Intellijel Zeroscope 1U User manual

Type
User manual

Intellijel Zeroscope 1U is a compact and versatile Eurorack oscilloscope, frequency meter, and tuner. With its dual-channel design, you can simultaneously monitor and analyze two signals, making it ideal for troubleshooting patches, tuning oscillators, or simply visualizing your sounds. The Zeroscope 1U offers precise waveform display, allowing you to measure voltages, periods, and frequencies with accuracy. Its advanced features include variable gain controls, trigger level adjustment, and multiple display modes, providing you with a comprehensive toolkit for exploring your modular system.

Intellijel Zeroscope 1U is a compact and versatile Eurorack oscilloscope, frequency meter, and tuner. With its dual-channel design, you can simultaneously monitor and analyze two signals, making it ideal for troubleshooting patches, tuning oscillators, or simply visualizing your sounds. The Zeroscope 1U offers precise waveform display, allowing you to measure voltages, periods, and frequencies with accuracy. Its advanced features include variable gain controls, trigger level adjustment, and multiple display modes, providing you with a comprehensive toolkit for exploring your modular system.

1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 1/13
 
Zeroscope 1U Manual
Zeroscope 1U
Dual Channel Oscilloscope, Frequency Meter and Tuner
Manual Revision: 2019.01.08
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 2/13
 
Zeroscope 1U Manual
 
Table of Contents
Table of Contents
Compliance
Installation
Before Your Start
Installing Your Module
Front Panel
Inputs & Outputs
Controls
Basic Operation
Advanced Operation
Scope Type
Using Tune Mode
Hz Display
Note Name Display
Reference Pitch and Averaging Time
Trigger Source
Trigger Detection Level
CH1 Offset
CH2 Offset
Grid Display
Technical Specifications
Firmware Changelog
1.2 (Jan 08, 2019)
 
Page 1
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 3/13
 
Zeroscope 1U Manual
 
Compliance
This device complies with Part 15 of the FCC Rules. Operation is subject to the
following two conditions: (1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that
may cause undesired operation.
Changes or modifications not expressly approved by Intellijel Designs, Inc. could
void the user’s authority to operate the equipment.
Any digital equipment has been tested and found to comply with the limits for a
Class A digital device, pursuant to part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference when the
equipment is operated in a commercial environment. This equipment generates,
uses, and can radiate radio frequency energy and, if not installed and used in
accordance with the instruction manual, may cause harmful interference to radio
communications.
This device meets the requirements of the following standards and directives:
EMC: 2014/30/EU
EN55032:2015 ; EN55103-2:2009 (EN55024) ; EN61000-3-2 ; EN61000-3-3
Low Voltage: 2014/35/EU
EN 60065:2002+A1:2006+A11:2008+A2:2010+A12:2011
RoHS2: 2011/65/EU
WEEE: 2012/19/EU
 
Page 2
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 4/13
 
Zeroscope 1U Manual
 
Installation
This module is designed for use within an
Intellijel-standard 1U row, such as
contained within the Intellijel 4U and 7U
Eurorack cases. Intellijel’s 1U specification
is derived from the Eurorack mechanical
specification set by Doepfer that is
designed to support the use of lipped rails
within industry standard rack heights.
Before Your Start
Intellijel Eurorack modules are designed to be used with a Eurorack-compatible case and power
supply. We recommend you use Intellijel cases and power supplies.
Before installing a new module in your case, you must ensure your power supply has a free
power header and sufficient available capacity to power the module:
Sum up the specified +12V current draw for all modules, including the new one. Do the
same for the -12 V and +5V current draw. The current draw will be specified in the
manufacturer's technical specifications for each module.
Compare each of the sums to specifications for your case’s power supply.
Only proceed with installation if none of the values exceeds the power supply’s
specifications. Otherwise you must remove modules to free up capacity or upgrade your
power supply.
You will also need to ensure your case has enough free space (hp) to fit the new module. To
prevent screws or other debris from falling into the case and shorting any electrical contacts, not
leave gaps between adjacent modules, and cover all unused areas with blank panels. Similarly,
do not use open frames or any other enclosure that exposes the backside of any module or the
power distribution board.
Page 3
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 5/13
 
Zeroscope 1U Manual
 
You can use a tool like ModularGrid to assist in your planning. Failure to adequately power your
modules may result in damage to your modules or power supply. If you are unsure, please
contact us before proceeding.
Installing Your Module
When installing or removing a module from
your case always turn off the power to the
case and disconnect the power cable. Failure
to do so may result in serious injury or
equipment damage.
Ensure the 10-pin connector on the power
cable is connected correctly to the module
before proceeding. The red stripe on the
cable must line up with the -12V pins on the
module’s power connector. The pins are
indicated with the label -12V, a white stripe
next to the connector, the words “red stripe”,
or some combination of those indicators.
Most modules will come with the cable
already connected but it is good to double
check the orientation. Be aware that some modules may have headers that serve other
purposes so ensure the cable is connected to the right one.
The other end of the cable, with a 16-pin
connector, connects to the power bus
board of your Eurorack case. Ensure the
red stripe on the cable lines up with the
-12V pins on the bus board. On Intellijel
power supplies the pins are labelled with
the label “-12V” and a thick white stripe:
If you are using another manufacturer’s
power supply, check their documentation
for instructions.
 
Page 4
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 6/13
 
Zeroscope 1U Manual
 
Once connected, the cabling between the module and power supply should resemble the
picture below:
Before reconnecting power and
turning on your modular
system, double check that the
ribbon cable is fully seated on
both ends and that all the pins
are correctly aligned. If the pins
are misaligned in any direction
or the ribbon is backwards you
can cause damage to your
module, power supply, or other
modules.
After you have confirmed all the
connections, you can reconnect
the power cable and turn on
your modular system. You should immediately check that all your modules have powered on
and are functioning correctly. If you notice any anomalies, turn your system off right away and
check your cabling again for mistakes.
 
Page 5
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 7/13
 
Zeroscope 1U Manual
 
Front Panel
Inputs & Outputs
1. CH 1 input: The input is DC coupled and accepts signals between -10V and +10V.
2. CH 2 input: This input, like CH1, is DC coupled and accepts signals between -10V and
+10V.
3. CH 1 output: The signal appearing at the CH1 input jack is normalled to the CH1
output jack, so signals pass unimpeded through the Zeroscope1U. This allows you to
monitor signals without needing to re-patch or use signal multipliers.
4. CH2 output: As with CH1, the signal appearing at the CH2 input jack is normalled to
the CH2 output jack, so signals pass unimpeded through the Zeroscope 1U.
Controls
A. Scope: In Scope or X-Y modes, this screen displays both the CH1 and CH2
waveforms, while in Tune mode, it displays their frequency by either Hz or note name.
B. Encoder: This push-function encoder serves multiple functions — controlling the
appearance of the scope, and enabling both the selection of parameters and the editing
of their values.
Page 6
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 8/13
 
Zeroscope 1U Manual
 
Basic Operation
In its basic mode, Zeroscope1U displays the waveforms for
both CH1 and CH2, and gives you control over the display
height of these two waveforms plus their displayed
periodicity.
1. Connect CV or audio signals to either or both
the CH1 and CH2 inputs.
Both inputs are DC coupled and accept signals
between -10V and +10V. Either can serve as a
trigger source, configured to detect either the rising
or falling edge of the waveform (as described later in
the Trigger Source discussion).
All inputs are normalled to the jacks below.
2. Press the encoder to cycle through the three
three basic display options.
Specifically, each press cycles between:
CH1 Display Value (volts/division)
CH2 Display Value (volts/division)
T (Time) Display Value (seconds/division, both channels)
3. Turn the encoder to change the option’s value.
Specifically:
CH1: With this option displayed, turning the encoder sets CH1’sgain (height of
displayed signal) between 0.5volts/division and 10.0volts/division.
CH2: With this option displayed, turning the encoder sets CH2’s gain (height of
displayed signal) between 0.5volts/division and 10.0volts/division. CH2’s gain
level is completely independent from CH1’s.
T: With this option displayed, turning the encoder sets the waveform’s time base
between 10 s/div and 10 µs/div with seventeen incremental values between
these two extremes.
NOTE: At a setting of 10µs/div, only CH1 is displayed. Also, because the
Zeroscope employs a simple anti-aliasing filter, an input signal running at a
frequency higher than the timebase may produce a misleading waveform.
Page 7
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 9/13
 
Zeroscope 1U Manual
 
Advanced Operation
Zeroscope1U offers several advanced display and triggering options, located within a
Advanced Settings menu.
1. Long-press (>1 sec) the encoder to open the Advanced Settings menu.
A small menu of options appears to the right of the display.
2. Turn the encoder to select and highlight which of six Advanced Settings options
you wish to edit.
There are six options accessible in this menu. These are, from top-to-bottom:
SCOPE type
TRIGGER Source
TRIGGER DETECTION LEVEL
CH1 offset
CH2 offset
GRID display
3. With the desired option highlighted, push-turn the encoder to edit its value.
That is, turn the encoder while it’s fully depressed to change values. The following
sections discuss each advanced operation in detail.
 
Page 8
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 10/13
 
Zeroscope 1U Manual
 
Scope Type
With this option selected, push-turn the encoder to select what you want displayed on the
Zeroscope. There are three options:
Scope Standard graphical display, with time displayed on the X-axis and voltage
displayed on the Y-axis.
X-Y Alternate graphical display with CH1 driving the X-axis and CH2 driving the
Y-axis.
Tune Alphanumerical tuning display, which you can configure to show either Hz
values or note names (see below).
Using Tune Mode
Tune Mode displays CH1 and CH2 frequency either by number of cycles-per-second
(Hz) or by Note Name. Once in Tune mode, simply press the encoder to toggle between
the two display options.
NOTE: The input signal must be at least 1V in order for the pitch detector to work.
Hz Display
The top line displays the frequency (in Hz)
of the signal appearing at CH1, and the
second line displays the frequency of the
signal appearing at CH2.
 
Page 9
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 11/13
 
Zeroscope 1U Manual
 
Note Name Display
The top line shows CH1, and the second
line shows CH2.
Zeroscope displays the note name closest
to the incoming pitch, plus a vertical line
indicating how far the pitch is from that
note (from -45 to +45 cents).
A note is sharp if the vertical line is to the
right of the note name. For example, in
this illustration the vertical line is to the right of the note name for CH2, indicating that
CH2 is playing a C3 (“middle C”), but that the note is sharp.
A note is flat if the vertical line is to the left of the note name.
A note is perfectly in tune if a box surrounds the note name. In this illustration, the box
around the note name for CH1 indicates that it’s playing a C1 note, and that it’s perfectly
in tune.
Reference Pitch and Averaging Time
Both the Hz Display and the Note Name Display have two additional options at the
bottom of the scope: ref and avg.
ref This displays the reference pitch used by the tuner (which displays the
number of Hz to which the A above middle-C is tuned). Most devices
default to 440Hz, but other standards are sometimes used.
avg This displays how long the Zeroscope spends averaging out frequency
variations before it recalculates and reports a new note name. There are
three options: slow; norm; and fast. Fast is very sensitive to fluctuations
on input, while slow is not. Norm is in the middle.
A small indicator arrow to the left of the ref or avg parameter indicates which will be
changed by rotating the encoder. To change whether the encoder edits ref or avg, switch
between Hz and Note Name displays. Specifically:
1. After entering Tune mode, wait five seconds for the little Tune mode indicator in
the upper right corner to extinguish.
2. With Hz display active, turn the encoder (without pushing it) to change the avg
setting (slow, normal, fast)
3. Switch to the Note Name display, and turn the encoder (without pushing it) to
change the ref tuning value.
Page 10
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 12/13
 
Zeroscope 1U Manual
 
Trigger Source
With this option selected, push-turn the encoder to select how you want to trigger the
Zeroscope. You have five triggering options:
Trg off The scope is untriggered
Trg 1 The scope triggers on the rising edge of the signal at CH1
Trg 1 The scope triggers on the falling edge of the signal at CH1
Trg 2 The scope triggers on the rising edge of the signal at CH2
Trg 2 The scope triggers on the falling edge of the signal at CH2
Trigger Detection Level
With this option selected, push-turn the encoder to set the level at which trigger detection
occurs. Trigger threshold can be offset ± 10V (-10000µV - +10000µV). At 0V (the default), a
typical bipolar waveform triggers when it crosses the 0-line.
Trigger levels are graphically indicated on the Zeroscope display.
If you set a trigger threshold beyond the voltage limits of the current waveform, you’ll see a
small T? icon appear in the lower left corner of the display, indicating that it failed to detect a
trigger.
CH1 Offset
With this option selected, push-turn the encoder to set a display offset for the CH1 waveform.
Waveforms can be offset ± 10V (-10000µV - +10000µV).
Offsetting waveforms is particularly useful
if you want to display CH1’s waveform
above (or below) CH2’s waveform. For
example, assume CH1 and CH2 are
both typical ± 5V waveforms. If you set
both the CH1 and CH2 Display Values to
5.0 V per division, then each waveform is
½ the entire height of the scope. If you set
the CH1 Offset to +5000 (+5V) and the
CH2 Offset (described below) to -5000
(-5V), then CH1’s waveform is displayed
in the top half of the scope, and CH2’s is
displayed in the bottom half.
Page 11
1/8/2019 Zeroscope 1U Manual - Google Docs
https://docs.google.com/document/d/11f-r1wXdM0R0YPFTe9TY-U185nvz2PQHx-hAHVGTYTk/edit#heading=h.jm4q10wdqtr1 13/13
 
Zeroscope 1U Manual
 
CH2 Offset
With this option selected, push-turn the encoder to set a display offset for the CH2 waveform.
Waveforms can be offset ± 10V (-10000µV - +10000µV).
Grid Display
With this option selected, push-turn the encoder to select between two options:
grd: on turns on the scope’s background grid
grd: -- turns off the scope’s background grid
Technical Specifications
Width
16 hp
Maximum Depth
26 mm
Current Draw
8 mA @ +12V
0 mA @ -12V
Firmware Changelog
1.2 (Jan 08, 2019)
Initial product release
Page 12
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13

Intellijel Zeroscope 1U User manual

Type
User manual

Intellijel Zeroscope 1U is a compact and versatile Eurorack oscilloscope, frequency meter, and tuner. With its dual-channel design, you can simultaneously monitor and analyze two signals, making it ideal for troubleshooting patches, tuning oscillators, or simply visualizing your sounds. The Zeroscope 1U offers precise waveform display, allowing you to measure voltages, periods, and frequencies with accuracy. Its advanced features include variable gain controls, trigger level adjustment, and multiple display modes, providing you with a comprehensive toolkit for exploring your modular system.

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI