CMA ML42m User guide

Type
User guide
PHSENSORML42M
USERSGUIDE
CENTRE FOR MICROCOMPUTER APPLICATIONS
http://www.cma-science.nl
2|ML42mpHSensorUser’sGuide
Shortdescription
The pH sensor ML42m measures the degree of acidity/pH value of a solution in the range
from0to14pH.ThepHelectrodeisconnectedtothesensorboxbymeansofacoaxcable
andaBNCconnector.
The pH sensor is an I2C digital sensor, which
gives calibrated values of the measured
quantity. This sensor can only be connected to special interfaces that support I2C digital
sensorsliketheCMAMoLabinterface.Thesensorcableneededtoconnectthesensortoan
interfaceisnotsuppliedwiththesensor(sensorcablesaresuppliedwithinterfaces).
Sensor
specifications
The pH sensor ML42m is a digital sensor that converts the measured pH value to a digital
valuevia16bitanalogtodigitalconversion.
Theresolutionofthesensoris±0.0002pH.
pHelectrode
ThepHelectrodeisagelfilled glasselectrodebuilt inaplastictube
withanopeningatthe
bottom side. The pH electrode has been designed to function in a reliable and accurate
mannerundervariedconditions.TheruggedepoxybodyprotectsthepHsensingglassbulb.
Thegelfilledreferenceelectrodeissealedanditneverneedstoberefilled.
The pH
electrode is supplied
inastoragebottlecontaining
a protective solution. When
it is not being used, it must
be kept in this storage
solution. During measure
mentstheelectrodemustbe
dipped in the solution for
about1cm.
The pH electrode can be
damaged beyond repair by improper use
such as keeping the electrode in distilled water
storage,usageoveritstemperaturerangeandlongtimeexposuretostrongacid/base.
ThepHelectrodehasalimitedoperationallifeandcanbeorderedseparately.
PreparingthepHelectrodeforuse
Carefully remove the electrode from the bottle of storagesolution
and rinse the tip of the
electrode with distilled water. If air bubbles are seen in the pH bulb, gently shake the
electrodedownward,likeaclinicalthermometer,untilthebubblesdisappear.Theelectrode
isreadyformeasurement.
Maintenanceandstorage
Maintain the level of pH electrode storage solution, the pH
sensitive membrane must be
keptwet.StoretheelectrodeinapH4buffer,towhichpotassiumchloride(KCl,10g/100ml)
hasbeenadded(pH4/KCl).
ML42mpHSensorUser’sGuide| 3
CAUTION:Neverstorethe electrodeindistilledwater.Usingorstoringtheelectrodeatvery
highorverylowtemperatures(near0°C)candamageitbeyondrepair.
Toprepareadditionalstoragesolution(pH4/KCl):
1. pH4.00buffer:add2.0mLof0.1MHClto1000mLof0.1potassiumhydrogenphthalate.
2. pH4/KClsolution:add10gofsolidpotassiumchloride(KCl)to100mLofpHsolution.
Cleaningoftheelectrode
A dirty but
mechanically intact electrode can often be restored by one of the following
procedures,ingeneral:soaktheelectrodein0.1MHClfor15minutes.
Deposit of proteins: soak the electrode in 0.1 M HCl to which 1% of Pepsin has been
added.
Deposit of inorganic material: rinse
the electrode with a 0.1 M EDTA tetrasodium
solution.
Depositofoilorgreasefilms:washtheelectrodeinamilddetergentorsolventknownto
attacktheparticularfilm(butnottheelectrodeitself!).
Aftercleaning,soaktheelectrodeinpH7bufferfor30minutes.Ifnoneof
theseprocedures
improveresponse,replacetheelectrode.
Donotusetheelectrodein:
base solution (pH > 10) for longer than a few hours. This can affect the glass of the
electrode,
hydrofluoricacidorinacidorbasesolutionwithaconcentrationgreaterthan1.0M,
solutionscontainingperchlorate,silver,orsulfideions.
The electrode can be used to measure the pH of sodium hydroxide solutions with a
concentration near 1.0 M, but should not be left in this concentration for periods of time
longerthan5minutes.
Whitecrustontheelectrode
The white crystals you
might find on the electrode are formed by the Potassium Chloride
(KCl) from the storage solution. Carefully remove the electrode from the bottle of storage
solution,rinsetheelectrodewithdistilledwaterandproceedasusual.
Collectingdata
ThispHsensorworksonlywithspecificinterfaces.Thesensorwillbe
automaticallydetected
when connected to such an interface. For detailed information about measurements with
sensorsconsulttheUserManualsoftheinterfaceandtheCoach6software.
Calibration
The pH sensor is supplied with a factory calibration and gives values in pH. The Coach 6
programallowsshiftingthepredefined
calibrationorcreatinganewthreepointcalibration
if needed. For routine measurements, one buffer suffices the calibration. For precise
calibrationoftheelectrodethreebuffersolutionsarerequired.Onebuffershouldbecloseto
4|ML42mpHSensorUser’sGuide
the desired measuring range. During calibration you need to
rinsetheelectrodewithdistilledwater.Theusercalibrationis
storedinnonvolatileusersensormemory.
Suggestedexperiments
ThepHsensorcanbeusedforvariousexperimentssuchas:
MeasurementsofpHofdifferentacidsandbases
Acidbase
titrationexperiments.
MonitoringpHduringchemicalreactions.
Investigationsofwaterqualityinstreamsandlakes.
TechnicalSpecifications
Sensorkind Digital:onsensordigitalconversion
(16bitsresolution,communicationviaI2C)
Measuringrange 0..14pH
Resolution ±0.0002pH
Uncertainty(accuracy) Typical±1.0%aftercalibrationat25°C
Temperaturerange 5°Cto80°C
(IfyouareusingthepHelectrodeatvarioustemperatures,
itiscertainlyneededtocalibratethesensorthroughthree
pointswithtemperaturecompensationbytheNernst
equationE=E
0
kT*pH)
Maximalsamplingrate 10Hz
Dimensions Sensorhousing:37x18x16mm
Connection 5pinsminijackplug
Warranty:
The pH sensor ML42m is warranted to be free from defects in materials and
workmanshipfor a periodof12monthsfromthedateofpurchaseprovidedthatithas
been used under normal laboratory conditions. This warranty does not a pply if the
sensorhasbeendamagedbyaccidentormisuse.

Rev. 17/11/2011
Note:Thisproductistobeusedforeducationalpurposesonly.Itisnotappropriatefor