin with glass content epoxy resin with glass content epoxy resin with glass content epoxy resin with glass content
class of protection to EN 60529 IP 67 IP 67 IP 67 IP 67 IP 67
norm conformity EN 60947-5-2 EN 60947-5-2 EN 60947-5-2 EN 60947-5-2 EN 60947-5-2
type of connection 4-pin M8 initiator plug 4-pin M8 initiator plug 4-pin M8 initiator plug 4-pin M8 initiator plug 4-pin M8 initiator plug
controls Teach-in push-button Teach-in push-button Teach-in push-button Teach-in push-button Teach-in push-button
indicators LED green (operation) LED green (operation) LED green (operation) LED green (operation) LED green (operation)
LED yellow (state of output) LED yellow (state of output) LED yellow (state of output) LED yellow (state of output) LED yellow (state of output)
synchronisation external external external external external
pulse width synchronisation signal ti>150 µs >150 µs >150 µs >150 µs >150 µs
cycle time synchronisation signal tp8 ms < tp < 1 s 10 ms < tp < 1 s 10 ms < tp < 1 s 16 ms < tp < 1 s 14 ms < tp < 1 s
operating temperature –25 to +70 °C –25 to +70 °C –25 to +70 °C –25 to +70 °C –25 to +70 °C
storage temperature –40 to +85 °C –40 to +85 °C –40 to +85 °C –40 to +85 °C –40 to +85 °C
weight 10 g 10 g 11 g 11 g 11 g
response time 50 ms 50 ms 50 ms 80 ms 70 ms
switch-off delay time <300 ms <300 ms <300 ms <300 ms <300 ms
order no. zws-15/CI/QS zws-24/CI/QS zws-25/CI/QS zws-35/CI/QS zws-70/CI/QS
analogue output 4 to 20 mA RL 500 ΩRL 500 ΩRL 500 ΩRL 500 ΩRL 500 Ω
rising/falling characteristic rising/falling characteristic rising/falling characteristic rising/falling characteristic rising/falling characteristic
order no. zws-15/CU/QS zws-24/CU/QS zws-25/CU/QS zws-35/CU/QS zws-70/CU/QS
analogue output 0 to 10 V RL 100 kΩ, short-circuit-proof, RL 100 kΩ, short-circuit-proof, RL 100 kΩ, short-circuit-proof, RL 100 kΩ, short-circuit-proof, RL 100 kΩ, short-circuit-proof,
rising/falling characteristic rising/falling characteristic rising/falling characteristic rising/falling characteristic rising/falling characteristic
*B6285*
MV-DO-052890-763146
  • Page 1 1
  • Page 2 2

Microsonic zws-15-CI-QS User manual

Type
User manual

The Microsonic zws-15-CI-QS ultrasonic proximity switch provides non-contact distance measurement and features one analog output. With a range of 4 cm to 35 cm, it offers a resolution of 0.20 mm and accuracy of ±0.15%. The device includes a Teach-in function for easy adjustment of the analog output and characteristic curve. It has two LEDs for operation and output status indication.