Warnings/Precautions
Federal Law in the U.S.A. restricts this device to sale by or on the order of a
physician.
This device is intended for use by trained medical professionals only.
Refer all service to Iradimed Corporation Authorized Service Representatives.
The 3860 Pump has been specifically designed for operation inside an MRI Magnet
Room, and is designed to operate normally in the presence of most frequently
encountered electromagnetic interference in the MRI environment. Under extreme
levels of interference, such as in close proximity to an electrosurgical generator,
cellular telephone or a 2 way radio normal pump operation may be interrupted.
Avoid use of this pump under these conditions.
Use only MRI-compatible patient access devices (e.g. needles, luer-ports, etc.) to
prevent any possible RF current from reaching the patient's skin.
The Remote Display Charger is intended for use in the MRI control room. Do not
operate the 3865 Remote Display/Charger inside the MRI Magnet Room.
For safe operation, use only Iradimed Corporation recommended MRI-compatible or
MRI-safe accessories.
The MRI pump must be mounted securely when used in the MRI scan room. Always
securely mount the pump using it’s integral pole mount clamp or other mechanical
fixing means.
Always secure the I.V. Pole wheel locks after positioning within the MRI Magnet
Room.
Avoid placing the I.V. Set adjacent to any electrical conductor within the MRI bore,
which can become heated during MRI scans.
The Alarm Sound Volume is adjustable for various clinical environments. Ensure the
alarm sound level is appropriate for the use environment in the MRI so that it can be
heard above the ambient noise level, especially during scanning.
Product damage may occur unless proper care is exercised during unpacking and
installation. Do not use the pump if it appears damaged in any way. The battery
should be charged before use.
Battery may not be fully charged upon receipt. Connect pump to AC power for at
least nine (9) hours before placing the pump into use.
Always connect the infusion pump power supply or Remote Display/Charger to a
properly grounded 3-wire power receptacle. Never use power extension cords. If the
quality of the earth grounding is in question, use battery power for the infusion
pump.
This equipment is not suitable for use in the presence of flammable anesthetic or
other gases. Do not use this system in the presences of flammable gases.