NDT Systems Curlin-Air User manual

Type
User manual
SAFETY PRECAUTIONS ............................................ 4
1. INTRODUCTION ................................................ 2
Basic System Components ...................................... 2
2.0 Specifications.................................................. 3
3.0 Operator Qualifications ........................................... 5
3.1 Operational Notations .................................... 5
KEYS ................................................. 5
VARIABLES............................................ 5
SCROLLING ........................................... 5
TOGGLING ............................................ 5
CURSOR .............................................. 5
4.0 Transducers and Reference Standards .............................. 6
5.0 MENU DESCRIPTIONS.......................................... 7
5.1 Special Menu & Keyboard ................................... 7
Help Screens ........................................... 7
Menu Flow ............................................. 7
5.2 MENUS ................................................. 8
5.2.1 MAIN MENU ...................................... 8
ENERGY......................................... 8
BLANK .......................................... 8
Example .................................... 8
BLNKADJ ........................................ 9
GAIN ............................................ 9
RANGE .......................................... 9
DELAY .......................................... 9
GATE ........................................... 9
AMPLITUDE................................. 9
DISTANCE .................................. 9
IF IN AMPLITUDE MODE............................ 9
ALARM POLARITY ........................... 9
START ..................................... 9
WIDTH .................................... 10
LEVEL .................................... 10
IF IN DISTance MODE ............................. 10
VELOCITY ................................. 10
ZERO ADJUST.............................. 10
BLOCK.................................... 10
LEVEL .................................... 10
AUDIO.......................................... 10
5.2.2 DISPLAY MENU ................................. 10
DISP ........................................... 10
HOLLOW .................................. 10
SOLID .................................... 10
SMOOTH ....................................... 10
GRID .......................................... 10
FREEZE................................... 11
BACK LIGHT .................................... 11
CONTRAST ..................................... 11
BEEP .......................................... 11
LOCK .......................................... 11
5.2.3 MISC Menu...................................... 11
BATTERY TEST.................................. 11
MODE.......................................... 11
UNITS.......................................... 11
SAVE SETUP .................................... 12
Rcall ........................................... 12
PRINT.......................................... 12
AlmTim ......................................... 12
PRF ........................................... 13
6.0 Curlin AIR SETUP AND OPERATING CONSIDERATIONS ............. 14
6.1 Safe Operation ........................................ 14
6.2 Safe Inspection Site .................................... 14
6.3 Ambient Ultrasonic Noise ................................ 14
6.4 Calibrate Before Performing Any Inspection !! ................ 14
6.5 Reference Samples..................................... 14
6.6 Operational Considerations ............................... 14
6.7 Curlin AIR Display Characteristics.......................... 15
6.8 Factory Setup Variables ................................. 15
7.0 Flaw Detection Procedures .................................... 16
7.1 Mechanical Assembly ................................... 16
7.1.2 Adjustment of Yoke ............................... 16
7.1.3 Cable Attachment to Curlin-AIR ...................... 16
7.2 Electronic Initialization & Checkout ......................... 16
7.2.1 Power-ON....................................... 16
7.2.2 Message Display ................................. 16
7.2.3 Automatic Initializing Set-up ......................... 16
7.2.4 Insert Material Into Beam ........................... 16
7.3 CALIBRATION ........................................ 17
7.3.1 Calibration On Reference Samples .................. 17
7.3.2 Reference Sample ................................ 17
7.3.3 Curlin-AIR Adjustment On Reference Sample ........... 17
7.3.4 Maximum Scan Speed Determination ................. 17
7.3.5 Edge Effects ..................................... 17
7.4 Keep Equipment Clean .................................. 18
7.5 TECHNICAL ASSISTANCE ............................... 18
8.0 Output and Optional Printer Technical Specifications ................. 19
8.1 RS-232C Output Specifications ............................ 19
8.2 Printers .............................................. 19
9.0 Battery Pack and Charger ....................................... 20
10 Optional Curlin AIR Accessories ................................. 21
11 WARRANTY .................................................. 22
22
1. INTRODUCTION
First and foremost we at NDT Systems, would like to thank you for your purchase of the ALL NEW
Curlin- AIR, portable airborne ultrasonic flaw detector.
The Curlin-AIR has been designed for airborne non-contact or dry-coupled contact ultrasonic flaw
detection applications when used in conjunction with a special AT1 ultrasonic transducer.
The Curlin-AIR’s application is for nondestructively inspecting low-density materials, such as
polyurethane foam and other porous materials. It is not intended for inspecting high-density materials,
such as metal castings, plates, welds, or thick solid plastic (NDT Systems, standard QUANTUM TE
flaw detector is a better choice for these applications).
The primary mode of operation for the Curlin-AIR is through-transmission, using separate transmitting
and receiving transducers mounted in opposition on a hand held “yoke” (for scanning the material).
Used, in this fashion, the ultrasonic beam is transmitted through the material and detected on the
opposite side. Flaws are typically indicated by their property to “block/reflect” the transmission of the
ultrasonic beam (they tend to greatly reduce the “normal” signal amplitude).
In addition to the detection of localized flaws, the Curlin-AIR should also respond to density variations
in the material (sensitivity a function of the material type and degree of variation).
This manual serves two purposes. 1) Quick Reference and 2) Tutorial.
C The Table of Contents of this manual is actually a TABLE/INDEX. Each menu item is mentioned
here for quick easy access to a description of the menu item you may have a question with.
C Anywhere you are in the Curlin AIR’s menu you can get a description of the feature and how to
use it with a simple keystroke. Just highlight the menu item you want to use or have question
with and then press the SEL/HELP key for about 1 second.
C Chapter 2 deals with the menu, structure, and description of the features. This would be
recommended reading. There are quite a few images to help you better understand the
concepts.
Again, thank you for your purchase of the Curlin -AIR.
Basic System Components
The basic through-transmission system consists of.
Curlin-AIR
Two coaxial cables (flexible, heavy duty, small diameter with BNC to BNC connectors)
One pair of identical AT1 transducers, interchangeable as a transmitter or receiver).
One adjustable yoke for manually scanning the transducers.
33
2.0 Specifications
Display Super Twist LCD, 4.8 x 2.4 inches ( 122 x 61mm), transflective;
backlit, adjustable contrast, battery status and mode icons, large
thickness/soundpath display. Hollow or filled waveform, reversed
field selectable.
Display Mode Full-wave rectified.
Graticule Graphically Generated, 10 x 10 major divisions
Receiver 0.5 to 25 MHZ Broadband, Tuned 0.5, 1.0, 2.25, Frequency 5.0,
10.0 MHZ selectable.
Gain 115 dB; 1.0dB steps
Pulser Pulsed Continuous Wave, selectable from 1- 15 cycles
Linear Reject Linear to 90% maximum.
Range 1.0 inch (25mm) FS to 200 inches (5077mm) FS (steel equivalent).
Delay 0 to 150 inches (3807mm) FS (steel equivalent).
Velocity 0.0500 to 0.7500 inch per microsecond
Zero Adjust 0 to 3,000,000 nanoseconds, 50ns increment
Synchronization IP (Initial Pulse)
Transducer Modes THRU-TRANS (primary mode of operation), Pulse Echo, Pitch-
Catch,.
Display Mode Full-wave rectified; full-wave smoothed.
Flaw Mode One gate selectable; positive or negative threshold. Visible and
audible alarms triggered at pulse repetition rate. Gate start variable
over entire displayed range. Gate width variable from gate start to
maximum displayed range. Gate level adjustable from 1% to 99%
full screen.
Distance Mode Single echo distance measurement. Large digital readout.
Resolution 0.001 inch (0.01mm).
Stored Setups 50 user defined setups.
Screen Print Print current screen, including waveform, or complete setup.
Input/Output RS232 (9 pin D style); transfer setups to and from PC.
Units Inches or millimeters
Battery Pack Discharge time - 3.5 to 7 hours (ea), typical, depending upon display
back light usage.
44
Transducer Cable BNC Connectors.
Connectors
Size
Weight (Including Approximately 6.3 pounds
Battery Pack)
55
3.0 Operator Qualifications
The Curlin AIR's concept and design makes it the most unique ultrasonic flaw detector available.
In order for the owner of this advanced technology instrument to fully benefit from the unique features
of Curlin AIR, the assigned operator(s) must be experienced and well-founded in the fundamentals of
ultrasonic testing. Operators should fully possess the qualifications of ultrasonic testing personnel as
defined in Recommended Practice No. SNT-TC-1A, Personnel Qualification and Certification in
Nondestructive Testing, available from the American Society for Nondestructive Testing (ASNT), 1711
Arlingate Lane, P.O. Box 38518, Columbus, OH 43228-0518, phone (614) 274-6003, FAX (614) 274-
6899, telex 245347.
3.1Operational Notations
In order to simplify these instructions, certain consistent notations are used. In some cases, a
single key stroke produces a desired change; other changes may require a sequence of key strokes or
require continued key activation to "scroll" changes. The following terms and operational notations will
be used throughout. The operator is advised to become familiar with all of them.
KEYS
On the sealed membrane keypad are two types of keys: the arrow keys and the direct access
keys. In general, the arrow keys are multi functional, being used to change from one highlighted
menu item to another, or to increase or decrease or change variables. The direct access keys
and SEL/HELP keys are also multi functional, being used to select, or enter, a variable, or to
access the HELP screens.
VARIABLES
On analog-type ultrasonic instruments, a variety of knobs are available to change continuous
variables such as range, delay, gate positions, alarm levels, and reject. The same variables are
addressable in Curlin AIR; instead of knobs, holding down a key changes the variable.
SCROLLING
When a key is continuously depressed to change a variable, the effect is referred to as "scrolling".
Scrolling produces either a visible change in the display, a digital readout of the changing value of
the variable, or both.
TOGGLING
Toggling refers to the kinds of changes made by "toggling" switches on an analog-type instrument.
Functions such as OFF/ON, SINGLE/DUAL, and other discrete, limited value operations are
examples of toggling. With the Curlin AIR, toggling requires depressing a key to change a
function from one state to another (OFF/ON, for example), and again depressing the key to further
advance or reverse the discrete function.
CURSOR
This term refers to the means for notifying the operator which menu item, function or variable can
be currently acted upon. In the Curlin AIR, the cursor location is indicated by highlighting the
menu item. Changing the location of the cursor is accomplished by using the arrow keys. In
some menus, when that menu is exited, and later re-entered, the cursored item will be the one last
selected. In other menus, for operator convenience, the cursor defaults to a "home" item, the item
most often used in that particular menu.
NOTE:
The Curlin AIR is a type of computer in operation, relying heavily on it’s extremely sophisticated
software for much of it’s operation. Should the unit ever seem non-responsive to the front panel
keys or if the operator would like to set the instrument in the factory default use the following
keystrokes. First, turn the Curlin AIR off using the on/off key. If the unit won’t turn off in this
77
Curlin AIR
Menu Structure
Introduction Screen
Press Select For BackLight
(For Use In Dark Areas)
Software Version Number
-- MAIN MENU --
Energy 1-15
Blank Off, Track, Stay
BlnkAdj 0.1 -10 (X of Energy)
Delay 0 - 150
Gate Ampl / Distance
Level (see Amp or Dist)
Audio
MODE
(Amplitude or
Distance) see following "setup"
IF GATE AMPLITUDE
Alarm Off, POS, NEG
Start 0 - 200
Width 0 - 200
Level 10 - 90%FS
Audio On / Off
DISPLAY
Reject 0 - 90% FS
Display Solid, Hollow
Smooth On, Off
Grid On, Off
Freeze On, Off
Backlite On, Off
Contrast 0 - 100%
Beep On, Off
Lock On, Off
Pressing the MENU key repeatedly, produces the following menu flow.
IF GATE DISTANCE
Velocity .005 - .750"/us
Zero 0 - 3,000,000ns
Block .001 - 200
Level 10 - 90% FS
Hold Last, Min, Max
-- MISC --
Battery Test
Mode Thru / Echo
Units Inch/mm
Save 1 - 50
Recall 1 - 50
Print Screen, Setup
AlmTim 0.1 - 60 sec
PRF 6 - 125
MAIN MENU
5.0 MENU DESCRIPTIONS
One of the significant features of the
Curlin AIR is the Direct Access Keypad.
You can perform a simple calibration
without the use of any menus
whatsoever! Range and Gain are just
a single color coded keystroke
away.They are also, just about the
only keys you’ll need to calibrate the
Curlin AIR!! Also present on the
front panel are keys permitting the
user single stroke access to HELP,
MENU & FLAWgate menus directly.
5.1 Special Menu & Keyboard
As you read through this manual, the
MENU and arrow keys allow you to navigate
through the menu structure. Depressing the MENU key
repeatedly will cycle you through the various available menus
within the Curlin AIR. If you press the MENU key and hold it for
approximately 1 second a full list of the current instrument setup condition will be displayed. Using
the UP & DOWN arrow keys will allow you to scroll to any item in the setup list and then change
the value. This is convenient if you are entering a full setup manually which may be listed in a
calibration procedure. This also eliminates “menu hopping” which can be tedious in these
instances.
Help Screens
On Line Help is available for any menu item at any time. To access help on a particular function
just highlight the menu item and then press and hold the SEL/HELP key to read up to a full screen
of text explaining what the function is and how it’s used! HELP is also available from the full
screen setup mode by moving the highlight to the item and pressing SEL/HELP.
Menu Flow
The following represents the menu flow of the Curlin AIR. Pressing the MENU key repeatedly will
sequence through the menus as shown.
88
One key point to keep in mind is that HELP is always just a keystroke away. No
matter the menu item you are on or have question with, just Hi-Light the menu item
then press and hold the select key. A menu item description and use will pop up for
review. It’s almost like having a mini-manual built in.
Figure 4 - Main Menu
Figure 5 - Blank Adjust @ 10
With limited space available, some menu items are abbreviated. None-the-less, abbreviations
and acronyms represent terms familiar to qualified ultrasonic NDT personnel. Curlin AIR is based
upon a microprocessor, combining in high-speed digital electronic technology and new high capability
ultrasonic instrument features. The following is a description of each Menu item(s) and its associated
sub-functions.
5.2 MENUS
Three Menu items are arranged vertically along the left-hand margin of the display as shown in
the following Figure. Whenever a Menu is displayed, an active A-trace is also displayed. The
highlighted menu item can either be selected (to reveal related sub-functions) or changed by
appropriate keying of the arrow keys. In some highlighted Menu items, variables can be changed
without having to select sub-functions.
5.2.1MAIN MENU
The functions grouped in the MAIN menu are
used to select Curlin AIR basic setup
parameters. These items are logically among
the first to be addressed during a new setup.
Last setup /factory default values are
automatically displayed. Even though the
defaults permit many kinds of ultrasonic tests
using many kinds of transducers to be
successfully performed, precise flaw detection
tests may require refinements to the default
variables to better match material characteristics.
ENERGY
Energy allows the operator to adjust the energy of the pulse, in number of cycles, delivered to the
transducer. Values available are 1 - 15 cycles.
BLANK
Noise blanking, when selected, will blank any unwanted
noise after the first detected signal. BLANK ADJUST sets
the width of the signal to be accepted
before blanking the
balance of the waveform.
Example
C The gage is set to THRU MODE (through transmission)
which is the default mode. In the THRU mode, only the
first received signal is significant. There will be no multiple echos as in pulse echo or pitch
catch. Given this fact much of the trailing information as in Fig 3 is useless and in fact,
maybe a little confusing. Compare Figures 3, 4 & 5. In Fig 3, the BLANKing is turned OFF
and there is a fair amount of “signal” following the first peek envelope. The information past
this point is of little value. In Fig 4, the BLANK mode is turned on to TRACK and BlnkAdj
(BLANK ADJUST) is adjusted to a value of 10. In Fig 5 (next page) the BlnkAdj is set to 4.
Notice the width of the echo envelope and the elimination of any signal past the initial echo
99
Figure 7
Figure 6 - Blank Adjust @ 4
envelope. This feature has helped “clean” up the screen for easier interpretation.
When blanking is set to TRACK, the leading edge is the reference point for the blank
function. This allows the waveform to move horizontally on the screen (due to variation in
velocity or transducer spacing) while still maintaining the blank function relative to the leading
edge of the waveform. When blanking is set to STAY, the blanking will stay at the last
TRACKed position. If the echo moves to the right on the screen (further in time) it will
disappear into the BLANKed zone. The BLANKing and TRACKing take on fixed properties of
the BLANK while it was last in the TRACK mode.
BLNKADJ
Blank Adjust is used to adjust the width of the first signal
accepted before blocking the rest. The number shown is
based on the pulse ENERGY. A number of 2.0, for
example, sets a width two times the time of the pulse.
Selections include:
GAIN
Gain is available from the front panel. This function is used
to adjust the amplitude of the A-trace display. Increasing or Decreasing the GAIN will cause the
A-trace signal to increase or decrease in amplitude.
RANGE
Range is available from the front panel, This function is used to determine how much time
(distance) is represented on the horizontal axis of the A-trace display. Increasing or Decreasing
the RANGE will cause the A-trace display to expand or contract.
DELAY
Used to adjust the start of the A-trace display along the horizontal axis. Gates that are
synchronized with the A-trace will also delay accordingly.
GATE
MEASUREMENT GATE is used to select the type of
inspection to be performed.
C AMPLITUDE mode enables the use of the flaw
amplitude gate and alarm(as seen in Fig 3). Note that
ALARM follows GATE on the menu in this mode.
C DISTANCE gate enables the distance measurement
gate and
calibration settings as seen in Fig 6. Note that
VELocity and ZERO follow DIST. Setting of GATE.
IF IN AMPLITUDE MODE (refer to Fig 7)
ALARM POLARITY
C OFF
: ALARM enables or disables the FLAW gate alarm.
C POSitive
alarm means alarm activates when the signal exceeds threshold in gate.
C NEGative
alarm means alarm activates when signal drops below gate threshold
START
GATE START allows adjustment of the horizontal position of the leading edge (start) of the gate.
Continuously adjustable from 0.000 to 200.0 inches (500mm).
WIDTH
GATE WIDTH allows adjustment of the horizontal position of the trailing edge (width) of the gate.
Continuously adjustable from 0.000 to 200.0 inches (500mm).
1010
Figure 8 - DISPLAY MENU
LEVEL
GATE level allows adjustment of the threshold level of the gate. Adjustable from 10% to 90% of
full screen height in 1% increment.
IF IN DISTance MODE (refer to Fig 8)
VELOCITY (VEL)
Sets the material velocity (inches/microsecond or mm/microsecond) used in calculations of
thickness. NOTE: The VELOCITY value will be decreased by approximately one-half from
reference velocity when ANGLE MODE is selected.
ZERO ADJUST (Zero)
A fine delay function which allows for compensation of transducer wearsurfaces, coupling
membranes and angle beam wedge. Adjustment is continuous from 0 to 300,000 nanoseconds.
BLOCK
BLOCKING GATE is used to Increase or Decrease the length of the gate that is used to block
out any unwanted signals after the Initial Pulse. Adjustment of the BLOCKING GATE will allow
proper setup for IP to first echo measurement in the distance mode. This is used if there is any
spurious noise between the IP and “real” received echo.
LEVEL
DISTance Gate Threshold allows Increase or Decrease the Measurement Threshold level. Any
echo with amplitude equal to or greater than the Thickness Gate Threshold will be measured.
The threshold gate is the bar on the left of the A-Trace frame extending to the first echo in the
figure. This is the vertical point on the echo envelope used for the DISTance measurement.
AUDIO
When AUDIO is On, the audible annunciator will sound when an alarm is activated. When
AUDIO is off, no sound is generated for alarms.
5.2.2 DISPLAY MENU
The functions grouped in the DISPLAY menu are primarily
used to select display related functions. Generally these will
be user preferences, with little, other than REJECT,
affecting calibration.
DISP
DISPlay - There are two selectable screen displays:
C HOLLOW When selected, displays a black lined, hollow A-
trace waveform against a light background.
C SOLID When selected, displays a black filled A-trace waveform against a light background.
SMOOTH
Smoothing, or averaging, is optionally performed on the waveform by setting each point to be the
average of itself and the point before and after it.
GRID
Turns the background grid on or off. Two types of grids can be displayed, Normal or Numbered.
1111
Figure 10 - Hollow DISP
Figure 11 - Hollow Smoothed
Figure 9 - Hollow, Smoothed, No Grid
Figure 13 - Solid, Smoothed, No Grid
Figure 12 - Hollow, Grid, Numbers
SAMPLES OF VARIOUS DISPLAY SCHEMES
FREEZE
When activated stops any screen activity and “holds” whatever was on the screen at the time the
freeze option was selected. Activate the function by using the left & right arrow keys.
BACK LIGHT
Allows On or Off selection of the Back light. Turn the Back light On to operate in low ambient
lighting conditions. The Back light can be switched Off during bright ambient lighting conditions
to save power.
CONTRAST (CONTR)
Adjusts the contrast of the LCD display for best possible viewing. Continually adjustable from full
white to full dark.
BEEP
When BEEP is On, the audible annunciator will sound when a key is pressed or when the user
inputs an erroneous value. If the beep is Off, no sounds are generated for these conditions.
LOCK
When selected, will lock out key pad functions from inadvertently being activated.
5.2.3MISC Menu
BATTERY TEST
Displays the Battery gage and informs the user of the battery charge status. Press SEL/HELP to
view current battery status
MODE
Used to select the correct distance display, according to the probe configuration. THRU shows
the correct distance when the transducers face each other ( through transmission ), and ECHO
shows the round-trip distance when they are side by side in a pitch catch configuration.
UNITS
Allows selection of the basic unit of measurement for the gage. Select between INCHES or
MILLIMETERS.
SAVE SETUP
1212
Figure 14 - Save Location
Stores the Ultrasonic setup currently in use. 1 - 50
setups can be stored. A user defined, 16 char
comment can be applied to each saved setup. When
the cursor is on the SAVE or RECALL field the
comment for that setup is displayed on the status line.
The parameter LIST allows the user to see and select
from a list off all stored setups with their comments. To
Save a setup, follow these stepes:
C Cursor to SAVE
C Press the right &/or left arrow keys to locate an
available position (1-50). Used setup locations will display a user defined setup name in the
status line (line available above A-Trace). The status line will normally display the receiver
MODE.
C If an attempt is made to save “over” an existing location, a message will be displayed asking
for verification to overwrite the existing setup.
C When an available location is found, a screen as shown in the previous figure prompting for
an alpha-numeric setup name. To name the setup, use the up, down, left & right arrow keys
to scroll to a character (for example, highlight menu item A - Z and press the right arrow key
until the letter C is displayed) and then press SEL/HELP to select the desired character.
C Continue this operation up to 16 characters to fill in the setup name line.
C Scroll down to SAVE and press SEL/HELP
C An exit from the save menu will be performed and the setup name will be displayed in the
status line.
Rcall
Recalls any ultrasonic setup stored in memory. This includes the Factory Default setup (NEW) or
any of the other 50 instrument setups. Note: If the gage is powered ON with the SEL/HELP
button depressed, the Factory Default setup will be loaded. When the cursor is on the SAVE or
RECALL field the comment for that setup is displayed on the status line. The parameter LIST
allows the user to see and select from a list off all stored setups. Use the left & right arrow keys
to move through the list in sequence or, set Rcall to list and press select. A list showing setup
number along with alpha numeric name will then be displayed. Scroll down through the list to
select the desired setup and press SEL/HELP
PRINT
SCREEN
Allows selection of PRINT SCREEN. PRINT SCREEN, when selected, toggles the screen
printing mode on or off. When the screen printing mode is on, any screen displayed may be
printed by pressing and holding the MENU button. To cancel the screen print function, press the
SEL/HELP button while the cursor is on the PRINT:SCREEN field
SETUP
When highlighted allows the user to print the current to a serial (RS-232) printer or to a file. When
the desired setup is loaded, press the SEL/HELP key.
AlmTim
ALARM TIME: Alarm ON time: This sets the amount of time ( in seconds ) the alarm will stay on
even after a single pulse triggers it. Adjustable from 0.1 seconds to 60 seconds.
PRF
Pulse Repetition Frequency : This is the frequency ( number of times per second ) at which the
system pulses the transducer, and collects thickness / amplitude readings.
1313
6.0 Curlin AIR SETUP AND OPERATING CONSIDERATIONS
6.1 Safe Operation
Always use this electrical equipment in a safe manner, including the practice of all the SAFETY
PRECAUTIONS listed earlier in this manual, as well as applicable local and company specific
safety regulations.
6.2 Safe Inspection Site
Always choose an inspection site (including it’s atmosphere) that is safe, in accordance with the
warnings presented in the SAFETY PRECAUTIONS section of this manual, as well as any
applicable safety regulations.
6.3 Ambient Ultrasonic Noise
The Curlin-AIR operates in the relatively low ultrasonic frequency range in order to achieve its
desired performance characteristics. In this regard, the Curlin-AIR has been electronically
designed to reduce or suppress interference effects that might be caused by ambient sonic and
ultrasonic noise occurring in the vicinity of the inspection site. Such interference can detrimentally
affect the inspection reliability performance (such as flaw detectability, alarms, flaw gate and
display).
Typically, it is unlikely that ambient ultrasonic noise should be a problem. However, it is highly
recommended that the selected inspection site be checked prior to performing any actual testing.
One good check method is to first “calibrate” the Curlin-AIR on a reference sample flaw (see prior
calibrating instructions). Next turn the instrument’s power OFF, disconnect the transmitter coaxial
cable from the front panel BNC connector and, then, turn the power ON again. Now, simply
watch the display and flaw alarms for any noticeable intermittent or continuous ultrasonic “static”
(noise pick-up). It is recommended that this setup be monitored long enough time wise to obtain
a reliable assessment of the inspection site for the possible presence of interfacing levels of
ambient ultrasonic noise.
It is not advisable to perform any inspection in the presence of unacceptable levels of ambient
ultrasonic noise. The primary problem with the excess noise levels is that the likelihood of false
indications will be high. In the unlikely event that such noise is observed, it will be necessary to
either move the inspection site to an acceptable, noise free location, acoustically shield the
vicinity adjacent to the ultrasonic beam or determine the source of and prevent the ultrasonic
noise.
6.4 Calibrate Before Performing Any Inspection !!
Always calibrate the Curlin-AIR system with a reference sample(s) in accordance with the
instructions previously given in this manual prior to actual inspection use. Also, check calibration
occasionally during inspection to be sure the system is operating properly.
6.5 Reference Samples
As with all nondestructive testing instrumentation, a satisfactory sample of the material being
tested, containing the minimum-size and type of flaws to be detected, must be used to properly
adjust the detection level of the Curlin-AIR prior to actual use. Operating without such reference
samples will only yield unknown and uncontrolled performance levels.
6.6 Operational Considerations
For experienced Level I or Level II ultrasonic inspectors, setting up and operating the Curlin AIR
will be little different from analog-type ultrasonic instruments. Instead of using knobs and
switches for control adjustments, the appropriate menu item must be found, then selected, then
controlled by using one or more of the keys.
For persons unfamiliar with fully micro computerized instruments, perhaps the most frustrating
part of learning is in the finding
the item they wish to control. In designing the menus for the
1414
Curlin AIR, particular attention has been paid to minimizing the number of menus and simplifying
the process of locating desired functions. To some extent, once the operator is familiar with the
operation, digitally controlled instruments are easier, more intuitive to operate than analog types.
6.7 Curlin AIR Display Characteristics
A difference noticed by experienced operators of analog-type instruments is in the "stair-step"
appearance of the display on instruments having digitally derived pixel-type displays, especially
at relatively short ranges of the time base. There are two aspects of this type of display that
should be considered:
1. The effect on determining vertical linearity, and
2. The effect on the appearance
of horizontal linearity.
Horizontally, the Curlin AIR A-Trace area, exclusive of the menu area is 170 pixels wide.
However, because of the way a smoothly curved analog waveform is sampled by analog-to-
digital circuitry, "stair-stepping" can produce the appearance of being less smoothly shaped.
This effect is visual only, being more pronounced when the time-base is at short range.
However, the determining factor regarding horizontal linearity is in the digital sampling rate, not in
the actual visible appearance. Because distance/thickness readouts in Curlin AIR are
determined from a Hi-Speed sampling or “digitizing” rate, the digital readout has inherent
accuracy and resolution far superior to that which can be measured visually from a conventional
CRT. Even at modestly short ranges, measurements made visually from Curlin AIR’s display are
essentially equivalent to those made from CRTs. Curlin AIR’s thickness readouts are precise,
regardless of the range or stair-stepped appearance of the A-trace display.
Within a short time of familiarization with Curlin AIR’s display and alarm systems, the operator
will adapt to the display and will especially appreciate the large area and visibility of the A-trace.
Most experienced operators of analog-type instruments can effectively operate the Curlin AIR
without reference to the Operator's Manual. However, some of the procedures outlined below
will guide the operator in quickly setting up and using the many features incorporated in Curlin
AIR that are not found in many "standard" ultrasonic instruments. The versatility of Curlin AIR
will be readily evident by following the procedures described below.
6.8 Factory Setup Variables
When the Curlin AIR is turned on by pressing the ON/OFF control, the LOGO screen is
displayed automatically. When the gage is operated for the first time, or if FACTORY DEFAULT
is selected from memory, all the control variables "default" to specific values (After that they will
display the last previous setup parameters). For many common thickness gaging applications, it
is only necessary to "fine tune" a few variables. Some applications can be started without any
adjustments. The FACTORY setup variables have values that match or nearly match a variety of
the most commonly used transducers for thickness gaging. By pressing any key to continue, then
by depressing and holding the MENU key the FACTORY setup default variables are displayed.
1515
7.0 Flaw Detection Procedures
7.1 Mechanical Assembly
Be sure power is turned off before you connect or disconnect the transducers from the
instrument.
7.1.1 Attachment of Transducers
Attached the AT1 transducers to the screw-on connectors or clamp assembly end of the yolk.
The transducers are identical and interchangeable.
7.1.2 Adjustment of Yoke
Adjust telescopic arms of yoke to fit width of material to be manually scanned. Adjust air gap
between the yoke arms so the gap is satisfactorily large enough to accommodate the thickness
of the material to be scanned. Align transducers so they are precisely opposite one another by
using the various rotation and linear adjustments provided by the yoke and adapters.
7.1.3 Cable Attachment to Curlin-AIR
With Curlin-AIR’s power turned OFF
, connect the cables to the BNC coax cable connectors
located on the lower right-hand corner of the Curlin-AIR’s front panel.
7.2 Electronic Initialization & Checkout
7.2.1 Power-ON.
Turn on the Curlin-AIR by pressing the green/red button in the lower left-hand corner of its front
panel.
7.2.2 Message Display
The first message on the display screen will ask you to select or not select to use a backlit
screen (for use in darkened areas). Follow the screen prompt/instructions.
7.2.3 Automatic Initializing Set-up
By pressing any button after the battery check screen appears will automatically adjust the
instrument so it displays a “reference air A-scan” pulse and Flaw Gate alarm on the screen.
This automatic “default” program assures you that the system is basically operational. At this
point, you can adjust (and practice) changing ultrasonic pulse signal by adjusting the GAIN and
RANGE controls on the front panel. Also, the DELAY menu function allows you to scroll the
pulse across the baseline without expanding or compressing the signal (in contrast to the
RANGE control). Move your hand or some object through the ultrasonic beam and note the
disruption of the signal on the screen and triggering alarms. Also note the beam diameter in a
similar fashion..
7.2.4 Insert Material Into Beam
Place the material under test into the ultrasonic beam, or move the yoke over the material.
Orient the ultrasonic beam (transducers) so it is as perpendicular
as possible to the surface of
the material. This is the required alignment for inspection and is called “NORMALIZING THE
BEAM”. Adjust (increase) the GAIN until you obtain a pulse signal of reasonably high amplitude
(greater than 75% of screen height). Adjust the RANGE (if necessary) and/or DELAY controls
to conveniently position the signal along the baseline For preliminary checkout, satisfactorily
position the horizontal location/width and the vertical location (alarm trigger level) of the FLAW
GATE on the signal (see adjusting the FLAW GATE). Slowly scan the ultrasonic beam across
the material to develop a “feeling” to maintain beam alignment (as perpendicular as possible)
and handling of the yoke. Watch the pulse on the display while scanning, keeping it “peaked”
by maintaining perpendicular beam alignment with the yoke. (In actual inspection work, the
FLAW GATE is adjusted to automatically trigger the alarm whenever a detected level of flaw
signal occurs, thus, eliminating the need to constantly watch the screen display).
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27

NDT Systems Curlin-Air User manual

Type
User manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI