Drawell TDZ5WS User manual

Type
User manual

The Drawell TDZ5WS is a benchtop low-speed centrifuge ideal for use in biochemistry, genetic engineering, and radioimmunology. With a maximum speed of 5000 rpm and a maximum RCF of 4390 xg, it can process samples in a variety of rotors, including fixed-angle and swinging-bucket rotors. The intuitive LCD display and easy-to-use controls allow for precise setting of speed, time, and temperature. The centrifuge also features safety features such as an electronic door lock and imbalance detection to ensure safe and reliable operation.

The Drawell TDZ5WS is a benchtop low-speed centrifuge ideal for use in biochemistry, genetic engineering, and radioimmunology. With a maximum speed of 5000 rpm and a maximum RCF of 4390 xg, it can process samples in a variety of rotors, including fixed-angle and swinging-bucket rotors. The intuitive LCD display and easy-to-use controls allow for precise setting of speed, time, and temperature. The centrifuge also features safety features such as an electronic door lock and imbalance detection to ensure safe and reliable operation.

Benchtop low-speed centrifuge
TDZ5WS
operation instruction
Before use, please read the operation
instruction handbook
Safety alerts and precautions:
1.Safety alerts
1.Before use, please read the operation instruction handbook.
2.Do not use in wet, hot or dusty locations to avoid instrument
failure.
3.When the instrument is running, it must not move the instrument
and must not open the door cover of the instrument.
4.The countertop of the installation instrument should be solid and
flat, there should be no shaking phenomenon, all four rubber
machine feet should be in full contact with the countertop and
uniform force.
5.Centrifuge tube added liquid as evenly as possible, if the
difference between the addition will produce large vibration, at this
time should be stopped to check, so that the addition meets the
requirements, centrifugal test tube must be placed in even
symmetry, not singular into or asymmetrical into the rotor.
6.After each use, open the instrument door cover, allow the water
in the centrifugal cavity to evaporate, and clean the inner cavity of
the instrument. The instrument must have a reliable ground wire
when using a power supply.
2.Precautions
1.Do not strongly collide, vibrate or re-throw the instrument to
avoid non-serviceable consequences of damage to the mechanical
or electrical parts of the instrument.
2.Do not use the instrument in wet or hot places to avoid extensive
damage to the electrical part of the instrument.
Product form picture
1
12
Benchtop low-speed centrifuges
Packing list
No
Name
Specification
Quantity
Note
1
Host
TDZ5WS
1
Wooden
box
2
Rotor
48x7ml
3
Rotor
4
Rotor
5
Test tube
6
Test tube
7
Test tube
8
Accessories/tools
9
Operation
instruction
1
10
Power cord
1
11
Packing list
1
Product NameCentrifuge ModelTDZ5WS
Product performance structure and compositionIt consists of a
control system, Centrifugal cavity, Drive system, Rotor,
Refrigeration system (If it is a refrigerated centrifuge) and safety
protection.
Scope of applicationThis product is suitable for use by trained
professional operators and technicians.
Taboo It is taboo to use without professional operators or
personnel without relevant operational qualifications.
PrecautionsSee above.
Alerts and suggestive contentStop prompts
Operating instructionsSee below
Label, packaging identification sample Please read the
instructions
Storage conditions/methods
Rooms with ventilation/drying/light avoidance/non-corrosive
gases should be stored.
1. Overview
This manual adapts to TDZ5WS benchtop low-speed
centrifuges can be widely used in biochemistry, Genetic
engineering, Radioimmunology and other fields, Is an ideal
experimental instrument for research institutes. This machine uses
DC variable frequency motor, microcomputer control, Liquid
crystal display, Lift speed, low noise, Small vibration, Electronic
door lock safe and reliable. (See Table 1/Table 2 for major
technical performance)
Table 1: Technical parameters
Wiring diagram
Model
TDZ5WS
Max. Speed
5000 rpm
Max. RCF
4390 xg
Max. Capacity
4 x 250 ml
Speed Accuracy
± 10 rpm
Presentation
LCD
Timer Range
1sec ~ 99min59sec
Motor
AC frequency conversion
Power Supply
AC110 V 60 Hz 10 A
Full Power
450 W
Noise
65 db
Dimensions(LxDxH)
21.3 x 16.7 x 14.4 in / 540 x 425 x 365 mm
Weight(w/o rotor)
92 lb / 41.7 kg
the fault display.
E-1Imbalance E-2Overspeed E-3Door
E-4Speed-hall error E-5Overvoltage (during shutdown)
E-6: Overcurrent E-7: No speed measured
E-8: Communication
E-9: The external input power supply fluctuates too much
E-10: Temperature sensor
Table 2: Instrument rotor parameters
Number
Name
Max. Speed
Max. RCF
Capacity
Tube
Size(mm)
NO.1
Fixed angle
rotor
5000 r/min
4390 xg
12 x 10 ml
Φ16 x 85
NO.2
5000 r/min
2540 xg
6 x 15 ml
Φ12.4 x 106
NO.3
5000 r/min
3080 xg
12 x 15 ml
Φ12.4 x 106
NO.4
5000 r/min
3515 xg
24 x 15 ml
Φ12.4 x 106
NO.5
5000 r/min
3830 xg
30 x 15 ml
Φ12.4 x 106
NO.6
5000 r/min
2520 xg
4 x 50 ml
Φ29 x 110
NO.7
5000 r/min
2850 xg
6 x 50 ml
Φ29 x 110
NO.8
5000 r/min
2630 xg
4 x 100 ml
Φ42 x 98
NO.9
4000 r/min
2980 xg
32 x 5 ml
Φ13 x 55
NO.10
Swinging-b
ucket rotor
4000 r/min
3016 xg
32/24/16 x 15 ml
Φ12.4 x 106
NO.11
4000 r/min
3215 xg
48 x 7 ml
Φ12.4 x 106
NO.12
4000 r/min
3215 xg
72 x 7 ml
Φ12.4 x 106
NO.13
4000 r/min
3016 xg
4 x 250 ml
Φ61 x 122
NO.14
Microplate
swinging
bucket rotor
4000 r/min
2310 xg
2 x 48 holes
2.Installed
A.The native power supply is a single-phase AC110V 60Hz 15A.
B.The power supply should have a safe and reliable wire, strictly
prohibit the ground wire instead of zero line, not to pull the
machine wire indiscriminately.
C.The placement of the site indoor environment non-corrosive gas
relative humidity shall not be greater than 85%, there should be no
other strong vibration source. The table surface of the unit should
be rigid and hard enough to support the instrument, the countertop
should be flat without any slope, so that the four rubber soles of the
machine and the table to maintain smooth contact.
3.Introduction to the interface and instructions for use:
1.The centrifuge placed on a solid countertop, four rubber machine
feet should be in good contact with the table surface, shake the
centrifuge with your hand, check whether the centrifuge is placed
smoothly.
2.Open the door cover, the centrifuge tube into the rotor test tube
hole, centrifuge tube must be placed in even symmetry (high-speed
centrifuge centrifuge test fluid should be weighed to join), pay
attention to the rotor cover tight, and re-check whether the test tube
is symmetrically placed, screws are tight.
3.Close the door cover, pay attention to make the door cover
locked, finished by hand to check whether the door cover is closed.
Plug in the power supply and press the power switch.
4.Set speed, temperature, time, rotor number, program, speed-up
gear, speed-down gear: (operator panel diagram below)
can be raised) otherwise long time on the shaft, may rust to death,
the rotor will not be taken out and cause the centrifuge machine
scrap.
6.The life of the native host 6 years, rotor service life of 5 years,
expired should replace the rotor. All rotors must not be used at
more than their maximum speed.
5.Troubleshooting/warranty coverage
Common troubleshooting methods:
Common failures
Possible reason
Elimination
method
Power on, the
display window is
not lit
No 110V power supply
Check power
supply
Fuse is broken
Replacement
Power plug-in is loose
Plug in
Press the start
button centrifuge
does not operate
Door cover is not closed
Close the door
cover
Wiring plug-in is loose
Plug in
Control board is damaged
Replacement
Motor has electricity does not
turn, the motor is damaged
Replace the
motor
Centrifuges
vibrate a lot
Centrifuge tube is not placed
symmetrically
Check
Centrifuge tube is broken
Replacement
Rotor is not rotating
Check
When the centrifuge fails, check the fault condition corresponding
to the fault number and resolve the failure of the centrifuge
according to the actual situation. Press the stop button to eliminate
5
8
When there is only one rotor, the rotor number adjustment is
skipped and the room temperature interface skips the temperature
adjustment. Press the up and down key to adjust the parameter
value, adjust the parameter to the expected value, press the
confirmation key to exit the set state and save the modified
parameters.
When the above steps are complete, press the "Confirm" key to
confirm the speed, temperature and time set above, and then press
the "Start" key to start the centrifuge (when pressing the "Start"
key, if the instrument cannot start, check that the door cover is
closed).
4.Precautions
1.The table of the centrifuge should be flat, four rubber machine
feet should be in contact with the table and uniform force, so as not
to produce vibration.
2.Centrifuge tube liquid should weigh the balance, if the difference
between the addition of liquid too large operation will produce
large vibration, at this time should be stopped to check, so that the
addition meets the requirements, centrifugal test tube must be
placed in even symmetry.
3.If there is a centrifugal test tube rupture, will cause large
vibration should be immediately stopped treatment.
4.After each shutdown and then power on the interval should not
be less than 5 minutes, so as not to block the compressor to turn
damage.
5.After each centrifuge is completed, open the instrument door
cover, let the water evaporate, and regularly take out the rotor,
(unscrew the rotor cover, with a handle into the rotor center hole
Introduction to keystrokes
Set: Switch adjustment parameters, this key is set for switching
adjustment speed, time, speed etc.
Up: Increase hypothesis parameter value; This pressed key
adjustment rotational speed value and time value establishment.
Down: Reduce the set parameter value, this button adjusts the
speed value and time value setting.
Enter: Exit the set state and save the setting parameters.
ShortPress this key centrifuge to start centrifugation, relax and
stop working.
Start: Start the motor, (press this key to start the centrifuge after
setting all parameters).
Stop/Open: Eliminate alarm when alarm, operating status press
can stop midway standby state open cover.
Parameter set:
Press the setting key to cycle through the set parameters, the
parameters set the cycle order (number of keys): 1: speed; 2: time
minutes; 3: time seconds; 4: rotor number; 5: program number; 6:
speed upshift (Larger the number, Faster the speed), 8: the lower
speed (Larger the number, Faster the descent)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7

Drawell TDZ5WS User manual

Type
User manual

The Drawell TDZ5WS is a benchtop low-speed centrifuge ideal for use in biochemistry, genetic engineering, and radioimmunology. With a maximum speed of 5000 rpm and a maximum RCF of 4390 xg, it can process samples in a variety of rotors, including fixed-angle and swinging-bucket rotors. The intuitive LCD display and easy-to-use controls allow for precise setting of speed, time, and temperature. The centrifuge also features safety features such as an electronic door lock and imbalance detection to ensure safe and reliable operation.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI