Rosemount NGA 2000 TFID Hydrocarbon Analyzer Module SW 3.4 Software-1st Ed. Owner's manual

Category
Water pumps
Type
Owner's manual
www.EmersonProcess.com
Instruction Manual
ETC00612
02/2001
Instruction Manual
Software Manual 3.4.x
Thermo Flame Ionization Detector
TFID Analyzer or Analyzer Module (combined with NGA 2000
Platform, MLT or CAT 200)
1
st
Edition 02/2001
TFID Software 3.4.x Instruction Manual
ETC00612
02/2001
Emerson Process Management
GmbH & Co. OHG
Industriestrasse 1
D-63594 Hasselroth
Germany
T +49 (0) 6055 884-0
F +49 (0) 6055 884-209
Internet: www.EmersonProcess.com
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Emerson Process Management (Rosemount Analytical) designs, manufactures and tests
its products to meet many national and international standards. Because these instruments
are sophisticated technical products, you MUST properly install, use, and maintain
them to ensure they continue to operate within their normal specifications. The following
instructions MUST be adhered to and integrated into your safety program when installing,
using and maintaining Emerson Process Management (Rosemount Analytical) products.
Failure to follow the proper instructions may cause any one of the following situations to
occur: Loss of life; personal injury; property damage; damage to this instrument; and warranty
invalidation.
Read all instructions prior to installing, operating, and servicing the product.
If you do not understand any of the instructions, contact your Emerson Process
Management (Rosemount Analytical) representative for clarification.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Inform and educate your personnel in the proper installation, operation, and
maintenance of the product.
Install your equipment as specified in the Installation Instructions of the appropriate
Instruction Manual and per applicable local and national codes. Connect all products
to the proper electrical and pressure sources.
To ensure proper performance, use qualified personnel to install, operate, update, program,
and maintain the product.
When replacement parts are required, ensure that qualified people use replacement parts
specified by Emerson Process Management (Rosemount Analytical). Unauthorized parts
and procedures can affect the product’s performance, place the safe operation of your
process at risk, and VOID YOUR WARRANTY. Look-alike substitutions may result in fire,
electrical hazards, or improper operation.
Ensure that all equipment doors are closed and protective covers are in place, except
when maintenance is being performed by qualified persons, to prevent electrical
shock and personal injury.
The information contained in this document is subject to change without notice. Misprints
reserved.
1
st
Edition 02/2001
©
2001 by Emerson Process Management
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 06/01
NGA 2000 I
Contents
1 Introduction 1 - 1
2 Menu Structure 2 - 1
3 Startup and Operation, General Notes and Main Menu 3 - 1
3.1 Starting and Initializing...................................................................................3 - 1
3.2 Display and Function......................................................................................3 - 2
3.3 "TAG" and Operating Keys ............................................................................3 - 2
3.4 Lines and Softkey Functionality .....................................................................3 - 3
3.5 Important Function Softkeys ..........................................................................3 - 4
3.6 Entering/Changing of Variables ..................................................................... 3 - 5
3.7 Executing a Function......................................................................................3 - 6
3.8 Main Menu .....................................................................................................3 - 7
4 Analyzer Basic Controls (Calibration) and Setup 4 - 1
4.1 Analyzer Channel Status................................................................................ 4 - 3
4.1.1 Status Details
e.g. Failures........................................................................4 - 5
e.g. Acknowledge and Clear Failures.................................4 - 9
4.1.2 Current Operation Parameters (Analyzer Operation Settings).......................4 - 13
4.2 Single Component Display - Change of Channel...........................................4 - 15
4.3 Multi Component Display - Change of Channel.............................................4 - 17
4.4 Calibration Procedure Status......................................................................... 4 - 19
4.5 Zero Calibration..............................................................................................4 - 21
4.6 Span Calibration/ Basic Parameters..............................................................4 - 25
4.7 Active Zero Gas, Span Gas, Sample Gas or Test Gas - Close all Valves.....4 - 29
4.8 Flow measurement.........................................................................................4 - 31
II NGA 2000
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 06/01
5 Analyzer and I/O, Expert Controls & Setup 5 - 1
5.1 Analyzer Module Setup (or Controls)..............................................................5 - 3
5.1.1 Calibration Parameters...................................................................................5 - 5
– Span gases.........................................................................................5 - 6
Tolerances ..........................................................................................5 - 8
– Calibration procedure setup................................................................5 - 10
– Time controlled calibration..................................................................5 - 13
Calibration...........................................................................................5 - 15
– Advanced calibration methods............................................................5 - 18
– Zero Gases .........................................................................................5 - 20
5.1.2 Alarm Parameters...........................................................................................5 - 21
5.1.3 Range parameters..........................................................................................5 - 25
– Begin and end of ranges.....................................................................5 - 27
– Response times (t
90
) ...........................................................................5 - 28
– Autoranging control.............................................................................5 - 29
5.1.4 Cross Interference Compensation..................................................................5 - 31
5.1.5 Linearization...................................................................................................5 - 33
5.1.6 Programmable Logic Control (PLC)................................................................5 - 37
5.1.7 Programmable Calculator...............................................................................5 - 45
5.1.8 Measurement Display Configuration...............................................................5 - 49
5.1.9 Acknowledgement of Status Reports .............................................................5 - 53
5.1.10 Concentration Measurement Parameters.......................................................5 - 55
5.1.11 Concentration Peak Measurement.................................................................5 - 57
5.1.12 Differential Measurement ...............................................................................5 - 59
5.1.13 Gasflow Setup................................................................................................5 - 61
5.1.14 Pressure Compensation.................................................................................5 - 62
5.1.15 Flow Measurement.........................................................................................5 - 64
5.1.16 Temperature Measurement............................................................................5 - 65
5.1.17 Load/Save Analyzer Module Configuration ....................................................5 - 66
5.1.18 Inputs and Outputs (Local SIO/DIO)...............................................................5 - 69
– Local SIO .............................................................................5 - 70
– Local DIO .............................................................................5 - 76
– Signal codes .............................................................................5 - 77
5.1.19 Delay and Average.........................................................................................5 - 81
5.1.20 Special Functions...........................................................................................5 - 83
5.1.21 AK-Protocol Communication ..........................................................................5 - 84
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 06/01
NGA 2000 III
5.2 I/O Module Controls (System or CM SIO/DIO).............................................5 - 85
5.2.1 SIO Module..................................................................................................5 - 86
– Analog output setup.................................................................5 - 87
– Serial interface setup ...............................................................5 - 91
– Configuration of Relay Outputs................................................5 - 92
5.2.2 DIO Module(s)..............................................................................................5 - 95
– Configuration of the 8 DIO module inputs................................5 - 96
– Configuration of the 24 DIO module outputs............................5 - 98
5.3 I/O Module Setup (Network I/O Modules) ....................................................5 - 101
6 System Configuration and Diagnostics 6 - 1
6.1 Diagnostic Menus...........................................................................................6 - 3
6.1.1 Control Module Diagnostics.................................................................6 - 4
6.1.2 Analyzer Module Diagnostics .............................................................. 6 - 5
6.2 Load/Save Module Configuration...................................................................6 - 6
6.3 Date and Time ...............................................................................................6 - 7
6.4 Security Codes...............................................................................................6 - 8
6.5 Network Module Binding................................................................................6 - 10
6.6 System Reset.................................................................................................6 - 12
7 Display Controls 7 - 1
Supplement:
System Calibration
Tables:
Response factors
Conversion factors ppm
mg/Nm³
Index
1 Introduction
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 02/01
NGA 2000
1 - 1
This software manual describes step by step how to operate successfully with the
NGA 2000 Series 19" TFID (Thermo-FID) analyzer modules and analyzers.
Chapter 2 shows the TFID software menu structure. Chapter 3 describes the analyzers
display and keyboard as well as the main menu and submenus. Chapter 4 describes the
basic controls incl. calibration with detailed illustrations. So you can easily compare the
actual analyzer (module) display with the illustrations of this manual.
Chapter 5 describes the expert configurations of the analyzer module and of the I
nput/
O
utput Modules (I/O modules). Chapter 6 describes the system configuration and
diagnostics. The layout of both chapters is not as detailed as of chapter four. By default,
the way through the menus is described giving the menu lines you have to enter
sequentially to reach a specific menu within the TFID software . At the end of this
description you will find an illustration of the final submenu screen followed by set up
instructions and explanations of functions and variables of each expert or system
configuration menu. Chapter 7 will give you some information about the display controls.
Not all sections of Chapter 5 (expert configurations) are important for each operator.
Which section is of interest depends on the NGA 2000 configuration with reference to the
following components:
Control Module CM
Analyzer Module AM
Input/Output Modules I/O's (SIO = Standard I/O, DIO = Digital I/O)
Network I/O Modules Analog Output with 3 Alarms I/O, Auto Calibration I/O,
System Auto Calibration I/O
You can distinguish the following system units and SIO/DIO configurations:
System Unit SIO/DIO-Configuration
Section
Page
TFID analyzer module (AM):
without front panel,
i.e. without control unit
can be combined with a platform, a
MLT analyzer, a TFID analyzer or a
customer developed control unit
1 local SIO and 1 local DIO (or 2
local DIO’s) can be installed in the
TFID/MLT analyzer module
SIO and DIO can be configured
for the MLT AM channels or the
TFID analyzer module only
5.1.18
p. 5-69
Platform (CM Software):
Control unit with front panel
Without measurement channels
1 SIO and up to 4 DIO's can be
installed in the platform (CM I/O)
SIO and DIO can be configured
for all MLT channels % AM’s
combined with the platform
5.2
p. 5-85
TFID analyzer (CM plus TFID AM
software = TCA software):
Analyzer with front panel
CM and AM software in the same
analyzer, i.e. all functions of the
control unit and of the AM are
combined in one controller board
1 SIO and 1 DIO (or 2 DIO’s) can
be installed in the TFID/MLT
analyzer (CM I/O)
SIO and DIO can be configured
for all MLT channels and AM’s
combined with the TFID/MLT
analyzer
5.2
p. 5-85
1 - 2
NGA 2000
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 02/01
The following illustrations explain the relationship between the hardware configuration and
the software setup of the modules:
Platform
MLT
CLD
PMD
1 SIO
4 DIO's
max.
Network I/O
(see 5.2.1)
(see 5.2.2)
NGA 2000 System containing a Platform
TFID
A
nalyzer
Modules (AM's)
Control
Module (CM)
System I/O
Modules
Local
I/O's
(additional
manuals)
(additional
manuals)
(see 5.1.18)
FID
(additional
manuals)
DIO
SIO
DIO
SIO
(see 5.1.18)
TFID Analyzer
TFID
CLD
PMD
1 SIO
1 DIO
NGA 2000 System containing an TFID Analyzer
FID
Analyzer
Modules (AM's)
Control
Module (CM)
System I/O
Modules
Local
I/O's
(additional
manuals)
(additional
manuals)
SIO
DIO
(additional
manuals)
MLT
(see 5.1.18)
DIO
SIO
(see 5.1.18)
(see 5.2.1)
(see 5.2.2)
1 Introduction
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 02/01
NGA 2000
1 - 3
Note:
This software manual describes the software of all TFID analyzers and TFID analyzer
modules combined with a platform, an FTID analyzer or an MLT analyzer.
It will not describe the software of TFID analyzer modules running at a customer control
unit.
The software of MLT analyzers / MLT analyzer modules is nearly identically to the TFID
software. An separate software manual is available containing the MLT specific functions.
MLT Analyzer
MLT
CLD
PMD
1 SIO
1 DIO
NGA 2000 System containing an MLT Analyzer
FID
Analyzer
Modules (AM's)
Control
Module (CM)
System I/O
Modules
Local
I/O's
(additional
manuals)
(additional
manuals)
SIO
DIO
(additional
manuals)
TFID
(see 5.1.18)
DIO
SIO
(see 5.1.18)
(see 5.2.1)
(see 5.2.2)
1 - 4
NGA 2000
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 02/01
2 Menu Structure
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 02/01
NGA 2000
2 - 1
Notes:
* Menu lines ending with three points (...) are followed by submenus with further functions and setups, lines ending with "!" allow to start
the function, lines ending with ":" display module parameters - some can be changed.
* All setups in the menu "Analyzer basic controls (calibration) & setup" are valid for the analyzer or analyzer module (AM).
* At each menu point of the "Analyzer and I/O, expert controls & setup" and "System configuration and diagnostics"
you can find which functionality will be setup:
AM (Analyzer Module), CM (Control Module), local, system or network I/O (Input/Output Module).
- Calibration procedure state...
- Start zero calibration procedure!
- Start span calibration procedure!
- Check calibration deviation:
- Range number:
- Span gas:
- Range upper limit:
- Valves => F5:
- Flow zero gas!
- Flow span gas!
- Flow sample gas!
- Flow test gas!
- Close all valves!
- Analyzer module controls...
(AM
incl.
local SIO/DIO)
- System & network I/O
module controls...
(System I/O's: SIO, DIO)
- Analyzer module setup...
(AM:
for
MLT/TFID
like
analyzer module
controls)
- System & network I/O
module setup...
(Network I/O's)
- System calibration...
(CM)
- Diagnostic menus...
(AM, CM, network I/O's)
- Load/save configuration
(CM/MCA)
...
- Date and time...
(CM)
- Security codes (PIN)...
(Lock any level)
- Network module managment...
(AM, CM, I/O's: SIO, DIO etc.)
- System reset...
(CM)
- System tag:
(CM)
- Brightness:
- Contrast:
- Switch automatically
to "Measure" after
Main Menu
Analyzer basic controls
(calibration) & setup...
Analyzer and I/O, expert
controls & setup...
System configuration
and diagnostics...
Display controls...
Section 4 Section 5 Section 6 Section 7
Section 3.8
2 - 2
NGA 2000
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 02/01
3
Startup and Operation, General Notes and Main Menu
ETC00612(1) [NGA-e (TFID Software 3.4.X)] 06/01
NGA 2000
3 - 1
3.1 Starting and Initializing (Refer to the operation manual, too !)
After switching on the TFID analyzer / analyzer module ( as part of a NGA network), the
initialization procedure will be performed. . A self test is started showing a sequence of
several screens with information about the initialization status, software revision notes and
the
tag:
The network is initialized while the screen shows the messages “Initializing network
interface” , “ Searching for nodes” and “Calculating binds”.
If you press the F1-key during the initializing, you will reset the LCD brightness and
contrast to factory settings (see also section 7). Pressing the F3-key will abort the
network initializing. There will be no connection to any analyzer module. Only the control
module menus (platform or TFID analyzer) will be available.
At the end of the initializing procedure you can see the single component display of
channel one (see illustration on next page). It is the starting point to all the other channel
displays, menus and submenus.
The instructions of the basic controls (chapter four) are all beginning at the single com-
ponent display. Since this screen is customer configurable the actual analyzer display
might differ from the one shown in this manual (see section 5.1.8 pg. 5-49 and section 7).
LCDReset
A
bort
F1 F2 F3 F4 F5
(C) 2000 FISHER-ROSEMOUNT Analytical
NGA-2000 Control-Module Rev. 3.4 /P010
Initializing Network
Initializing network interface
Language:P010/01/00
3 - 2
NGA 2000
ETC00612(1) [NGA-e (TFID Software 3.4.X)] 06/01
3.2 Display and Function
The LCD screen shows all analyzer measurement values and all customer instructions.
You can operate with five function keys, four arrow keys (cursors) and the enter key. The
function of each key depends on:
the type of analyzer/analyzer module used
the optional auxiliary modules (e.g. I/O boards) used
the individual menu displayed
In case of power failure all customer specific module parameters are saved in a battery-
powered buffer.
3.3 "TAG" and Operating Keys
Display Status... Main... Channel
BasicCal
TAG
F1 F2 F3 F4 F5
Failures: No
Maintenance-Requests: No
Temperature: 20.0 C
Operation: Ready
Range: 1
37.50 ppm CH4
0.00 50.00
0.0 100.0
Function Keys:
Keys without defined function
The current function depends on
the menu selected
The softkey legend is shown on
the display right above the key
At the top left of each menu page you will find the tag of the current channel.
Typical tags:
TFID/R1: T
hermo FID Analyzer or Analyzer Module / Range 1
MLT/CH1/R1: MLT Analyzer or Analyzer Module / CHannel 1 / Range 1
By default screen dumps in this manual will show "TAG". Specific TFID menu pages will use
"TFID" instead to show this property.
Enter Key:
To confirm a previously entered value
(variable)
To start a selected function
(alternative
:
-key)
To enter a menu (via menu line)
Cursor keys:
-key /
-key:
Line up / line down
within the same menu
Alteration of numbers,
variables or digits
-key /
-key:
Moving back/forth
between the pages of
a menu
Selection of digits
Start a selected
function
3
Startup and Operation, General Notes and Main Menu
ETC00612(1) [NGA-e (TFID Software 3.4.X)] 06/01
NGA 2000
3 - 3
3.4 Lines and Softkey Functionality
Lines can be selected by using the
-key or the
-key. The selected line is displayed
white on black (highlighted). There are four different types of lines in the menus:
Menu line... / Menu Softkey...
Line/Softkey description ending with three dots.
You will enter a submenu/next menu by pressing the softkey resp. by pressing the
-key or the
-key while the menu line is highlighted.
Function line / Function Softkey !
Line/Softkey description ending with an exclamation-mark.
You will start a function (e.g. calibration) by pressing the softkey resp. by pressing the
-key or the
-key while function line is highlighted.
Line of variables:
Line ending with a colon.
View module parameters (variables).
Some parameters can be changed (e.g. begin of range). Other parameters showing a
status (e.g. temperature) are for information only and cannot be changed. They are
displayed below a separation line within the menu.
Text line
Line without any punctuation marks.
Giving additional information only (such as headlines a.s.o.).
The following illustrations explain the function of lines and softkeys as described above:
Measure Status... Channel Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Basic Controls and Setup --
Calibration procedure status...
Start s
p
an calibration
p
rocedure!
Operation status: Ready
Start zero calibration procedure!
Range upper limit: 100 ppm
Valves...
Span gas: 100 ppm
Range number:
1
Check calibration deviation:
Enabled
Menu Line
Function Lines
Current channel measurement signal (see TAG).
It is always shown on the single component display.
Lines of Variables
(The last one can not
be changed – belo
w
the separation line)
Menu Headline
3 - 4
NGA 2000
ETC00612(1) [NGA-e (TFID Software 3.4.X)] 06/01
3.5 Important Function Softkeys
Display
Switch from the single component display to the multi component display.
F1 in the single component display.
Measure
Switch from any menu or submenu to the single component display of the selected
channel.
F1 in any screen except the single component display.
Status (see section 4.1 page 4-3!)
Switch to the menu "Analyzer Channel Status":
Shows the most important parameters and information about the status of the current
channel or module.
If available: F2.
Main (see section 3.8 page 3-7!)
Switch from the single component display to the main menu.
F3 in the single component display.
Channel
Scroll through the channels in the current menu. In the main menu and the single
component display you may scroll through all channels of the connected analyzer /
analyzer modules. In submenus this key is without function although it is displayed (only
activated in MLT software). TFID is a single channel analyzer or AM only!
If available: F3 (F4 in the single component display).
Measure
Cancel !
Back...
Mor e...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Calibration Procedure State --
Procedure state:
Ready
Valve position:
Samplegas
Concentration in span gas units:
37.50 ppm
Last zero calibration:
Success
Last span calibration:
Success
Last span calibration was:
Fri 05-29-1998 13:37:23
Successful zero+span calibrated ranges:
1+2+3+4
Maximum remaining procedure time:
0 s
Last zero calibration was:
Fri 05-29-1998 13:32:06
Channel
Text Line
Menu Softkeys
Function Softkey
3
Startup and Operation, General Notes and Main Menu
ETC00612(1) [NGA-e (TFID Software 3.4.X)] 06/01
NGA 2000
3 - 5
Lock
Lock all three operation levels at the main menu if a security code is activated in the
system configuration and diagnostic menu (see section 6.4 pg. 6-8).
F4 in the main menu.
BasicCal (see section 4.4 pg. 4-19 and 5.1.1 pg. 5-15!)
Switch from the single component display to the menu "Analyzer module calibration".
F5 in the single component display.
MFG Data (see section 3.8 pg. 3-7/8!)
Switch from the main menu to the menu "Module Manufacturing Data":
Additional submenus are available with information about the control module and
analyzer module data, such as manufacturer address , modules serial numbers or
software and hardware revisions.
F5 in the main menu.
Back
Return to the previous menu page (alternative:
-key)
reset a changed but not confirmed parameter to the previous value.
If available: F4 for returning back, F2 for reset.
More
Entering the next page of the current menu.
If available: F5.
3.6 Entering/Changing Variables
-key
If you have already selected a line of variables (highlighted) and press the
-key, the
parameter will be selected and may be edited.
If you press the
-key again, the new value will be confirmed.
-key /
-key
Function depends on the variable selected: - Changing the parameter values
- Scrolling among variables selected
- Changing single digits or characters
- Increasing or decreasing numbers.
-key /
-key
Selects single digits within a number.
The number of digits or characters of some variables may be changed.
3 - 6
NGA 2000
ETC00612(1) [NGA-e (TFID Software 3.4.X)] 06/01
3.7 Executing a Function
If you press the
-key or the
- key while a function line is highlighted the software will
ask for a confirmation:
If you press the F2 -key the function will start immediately.
Pressing the F4 -key will return to the previous menu page.
Note:
Confirmation of function line entries may be disabled:
Select “Measurement Display Configuration” from the Analyzer and I/O, expert controls
& setup menu (see section 5.1.8, page 5-49).
Change “Display confirmation menus” to “No”.
Now the selected function will start immediately after entering. No confirmation is
required.
Yes Back...
TAG
F1 F2 F3 F4 F5
37.50 ppm
-- Confirmation Required --
Do you really want to do this ??
Press "Yes" or "Back..."
3
Startup and Operation, General Notes and Main Menu
ETC00612(1) [NGA-e (TFID Software 3.4.X)] 06/01
NGA 2000
3 - 7
3.8 Main Menu
If you press the F3-key (Main...) or the
-key in any single component display, you will
switch to the "Main Menu". From there you can switch to all operating levels of your
TFID/MLT analyzer / analyzer module to set up and control the parameters of
measurement, calibration and data transfer!
Via the F5-key (MFG Data) you may enter several submenus where you will find a lot of
important data about the control module (TFID/MLT analyzer / platform) and the analyzer
module, such as service address or serial number!
Ramifications from the Main Menu:
F1
Enter the current channel’s single component
display
Section 3.1, pg. 3-1/2 !
F2
Enter the current channel’s "Analyzer Channel
Status" menu.
Section 4.1, pg. 4-3 !
F3
Scroll through all channels of connected
analyzers / analyzer modules
See channel tag !
F4
Lock any operating level by security code
Section 6.4, pg. 6-8 !
F5
Enter the menu "Manufacturing Data"
See next pages !
See chapter 4 !
See chapter 5 !
See chapter 6 !
See chapter 7 !
Setup: see 6.3 pg. 6-7
Factory setting
Measure Status... Channel Lock... MFG Data
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Main Menu --
Analyzer basic controls (calibration) & setup...
Analyzer and I/O, expert controls & setup...
System configuration and diagnostics...
Display controls...
Time & Date: 14:01:45 25 May 1999
System tag: Fisher-Rosemount
3 - 8
NGA 2000
ETC00612(1) [NGA-e (TFID Software 3.4.X)] 06/01
Ramifications from the menu "Manufacturing Data":
1. Control module data:
2. Analyzer module data:
Measure Or... Back... More...
TAG
F1 F2 F4 F5
95.00 ppm
Manufactured by:
Fisher-Rosemount GmbH & Co
Industriestrasse 1
D-63594 Hasselroth / Germany
FAX. (+49) 6055 884-209
(C) Copyright Fisher-Rosemount GmbH & Co, 2001
Tel. (+49) 6055 884-0
F3
Me a su r e
<<<
Back...
>>>
TAG
F1
F2
F3
F4
F5
95.00 ppm
-- Manufacturing Data --
Control modul e manufacturing data...
Analyzer module manufacturing data...
Measure Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
Manufactured by:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, CA 92807-1802 /USA
FAX: (714) 577-8739
(C) Copyright Fisher-Rosemount Analytical Inc., 2001
Tel: (714) 986-7600
Measure Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Control Module Version Information --
Serial number: CM 4711
Hardware revision: ACU02 R: 3.2.4, Final D:Jan 1
Software revision: 3.4/P010
Revision date: Nov 2 2000
Revision time: 16:09:49
Phrase dictionary version:
P010/01/00
Manufacturing date: 14.02.2001
Language: English
Measure
<<<
Back...
>>>
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Module Manufacturing Data --
Control module data...
Analyzer module data...
Measure Back... More...
TAG
F1 F2 F3 F4 F5
95.00 ppm
Manufactured by:
Fisher-Rosemount GmbH & Co
Industriestrasse 1
D-63594 Hasselroth / Germany
FAX. (+49) 6055 884-209
(C) Copyright Fisher-Rosemount GmbH & Co, 2001
Tel. (+49) 6055 884-0
Measure Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Analyzer Module Version Information --
Serial number: AM 4712
Hardware revision: ACU02 R: 3.2.4, Final D:Jan 1
Software revision: 3.4.1 / P010/Ch1
Revision date: Nov 2 2000
Revision time: 16:25:28
Manufacturing date: 14.02.2001
More...
Measure Back...
TAG
F1 F2 F3 F4 F5
95.00 ppm
-- Hardware Configuration --
Measurement system: PSV-System
Serial interface adapter: RS-232
Heater installed: No
Local DIO module installed: 2
Sensor system revision:
RAM-memory: 763296 Bytes
Channel
Local SIO module installed: Enabled
Sensor system serial number:
04011999
311976
4 Analyzer Basic Controls (
Calibration) & Setup
ETC00612(1) [NGA-e (TFID SW 3.4.X)] 02/01
NGA 2000
4 - 1
Chapter 4 "Analyzer basic controls (calibration) & setup" describes the most important
measurement and calibration functions of your TFID/MLT analyzer / analyzer module.
All steps are figured with detailed illustrations and operation instructions. In the left column
you can see display and keyboard of the NGA front panel. The keys you have to press are
illustrated in black. In the right column you can read the instructions and notes. All
instructions will begin with any single component display and will end with the correspond-
ing single component display after the setups are done. So you can easily compare the
actual display of the analyzer / analyzer module with the illustrations of this manual.
Example:
You want to change from the single component display of the TFID (propane) to the single
component display of the MLT channel 2 (CO).
Picture one shows the starting situation: single component display of propane.
Picture two shows the result you get if you press the F4 -key (Channel):
single component display of CO.
Left column: Right column:
Display and keyboard Instructions and notes
Display Status... Main... Channel BasisCal
TFID-R1
F1 F2 F3 F4 F5
Failures: No
Maintenance- No
Temperature: 20.0 C
Operation: Ready
Range: 1
37.50 ppm Propane
0.00 50.00
0.0 100.0
Display Status... Main... Channel BasisCal
MLT25/CH2/R2
Temperature: 25.0 C
Maintenance-
No
F1 F2 F3 F4 F5
95.00 ppm CO
0 250Range: 2
0.0 100.0
Any_Alarms: No
Operation: Ready
Switch to the single
component display
of another channel
Press
F4
Notice the tag in the
display's upper left corner:
It changed from TFID to
MLT2 indicating a different
analyzer / analyzer module.
Next instruction
or step may be
entered.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200
  • Page 201 201
  • Page 202 202
  • Page 203 203
  • Page 204 204
  • Page 205 205
  • Page 206 206
  • Page 207 207
  • Page 208 208

Rosemount NGA 2000 TFID Hydrocarbon Analyzer Module SW 3.4 Software-1st Ed. Owner's manual

Category
Water pumps
Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI