Mettler Toledo O2 Sensors InPro6850i, InPro6900i and InPro6950i Operating instructions

Type
Operating instructions
InPro 6850 i / 6900 i / 6950 i
52 206 349 F
InPro 6850 i, 6900 i & 6950 i
Series O
2
Sensors
Instruction manual
Bedienungsanleitung
Instructions d’utilisation
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
2 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
English Page 3
Deutsch Seite 42
Français Page 82
InPro is a registered trademark of the METTLER TOLEDO Group in
Switzerland and a further twelve countries.
InDip, InFit, ISM and InTrac are registered trademarks of the
METTLER TOLEDO Group.
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 3
InPro 6850 i, 6900 i & 6950 i
Series O
2
Sensors
Instruction manual
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
4 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
Contents
1. Introduction 5
2. Important notes 6
2.1 Notes on operating instructions 6
2.2 Intended use 6
2.3 Safety instructions 7
2.4 Environmental protection 8
2.5 Examples of some typical applications 9
2.6 Use in Ex-zones 9
2.7 Ex-classicationATEX 10
2.7.1 Introduction 10
2.7.2 Rated data 10
2.7.3 Special conditions 11
2.8 Ex-classicationFMapproved 12
2.8.1 Introduction 12
2.8.2 Rated data 13
2.8.3 Special conditions 13
2.8.4 Appliedstandards: 14
3. Product description 15
3.1 General information 15
3.2 Principle 15
3.3 Scope of delivery 16
3.4 Equipment features 17
4. Installation 18
4.1 Mounting the sensor 18
4.2 Connection 18
4.2.1 ConnectingthesensortoanAK9cable 18
4.2.2 ConnectingtheAK9cabletothetransmitter 19
5. Operation 20
5.1 Start-up and polarizing 20
5.2 Calibration 21
5.2.1 Purposeofcalibration 21
5.2.2 Whatyouhavetoknowforcalibration 21
5.2.3 Singlepointcalibration 22
5.2.4 Dualpointcalibration 22
6. Maintenance 23
6.1 Inspection of the sensor 23
6.1.1 Visual inspection 23
6.1.2 Testing the METTLER TOLEDO O
2
sensor master 24
6.1.3 Testing the sensor via a transmitter 25
6.1.4 ISM design 26
6.2 Changing the electrolyte,
themembranebodyortheinteriorbody 26
7. Storage 29
8. Productspecication 29
8.1 Certicates 29
8.2 Specications 30
9. Ordering information 32
9.1 Sensors with ISM functionality 32
9.2 Accessories 32
9.3 Spare parts 32
9.4 Recommended transmitters 33
9.5 Recommended housings 34
10. Theory of the polarographic sensor 35
10.1 Introduction 35
10.2 Principle of the design of an oxygen electrode 35
10.3 Parameters determining current 39
10.4 Polarization voltage 39
10.5 Temperature 40
10.6 Dependenceonow 40
10.7 Oxygenpartialpressure–oxygenconcentration 40
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 5
1. Introduction
ThankyouforbuyingthisInPro
®
6850 i / 6900 i /
6950 i sensor from METTLER TOLEDO.
The construction of the InPro series employs lead-
ing edge tech nology and complies with safety
regulations currently in force. Notwithstanding this,
improper use could lead to hazards for the user or
a third-party, and/or adverse effects on the plant or
other equipment.
a
Therefore, the operating instructions must be read
and understood by the persons involv ed before
work is started with the sensor.
Thisinstructionmanualmustalwaysbestoredclose
athand,inaplaceaccessibletoallpeopleworking
with the InPro sensor.
Ifyouhavequestions,whicharenotorinsufciently
answered in this instruction manual, please contact
yourMETTLERTOLEDOsupplier.Hewillbegladto
assist you.
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
6 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
2. Important notes
2.1 Notes on operating instructions
These operating instructions contain all the in-
formation needed for safe and proper use of the
InPro 6850 i / 6900 i / 6950 i sensor.
The operating instructions are intended for personnel
entrusted with the operation and maintenance of the
sensor. It is assumed that these persons are familiar
with the equipment in which the sensor is installed.
Warning notices and symbols
Thisinstructionmanualidentiessafetyinstructions
andadditionalinformationbymeansofthefollowing
symbols:
a
This symbol draws attention to safety instruc-
tions and warnings of potential danger which, if
neglected, could result in injury to persons and/or
damage to property.
h
Thissymbolidentiesadditionalinformationand
instructions which, if neglected, could lead to
defects,inefcientoperationandpossiblelossof
production.
2.2 Intended use
METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i sen-
sors are intended solely for inline measurement of
the oxygen partial pressure in liquids and gases,
as described in this instruction manual.
Anyuseofthesesensorswhichdiffersfromor
exceedsthescopeofusedescribedinthisinstruc-
tionmanualwillberegardedasinappropriateand
incompatiblewiththeintendedpurpose.
Themanufacturer/supplieracceptsnoresponsibil-
ity whatsoever for any damage resulting from such
improperuse.Theriskisborneentirelybytheuser/
operator.
Otherprerequisitesforappropriateuseinclude:
compliance with the instructions, notes and
requirements set out in this instruction manual.
– acceptanceofresponsibilityforregularinspec-
tion, maintenance and functional testing of all
associ ated components, also including compli-
ance with local operational and plant safety
regulations.
compliance with all information and warn-
ings given in the documentation relating to the
products used in conjunction with the sensor
(housings, transmitters, etc).
– observanceofallsafetyregulationsgoverning
the equipment in which the sensor is installed.
correct equipment operation in conformance with
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 7
theprescribedenvironmentalandoperational
conditions,andadmissibleinstallationpositions.
– consultationwithMettler-ToledoProcessAnalyt-
ics in the event of any uncertainties.
2.3 Safety instructions
a
– Theplantoperatormustbefullyawareofthe
potential risks and hazards attached to opera-
tion of the particular process or plant. The opera-
torisresponsibleforcorrecttrainingofthework-
force, for signs and markings indicating sources
ofpossibledanger,andfortheselectionofap-
propriate, state-of-the-art instrumentation.
It is essential that personnel involved in the
commissioning, operation or maintenance
of these sensors or of any of the associated
equipment(e.g.housings,transmitters,etc.)be
properly trained in the process itself, as well as
in the use and handling of the associated equip-
ment. This includes having read and understood
this instruction manual.
The safety of personnel as well as of the plant
itselfisultimatelytheresponsibilityoftheplant
operator. This applies in particular in the case of
plants operating in hazardous zones.
The oxygen sensors and associated components
have no effect on the process itself and cannot
inuenceitinthesenseofanyformofcontrol
system.
Maintenance and service intervals and schedules
depend on the application conditions, composi-
tion of the sample media, plant equipment and
signicanceofthesafetycontrolfeaturesofthe
measuringsystem.Processesvaryconsiderably,
sothatschedules,wheresucharespecied,can
onlyberegardedastentativeandmustinany
casebeindividuallyestablishedandveriedby
the plant operator.
– Wherespecicsafeguardssuchaslocks,labels,
or redundant measuring systems are necessary,
thesemustbeprovidedbytheplantoperator.
– Adefectivesensormustneitherbeinstallednor
put into service.
– Onlymaintenanceworkdescribedinthisoperat-
inginstructionmaybeperformedonthesensors.
When changing faulty components, use only
originalsparepartsobtainablefromyour
METTLER TOLEDO supplier (see spare parts list,
“Section 9.3).
– Nomodicationstothesensorsandtheacces-
sories are allowed. The manufacturer accepts no
responsibilityfordamagescausedbyunauthor-
izedmodications.Theriskisborneentirelyby
the user.
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
8 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
Whenusingcablesforsensorconnectionnot
suppliedandrecommendedbythemanufacturer
thefollowingstandardsneedtobeconsidered:
a
US National Electrical Code
®
(ANSI/NFPA70
[NEC
®
]),whereapplicable.
Canadian Electrical (CE) Code
®
(CEC Part 1,
CAN/CSA-C22.1),whereapplicable.
− ForguidanceonUSinstallations,seeANSI/
ISA-RP12.06.01,InstallationofIntrinsicallySafe
SystemsforHazardous(Classied)Locations.
− Controlroomequipmentconnectedtointrinsi-
cally safe associated apparatus in the US should
notuseorgeneratemorethanthespeciedUm
of the associated apparatus.
− Caremustbetakenduringinstallationtoavoid
impacts or friction that could create an ignition
source.
− Tamperingandreplacementwithnon-factory
components may adversely affect the safe use of
the system.
− Insertionorwithdrawalofremovableelectrical
connectorsormodulesistobeaccomplished
onlywhentheareaisknowntobefreeofam-
mablevapors.
− WARNING
INTRINSICALLYSAFEAPPARATUSCANBEA
SOURCEOFIGNITIONIFINTERNALSPACINGS
ARESHORTEDORCONNECTIONSOPENED.
− WARNING
DONOTOPENWHENANEXPLOSIVE
ATMOSPHEREISPRESENT.
− WARNING
SUBSTITUTIONOFCOMPONENTSMAYIMPAIR
INTRINSICSAFETY.
− WARNING
SUBSTITUTIONOFCOMPONENTSMAYIMPAIR
SUITABILITYOFTHEEQUIPMENT
− WARNING
FORCONNECTIONONLYTONON-FLAMMABLE
PROCESSES.
− WARNING
TOMAINTAINTHEENCLOSUREIP66RATING,
THECONNECTORMUSTBEFULLYENGAGED.
2.4 Environmental protection
Wasteelectricalproductsshouldnotbe
disposed of with household waste. Please
recycle where facilities exist. Check with
yourLocalAuthorityorretailerforrecycling
advice.
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 9
2.5 Examples of some typical applications
Belowisalistofexamplesoftypicaleldsofap-
plication for the oxygen sensors. This list is not
exhaustive.
Measurement in liquids:
– Biotech
Chemical applications
– Brewing
– Beverageltration
Filling stations
Measurement in gases:
– CO
2
recovery
– CO
2
purity
Product storage
Inert production
2.6 Use in Ex-zones
a
Attention! For an installation in Ex-zones please
readtheguidelinesfollowinghereafter:
1258
Ex-classicationATEX:
x
ExiaIICT6/T5/T4/T3Ga/Gb
x
ExiaIIICT69°C/T81°C/T109°C/T161°CDa/Db
Numberofthetestcerticate:
SEV14ATEX0169X
IECExSEV14.0026X
Ex-classicationFMapproved:
IS/I,II,III/1/ABCDEFG/T6Ta=60°C
– 53 800 002; Entity
ProjectID:3021227
FMcerticatenumber:FM16US0034X
FM18CA0021X
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
10 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
2.7 Ex-classicationATEX
2.7.1 Introduction
– AccordingtoDirective2014/34/EU(ATEX95)
Appendixl,lnPro6XXX*/*/*/*/*oxygensensors
are devices group ll, category 1/2G and accord-
ingtoRL99/92/EG(ATEX137)maybeusedin
zones0/1or1/2or1or2andgasgroupsllA,
llBandllCthatarepotentiallyexplosivedueto
combustiblesubstancesinabovestatedtem-
perature classes.
For use / installation, the requirements of EN
60079-14mustbeobserved.
– AccordingtoDirective2014/34/EU(ATEX95)
Appendixl,lnPro6XXX*/*/*/*/*oxygensensors
are devices group ll, category 1/2D and accord-
ingtoRL99/92/EG(ATEX137)mayalsobe
used in zones 20/21 or 21 or 22 that contain
combustibledusts.
The digital sensor circuit is part of a common
intrinsically safe system and is for operation con-
nectedtoaseparatelycertiedtransmitter.
The digital sensor circuit as part of an intrinsi-
cally safe system is isolated from the not-intrinsi-
cally safe electric circuits up to a maximum rated
voltage of 375 V and from grounded parts up to
a maximum rated voltage of 30 V.
2.7.2 Rated data
Measuring circuit
Fail-safe ignition protection class Ex ia IIC only for
connectiontoacertiedfail-safecircuit.
Digital O
2
oxygen sensor maximum values
Two-wirecurrentcircuit:
U
i
≤16V
I
i
≤30mA
P
i
≤50mW
L
i
=negligible
C
i
=negligible
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 11
2.7.3 Special conditions
– Therelationshipbetweenthemaximumper-
missibleambientormediatemperatureand
temperature class, for category 1G applications,
zone0,isshowninthefollowingtable:
Temperature class Max. ambient or
media temperature
T 6 68 °C
T 5 80 °C
T 4 108 °C
T 3 160 °C
– Therelationshipbetweenthemaximumper-
missibleambientormediatemperatureand
temperature class, for category 1D applications,
zone20,isshowninthefollowingtable:
Temperature class Max. ambient or
media temperature
T 69 °C 68 °C
T 81°C 80 °C
T 109 °C 108 °C
T 161°C 160 °C
The capacitance and inductance of the connect-
ingcablehastobeconsidered.
The O
2
oxygensensortypeInPro6XXXcan
beusedin/withthettingsInFit
®
76*-***or
InTrac
®
7**-***,orin/withothersuitablettings
in potentially explosive areas.
– ThemetalbodyoftheO
2
oxygen sensors, or the
ttingsInFit76*-***orInTrac7**-***,orother
appropriatettingisoptionallyincludedinthe
routine pressure test of the system.
– Theindependentttingusedforinstallationofthe
O
2
oxygensensormustbeconductivelycon-
nectedtotheequipotentialbondingsystem.
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
12 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
2.8 Ex-classicationFMapproved
2.8.1 Introduction
The following FM control drawing and the standards
listedinsection2.8.4mustbeobserved,whereap-
plicable:
aapppprroovveeddaapppprroovveedd
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 13
2.8.2 Rated data
Forrateddatapleaseobservesection2.7.2
2.8.3 Special conditions
In intrinsicallysafeapparatus,theanalyticalprobes
(InPro6000SerieswithInFit76X/YandInTrac7XX/
YYSerieshousings)equipmentisdesignatedwith
thefollowingspecicconditionsofuse.
1.Theoxygensensorshallbeinstalledincompli-
ance with the enclosure, mounting, spacing and
segregation requirements of the ultimate applica-
tion,includingatoolremovablecover,andis
suitableforusewithFMApprovedInFit76X/Yand
InTrac7XX/YYSerieshousings.
2. Process temperature no greater than +130 ºC.
3.Maximumpermissibleworkingpressureis
12barg(174psig).
4. Potential Electrostatic Charging Hazard –
To prevent the risk of electrostatic sparking, the
non-metallicsurfaceshouldonlybecleanedwith
a damp cloth.
5. Enclosures containing titanium constitute a
potentialriskofignitionbyimpactorfriction.
Caremustbetaken during installation and use to
prevent impact or friction.
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
14 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
2.8.4 Applied standards:
United States Standards
Title Number Issue
Date
ApprovalStandardfor
Electrical Equipment
for Use in Hazardous
(Classied)Locations–
General Requirements
FM Class 3600 2011
ApprovalStandard
for Intrinsically Safe
Apparatusand
AssociatedApparatus
for Use in Class I, II & III,
Division 1, Hazardous
(Classied)Locations
FM Class 3610 2015
ApprovalStandardfor
Electrical Equipment for
Measurement, Control
andLaboratoryUse
FM Class 3810 2005
ExplosiveAtmospheres–
Part0:Equipment–
General Requirements
ANSI/ISA-60079-0
(12.00.01)
2005
ExplosiveAtmospheres–
Part11:Equipment
ProtectionbyIntrinsic
Safety “i”
ANSI/ISA-60079-
11
(12.02.01)
2009
Safety Requirements for
Electrical Equipment
for Measurement,
Control,andLaboratory
Use–Part1:General
Requirements
ANSI/ISA-61010-1
(82.02.01)
2004
Canadian Standards
Title Number Issue
Date
ExplosiveAtmospheres–
Part0:Equipment–
General Requirements
CAN/CSA-C22.2
No. 60079-0
2005
ExplosiveAtmospheres–
Part11:Equipment
ProtectionbyIntrinsic
Safety “i”
CAN/CSA-C22.2
No. 60079-11
2009
Safety Requirements for
Electrical Equipment
for Measurement,
Control,andLaboratory
Use–Part1:General
Requirements
CAN/CSA-C22.2
No. 61010-1
2004
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 15
3. Product description
3.1 General information
The oxygen sensor series InPro 6850 i / 6900 i /
6950 i withintegratedtemperatureprobeareused
for meas ure ment of oxygen at low and medium
concentrations.
The sensors are sterilizable, and in the majority of
cases arealsoautoclavable.They are compatible
with CIP (Cleaning-In-Place).
InPro 6xxx i sensors with ISM
®
functionality offer
Plug and Measure as well as enhanced diagnostics
features.
3.2 Principle
Amperometric oxygen sensors:
The amperometric oxygen sensors of METTLER
TOLEDOareallbasedontheClarkmeasuringprinci-
ple. However different series are offered, which differ
inthenumberandarrangementoftheirelectrodes
andthusintheirspecications:
The measurement system of the InPro sensors
consists of a working electrode (cathode), a
counter electrode (anode), a reference electrode
and a guard ring electrode. The measurement
system is separated from the process medium
byanoxygenpermeablemembrane.
The measuring technique in principle is alike with
all sensors.
The transmitter supplies a constant voltage
betweencathodeandanode.
The electrolyte creates a conductive connection
betweentheelectrodes.
The oxygen molecules migrate from the mea-
surementmediumthroughthemembranetothe
cathode to which the voltage is supplied and are
reduced.Atthesametime,oxidationtakesplace
at the anode.
• Thiscausesacurrenttoowbetweentheanode
and cathode which is directly proportional to the
partial pressure of oxygen (pO
2
) in the process
medium. With digital ISM sensors, the sensor con-
verts the current into oxygen concentration and
communicates the value to the transmitter.
The guard ring electrode of the InPro 6900 i and
6950ibuiltintothesensorreducestheoxygen
that migrates to the side of the cathode and that
can distort the measurement. The guard ring
electrodethereforeenablestracesofoxygento
bepreciselydeterminedateventhelowestcon-
centrations.
h
Note: Please refer to “Section 10 – Theory of the
polarographic sensor” for further information.
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
16 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
ISM sensors:
Alloxygensensorswiththeindex“i”(6850i,6900i,
6950 i) are equipped with ISM.
Principle:Inthesensorheadachipisintegrated,
which takes over the entire monitoring and control of
thesensorand,beyondthat,storesallsensordata.
Thefollowingdataareavailablepermanentlyinthe
sensor:
type of sensor
serial no.
software version
hardware version
order no.
operating time
– calibrationtimeandcalibrationdate
– calibrationtable
To check the system, the following indicators are
supervised:
– temperature
– slope
zero current
air current
polarization voltage
Basedonthisinformation,thetransmittercalculates
a wear monitor and displays it depending upon
transmitter type. (see the respective instruction
manuals)
ISMenablestheconnectionofthedigitalsensorto
iSense. This user-friendly software allows manage-
ment of all sensor information and storage of it in a
database.Furthermore,thesensorcanbecalibrated
via the iSense software.
3.3 Scope of delivery
Eachsensorissuppliedfullyassembledand
factory-testedforcorrectfunctiontogetherwith:
– aqualitycontrolcerticate
– inspectioncerticates3.1
(complying with EN 10204)
METTLER TOLEDO ISM O
2
sensors are supplied
withttedmembranebodybutwithoutelectrolyte,
coveredwiththeprotectioncap.Theyhavebeen
checked for proper functionality. See electrolyte order
details at section “Spare Parts“, 9.3.
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 17
3.4 Equipment features
12 mm sensor
Anode (Pt)
Cathode and Guard Ring
(InPro 6900i;6950i)
Protection Cap
Protection Cap Sleeve (N-type)
O-ring (silicone FDA/USP VI)
Membrane Body
Retainer Nut
Reference (Ag/AgCl)
Interior Body
O-ring
(9.0 1.0 mm, silicone FDA/USP VI)
Pg 13.5 Threaded Sleeve
AK9 Connector
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
18 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
4. Installation
4.1 Mounting the sensor
a
Important! Remove the protection cap before
mount ing the sensor.
Mounting the sensor in a housing
Please refer to the instruction manual of your hous-
ing explaining on how to mount the sensor in place.
Mounting the sensor directly in a pipe or a vessel
The12mmsensorscanbemounteddirectly
through a socket with inside thread Pg 13.5 and
securely tightened via the Pg 13.5 threaded sleeve.
4.2 Connection
4.2.1 Connecting the sensor to an AK9 cable
h
ThesensorisconnectedtothetransmitterviaanAK9
cable.TheAK9cableensuresasecureconnection
betweenthetransmitterandthesensorunderharsh
industrialconditions.TherobustwatertightIP68con-
nector housing ensures maximum process safety.
a
Do not touch the sensor at the AK9 connector plug!
Tightly screw the plug to fasten the two parts.
Admissible mounting position
Zulässige Einbaulage
Positionement de montage admis
Inadmissible angle
Unzulässiger Winkel
Angle non admissible
52910094ME
15°1
© 11 / 2020 METTLER TOLEDO InPro 6850 i / 6900 i / 6950 i
Printed in Switzerland 52 206 349 F
InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm 19
4.2.2 Connecting the AK9 cable to the transmitter
O
2
Transmitter
AK9 or VP Cable
h
Note: ForconnectingtheAK9cabletotheterminals
of the transmitter, please refer to the instructions
given in the METTLER TOLEDO transmitter manual.
ISM Sensor
AK9 Connector
Plug
AK9 Cable for
Standard Use
InPro 6850 i / 6900 i / 6950 i © 11 / 2020 METTLER TOLEDO
52 206 349 F Printed in Switzerland
20 InPro 6850 i / 6900 i / 6950 i O
2
Sensor 12 / 25 mm
5. Operation
a
Important!Beforeusingthesensorfortherst
time, electrolyte must be added to the membrane
body (see “Chapter 6.2).
5.1 Start-up and polarizing
a
Important! The protection cap must be removed
before mounting the sensor in the process.
Whenthesystemisoperatedforthersttimeorif
thesensorhasbeendisconnectedfromthevoltage
source (transmitter or O
2
sensor master) for longer
than5minutes,thesensorhastobepolarizedprior
tocalibrationbyconnectingittotheoperatingO
2
transmitterortoasensormaster.After6hours,the
sensor is fully polarized and ready for operation.
h
Attention: The InPro 6950 i should never be
polarized in air!
During this time, it is recommended to keep the pro-
tectioncaplledwithfreshconditioningsolutionon
the sensor (see “Section 7 – Storage”), especially if
thesensorisbeingpolarizedforlongerthan6hours
(Exemption:theISMversionInPro6850ireaches
polarization after 1h and is then ready for operation).
Ashorterpolarizationperiodissufcientifthesensor
hasbeendisconnectedforonlyafewminutes.The
followingtableservestoestablishthecorrectpolar-
ization time in relation to the depolarization time.
Depolarization time
1
t
depol
[Min.]
Minimum required
polarization time
2
[Min.]
InPro 6900 i;
InPro 6950 i
InPro 6850 i
> 30 360 60
15 to 30 6 × t
depol
45
5 to 15 4 × t
depol
30
< 5 6 × t
depol
0
1 Depolarizationtime:Timespaninwhichthepo-
larization voltage is cut off from the sensor. This
isthecase:
–duringthetimethecableisdisconnectedorno
transmitter or sensor master is connected to
thecable,orthetransmitterhasbeendiscon-
nected from the current supply.
–afterchangingtheelectrolyteormembrane
body.Inthiscaseatleast6hrs(InPro6850i;
1 h) of polarization must follow.
2 Polarizationtime:Timespanduringwhichthe
sensor is under a polarization voltage.
a
Important! Setting of the polarization voltage on
the transmitter for correct measurements:
– StandardapplicationsInPro6850i:– 675 mV
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124