Mettler Toledo conductivity transmitter Cond7050e Operating instructions

  • Hello! I am an AI chatbot trained to assist you with the Mettler Toledo conductivity transmitter Cond7050e Operating instructions. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
Press , display shows
Press to scroll to s.CD. 22.
Press , display shows
Press to select desired function.
ENTER
ENTER
Conductivity Transmitter Con 7050 e
Short Instructions
How to enter in set-up menu
How to return to measurement
TC
000
S.CD.
Security Codes:
Set-up (operation parameters) S.CD 22
Calibration C.CD 11
Subject to technical changes.
© Mettler-Toledo GmbH 09/02
Printed in Singapore. 52 120 986
Mettler-Toledo GmbH, Process Analytics
Industrie Nord, CH-8902 Urdorf
Tel. +41 1 736 22 11
Fax +41 1 736 26 36
Press together, escape function.
Kurza_7050e_E_52 120 986 24.09.2002 17:11 Uhr Seite 1
Calibration
Set ATC or manual temperature compensation
Entering set points
SETUP
HOLD
6
CAL
Con
SETUP CAL
HOLD
6
0.100
CEL
SETUP CAL
HOLD
6
14.3
25.0
SETUP CAL
HOLD
6
1.01
FCt
SETUP
HOLD
SEt
°C °F
SETUP
HOLD
On
AT C
SETUP
HOLD
OFF
AT C
SETUP
HOLD
SP1
SETUP
HOLD
200
SP1
µS
SETUP
HOLD
LO
SP1
How to enter in calibration menu
µS
°C
AT C
Enter calibration menue and
press to start calibration.
CAL
Adjust cell constant with
keys and press .
ENTER
Put sensor in calibration solu-
tion and use to match
measured value (µS/mS) to
standard solution, press .
ENTER
The display shows the revised
factor. Press to return to
measurement.
CAL
Press , display shows
Press to scroll to C.CD. 11.
Press , display shows
ENTER
CAL
000
C.CD.
Enter Setup menu with S.CD 22
and press until display.
shows “SEt °C °F”, press .
ENTER
Select unit °C or °F with
keys and press .
Select ATC ON or ATC OFF with
keys and press .
If ATC OFF is selected, set manu-
ally calibration and process
temperature.
Press to confirm.
ENTER
ENTER
ENTER
Enter in Setup menu and scroll
with to show SP1
Press .
ENTER
Scroll with to
desired value (µS/ms)
Press .
ENTER
Set relay in LO or HI mode.
Press . Proceed to adjust
hysteresis and delay timer.
Continue same with SP2.
ENTER
CAL
Con
SETUP HOLD