Hand-Held Pressure Indicator model CPH6300 GB
Hand-Held Druckmessgerät Typ CPH6300 D
Hand-Held Pressure Indicator model CPH6300
Operating instructions
Betriebsanleitung
2 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
GB Operating instructions CPH6300 Page 1 -30
D Betriebsanleitung Typ CPH6300 Seite 31 - 60
© 2012 WIKA Alexander Wiegand SE & Co. KG
All rights reserved. / Alle Rechte vorbehalten.
WIKA® is a registered trademark in various countries.
WIKA® ist eine geschützte Marke in verschiedenen Ländern.
Prior to starting any work, read the operating instructions!
Keep for later use!
Vor Beginn aller Arbeiten Betriebsanleitung lesen!
Zum späteren Gebrauch aufbewahren!
Contents
WIKA Operating Instruction, Model CPH6300 3
14043054.01 07/2012 GB/D
C
C
Co
o
on
n
nt
t
te
e
en
n
nt
t
ts
s
s
1 General Information ............................................................................................................. 4
2 Safety .................................................................................................................................. 5
2.1 Intended use ........................................................................................................................ 6
2.2 Personnel qualification ......................................................................................................... 6
2.3 Special hazards ................................................................................................................... 7
3 Specifications ...................................................................................................................... 8
4 Design and Function .......................................................................................................... 11
4.1 Kurzbeschreibung / Beschreibung ..................................................................................... 11
4.2 Scope of delivery ............................................................................................................... 11
5 Transport, packaging and storage ..................................................................................... 11
5.1 Transport ........................................................................................................................... 11
5.2 Packaging .......................................................................................................................... 11
5.3 Storage .............................................................................................................................. 11
6 Commissioning, operation ................................................................................................. 12
6.1 Commissioning .................................................................................................................. 12
6.1.1 Display elements ............................................................................................................... 12
6.1.2 Keyboard ........................................................................................................................... 12
6.2 Operating........................................................................................................................... 13
6.3 Configuration ..................................................................................................................... 13
6.4 Special Features ................................................................................................................ 18
6.4.1 Different Pressure Units ..................................................................................................... 18
6.4.2 Different measuring rates .................................................................................................. 18
6.4.3 Sea-level correction for absolute pressure sensors ........................................................... 19
6.4.4 Averaging function ............................................................................................................. 19
6.4.5 CALC: Pressure change per time dP/dt (only for sensor 1) ............................................... 20
6.5 Data Logger ....................................................................................................................... 21
6.5.1 Manual Recording (“Func-Stor”) ........................................................................................ 21
6.5.2 Automatic Recording with Selectable Cycle Time (“Func CYCL“) ...................................... 23
6.6 Universal Output ................................................................................................................ 26
6.6.1 USB Interface .................................................................................................................... 26
6.6.2 Analog output .................................................................................................................... 26
6.7 Input adjustment ................................................................................................................ 26
6.8 Real Time Clock („CLOC“) ................................................................................................. 26
6.9 Pressure connection to the sensors ................................................................................... 27
6.10 Alarm („AL.“) ...................................................................................................................... 27
6.11 Handling the tilt stand ........................................................................................................ 27
7 Maintenance, cleaning and servicing ................................................................................. 28
7.1 Maintenance ...................................................................................................................... 28
7.2 Replacing Batteries ........................................................................................................... 28
7.3 Cleaning ............................................................................................................................ 28
7.4 Recalibration ..................................................................................................................... 29
8 Faults ................................................................................................................................ 29
9 Dismounting, return and disposal ...................................................................................... 30
9.1 Dismounting ...................................................................................................................... 30
9.2 Return ............................................................................................................................... 30
9.3 Disposal ............................................................................................................................ 30
Declarations of conformity can be found online at www.wika.com.
1 General Information
4 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
1 General Information
The Hand-Held Pressure Indicator model CPH6300 described in the operating instructions has
been manufactured using state-of-the-art technology.
All components are subject to stringent quality and environmental criteria during production. Our
management systems are certified to ISO 9001 and ISO 14001.
These operating instructions contain important information on handling the Hand-Held Pressure
Indicator model CPH6300. Working safely requires that all safety instructions and work
instructions are observed.
Observe the relevant local accident prevention regulations and general safety regulations for the
range of use of the Hand-Held Pressure Indicator model CPH6300.
The operating instructions are part of the instrument and must be kept in the immediate vicinity
of the Hand-Held Pressure Indicator model CPH6300 and readily accessible to skilled personnel
at any time.
Skilled personnel must have carefully read and understood the operating instructions, prior to
beginning any work.
The manufacturer's liability is void in the case of any damage caused by using the product
contrary to its intended use, non-compliance with these operating instructions, assignment of
insufficiently qualified skilled personnel or unauthorised modifications to the CPH6300.
The general terms and conditions, contained in the sales documentation, shall apply.
Subject to technical modifications.
Factory calibrations/DKD/DAkks calibrations are carried out in accordance with international
standards.
Further information:
- Internet address: www.wika.de / www.wika.com
- Relevant Data Sheet: CT 12.01
- Application consultant: Tel.: (+49) 9372/132-9986
Fax: (+49) 9372/132-8767
GB
2 Safety
WIKA Operating Instruction, Model CPH6300 5
14043054.01 07/2012 GB/D
Explanation of symbols
WARNING!
... indicates a potentially dangerous situation, which can result in serious
injury or death, if not avoided.
CAUTION!
... indicates a potentially dangerous situation, which can result in light
injuries or damage to equipment or the environment, if not avoided.
Information
... points out useful tips, recommendations and information for efficient and
trouble-free operation.
2 Safety
WARNING!
Before installation, commissioning and operation, ensure that the appropriate
reference pressure sensor has been selected in terms of measuring range, design and
specific measuring conditions. Serious injuries and/or damage can occur should these
not be observed.
Further important safety instructions can be found in the individual chapters of these
operating instructions.
2 Safety
6 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
2.1 Intended use
The Hand-Held pressure indicator CPH6300 gauge can be used as a calibration instrument and
also for any application which requires accuracy pressure measurement.
The Hand-Held pressure indicator CPH6300 has been designed and built solely for the intended
use described here, and may only be used accordingly.
The technical specifications contained in these operating instructions must be observed.
Improper handling or operation of the Hand-Held pressure indicator CPH6300 outside of its
technical specifications requires the instrument to be shut down immediately and inspected by an
authorised WIKA service engineer.
Handle electronic precision measuring instruments with the required care (protect from strong
magnetic fields, static electricity and extreme temperatures, do not insert any objects into the
instrument or its openings). Plugs and sockets must be protected from contamination.
If the Hand-Held pressure indicator CPH6300 is transported from a cold into a warm environment,
the formation of condensation may result in the instrument malfunctioning. Before putting it back
into operation, wait for the instrument temperature and the room temperature to equalise.
The manufacturer shall not be liable for claims of any type based on operation contrary to the
intended use.
2.2 Personnel qualification
WARNING!
Risk of injury should qualification be insufficient!
Improper handling can result in considerable injury and damage to equipment.
The activities described in these operating instructions may only be carried out by
skilled personnel who have the qualifications described below.
Skilled personnel
Skilled personnel are understood to be personnel who, based on their technical training,
knowledge of measurement and control technology and on their experience and knowledge of
country-specific regulations, current standards and directives, are capable of carrying out the work
described and independently recognising potential hazards.
Special operating conditions require further appropriate knowledge, e.g. of aggressive media.
GB
2 Safety
WIKA Operating Instruction, Model CPH6300 7
14043054.01 07/2012 GB/D
2.3 Special hazards
WARNING!
If it is obvious that safe operation of the instrument is no longer possible, the
instrument must be put out of operation and marked in order to secure it against
inadvertent use.
The measurement signal of the reference (or test sample) can be influenced by
massive electromagnetic irradiation and/or the indication of the signal could be
completely absent.
If the Hand-Held pressure indicator CPH6300 is used in applications with oil as
a pressure medium, make sure it will not be used with flammable material or
gases directly afterwards since it can lead to dangerous explosions and danger
to personnel and machinery.
Pressure sensor CPT6200 should only be fitted or removed when the system is
free from pressure.
Observe the working conditions in accordance with chapter “3. Specifications”.
Always operate the pressure indicator within its overload limits.
This device must not be used at potentially explosive areas! The usage of this
device at potentially explosive areas increases danger of deflagration, explosion
or fire due to sparking.
When connecting the device to other devices the connection has to be designed
most thoroughly as internal connections in third-party devices (e.g. connection
GND with protective earth) may lead to undesired voltage potentials that can
lead to malfunctions or destroying of the CPH6300 and the connected devices.
3 Specifications
8 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
3 Specifications
GB
1)
Possibly an error up to 1% can occur under influence of strong magnetic fields. Keep sending devices or mobile phones away.
1)
3 Specifications
WIKA Operating Instruction, Model CPH6300 9
14043054.01 07/2012 GB/D
GB
(Sensor and Connector)
3 Specifications
10 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
For further specifications see WIKA data sheet CT 12.01 and the order documentation.
GB
4 Design and Function / 5 Transport, packaging and storage
WIKA Operating Instruction, Model CPH6300 11
14043054.01 07/2012 GB/D
4 Design and Function
4.1 Kurzbeschreibung / Beschreibung
Because of the robust and waterproof design of the Hand-Held pressure indicator CPH6300, it is
suitable for though conditions. Due to the high accuracy of the Hand-Held pressure indicator
CPH6300 it can be used for precision measurements and calibration.
4.2 Scope of delivery
CPH6300 with 2 AAA batteries
One sensor connection cable per channel
Calibration report 3.1according DIN EN 10204 for pressure sensor
Pressure sensors as selection
Cross-check scope of delivery with delivery note.
5 Transport, packaging and storage
5.1 Transport
Check the Hand-Held Pressure Indicator model CPH6300 for any damage that may have been
caused by transport. Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in
installation site, sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
Storage temperature: -25 ... +70 °C
Relative Humidity: 0 ... 95 % r. H. (non-condensing)
Avoid exposure to the following factors:
Direct sunlight or proximity to hot objects
Mechanical vibration, mechanical shock (putting it down hard)
Soot, vapour, dust and corrosive gases
Potentially explosive environments, flammable atmospheres
Store the gauge in its original packaging in a location that fulfills the conditions listed above. If the
original packaging is not available, pack and store the gauge as described below:
1. Wrap the gauge in an antistatic plastic film.
2. Place the gauge, along with shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag, containing a
desiccant, inside the packaging.
WARNING!
Before storing the gauge (following operation),
remove any residual media. This is of particular importance if the medium
is hazardous to health, e.g. caustic, toxic, carcinogenic, radioactive, etc.
GB
6 Commissioning, Operation
12 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
6 Commissioning, operation
6.1 Commissioning
6.1.1 Display elements
6.1.2 Keyboard
GB
6 Commissioning, Operation
WIKA Operating Instruction, Model CPH6300 13
14043054.01 07/2012 GB/D
6.2 Operating
Connect sensors, turn device on via key.
After segment test the device displays some configuration:
if standard measurement is selected
if fast measurement is selected
if peak value detection is selected
After that the device is ready for measuring.
6.3 Configuration
Some menu points depend on current device settings (e.g. some menu points are
locked, if the logger memory contains data sets).
To change device settings, press MENU for 2 seconds. This will activate the configuration
menu (main display: “SEt”). Pressing “MENU changes between the menus points, pressing
jumps to the referring parameters. The parameters can be changed with or . (Switch
through the parameters with .)
Pressing MENU again jumps back to the main configuration menu and saves the settings.
“STORE finishes the configuration and returns to standard measuring operation.
Pressing “MENU” and “STORE at the same time for more than 2 seconds will reset
the device to factory defaults.
If there are data sets stored and logger is set to “manual recording” (“Func Stor”) the
first menu point displayed is “rEAd Logg” (see chapter 6.5 “Data Logger”)
If no key is pressed for more than 2 minutes the configuration will be aborted. All
changes will not be saved!
GB
6 Commissioning, Operation
14 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
Menu Parameter Value Description
or
Set Configuration: General configurations
Unit: Select measuring unit
*,**
Arrow to bar,
mbar
Measuring value is displayed in corresponding unit, the
selectable units depend on connected sensor
Arrow to
“user”
Measuring value is displayed in freely adjustable user-unit
bar, mbar
Unit “user”: base unit
0000, 000.0
… .0000 Unit “user”: decimal point setting (only if user-unit selected)
-19999 ...
19999
Unit “user”: multiplication factor
Sea Level: Sea-level correction
*,**
oFF
Sea-level correction off
on
Sea-level correction on (see6.4.3)
-2000 ... 9999
Sea-level in [m]
*,**
Rate: Measuring rate
*
Slo
Slow measuring rate (4 Hz filtered, low power consumption)
FASt
Fast measuring rate, filtered (1000 Hz)
P.dEt
Peak detection: fast measuring rate, unfiltered (1000 Hz)
Averaging Filter
*
1 … 120
Averaging period in seconds
oFF
Averaging function is deactivated
Auto Power-Off: Select power-off delay
1 ... 120
Power-off delay in minutes.
Device will be automatically switched off as soon as this time has
elapsed if no key is pressed/no interface communication takes
place.
oFF
Automatic power-off function deactivated (continuous operation)
Background illumination
(factory settings: 5 s)
oFF
Illumination deactivated
5 … 120
Turn off illumination after 5 … 120s
on
Illumination always on
GB
6 Commissioning, Operation
WIKA Operating Instruction, Model CPH6300 15
14043054.01 07/2012 GB/D
Menu
Parameter
Value
Description
or
Set Calculation: Calculated display values
*
Choice of function
oFF
No calculated display
dP:dt
Pressure change per time
(leak rate measurements, ...)
SPEd
Flow speed measurement with pitot tube etc.
FLo
Volumetric flow rate measurement with pitot tube etc.
“dP:dt”: Settings
Time unit of pressure change rate
PEr.S
Pressure change rate unit per second, e.g. mbar/s
PEr.n
Pressure change rate unit per minute, e.g. mbar/min
PEr.h
Pressure change rate unit per hour, e.g. mbar/h
Test period, for which the pressure change rate is calculated
0:01 ... 1:00 Period in [minutes:seconds], ex works: 1s
“SPEd” or “Flo”: Settings
SPEd: speed unit
n:S
meter per second, m/s
kn:h
kilometer per hour, km/h
nPh
miles per hour, mph
knot
knots
FLo: Volumetric flow unit
n3:s
cubic meter per second, m3/s
n3:n
cubic meter per minute, m3/min
n3:h
cubic meter per hour, m3/h
L:S
liter per second, l/s
L:n
liter per minute, l/min
FLo: Volumetric flow - cross section
10.01999.9,
2000 14000
cross section in cm²
-100.0 ...
1000.0 °C
air flow temperature
100.0 ... 1999.9
mbar
ambient pressure
0000, 000.0,
00.00
decimal point setting
oFF, 1 … 1000 minimum display, below always display 0
6 Commissioning, Operation
16 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
Menu
Parameter
Value
Description
or
Set Output: Universal output adjustments
oFF
Interface and analog output off
SEr:
Serial interface activated
dAC:
Analog output activated
01,11..91
Base address for serial interface communication
(only at Out = SEr)
CH 1, CH 2 or
DIF CH
Channel that should be used for analog output
(only at Out = dAC)
***
sensor dep.
e.g.-5.00..5.00 mbar
Zero-point setting for Out = dAC: Measuring value which
should correspond to output 0 V
sensor dep.
e.g.-5.00..5.00 mbar
Slope setting for Out = dAC: Measuring value which should
correspond to output 1 V
Set Corr: Input adjustment (1-channel version)
*
Zero adjustment/offset of sensors
oFF
No zero adjustment of sensors
sensor dep.
e.g.-5.00..5.00 mbar The offset of sensor will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.
Slope adjustment of sensors
oFF
No slope adjustment of sensor
-2.000 ... 2.000 The slope of sensor will be changed by this factor [%] to
compensate deviations of probe or measuring device
Set Corr: Input adjustment (2-channel version)
*
***
Zero adjustment/offset of sensors 1
oFF No zero adjustment of sensors 1
sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor 1 will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.
Slope adjustment of sensors 1
oFF No slope adjustment of sensor 1
-2.000 ... 2.000 The slope of sensor 1 will be changed by this factor [%] to
compensate deviations of probe or measuring device
Zero adjustment/offset of sensors 2
oFF No zero adjustment of sensors 2
sensor dep.
e.g.-5.00..5.00 mbar
The offset of sensor 2 will be displaced by this value to
compensate for deviations of the probe or of the measuring
device.
Slope adjustment of sensors 2
oFF
No slope adjustment of sensor 2
-2.000 … 2.000 The slope of sensor 2 will be changed by this factor [%] to
compensate deviations of probe or measuring device
GB
6 Commissioning, Operation
WIKA Operating Instruction, Model CPH6300 17
14043054.01 07/2012 GB/D
Menu
Parameter
Value
Description
or
Set Alarm: Settings for alarm function
On/No.So
Alarm Sensor 1 on, with/without buzzer
OFF
No alarm function for Sensor 1
Sensor1-Min
... AL.1-Hi
Min-alarm-limit sensor 1 (not if AL.1 = oFF)
Sensor1-Min is lower display range limit of sensor 1
AL.1-Lo ...
Sensor1-Max
Max-alarm-limit sensor 1 (not if AL.1 = oFF)
Sensor1-Max is upper display range limit of sensor 1
On/No.So
Alarm Sensor 2 on, with/without buzzer
OFF
No alarm function for Sensor 2
Sensor2-Min
... AL.2-Hi
Min-alarm-limit sensor 2 (not if AL.2 = oFF)
Sensor2-Min is lower display range limit of sensor 2
AL.2-Lo ...
Sensor2-Max
Max-alarm-limit sensor 2 (not if AL.2 = oFF)
Sensor2-Max is upper display range limit of sensor 2
On/No.So
Alarm on, with/without buzzer (1-channel version) or
Alarm sensor difference on with/without buzzer
(2-channel version)
OFF No alarm function (1-channel version) or
No alarm function for sensor difference (2-channel vers.)
-19999
AL.DIF-Hi
Min-alarm-limit difference (not if AL.DIF = oFF) ***
Sensor-Min
... AL.Hi Min-alarm-limit sensor (not if AL. = oFF)
Sensor-Min is lower display range limit of sensor
AL.DIF-Lo
19999
Max-alarm-limit difference (not if AL.DIF = oFF) ***
AL.Lo ...
Sensor-Max Max-alarm-limit sensor (not if AL. = oFF)
Sensor-Max is upper display range limit of sensor
Set Logger: Settings for logger function
*
Selection of logger function
*
CYCL
Cyclic: cyclic logger
Stor
Store: manual recording
oFF
No logger activated
0:01 ... 60:00 Cycle time in [minutes: seconds] (for cyclic logger)
*
on/oFF Low-power-logger with low power consumption
(only for cyclic logger and slow measuring rate)
*
Set Clock: Settings for real time clock
HH:MM Clock: set time hours: minutes
YYYY Year: set year
TT.MM Date: set date day. month
rEAd Logg: Read calibration data:
seechapter 6.5.1
(*) If logger memory contains data sets parameters marked with (*) cannot be called. You have to
clear memory to change these parameters! (see chapter 6.5)
(**) Parameter can only be called if corresponding sensor is plugged to connection 1. The settings
are applied to the sensor at connection 2 (if any connected).
(***) Only 2-channel version
GB
6 Commissioning, Operation
18 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
6.4 Special Features
6.4.1 Different Pressure Units
Depending on the connected sensors different units can be selected via menu (Unit). The
measuring range of the sensors may restrict the choice!
User-Unit
For pressure units, which are not covered by the ones printed on the display, there can be done a
manual setting for a “user-defined” unit.
Example:
To enable a reference pressure sensor CPT6200 to display the unit kg/cm2, following settings
(bold) have to be entered:
kg/cm2
Torr
atm
at
bASE user:
bar
mbar
bar
bar
DP user:
.0000
.0000
.0000
.0000
FACt user:
1.0197
.7433
0.9869
1.0197
6.4.2 Different measuring rates
Three different kinds of pressure measuring are supported. Two of them (P.dEt and FASt) are
working with high measuring frequency of more than 1000 measurings per second.
6.4.2.1 Standard measuring (slow)
Measuring rate 4 Hz, averaging and filter functions are active.
Application: Measuring of slowly changing or static pressures, e.g. leak testing,
atmospheric pressure, etc.
Highest accuracy, high noise immunity, low power consumption.
6.4.2.2 Peak detection (Peak detection)
Measuring rate >1000 Hz, the value is displayed unfiltered.
Application with logger function: Measuring of short pressure peaks or fast changing
pressures with a resolution of <1 ms. The cyclic logger function records the arithmetic
mean value, the highest and the lowest peak of the referring time interval.
This measuring is sensitive to noise (EMI, ...) and power consumption is
enhanced.
6.4.2.3 Fast filtered measuring (fast)
Measuring rate >1000 Hz, but the value is filtered (higher noise immunity than P.dEt,
short peaks will be filtered out), apart from that identical behavior like rAtE-P.dEt.
GB
6 Commissioning, Operation
WIKA Operating Instruction, Model CPH6300 19
14043054.01 07/2012 GB/D
6.4.3 Sea-level correction for absolute pressure sensors
The device displays the absolute pressure. This is not necessarily the same like the values given
by weather stations! The weather stations are displaying the pressure at sea level. Usually the
sensor is placed above sea level and therefore the pressure loss resulting from the actual level
above sea level has to be considered, if the value at sea level (zero) should be measured! To
correct the measuring display, activate the “Sea-Level-Function” (SL, see chapter 6.3, setting is
only possible, if an absolute pressure sensor is connected to sensor connection 1). Then enter the
altitude above sea level of the sensor’s location in meters (Alti, see chapter 6.3). When activated,
the display shows the SL-arrow and the device displays the pressure value at sea level.
When two absolute pressure sensors are connected, the sea level function for both
corresponds to the setting of sensor 1
(Only 2-channel version)
6.4.4 Averaging function
The averaging function concerns the display values (LCD and interface).
It is completely independent from the averaging of the logger function FASt and
P.dEt.
The averaging integrates the measuring values during a selectable period of time and then
calculates the average display value. It is independent from the selected kind of measuring (slow,
fast, peak detect).
As long as not enough values have been collected i. e. for the selected averaging time (seconds)
to calculate an average value, the upper display shows “----, the lower display a ‘countdown‘.
During an active low-power-logging procedure the averring is always deactivated.
Function of min/max-value memory during averaging:
If averaging is activated and slow measuring is selected (rAtE-Slo), the min-/max-value memory
refers to the average display value.
If averaging is activated and fast measuring is selected (rAtE-FASt or P.dEt), the min-/max-
value memory refers to the internal measured values (fast peaks can be detected).
GB
6 Commissioning, Operation
20 WIKA Operating Instruction, Model CPH6300
14043054.01 07/2012 GB/D
6.4.5 CALC: Pressure change per time dP/dt
(only for sensor 1)
The CALC function “dPdt” allows displaying pressure changes per time directly. The corresponding
measuring intervals (t.int) can be freely chosen. The device automatically saves pressure values
during this period and calculates the pressure change rate (pressure change / time) from the
current measuring value and historical data.
TARE-button: Historical data is deleted, calculation restarts.
Example: Tightness testing for sewage ducts with requirements less than -3 mbar/min in a
5 minutes test duration at a test pressure of 250 mbar.
Equipment/presetting:
Sealing bladder and pressure pump
Sensor: CPT6200 (measuring range 0 … 1 bar)
bASE= Pet.n (minutes) time unit for pressure change rate
t.int = 1:00 (1 minutes) test duration, for which the pressure change rate is calculated
Logger: Cycl, 5 sek
Realising:
Connect all parts, arrange all necessary sealings
Start logger
Charge with pressure, wait for sufficient time
Restart calculation with button “TARE”
Current pressure value is displayed, button “SET” changes to dP:dt display
The display can give some preliminary information about the test result
(will it fail the test?)
Release pressure after 5 minutes
Stop logger
Evaluation:
Read out data logger with GSOFT, input at register “comment”:
Client, place of inspection, date, time, address / duct number, as-built data (duct type,
nominal size, etc), test instructions, pressures, tolerances, etc. as well as test result.
Therefore a comment sample can be prepared as file, company logo etc. can be added for
printing.
t.int- setting: If there are set high values the displayed values can be easily
misinterpreted, because the displayed value responds very slowly.
The displayed value is calculated: P(Tcurrent) P(Tcurrent -t.int) / T.int
If there are no data for the whole t.int-period yet the values are extrapolated with the
already available data.
GB
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60