Start-Up Procedure 7
the revision of
the instrument control code. The following table list the meaning of the different status
display states.
Instrument status
display code
Instruments prior to October 1996
Revision A
Instruments prior to April 1997
Revision B
Instruments after April 1997
Revision C
0
start code
start code
start code
1
laser operational
laser operational
laser operational
2
scan motors operational
scan motors operational
scan motors operational
3
initiate ZPD search
scan displacement large enough
for 1 cm
-1
operation
sc
an displacement large
enough for 1 cm
-1
operation
4
ZPD position same as previous
(requires two consecutive identical
ZPD positions)
initiate ZPD search
initiate ZPD search
5
ZPD position not same as previous
(two different consecutive ZPD
positions)
ZPD
position same as previous
(requires two consecutive
identical ZPD positions)
ZPD position same as
previous (requires two
consecutive identical ZPD
positions)
6
ZPD located (requires three
consecutive identical ZPD
positions)
Invalid ZPD position (A ZPD
p
osition must be clearly defined)
Invalid ZPD position (A ZPD
position must be clearly
defined)
7
fringe counter initialisation with
ZPD location successful, normal
scanning
ZPD position not same as
previous (two different
consecutive ZPD positions)
ZPD po
sition not same as
previous (two different
consecutive ZPD positions)
8
ZPD located once (requires three
consecutive identical ZPD
positions)
ZPD located once (requires
three consecutive identical
ZPD positions)
9
fringe counter initialisation
successf
ul, ZPD located three
times at fringe counter zero
position, normal scanning
fringe counter initialisation
successful, ZPD located three
times at fringe counter zero
position
A
normal scanning
The ACTIVE LEDs on the instrument and on the DSP board indicate that the fibre optic link is
functioning properly and that there is communication between the DSP board, located in the computer,
and the unit itself. These LEDs will only turn on when there is power on both the MR200 and the host
computer and when
the fibre optic cable is properly connected.
Please note that the fibre optic data cable can be connected or disconnected at any time whether the unit
is turned on or not. Doing so will not harm the equipment.
For both MR100 and MR200, the other end of t
he data connects to the DSP board which must be installed in the
host computer. The DSP board differs for the MR100 and MR200.