Teledyne 6400EH User manual

  • Hello! I am an AI chatbot trained to assist you with the Teledyne 6400EH User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Teledyne Analytical Instruments
OPERATING INSTRUCTIONS FOR
Model 6400EH
UV Fluorescence SO
2
Analyzer
DANGER
Dependingon your application, toxic gases may be present in this monitoring
system.
Personal protective equipment may be required when servicing this instrument.
Hazardous voltages exist on certain components internally which may persist
for a time even after the power is turned off and disconnected.
Only authorized personnel should conduct maintenance and/or servicing.
Before conducting any maintenance or servicing, consult with authorized
supervisor/manager.
P/N M77713
Model 6400EH
Teledyne Analytical Instruments ii
Copyright © 2008 Teledyne Instruments/ Analytical Instruments
All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed,
stored in a retrieval system, or translated into any other language or computer language in
whole or in part, in any form or by any means, whether it be electronic, mechanical,
magnetic, optical, manual, or otherwise, without the prior written consent of Teledyne
Instruments/ Analytical Instruments, 16830 Chestnut Street, City of Industry, CA 91749-
1580.
Warranty
This equipment is sold subject to the mutual agreement that it is warranted by us free from
defects of material and of construction, and that our liability shall be limited to replacing or
repairing at our factory (without charge, except for transportation), or at customer plant at
our option, any material or construction in which defects become apparent within one year
from the date of shipment, except in cases where quotations or acknowledgements provide
for a shorter period. Components manufactured by others bear the warranty of their
manufacturer. This warranty does not cover defects caused by wear, accident, misuse,
neglect or repairs other than those performed by TI/AI or an authorized service center. We
assume no liability for direct or indirect damages of any kind and the purchaser by the
acceptance of the equipment will assume all liability for any damage which may result from
its use or misuse.
We reserve the right to employ any suitable material in the manufacture of our apparatus,
and to make any alterations in the dimensions, shape or weight of any parts, in so far as
such alterations do not adversely affect our warranty.
Important Notice
This instrument provides measurement readings to its user, and serves as a tool by which
valuable data can be gathered. The information provided by the instrument may assist the user
in eliminating potential hazards caused by his process; however, it is essential that all
personnel involved in the use of the instrument or its interface, with the process being
measured, be properly trained in the process itself, as well as all instrumentation related to it.
The safety of personnel is ultimately the responsibility of those who control process
conditions. While this instrument may be able to provide early warning of imminent
danger, it has no control over process conditions, and it can be misused. In particular, any
alarm or control systems installed must be tested and understood, both as to how they
operate and as to how they can be defeated. Any safeguards required such as locks, labels,
or redundancy, must be provided by the user or specifically requested of TI/AI at the time
the order is placed.
Therefore, the purchaser must be aware of the hazardous process conditions. The purchaser
is responsible for the training of personnel, for providing hazard warning methods and
instrumentation per the appropriate standards, and for ensuring that hazard warning devices
and instrumentation are maintained and operated properly.
Teledyne Instruments/ Analytical Instruments, the manufacturer of this instrument, cannot
accept responsibility for conditions beyond its knowledge and control. No statement
expressed or implied by this document or any information disseminated by the
manufacturer or its agents, is to be construed as a warranty of adequate safety control under
the user’s process conditions.
UV Fluorescence SO2 Analyzer
Teledyne Analytical Instruments iii
Specific Model Information
Instrument Serial Number: _______________________
Instrument Range: _______________
Calibrated for: _______________
Background Gas: _______________
Zero Gas: _______________
Span Gas: _______________
Model 6400EH
Teledyne Analytical Instruments iv
Safety Messages
Your safety and the safety of others is very important. We have
provided many important safety messages in this manual. Please read
these messages carefully.
A safety message alerts you to potential hazards that could hurt you
or others. Each safety message is associated with a safety alert symbol.
These symbols are found in the manual and inside the instrument. The
definition of these symbols is described below:
GENERAL WARNING/CAUTION: Refer to the instructions
for details on the specific danger. These cautions warn of
specific procedures which if not followed could cause bodily
Injury and/or damage the instrument.
CAUTION: HOT SURFACE WARNING: This warning is
specific to heated components within the instrument. Failure
to heed the warning could result in serious burns to skin and
underlying tissue.
WARNING: ELECTRICAL SHOCK HAZARD: Dangerous
voltages appear within this instrument. This warning is
specific to an electrical hazard existing at or nearby the
component or procedure under discussion. Failure to heed this
warning could result in injury and/or death from
electrocution.
Technician Symbol: All operations marked with this symbol
are to be performed by qualified maintenance personnel only.
CAUTION: THE ANALYZER SHOULD ONLY BE USED FOR THE
PURPOSE AND IN THE MANNER DESCRIBED IN
THIS MANUAL.
IF YOU USE THE ANALYZER IN A MANNER OTHER
THAN THAT FOR WHICH IT WAS INTENDED,
UNPREDICTABLE BEHAVIOR COULD RESULT
POSSIBLY ACCOMPANIED WITH HAZARDOUS
CONSEQUENCES.
UV Fluorescence SO2 Analyzer
Teledyne Analytical Instruments v
This manual provides information designed to guide you through
the installation, calibration operation and maintenance of your new
analyzer. Please read this manual and keep it available.
Occasionally, some instruments are customized for a particular
application or features and/or options added per customer requests.
Please check the front of this manual for any additional information in
the form of an Addendum which discusses specific information,
procedures, cautions and warnings that may be peculiar to your
instrument.
Manuals do get lost. Additional manuals can be obtained from
TI/AI at the address given in the Appendix. Some of our manuals are
available in electronic form via the internet. Please visit our website at:
www.teledyne-ai.com.
Model 6400EH
Teledyne Analytical Instruments vi
Table of Contents
List of Figures .............................................................................. ix
List of Tables ................................................................................ xi
Documentation .............................................................................. 1
1.1. Using This Manual Addendum 2
Specifications, Approvals & Warranty ........................................ 5
2.1. Specifications 5
2.2. CE Mark Compliance 6
2.3. Warranty 7
Getting Started .............................................................................. 9
3.1. Unpacking and Initial Setup 9
3.1.1. Electrical Connections: 10
3.1.1.1. External Pump 10
3.2 Pneumatic Connections 11
3.2.2 Pneumatic Connections to the 6400EH 11
3.2.2.1 Pneumatic Connections to 6400EH Basic
Configuration 11
3.2.2.2 Connections with Internal Valve Options Installed 12
3.2.3 6400EH Layout 13
3.3 Initial Operation 15
3.3.1 Warning Messages 15
3.3.2 Test Functions 15
3.3.3 Interferents for SO2 Measurements 15
Frequently Asked Questions ...................................................... 17
Optional Hardware and Software ............................................... 19
5.1 Zero/Span Valves (Option 50) 19
5.2 Internal Zero/Span Gas Generator (Option 51) 20
UV Fluorescence SO2 Analyzer
Teledyne Analytical Instruments vii
5.3 Zero and Two Span Point Valve Option (OPT 52) 20
5.4 Hydrocarbon Kicker Option (OPT 65) 22
Operating Instructions ................................................................ 25
6.1 Warning Messages 25
6.2 Test Functions 25
6.2.1 Test Channel Output 26
6.2.2 Range Units 26
6.2.3 Using the 6400EH with a Hessen Protocol Network 26
6.2.4 Default iDAS Channels 28
6.2.5 Remote Operation Using the External Digital I/O 28
6.2.5.1 Status Outputs 28
6.2.5.2 Control Inputs 29
Calibration and Calibration Check Procedures ........................ 31
7.1 Manual Calibration with the Zero and Two Span Point
Valve Option (OPT 52) installed 31
7.2 Manual Calibration Check with the Zero and Two Span
Point Valve Option (OPT 52) installed 33
Instrument Maintenance ............................................................. 35
8.1 Maintenance Schedule 35
8.2 Predictive Diagnostics 35
Theory of Operation .................................................................... 37
9.1 The UV Light Path 37
9.1.1 The Reference Detector 38
9.1.2 Direct Measurement Interferences 38
9.2 Pneumatic Operation 39
9.2.1 Sample Gas Flow 39
9.2.2 Pneumatic Sensors 40
9.2.2.1 Sample Pressure Sensor 40
9.2.2.2 Vacuum Pressure Sensor 40
9.2.2.3 Sample Flow Calculation 40
9.3 Electronic Operation 41
Troubleshooting & Repair .......................................................... 43
Model 6400EH
Teledyne Analytical Instruments viii
10.1.1 Fault Diagnosis with Warning Messages 43
10.1.2 Fault Diagnosis with Test Functions 43
10.2 Subsystem Checkout 43
10.2.1 Pneumatic Sensor Assembly 43
10.3 Repair Procedures 44
10.3.1 Repairing the Sample Gas Flow Control Assembly 44
10.3.2 Sensor Module Repair & Cleaning 46
10.3.2.1 Adjusting the UV Lamp (Peaking the Lamp) 47
10.3.2.2 PMT Hardware Calibration (FACTORY CAL) 49
10.4 Technical Assistance 52
Appendix A — Version Specific Software Documentation ...... 53
Appendix A-1: 6400EH Software Menu Trees, Revision C.0 54
Appendix A-2: Setup Variables For Serial I/O, Revision C.0 62
Appendix A-3: Warnings and Test Functions, Revision C.0 69
Appendix A-4: 6400EH Signal I/O Definitions, Revision C.0 72
Appendix A-5: 6400EH iDAS Functions, Revision C.0 77
Appendix A-6: Terminal Command Designators, Revision C.0 79
Appendix B - 6400EH Spare Parts List ...................................... 81
B-1: 6400EH Spare Parts List 81
B-2: Recommended Spare Parts Stocking Levels Model
6400EH 83
B-3: 6400E/EH Expendables Kit 84
Appendix C Warranty Questionnaire ......................................... 85
Appendix D - Electronic Schematics ......................................... 89
UV Fluorescence SO2 Analyzer
Teledyne Analytical Instruments ix
List of Figures
Figure 3-1: Example of Pneumatic Connections to 6400EH
External Pump ............................................................ 12
Figure 3-2: Pneumatic Connections to 6400EH with Zero and Two
Span Point Valve Option ............................................. 12
Figure 3-3: Internal Pneumatic flow for 6400EH in Basic
Configuration .............................................................. 13
Figure 3-4: 6400EH Layout (Basic Unit – No Valve Options) ........ 14
Figure 3-5: Available Functions in the Model 6400EH ................... 16
Figure 5-1: Pneumatic Diagram of the 6400EH With Z/S Option
Installed. ..................................................................... 19
Figure 5-2: Pneumatic Diagram of the 6400EH With 2-Span Point
Option Installed ........................................................... 21
Figure 5-3: Hydrocarbon Scrubber (Kicker) – OPT 65 .................. 23
Figure 5-4: 6400EH Internal Pneumatic Diagram with
Hydrocarbon Scrubber Installed ................................. 23
Figure 9-1: UV Light Path .............................................................. 37
Figure 9-2: Pneumatic Diagram of the6400EH – Base
Configuration .............................................................. 39
Figure 9-3: 6400EH Electronic Block Diagram .............................. 41
Figure 9-4: Model 6400EH Power Distribution Block Diagram ...... 42
Figure 10-1: Flow Control Assembly ............................................. 45
Figure 10-2: Sensor Module Wiring and Pneumatic Fittings .......... 46
Figure 10-3: Shutter Assembly - Exploded View ........................... 47
Figure 10-4: Location of UV Reference Detector Potentiomete
r .... 49
Figure 10-5: Pre-Amplifier Board Layout ....................................... 50
Figure A-1: Basic Sample Display Menu ....................................... 54
Model 6400EH
Teledyne Analytical Instruments x
Figure A-2: Sample Display Menu - Z/S Valve Option installed .... 55
Figure A-3: Primary Setup Menu (Except iDAS) ........................... 56
Figure A-4: Primary Setup Menu (iDAS) ....................................... 57
Figure A-5: Secondary Setup Menu (COMM & VARS) ................. 58
Figure A-6: Secondary Setup Menu (COMM Menu with Ethernet
Card) .......................................................................... 59
Figure A-7: Secondary Setup Menu - HESSEN Submenu ............ 60
Figure A-8: Secondary Setup Menu (DIAG) .................................. 61
UV Fluorescence SO2 Analyzer
Teledyne Analytical Instruments xi
List of Tables
Table 2-1: Model 6400EH Basic Unit Specifications ....................... 5
Table 3-1: Inlet / Outlet Connector Nomenclature ......................... 11
Table 3-2: Possible Warning Messages at Start-Up ...................... 15
Table 5-1: Two-Point Span Valve Operating States ...................... 22
Table 6-1: Additional 6400EH Warning Messages ........................ 25
Table 6-2: Additional 6400EH Test Functions ............................... 25
Table 6-3: Additional 6400 EH Test Parameters Available for
Analog Output A3 ......................................................... 26
Table 6-4: 6400EH Default Hessen Status Bit Assignments ......... 27
Table 6-5: Status Output Signals ................................................... 28
Table 6-6: Control Input Signals .................................................... 30
Table 8-1: Predictive Uses for Test Functions ............................... 35
Table 10-1: Warning Messages - Indicated Failures ..................... 43
Table 10-2: Test Functions - Possible Causes for Out-Of-Range
Values ......................................................................... 43
Table 10-3: Example of HVPS Power Supply Outputs .................. 48
Table A-1: 6400EH Setup Variables, Revision C.0 ....................... 62
Table A-2: 6400EH Warning Messages, Revision C.0 .................. 69
Table A-3: 6400EH Test Functions, Revision C.0 ......................... 70
Table A-4: 6400EH Signal I/O Definitions, Revision C.0 ............... 72
Table A-5: 6400EH DAS Trigger Events, Revision C.0 ................. 77
Table A-6: 6400EH iDAS Functions, Revision C.0 ........................ 77
Table A-7: Terminal Command Designators, Revision C.0 ........... 79
Table B-1: Model 6400EH Spare Parts List ................................... 81
Table D-1: List of Included Electronic Schematics ........................ 89
Model 6400EH
Teledyne Analytical Instruments xii
DANGER
COMBUSTIBLE GAS USAGE
WARNING
This is a general purpose instrument designed for use in a
non-hazardous area. It is the customer's responsibility to
ensure safety especially when combustible gases are being
analyzed since the potential of gas leaks always exist.
The customer should ensure that the principles of operating
this equipment are well understood by the user. Misuse of
this product in any manner, tampering with its components,
or unauthorized substitution of any component may
adversely affect the safety of this instrument.
Since the use of this instrument is beyond the control of
Teledyne Analytical Instruments, referred as TAI, no
responsibility by TAI, its affiliates, and agents for damage or
injury from misuse or neglect of this equipment is implied or
assumed.
UV Fluorescence SO2 Analyzer Documentation
Teledyne Analytical Instruments 1
Documentation
Note: Throughout this manual, words printed in capital, bold letters,
such as SETUP or ENTR represent messages as they appear
on the analyzer’s front panel display
Note: The flowcharts in this manual contain typical
representations of the analyzer’s display during the
various operations being described. These representations
are not intended to be exact and may differ
slightly from the actual display of your instrument.
Thank you for purchasing the Model 6400EH UV Fluorescence
SO
2
Analyzer.
The electronic manual is in Adobe
®
Systems Inc. “Portable
Document Format”. The Adobe
®
Acrobat Reader® software, which is
necessary to view these files, can be downloaded for free from the
internet at http://www.adobe.com/.
The electronic version of the manual has many advantages:
Keyword and phrase search feature
Figures, tables and internet addresses are linked so that clicking on
the item will display the associated feature or open the website.
A list of chapters and sections as well as thumbnails of each page
are displayed to the left of the text.
Entries in the table of contents are linked to the corresponding
locations in the manual.
Ability to print sections (or all) of the manual
Additional documentation for the Model 6400EH UV Fluorescence
SO
2
Analyzer is available from Teledyne Instruments’ website at
http://www.teledyne-ai.com/manuals/
Documentation Model 6400EH
Teledyne Analytical Instruments 2
APICOM software manual
Multi-drop manual
DAS Manual
1.1. Using This Manual Addendum
This manual addendum has the same overall structure as that of the
6400E operator’s manual, to simplify referring between the two. The
manual has the following sections:
Table of Contents:
Outlines the contents of the addendum in the order the information
is presented. This is a good overview of the topics covered in the
manual. There is also a list of tables, a list of figures and a list of
appendices.
Specifications and Warranty
This section contains a list of the analyzer’s performance
specifications, a description of the conditions and configuration under
which EPA equivalency was approved and Teledyne Instrument’s
warranty statement.
Getting Started:
A concise set of instructions for setting up, installing and running
your analyzer for the first time.
FAQ:
Answers to the most frequently asked questions about operating the
analyzer.
Optional Hardware & Software:
A description of optional equipment to add functionality to your
analyzer.
Operation Instructions:
This section includes step by step instructions for operating the
analyzer and using its various features and functions.
Calibration Procedures:
UV Fluorescence SO2 Analyzer Documentation
Teledyne Analytical Instruments 3
General information and step by step instructions for calibrating
your analyzer.
Instrument Maintenance:
Description of preventative maintenance procedures that should be
regularly performed on you instrument to assure good operating
condition.
Theory of Operation:
This section describes the aspects of 6400EH operation that differ
from the 6400E manual.
Maintenance & Troubleshooting Section:
This section includes pointers and instructions for diagnosing
problems that are specific to the 6400EH. The 6400E manual has a more
complete troubleshooting section, most of which also applies to the
6400EH.
Appendices:
For easier access and better updating, some information has been
separated out of the manual and placed in a series of appendices at the
end of this addendum. These include: software menu trees, warning
messages, definitions of iDAS & serial I/O variables, spare parts list,
repair questionnaire, interconnect listing and drawings, and electronic
schematics.
Documentation Model 6400EH
Teledyne Analytical Instruments 4
UV Fluorescence SO2 Analyzer Specifications/Warranty
Teledyne Analytical Instruments 5
Specifications, Approvals & Warranty
2.1. Specifications
Table 2-1: Model 6400EH Basic Unit Specifications
Min/Max Range
(Physical Analog Output)
In 1 ppb increments from 10ppm to 5,000 ppm, dual
ranges or auto ranging
Measurement Units ppm, mg/m3 (user selectable)
Zero Noise
1
0.05 ppm rms
Span Noise
1
< 0.5% of reading (above 50 ppm)
Lower Detectable Limit
2
0.1 ppm rms
Zero Drift (24 hours) < 1 ppm
Zero Drift (7 days) < 2 ppm
Span Drift (7 Days) < 0.5% FS
Linearity 1 % of full scale
Precision 0.5% of reading1
Temperature Coefficient < 0.1% per oC
Voltage Coefficient < 0.05% per V
Lag Time
1
5 sec
Rise/Fall Time
1
95% in < 30 sec
Sample Flow Rate 700 cm
3
/min. ±10%
Temperature Range 5-40oC
Humidity Range 0 - 95% RH, non-condensing
Dimensions H x W x D 7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)
Weight, Analyzer
(Basic Configuration)
45 lbs (20.5 kg) w/internal pump
Weight, Pump Pack 16 lbs (7 kg)
AC Power Rating
100 V, 50/60 Hz (3.25A); 115 V, 60 Hz (3.0 A);
220 – 240 V, 50/60 Hz (2.5 A)
Environmental
Installation category (over-voltage category) II; Pollution
degree 2
Analog Outputs Three (3) Outputs
Specifications/Warranty Model 6400EH
Teledyne Analytical Instruments 6
Analog Output Ranges
100 mV, 1 V, 5 V, 10 V, 2-20 or 4-20 mA isolated
current loop.
All Ranges with 5% Under/Over Range
Analog Output Resolution 1 part in 4096 of selected full-scale voltage
Status Outputs 8 Status outputs from opto-isolators
Control Inputs 6 Control Inputs, 3 defined, 3 spare
Serial I/O
One (1) RS-232; One (1) RS-485 (2 connecters in
parallel)
Baud Rate : 300 – 115200: Optional Ethernet Interface
Certifications
EN61326 (1997 w/A1: 98) Class A,
FCC Part 15 Subpart B Section 15.107 Class A,
ICES-003 Class A (ANSI C63.4 1992) & AS/NZS 3548
(w/A1 & A2; 97) Class A.
IEC 61010-1:90 + A1:92 + A2:95,
1 As defined by the USEPA.
2 Defined as twice the zero noise level by the USEPA.
2.2. CE Mark Compliance
Emissions Compliance
The TAI UV Fluorescence SO
2
Analyzer 6400EH was tested and
found to be fully compliant with:
EN61326 (1997 w/A1: 98) Class A, FCC Part 15 Subpart B Section
15.107 Class A, ICES-003 Class A (ANSI C63.4 1992) & AS/NZS
3548 (w/A1 & A2; 97) Class A.
Tested on 21 February 2003 - 08 March 2003 at CKC Laboratories,
Inc., Report Number CE03-021 A.
Safety Compliance
The TAI UV Fluorescence SO
2
Analyzer 6400EH was tested and
found to be fully compliant with:
I EC 61010-1:90 + A1:92 + A2:95,
Issued by CKC Laboratories on 4 April 2003, Report Number WO-
80146.
UV Fluorescence SO2 Analyzer Specifications/Warranty
Teledyne Analytical Instruments 7
2.3. Warranty
Warranty Policy (02024d)
Prior to shipment, TAI equipment is thoroughly inspected and
tested. Should equipment failure occur, TAI assures its customers that
prompt service and support will be available.
Coverage
After the warranty period and throughout the equipment lifetime,
TAI stands ready to provide on-site or in-plant service at reasonable
rates similar to those of other manufacturers in the industry. All
maintenance and the first level of field troubleshooting is to be
performed by the customer.
NON-TAI MANUFACTURED EQUIPMENT
Equipment provided but not manufactured by TAI is warranted and
will be repaired to the extent and according to the current terms and
conditions of the respective equipment manufacturers warranty.
GENERAL
During the warranty period, TAI warrants each Product
manufactured by TAI to be free from defects in material and
workmanship under normal use and service. Expendable parts are
excluded.
If a Product fails to conform to its specifications within the
warranty period, TAI shall correct such defect by, in TA's discretion,
repairing or replacing such defective Product or refunding the purchase
price of such Product.
The warranties set forth in this section shall be of no force or effect
with respect to any Product: (i) that has been altered or subjected to
misuse, negligence or accident, or (ii) that has been used in any manner
other than in accordance with the instruction provided by TAI, or (iii)
not properly maintained.
THE WARRANTIES SET FORTH IN THIS SECTION AND THE REMEDIES
THEREFORE ARE EXCLUSIVE AND IN LIEU OF ANY IMPLIED
WARRANTIES OF MERCHANTABILITY, FITNESS FOR PARTICULAR
PURPOSE OR OTHER WARRANTY OF QUALITY, WHETHER EXPRESSED
OR IMPLIED. THE REMEDIES SET FORTH IN THIS SECTION ARE THE
EXCLUSIVE REMEDIES FOR BREACH OF ANY WARRANTY CONTAINED
HEREIN. TAI SHALL NOT BE LIABLE FOR ANY INCIDENTAL OR
Specifications/Warranty Model 6400EH
Teledyne Analytical Instruments 8
CONSEQUENTIAL DAMAGES ARISING OUT OF OR RELATED TO THIS
AGREEMENT OF TAI'S PERFORMANCE HEREUNDER, WHETHER FOR
BREACH OF WARRANTY OR OTHERWISE
Terms and Conditions
All units or components returned to TAI should be properly packed
for handling and returned freight prepaid to the nearest designated
Service Center. After the repair, the equipment will be returned, freight
prepaid.
/