Pharm Pump II
Frequently Asked Questions
1
My pump is pre-programmed. How do I select
the pre-programmed option I need?
2
How do I program the pump myself?
Begin program selection by pressing #2 or #8. The options will
appear in alphabetical order with volume amount displayed.
Use the arrow keys to highlight your selection.
To start, press the arrow with SETUP above it. This is your main
arrow throughout the programming.
Press the TYPE function arrow key at this time to change to the
function of your choice. For example: VOLUME, AMBUL, TRANS,
USAGE-1, USAGE-2, CUSTOM and NAME SYRINGE.
Enter volume, set up speed and press QUIT. You may enter other
pump functions as necessary when prompted or enter QUIT to
save your program.
Select #2 or #8
Arrow key with SETUP in
the digital display window
3
How do I adjust the volume?
4
What is CW and CCW?
5
How do I change the speed?
6
How do I prime the pump?
To change the desired volume, press the SETUP key
and use the numeric key pad.
Key in the volume and then press ENTER. Once you
have chosen the desired volume, press QUIT.
CW turns the rotor clockwise. CCW turns the rotor counter-
clockwise. To change the existing setting, press the SETUP
key and select either CW or CCW. Press NEXT and then QUIT.
Press the SETUP key, then press NEXT. When you see
SPEED, enter the required RPM between 30 to 300 and
press ENTER. Press NEXT and then QUIT.
Press the START function key. Note: This is NOT a set priming
volume. When you verify the tubing is primed, press the STOP
function key, followed by the ABORT function key.
Access the menu
Select rotor direction
Enter required RPM
Press the START
function key
Prime complete
©Acacia
M-0151 031913
7
How do I copy setup?
9
How do I clean the pump?
8
How do I calibrate the pump?
Press the COPY or QUIT function key to copy parameters
into the memory bank you are using.
Press #2 or #8 to move from one memory bank to another.
The pump can be cleaned easily by using Isopropyl Alcohol and a clean soft cloth. Apply alcohol directly
to the cloth and wipe down the outside of the housing. The area around the keypad and the motor housing
are sealed so alcohol will not seep into the inner workings of the pump.
Select COPY
Press the START button
Press the QUIT key
to resume
Connect the distal end of the tubing for the device you are going to
use i.e., syringe, bag or bottle. Press the START button and after
dispensing approximately 70% press the STOP button. Calculate the
volume that was dispensed; press the CAL key on the right side.
The LCD Screen will display:
Enter the volume that was dispensed by using the numeric keypad
and follow it by pressing ENTER, then press QUIT key to resume.
The pump will resume from the amount you entered and stop at the
amount required.
Note: Whenever QUIT appears on the screen you may press the
function key beneath it and return to the main menu.
Foot switch may be used in lieu of start/stop button for dispensing
by plugging into round receptacle on rear of pump.
Call 1-800-486-6677 for assistance.
Refer to the pump instruction manual for
additional manual operation details.
785 Challenger Street, Brea, CA 92821
Ph: 714.257.0470 Fax: 714.257.0513
www.acaciainc.com
Acacia
Life Improving Infusion Solutions
®
/