ABB Advance Optima Uras 14 User manual

Type
User manual
Advance Optima
Module Uras 14
Service Manual 43/24-1005-1 EN
SHB_U14 Contents
Table of contents
Page
Chapter 1: Description of functions 1-1
Chapter 2: Module versions 2-1
Chapter 3: Module components 3-1
Chapter 4: Troubleshooting 4-1
Chapter 5: Testing 5-1
Chapter 6: Component replacement 6-1
Chapter 7: Configuration 7-1
Chapter 8: Calibration 8-1
Attachment Parts catalog
Contents SHB_U14
Contents
Chapter 1:
Description of functions
Page
Physical principle 1-2
Determination of influence values 1-6
Ex Concept 1-7
Chapter 2:
Module versions
Summary 2-2
Module versions (principles) 2-5
Module versions ( plastic tubing connections) 2-7
Module versions (stainless steel tubing) 2-12
Ex analyzer module 2-14
Chapter 3:
Module components
Emitter 3-2
Modulator 3-3
Apertures 3-6
Sample Cell 3-7
Filter Cell 3-11
Optical Filter 3-12
Calibration Unit 3-13
Calibration Cell 3-15
Infrared detector 3-16
IR module circuit board 3-19
Sensor-CPU circuit board 3-23
Pressure sensor circuit board 3-28
Thermostat circuit board 3-29
Hood with supplement heater 3-30
Connecting cable 3-31
Fortsetzung auf der folgenden Seite
SHB_U14 Contents
Contents, Fortsetzung
Chapter 4:
Troubleshooting
Status messages 4-2
No measurement signal 4-4
Unstable measurement value reading 4-5
Zero drift / Measured value drift 4-6
Negative measured signal 4-7
Measurement signal not enough sensitive 4-8
Excessive measurement signal preparation time 4-9
Temperature error 4-10
Continued on next page
Contents SHB_U14
Contents, continued
Chapter 5:
Testing
Page
Electrical connections 5-2
IR Detector measurement signal 5-4
IR detector preamplifier 5-5
Emitter insert 5-6
Modulator with emitter circuit board 5-7
Sample cell 5-8
Thermostat temperature 5-9
Heater 5-10
Temperature sensor 5-11
Thermal link 5-12
Calibration cell 5-13
Calibration unit 5-14
Pressure sensor 5-15
IR module circuit board 5-16
Sensor –CPU circuit board 5-18
Chapter 6:
Component replacement
Summary 6-2
Remove / install Uras14 analyzer module 6-3
Change emitter insert 6-5
Change modulator 6-6
Change sample cell 6-7
Change calibration cell and calibration unit 6-9
Change infrared detector 6-10
Change IR module and Sensor –CPU circuit boards 6-11
Change thermal link 6-12
Fortsetzung auf der folgenden Seite
SHB_U14 Contents
Contents, Fortsetzung
Chapter 7:
Configuration
Hardware configuration 7-2
Software configuration with TCT 7-4
General data 7-6
IR detectors 7-8
Temperature detectors 7-31
Pressure detector 7-36
Flow detector 7-40
Calibration cell 7-45
Solenoids 7-46
Pump 7-48
Chapter 8:
Calibration
Overview 8-2
Optical Alignment 8-3
Phese Alignment 8-6
Initial Calibration of Uras 14 IR detectors 8-9
Initial Zero Calibration 8-12
Postlinearisation 8-14
Cross-sensitivity alignment 8-17
Delta Phase 8-19
Calibration Cell measurement 8-22
Pressure Detector initial calibration 8-24
Flow detector initial calibration 8-26
Temperature detector initial calibration 8-31
Attachment
Parts catalog
SDE_U14 Chapter 1: Description of functions 1-1
Chapter 1: Description of functions
Overview
This chapter ... describes the underlying physical principles and provides information on the
determination of influence values.
˝
Chapter Contents In this chapter you will find the following information:
˝
Subject See Page
Physical principle 1-2
Determination of influence values 1-6
Ex Concept 1-7
˝
1-2 Chapter 1: Description of functions
SDE_U14
Physical principles
Measurement
principle
NDIR technique
(Nondispersive infrared analysis)
The measurement effect is based on resonance absorption of gas-specific
vibration-rotation bands of differing atom gas molecules in the median infrared
spectrum at wavelengths between 2.5 and 8 µm.
The individual gases to be measured are identified by their specific absorption
bands. Each gas has such an absorption spectrum (fingerprint).
Exceptions:
- Monoatomic gases, such as inert gases
- Symmetrical gases, such as N
2
, O
2
and H
2
- These types of gases cannot be measured with this method.
The relationship between measured infrared radiation absorption and the
measurement component is based on the LAMBERT-BEER law:
A = (I0 - I1) / I0 = 1 - e
-ε(λ)⋅ρ⋅l
where
A= Absorption
I0 = Radiation entering the cell
I1 = Radiation leaving the cell
ε(λ) = Sample component extinction factor
ρ = Sample component density
l = Sample cell length
The relationship between sample component density
ρ and its volumetric
concentration c is
ρ = ρ
0
c p/p
0
T
0
/T
where
ρ
0
= Pure gas density
p
0
= Pressure
T
0
= Temperature
under standard conditions (1013 hPa, 0°C).
The second equation shows that the sample component's volumetric
concentration depends on the sample cell pressure and temperature.
The first equation finds a non-linear relationship between absorption and
volumetric concentration.
Fortsetzung auf der folgenden Seite
SDE_U14 Chapter 1: Description of functions 1-3
Physical principles, Fortsetzung
˝
Basic design
The Uras 14 analyzer module is a twin-beam NDIR process photometer with no
dispersive elements.
The module consists of a completely self-contained optical unit with the
following elements:
Infrared source (emitter)
Aperture wheel (chopper)
Emitter aperture
Sample cell with measurement and reference chambers
Infrared detector with diaphragm capacitor (receiver)
Figure 1-1
Measurement
principle
CO
N
2
A
D
µP
RS232 Int.Bus
A
D
CO
Sensor-CPU
IR-Elektronic
Uras14_1
1 Emitter
2 Emitter aperture
3 Aperture wheel
4 Sample cell
5 Measurement chamber
6 Reference chamber
7 Infrared detector
8 Diaphragm capacitor
Fortsetzung auf der folgenden Seite
1-4 Chapter 1: Description of functions
SDE_U14
Physical principles, Fortsetzung
IR radiation
Generated by broad-band emitter
Emitted as a beam package alternately in the form of a sample and reference
beam through the sample and reference chambers of the sample cell and is
partially absorbed by the sample component molecules
Counterphase modulation by means of a motorized aperture wheel
Both modulated beam packages appear alternately at the infrared detector
Apertures Created by applicable regulation of the sample and reference beam balance
Sample cell Depending on the application, the sample chamber receives a sample, zero-point
or end-point gas flow so that a part of the infrared radiation is absorbed in a
concentration-dependent manner
˝
The radiation passes unhindered if the reference chamber is filled with a gas that
does not absorb infrared (N
2
).
Infrared detector A two-part transmission detector with front and rear chambers filled with the gas
components to be measured, with selectivity being determined by the infrared
detector. The two chambers are separated by an infrared-transparent window.
Additionally, the two chambers are separated by a stressed metal membrane with
counterelectrodes. This unit is known as the diaphragm capacitor.
It reacts in the following manner in the presence of the sample component:
IR radiation is weakened in the sample cell's sample chamber and enters the
receiver's front chamber.
The equilibrium between the sample and reference beams initially established
by calibration and the aperture is now disturbed.
There is an energy difference (temperature change) in the form of reduced
pressure in the front chamber.
This pressure reduction is transformed into a capacitance change in the
membrane capacitor by deflecting the metal diaphragm.
Since the diaphragm capacitor is connected to a high-impedance DC voltage, a
corresponding periodic AC signal is generated.
Fortsetzung auf der folgenden Seite
SDE_U14 Chapter 1: Description of functions 1-5
Physical principles, Fortsetzung
˝
Figure 1-2
Signal flow schematic
A
D
µP
RS232 Int.Bus
A
D
CO
N
2
N
2
N
2
N
2
CO
M
M
IR-Modul Sensor-CPU
p
t
R
t
t
Uras14_2
Figure 1-3
Block diagram as
fully equipped
N
2
CO
SO
2
NO
N
2
N
2
N
2
N
2
CO
SO
2
N
2
N
2
N
2
NO
CO
2
CO
2
M
M
Uras14_3
1-6 Chapter 1: Description of functions
SDE_U14
Determination of influence values
Associated gas
effects
The sample gas is a mixture of the sample component(s) and associated gas
components. If the infrared absorption bands of one or more associated gas
components overlap the sample components bands, the sample results will be
affected.
The influence of interfering gas components is termed cross sensitivity or carrier
gas dependence.
Cross sensitivity is determined by connecting an inert gas (e.g. N
2
) which is mixed
with the interfering gas components (corresponding to the sample gas).
The influence acts on the zero-point measurement value indication.
Carrier gas dependence, which is rarely observed. Occurs when the physical
properties of the sample gas differ markedly from those of the sample gas. This
interference changes the slope of the devices characteristic curve. This curve is
corrected at the end-point.
The Uras 14 has the following methods available for interference correction:
Interference filter
Filter cells
Internal electronic cross-sensitivity correction
Internal electronic carrier gas correction
Pressure In accordance with the gas laws, the sample cells volumetric concentration
depends on the pressure in the sample cell and is thus dependent on the process
gas and air pressure. This effect acts on the end-point and amounts to approx. 1%
of the measurement value per 1% of pressure change (therefore, per 10 hPa).
An internal pressure sensor reduces this effect to 0.2%.
Flow rate The flow rate affects pressure in the sample cell and the modules T
90
times.
The flow rate should be between 20 and 100 liters/hour.
Temperature Temperature has a markedly different effect on all optical components in the beam
path. This effect is reduced by:
either:
Temperature compensation
A temperature sensor in the first infrared detectors preamplifier measures the
temperature in the module.
This signal is used for electronic correction.
Zero-point effect:
1% of span per 10°C
End-point effect:
3% of measurement value per 10°C
or
Thermostat (optional)
The temperature influence is further reduced by controlled heating of the optical
components to 55°C.
Zero-point effect:
1% of span per 10°C
End-point effect:
1% of measurement value per 10°C
˝
SDE_U14 Chapter 1: Description of functions 1-7
Ex concept
˝
Application The Uras 14-Ex analyzer module is capable of measuring non-combustible and
combustible gases under atmospheric conditions which can form an explosive
environment (Zone 1).
Design For use in Zone 1 and Zone 2 (combustible measurement gas) the analyzer module
is installed in a pressure-tight cylinder.
The analyzer module consists of the following components:
Flange on which the analyzer is mounted
Cylinder with threaded cap, the cylinder screws into threads in the flange.
All electrical and gas connections are made through the flange.
Power supply and system bus lines are routed through pressure-tight cable
openings.
Gas inlet and outlet lines are routed through flame traps.
Figure 1-4
Ex module
Housing purge To protect the analyzer module sensor electronics against the entry of an
aggressive atmosphere or corrosive sample gas components, a purge gas can flow
through the pressure-tight cylinder.
The purge gas is routed in and out via two flame traps.
Identification EEx d IIC T4
Certification of
compliance
BVS 97.D.2021 X
SHB_U14 Chapter 2: Analyzer module versions 2-1
Chapter 2: Analyzer module versions
Overview
Introduction
This chapter describes the individual module versions.
Chapter contents
In this chapter you will find the following information:
˝
Subject See Page
Summary 2-2
Module versions (principles) 2-5
Module versions (plastic tubing connections) 2-7
Module versions (stainless steel tubing) 2-12
Ex analyzer module 2-14
2-2 Chapter 2: Analyzer module versions SHB_U14
˝
Summary
˝
General
Depending on the measurement task to be carried out, the Uras 14 can be fitted
with the following primary components:
1 to 4 infrared detectors
1 to 2 beam paths
Up to 2 infrared detectors per beam path
Modulator with three possible versions
1 emitter in first beam path (second beam path not installed)
1 emitter in second beam path (first beam path not installed)
2 emitter units
The following components are permanently installed
Thermostat-controlled version with heater and controller (option, must be confi.)
IR module circuit board
Sensor-CPU circuit board
Pressure sensor circuit board
Any additional components depend on the measuring task or desired equipment
level.
All versions of the module can be installed in a 19" rack or wall-mount housing
without any need for conversion.
The gas module and oxygen sensor can be combined in an analyzer module gas
path. These components are included in the sensor electronics package.
Special components
According to the measurement task and options ordered, the following components
can be installed:
˝
1 to 2 calibration units
1 to 2 filter cells
1 to 2 optical filters
Gas paths
FPM tubing
PTFE tubing
Stainless steel tubing
Continued on next page
SHB_U14 Chapter 2: Analyzer module versions 2-3
Summary, continued
Figure 2-1
Analyzer
Uras 14
1 Connection plate with electrical and gas connections
2 Sensor electronics
3 Optical analyzer components with heat shield
˝
Continued on next page
2-4 Chapter 2: Analyzer module versions SHB_U14
Summary, continued
Figure 2-2
Analyzer module
Uras 14
(open)
1 Calibration unit 2
2 Measurement cell 2
3 Main frame with aperture control
4 Beam path 2 aperture adjustment screw
5 Modulator with emitters (not visible)
6 Calibration unit 1
7 Infrared detector 1
8 Beam path 1 aperture adjustment screw
9 Measurement cell 1
10 Infrared detector 2
11 Stop disk (light or dark)
12 Infrared detector 3
13 Gas connections
14 24-VDC power supply, external connection
15 System bus, external connection
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122
  • Page 123 123
  • Page 124 124
  • Page 125 125
  • Page 126 126
  • Page 127 127
  • Page 128 128
  • Page 129 129
  • Page 130 130
  • Page 131 131
  • Page 132 132
  • Page 133 133
  • Page 134 134
  • Page 135 135
  • Page 136 136
  • Page 137 137
  • Page 138 138
  • Page 139 139
  • Page 140 140
  • Page 141 141
  • Page 142 142
  • Page 143 143
  • Page 144 144
  • Page 145 145
  • Page 146 146
  • Page 147 147
  • Page 148 148
  • Page 149 149
  • Page 150 150
  • Page 151 151
  • Page 152 152
  • Page 153 153
  • Page 154 154
  • Page 155 155
  • Page 156 156
  • Page 157 157
  • Page 158 158
  • Page 159 159
  • Page 160 160
  • Page 161 161
  • Page 162 162
  • Page 163 163
  • Page 164 164
  • Page 165 165
  • Page 166 166
  • Page 167 167
  • Page 168 168
  • Page 169 169
  • Page 170 170
  • Page 171 171
  • Page 172 172
  • Page 173 173
  • Page 174 174
  • Page 175 175
  • Page 176 176
  • Page 177 177
  • Page 178 178
  • Page 179 179
  • Page 180 180
  • Page 181 181
  • Page 182 182
  • Page 183 183
  • Page 184 184
  • Page 185 185
  • Page 186 186
  • Page 187 187
  • Page 188 188
  • Page 189 189
  • Page 190 190
  • Page 191 191
  • Page 192 192
  • Page 193 193
  • Page 194 194
  • Page 195 195
  • Page 196 196
  • Page 197 197
  • Page 198 198
  • Page 199 199
  • Page 200 200