1 Introduction
Which Ion Source—ESI or APCI—Is Better for Analyzing My Samples?
Thermo Electron Corporation Finnigan LXQ Getting Started 5
In ESI, the buffer type and buffer concentration both have a noticeable
effect on sensitivity. Therefore, it is important to choose these variables
correctly.
The ESI process is affected by aerosol droplet size, liquid surface tension,
solvent volatility, surface charge, ion solvation strength and solution
conductivity. Specifically, ESI is not favored by large droplets with high
surface tension, low volatility, low surface charge, strong ion solvation and
high conductivity. Conversely, ESI is favored by small droplets with low
surface tension, high volatility, high surface charge, weak ion solvation and
low conductivity.
Mixed organic/aqueous solvent systems that include organic solvents such
as methanol, acetonitrile, and isopropyl alcohol are recommended for ESI.
Using only water as a solvent is not recommended. Volatile acids and bases
give good results. Salt solutions with concentrations above 10 mM are not
recommended. Strong mineral acids and bases can damage the instrument
and should not be used.
To obtain good ESI follow these rules:
• Keep non-volatile salts and buffers out of the solvent system. For
example, avoid the use of salts containing sodium, potassium or
phosphate. Use ammonium or acetate salts instead.
• Use organic/aqueous solvent systems and volatile acids and bases.
• If possible, optimize the pH of the solvent system for your analyte of
interest. For example, if your analyte of interest contains a primary or
secondary amine, your mobile phase should be slightly acidic
(pH 2 to 5).
Using APCI/MS Like ESI, APCI is a soft ionization technique. APCI provides molecular
mass information for compounds of medium polarity that have some
volatility. APCI is typically used to analyze small molecules with molecular
masses up to about 2000 Da.
APCI is a gas phase ionization technique. Therefore, the gas phase acidities
and basicities of the analyte and solvent vapor play an important role in the
APCI process.
APCI is a very robust ionization technique. It is not affected by minor
changes in most variables such as changes in buffer type or buffer strength.
The rate of solvent flowing from the LC into the MS detector in APCI
mode is typically high (between 0.2 and 2 mL/min). Refer to Table 3,
Guidelines for setting operating parameters for LC/APCI/MS.