Reed Instruments 9802 Operating instructions

Type
Operating instructions
OPERATION MANUAL
3 IN ONE DATALOGGER PRINTER
MODEL: 9801 9802
Thank you for purchasing this 3 in one Hand Held
Datalogger Printer. This portable device is the
first 3 in one datalogger printer on the market.
FEATUREFEATURE
Datalogger programmer and printer
Print in graphic (2 modes) or text mode
USB port to PC link
Saving max. 4 logger memories
Powered by 4pcs AAA batteries
Optional 9VDC adaptor for long time use
Low battery warning
Datalogger printer
4pcs AAA batteries
One thermo paper roll
Instruction manual
Logger link cable
Software +USB cable
Optional:
-9VDC adaptor
-Datalogger
MATERIAL SUPPLYMATERIAL SUPPLY
PAPER FEEDINGPAPER FEEDING
Open the printer cover Insert the sharp-cut to
the printer Press if you cant
easily pull out the sharp-cut Done!
cartridge
POWER SUPPLY & PORT ILLUSTRATIONPOWER SUPPLY & PORT ILLUSTRATION
The meter is powered by 4pcs AAA batteries
or one 9VDC (>0.5A) adaptor.
When "Low Battery" appears on LCD: Turn off
the printer Replace the battery with correct
polarity and make good contact Done!
USB port
Connect to PC to download
the printer memories.
(Disabled in
9801/02)
9VDC
USB
Contact P oints
Contact Points
For loggers with two metal contacts to connect
with printer. Including: 88128/8828/8829/8813/
8814/8815/8833/8834/8835/88335/88345/88355.
KEYPADKEYPAD
Menu
Up
Down Enter Feed
LCD
display
Normal: Select the pervious/next option.
Modified Mode: increase/decrease the the value.
You may press and hold for quick jump.
Power On: Press to turn on the printer.
Normal : To turn to the next page.
Escape : Escape without saving.
Printing : Press the button to stop printing.
Power Off: Press for 3 seconds to turn off.
Normal : To enter modification mode or run a
command.
Modification Mode: Press to enter next digit or
end modification with saving.
Feed the paper when it's inactive except printing.
PRINT PAGE (PRN)
To custom made the
printing content you
need, press /
to highlight the
field which need to be
configured and then
press to enter the modification.
NOTE: If you want to print the data of logger
which is plugged in, the Download command
must run first.
CONTRAST
Valid values for Contrast
: 1 to 9 (light to heavy).
During printing, you can
press / to
change the contrast
any time.
To set the contrast, highlight the contrast field and
press to enter.
Press / to increase/decrease the value
of contrast.
Press to end the modification with saving
the value, or press to end without saving.
ALARM
This field allows you to set the alarm critical value
for printing. To set the alarm, highlight the Alarm
field and press to enter.
Press / to enable/disable the alarm.
Press to save the modification or press
to escape without saving.
If you select enable mode, alarm setting must be
edited.
Press / to
increase/decrease the
highlighted value, press
to highlight the
next. The alarm setting
will exit and save when
you press in the
last value.
Valid values for Air temp.
/external temp./RH%:
Air Hi:-40.0 to185
Air Lo: -40.0 to 185
Ext. Hi: -40.0 to 212
Ext. Lo: -40.0 to 212
%RH Hi: 0.0 to 100
%RH Lo: 0.0 to 100
MODE
This field allows you to set the printing mode. To
set the printing mode, highlight the Mode field and
press to enter modification.
Press / to select the printing mode.
Press to save or press to escape
the modification without saving. You can choose
Graph or Report (Summary) or Text mode.
Report Mode: Print logger summary information
and compact chart of all data.
Graph Mode: Print logger detailed chart page by
page. Data range is selected in
"Content" option before printing.
Text Mode: Print logger detailed data one by
one. Data range is selected in
"Content" option before printing.
If text is printed with underline
means the value is over the printing
alarm (see page 7)
124
678
Poin t
1 - 100
Text
LNK
Poin t
1 - 100
Text
LNK
Poin t
1 - 100
Text
LNK
Poin t
1 - 100
Text
Enable
LNK
1 - 1
Page
Prin t
Graph
Prin t
Text
1 - 1
Poin t
LNK LNK
IrDA
LOGGER
LINK
LOGGER
LINK
Logger Link
For loggers without two metal contacts to connect
with printer. Including: 88193/88195/88393/88394
/88395/88396/88397/88398/88378.
NOTE:
88375 & 9801printer are not compatible.
Dual stereo phone jack cable is offered in package.
The logger side phone jack size is dia. 2.5 mm.
The printer side phone jack size is dia. 3.5 mm.
NOTE:
All loggers can be programmed through 9801
printer since it has two different contact ports
for all loggers. However, be sure to avoid using
two contact ports at the same time.
Example: DO NOT use logger link port while
operating 8829 logger.
Example: DO NOT use contact port while
operating 88395 logger.
OPERATIONOPERATION
Next Page
Current Page
Press to access
the Message Page. If
no datalogger is
plugged, "Logger
Offline" and clock are
displayed on the LCD.
(2) Datalogger is plugging on
(1) Datalogger is not plugged on
While logger is plugging,
the LCD displays the
logger's ID, total record,
rate and unit.
After downloading, the LCD will turn to print page
automatically. Press to print or escape by
pressing . If any error is happened, the LCD
will display the error information.
NOTE:You can't change the logger setting
in this MSG mode.
Download function is
now selected. Press
to execute.
MESSAGE PAGE (MSG)
5
3
Menu
Feed
Enter
Up Down
Feed
Menu
Enter
Enter
Menu
Up Down
Enter
Enter
Enter
Menu
Up Down
Up
Down
Enter
Enter
Enter
Enter
Up Down
Up Down
Menu
Enter
Enter
Up Down
Menu
These value is based on printing ala rm setting
When select Report mode, you will see following:
NOTE: (x 512/Div) means one division
incl. 512 records .
When select Graph mode, you will see following :
NOTE: (x 16/Div) means one division
incl. 16 records .
These value is calculated from selected pages
When select Text mode, you will see following :
This field allows you to
set the sleep mode of
datalogger. To enable
the sleep mode, select
Enable. Otherwise
select Disable .
SLP MODE
ID
This field allows you to
set ID of datalogger.
To set the identification,
highlight the ID field and
press to enter
modified mode. Press / to set the first
character of identification then press to save
and enter the next. Max. 8 digits for editing, valid
characters for the identification are 0 to 9, A to Z ,
- , .
NOTE: When logger is removed, this page will
be skipped.
Select right after the last character, press
to save.
LOGGER PAGE (LOG)
This page enable you
to program the logger.
This field allows you to set the clock of the logger.
YY, MM, DD, hh, mm, ss stands for year, month,
date, hour, minute, second. Valid values for these
are: YY (2000 to 2099), MM (1 to 12), DD (1 to 31),
hh (0 to 24), mm (0 to 59), ss (0 to 59).
LOG'S CLOCK SETTING
Next LOG Setting
This allows you to set
more datalogger related
parameters, such as:
sample points, rate, unit,
start mode and alarm.
- Point
Point stands for the maximum record. Valid values
: 1000, 2000, 4000, 8000,12000, 16000 (You won't
see all of the points due to the printer will recognize
the available max. points of each datalogger)
-Unit
Unit stands for the unit of
temperature. Valid unit is °C or °F
-Mode
Mode stands for the start
mode of loggers. Valid
modes are: Immediate,
Magnetic, Schedule,
On-line, Key Start,
Repeat.
-Rate
Rate stands for the
sampling interval. Valid
value are 1 to 7200
seconds.
1. Immediate stands for starting right away.
2. Magnetic means you need to use magnet to
slide over the rear side of logger to start.
3. Schedule means logger will start at preset date
& time. The schedule time must be edited.
4. On-line means logger should connect to PC
while logging, so the logging point capacity
is unlimited and data are sent to PC real time.
5. Key Start means you could press the button to
start or stop the logger. In "non-logging" status,
press key for 1sec. to turn on/off the logger as
temp. monitor. Press key for 2 seconds to start/
stop the logging. Once stopping logging, you
could not press key to activate logging, it's
needed to use printer/PC to set the logger again.
6. Repeat means logger will auto-start again after
downloading the data to printer or PC.
Immediate Magnetic Schedule On-line Key Start Repeat
88 128 /88 28
/88 29
88 13/88 1 4
/88 15
88 33/34 /35 /
88 335/3 45
/35 5
Alarm Settin g
Alarm Type :
%RH Hi :
%RH Lo :
%RH
This field allows you to set the Temp. or Humidity
alarm thresholds of logger. Logger red LED will
flash while over alarm
limit.
-Alarm Setting
Choose the parameters
(Air or Ext. or%RH)
first before setting the
value.
Valid value for Air temp.
/Ext. Temp./RH% are:
10 11
13
12
14 15 16
SLP Mode
LOG's Clock Sett ing
Next LOG Sett ing
LNK
SLP Mode Enable
LOG's Clock Sett ing
Next LOG Sett ing
LNK
SLP Mo de Enable
Next LOG Setting
LOG's Clock Setting
SETTING CLOCK
MM-DD-YY hh:m m:ss
-2 5- 03 13 :4 4 : 59
0 4
LNK LNK
SLP Mode Enable
LOG's Clock Setting
Next LOG Setting
LNK
Poin t : 1000
o
Rate : 10 Unit : C
Mode : Im media te
Next LOG Setti ng
Poin t : 1000
o
Rate : 10 Unit : C
Mode : Schedule
xx-x x-xx xx:x x:x x
Next LOG Setti ng
LNK
LNK
LNK
CONTENT
(Ex: 1-10)
(Ex: 200-1000)
the most page
within recorded point
This field allows you to select data range in Graph
and Text printing mode. To select: Highlight the
page or point field and press to enter.
Mode : Report
Content : ( Blank )
Mode : Graph
Content : Page
xx - xx
Mode : Text
Content : Point
xx - xx
Press / to set the data range. Press
to save or press to escape without
saving.
Valid printing ranges are depended on the data of
logger.
One page includes 1000 temperature and humidity
(if available) records.
The next pages show the examples at
Report, Text, Graph printing mode.
9
88 193/8 81 95
88393/8839
4/88395
88 396/3 97 /
398
88378
Enter
Enter
Up Down
Menu
Up
Enter
Down
Enter
Enter
LNK CFG MSG
PC < > Logger
PC < Printer
PC CONNECTION
USB1.0
USB1.0
PC LINK (LNK)
This printer can connect
with PC via USB.
Select the function as
"LNK". Two PC Communication included in
"LNK" .
Use / to select the needed
function then press to confirm and enter.
While selecting , you only can
download the stored data to PC. This is an one
way communication. Under Printer to PC,
"Logger Setting" is disabled. After downloading
PC < Printer
the stored data to software, you can use this
software to review and analyse the stored data.
While selecting , you can use the
printer as a regular interface to set & download
the logger through PC. This is a two way
communication so the printer works like 8818P,
8828P, 8829P, 8824P interface. Under this
function, "Logger Setting" will not be disabled.
PC < > Logger
PC < Printer
PC < > Logger
This printer has
storage capacity for
4 loggers and then
download the data to
PC. Besides, 9801 can
also be treated as
standard interface to
upload the command
from PC to logger or to
download the data
from logger to PC.
4 LOGGER’S MEMORY
Printer can store 4 loggers' data. Before each
printing, uploading and downloading, it is
needed to highlight LOGGER 1 .. or LOGGER 4
file number and activate the selected storage
file.
DOWNLOADING
If you want to store
more than one logger,
select the file number
is important. To load
more than one logger
without selecting
another file number,
the LOGGER 1 will be
overwrite.
When execute the download command (see
page5), "LOG STO Setting" will display on the
LCD.
LOG STO Setting
1 LOGGER1
2 LOGGER2
3 LOGGER3
4 LOGGER4
The file indicator
points to current
storage file. After
downloading, the ID of each datalogger will
replace the original file name.
LOG STO Setting
LNK
1 Your own ID
2 LOGGER2
3 LOGGER3
4 LOGGER4
PRINTING & UPLOADING TO PC
Same as executing download command,
"LOG STO Setting" will display on the LCD.
Select a suitable file number and press
to confirm.
Press /
key to highlight the file
number and press
to confirm.
TROUBLE SHOOTINGTROUBLE SHOOTING
1).
Problem: Press to run print command
but skip to MSG page with nothing
printed.
Reason: Battery has less energy.
Solution: Open the rear battery cover to replace
with new battery ( ).
NOTE: Alkaline battery is better.
4pcs AAA
When screen displays "Low Battery
,though you might still print while
"Low Battery" is shown, strongly
recommend to replace batteries right
away.
24
2).
Problem: Press to run print command
but nothing displayed and no any
information.
Paper Jam !
Please Clear Paper !
Press to
continue...
3).
Problem: Press , but not working
Low Battery !
Reason: Paper is jammed.
Solution: Clear paper jammed in the printer.
Reason: Paper is jammed.
.
Solution: Clear paper jammed in the printer.
4).
Problem: Low Battery display on LCD
Reason: Power is not enough for operation.
Solution:
Electronic Characteristics:
a) Power : DC9V 500mA or
Alkaline AAA 1.5V × 4
(Recommend Alkaline)
b) Consumption :
< 3.2 A : Peak value when printing
< 300 mA : Average value when printing
0.5 ~ 2 mA: Quiescent
< 80 uA : Power down (Typ.: 40 uA)
c) Auto power off after 5 min. of inactivity
o o
d) Operating Temperature : 0 C ~ 50 C
Printing characteristics:
a) Voltage compensating to stabilize the contrast
of printing.
b) While paper jam, printer will stop feeding to
protect printer hand.
c) Feed paper
d) Printing Contrast : 1 ~ 9
e) Print type by report, graphic or text.
(Font : Roman 8 )
f)
g)
h)
LCD Contrast : 1~9
Thermo-paper : 58mm x 20M ( length)
Dimension : 120mm x 150mm x 48 mm
( L x W x T )
SPECIFICATIONSPECIFICATION WARRANTY & RETURN AUTHORIZATIONWARRANTY &
Authorization must be obtained from the
supplier before returning items for any reason.
When requiring a RA (Return Authorization),
please include data regarding the defective
reason, the meters are returned along with good
packing to prevent any damage in shipment and
insured against possible damage or loss.
Proof of purchase is required for warranty
repairs. Warranty is void if the meter has been
opened.
The meter is warranted to be free from defects
in material and workmanship for a period of one
year from the date of purchase. This warranty
covers normal operation and does not cover
battery, misuse, abuse, alteration, tampering,
neglect, improper maintenance or damage
resulting from leaking batteries.
18 19
21
20
22 23 24
NOTE: The calibration of 88335~88355
could only be set by software instead of
9801 printer.
NOTE: Press to exit PC communi-
cation mode.
MSG
Air Hi:-40 to 185 Air Lo: -40 to 185
Ext. Hi:-40 to 212 Ext. Lo:-40 to 212
%RH Hi:0 to 100 %RH Lo:0 to 100
NOTE: Low value isn't allowed to exceed high
alarm setting.
-DATE/TIME SETTING
To edit the clock of printer. Valid values are:
YY(2000 to 2099), MM(1 to 12), DD(1 to 31),
hh(0 to 24), mm (0 to 59), ss(0 to 59).
LCD Cont:
Baudra te: 9600
Date : MM-D D-Y Y
Date /Time Setti ng:
5
PRN
-LCD CONTRAST
To select the printer LCD contrast from 1 to 9.
-BAUDRATE
This field indicates the baudrate of communication.
It is fixed at 9600. Press to skip.
-DATE
To select the format of date. There are: MM-DD-YY
, DD-MM-YY, YY-MM-DD.
CONFIGURATION (CFG)
You can set the configuration of the printer.
SETTING CLOCK
MM-DD-YY hh:m m:ss
-2 5- 03 13 :4 4 : 59
0 4
17
2013.08 ver.3
Down Down
Up
Enter
Menu
Down
Up
Enter
Enter
Enter
Enter
Feed
Enter
Press to
continue...
Enter
  • Page 1 1
  • Page 2 2
  • Page 3 3

Reed Instruments 9802 Operating instructions

Type
Operating instructions