Plot of Sensorex DC Output Versus Conventional
Conductivity Meter Output
y = 6.0498x - 5.435
R
2
= 1
0
5
10
15
20
123
Sensorex Output (V)
Conductivity Meter Output
(mS)
Series1
Linear
Series1
After the sensor is properly connected, the system should
be calibrated. The typical calibration procedure uses a “low”
and a “high” known standard conductivity solution. The "low"
solution is often DI water or air and is used to calibrate the
zero point of the controller. Plotting these two points will cre-
ate a straight line, which can be used to find the conductivity
valueof any solution in the range. Make sure probe is im-
mersed in the calibration fluid such that the toroids are totally
submerged. A sample plot is provided in FIG. 5 to show the
approximate values, which will be encountered during calibra-
tion.
Note: The values in the plot are arbitrary, each probe will
require unique linear equation values.
If the sensor is to be used in a submersion application, cali-
brate the sensor in a large glass or plastic beaker with all sides
of the sensor at least one inch away from the wall
(FIG. 6). If the production installation is a pipe (plastic or metal),
calibration should be performed in a similar pipe arrangement.
All Electrodeless (Toroidal) sensors have a wall effect, which
must be taken into account during calibration. If the non-con-
ductive (plastic) wall is within 1 inch of the sensor, the sensor’s
reading will be reduced due to the insulator interaction with
the current path. If the sensor is within 1 inch of a conductive
(metal) wall, the sensor’s reading will be increased due to the
shorting effect of the conducting wall. These wall effects can
be calibrated out of the system by simulating the application’s
mounting configuration. A plot showing the effects of insulat-
ing and conducting walls on the output can be seen in FIG.7.
During calibration and production installation (especially in
a submersion environment), it is important to dislodge any
air bubbles, paying special attention to the center hole of the
toroids. Also make sure toroids are totally covered with fluid
when calibrating (FIG. 6).
Please refer to your controller's manual for specific calibration
instructions.
Sensor Calibration
FIG. 5
Plot of Normalized Output as a Function of Probe
Distance from Wall
2.15
2.2
2.25
2.3
2.35
2.4
2.45
2.5
2.55
024681012
Wall Distance (In)
Normalized Output (V/V)
Conducting wall
Insulatin
Wall
Liquid must totally
cover toroids (be
above this line)
> 1”
> 1”
FIG. 6
FIG. 7
Sensor Maintenance
The major advantage of the Electrodeless (Toroidal) Sensor
is almost no maintenance is required. The only maintenance
required during normal operational life of the sensor is to
prevent the toroidal opening from being plugged with debris.
Use a soft brush or rag to remove any debris in the core open-
ing. If that does not work, try a mild detergent or weak acid
(5-10% HCl).
PRODUCT INSTRUCTION SHEET
CDE-45 Page 3 of 4
®
DC Output vs. Conventional Conductivity Meter