Rosemount NGA 2000 NDIR Analyzer Module SW 3.3-Rev A Owner's manual

Category
Carbon monoxide (CO) detectors
Type
Owner's manual
N
ON
-D
ISPERSIVE
I
NFRARED
A
NALYZER
M
ODULE
NGA2000
R
OSEMOUNT
A
NALYTICAL
748417-A
The information contained in this document is subject to change without notice.
This manual is based on the production version of the Non-Dispersive Infrared Analyzer Module. Hardware
and/or software changes may have occurred since this printing.
Rosemount Analytical's NGA 2000 system of Modular Gas Analyzers and Controllers are patented,
under U.S. Patent 5.787.015.
Manual Part Number 748417-A
August 1999
Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
NOTICE
Pyrex
®
is a registered trademark of Corning Glass Works
Irtran
®
is a registered trademark of Eastman Kodak Co.
Teflon
®
and Viton
®
is a registered trademark of E.I. duPont de Nemours and Co., Inc.
Ty-Rap
®
is a registered trademark of Thomas & Betts Corp.
CONTENTS
748417-A
August 1999
i
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
P
REFACE
Purpose/Safety Summary.................................................................................1
Glossary ....................................................................................................3
Specifications - General...................................................................................5
Specifications - Sample....................................................................................6
Specifications - Physical...................................................................................6
Customer Service, Technical Assistance and Field Service.............................7
Returning Parts to the Factory..........................................................................7
Training ....................................................................................................7
Documentation.................................................................................................8
Compliances ....................................................................................................8
S
ECTION
1. I
NTRODUCTION
1.1 Overview....................................................................................................1
1.2 Typical Applications...................................................................................1
1.3 Theory of Technology................................................................................2
1.4 Sample Requirements...............................................................................3
1.5 Purge Kits..................................................................................................4
1.6 Features....................................................................................................4
CONTENTS
748417-A
August 1999
ii
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
S
ECTION
2. I
NSTALLATION
2.1 Unpacking................................................................................................. 7
2.2 Assembly.................................................................................................. 7
2.3 Location ................................................................................................... 8
2.4 Gases ................................................................................................... 9
2.4.1 Specifications................................................................................ 9
2.4.2 Connections.................................................................................. 10
2.5 Electrical Connections .............................................................................. 11
S
ECTION
3. S
TARTUP AND
O
PERATION
3.1 Overview................................................................................................... 15
3.2 Displays SCREENS.................................................................................. 15
3.2.1 Run Mode Display......................................................................... 15
3.2.2 Menu Displays............................................................................... 15
3.2.3 Help Displays ................................................................................ 16
3.3 Startup Procedure..................................................................................... 16
3.4 Binding ................................................................................................... 19
3.5 Calibration................................................................................................. 19
3.6 Linearization.............................................................................................. 22
3.7 Routine Operation..................................................................................... 25
CONTENTS
748417-A
August 1999
iii
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
S
ECTION
4. M
AINTENANCE AND
T
ROUBLESHOOTING
4.1 Overview....................................................................................................27
4.2 PCB Replacement.....................................................................................28
4.3 Power Fuse Replacement.........................................................................29
4.4 Module Fan Replacement.........................................................................30
4.5 Chopper Motor Replacement ....................................................................31
4.6 Source Replacement.................................................................................31
4.7 Detector Removal......................................................................................32
4.8 Flow Sensor Replacement ........................................................................33
4.9 Case Temperature Sensor Replacement..................................................33
4.10 Thermal Fuse Replacement....................................................................33
4.11 Oscillator Tune/Source Balance Shutter Adjustment ..............................33
4.12 Cleaning Cells .........................................................................................36
4.13 Cell Desiccant..........................................................................................38
4.14 Modulation Check....................................................................................39
S
ECTION
5. R
EPLACEMENT
P
ARTS
5.1 Replacement Parts....................................................................................41
CONTENTS
748417-A
August 1999
iv
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
A
PPENDIX
A. NDIR I
DENTIFICATION
M
ATRIX
General Precautions For Handling and Storing High Pressure Gas Cylinders
Warranty
Field Service and Repair Facilities
F
IGURES
Figure 1-1. NDIR Technology ........................................................................... 3
Figure 1-2. NGA 2000 NDIR Analyzer Module (Typical - Actual
Configuration May Vary)............................................................... 5
Figure 2-1. Analyzer Module Installation into Instrument Platform.................... 7
Figure 2-2. NDIR Back Panel............................................................................ 10
Figure 2-3. NDIR Front Panel Electrical Connections....................................... 11
Figure 2-4. NDIR Wiring Diagram..................................................................... 12
Figure 2-5 Outline and Mounting Dimensions................................................... 13
Figure 3-1. Run Mode Display .......................................................................... 17
Figure 3-2. Main Menu Display......................................................................... 17
Figure 3-3. Basic Controls Menu....................................................................... 17
Figure 3-4. Expert Controls and Setup Menu.................................................... 18
Figure 3-5. Technical Level Configuration Menu .............................................. 18
Figure 3-6. Typical Help Screen........................................................................ 18
Figure 3-7. Typical Linearization Curve, Linearizer OFF................................... 24
Figure 3-8. Operator-Determined Linearization Curve (Normalized)................. 24
Figure 4-1. Printed Circuit Board Fold-Out Panel Views................................... 28
Figure 4-2. Power Fuse Location...................................................................... 29
Figure 4-3. Fan Assembly................................................................................. 30
Figure 4-4. Motor/Source Assembly.................................................................. 31
Figure 4-5. Cell, PCB Assembly – Exploded View............................................ 32
Figure 4-6. Oscillator Tune, Source Balance Shutter Adjustments................... 35
Figure 4-7. Detector Block (Exploded View) ..................................................... 35
Figure 4.7. Cell Disassembly ............................................................................ 37
CONTENTS
748417-A
August 1999
v
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
T
ABLES
Table 2-1. Cell Purging Times at Atmospheric Sample Pressure......................8
Table 3-1. NDIR Analyzer Module Alarms.........................................................21
Table 4-1. Cell Desiccant ..................................................................................38
CONTENTS
748417-A
August 1999
vi
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
NOTES
748417-A
August 1999
P1
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
PURPOSE/SAFETY SUMMARY
The purpose of this manual is to provide information concerning the components, functions,
installation and maintenance of this particular NGA 2000 module.
Some sections may describe equipment not used in your configuration. The user should
become thoroughly familiar with the operation of this module before operating it. Read this
instruction manual completely.
To avoid explosion, loss of life, personal injury and damage to this equipment
and on-site property, all personnel authorized to install, operate and service this
equipment should be thoroughly familiar with and strictly follow the instructions
in this manual. Save these instructions.
If this equipment is used in a manner not specified in these instructions,
protective systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe personal
injury, death, or substantial property damage if the warning is ignored.
WARNING is used to indicate the presence of a hazard which can cause severe personal
injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor
personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation or maintenance information which is
important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Operate this equipment only when covers are secured. Servicing requires
access to live parts which can cause death or serious injury. Refer servicing to
qualified personnel. For safety and proper performance, this module must be
connected to a properly grounded three-wire source of electrical power.
748417-A
August 1999
P2
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
WARNING: POSSIBLE EXPLOSION HAZARD
This equipment is not designed for and should not be used in the analysis of
flammable samples. Use of this equipment in this way could result in explosion
and death.
Ensure that all gas connectors are made as labeled and are leak free. Improper
gas connections could result in explosion or death.
WARNING: OVER-VOLTAGE SPIKING
If this Analyzer Module is used with a non-Rosemount Analytical power supply,
adding Rosemount Analytical PN 903341 Current Protector in series with the
24V positive power line will prevent over-voltage spiking and resultant fuse
blowing when powering up the instrument.
CAUTION: PRESSURIZED GAS
This module requires periodic calibration with a known standard gas. See
General Precautions for Handling and Storing High Pressure Gas Cylinders at
the rear of this manual.
CAUTION: HAND INJURY HAZARD
Dropping the front panel of the Platform while hand or fingers are inside either
case handle can cause serious injury.
CAUTION: PARTS INTEGRITY
Tampering with or unauthorized substitution of components may adversely
affect safety of this product. Use only factory approved components for repair.
CAUTION: OVERBALANCE HAZARD
This Analyzer Module may tip instrument over if it is pulled out too far and the
Platform is not properly supported.
NOTE
Apply leak test liquid to cell or detectors only as a last resort.
748417-A
August 1999
P3
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
GLOSSARY
Analyzer Module
The module that contains all sensor/detector components for development of a Primary
Variable signal; includes all signal conditioning and temperature control circuitry.
Backplane
The interconnect circuit board which the Controller Board, Power Supply, Analyzer Module
power and network cables, I/O Modules and Expansion Modules plug into.
Control Module
The Operator Interface plus the Controller Board.
Controller Board
The computer board that serves as the Network Manager and operates the Display and
Keypad.
Distribution Assembly
The Backplane and the card cages that hold I/O and Expansion Modules.
Expansion Module
A circuit board that plugs into the Backplane from the front of the Platform and performs
special features not related to I/O functions.
I/O Module
A circuit board that plugs into the Backplane from the rear of the Platform. Has a connector
terminal for communication with external data acquisition devices and provides an
input/output function.
Operator Interface
The Display and Keyboard.
Platform
Any workable collection of the following: Controller Board, Power Supply, Distribution
Assembly, Enclosure and Operator Interface.
748417-A
August 1999
P4
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
Power Supply
Any of a variety of components that provides conditioned power to other NGA 2000
components, from the Power Supply Board that plugs into the front of the Backplane in a
stand-alone instrument to several larger ones that can power larger collections of modules
and components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
Data placed on the network by a module regarding current status, e.g., sample flow, source
voltage and other diagnostic information.
Softkeys
The five function keys located below the front panel display; they assume the function
displayed directly above each on the display, a function dictated by software.
System
Any collection of Analyzer Module(s), Platform(s), I/O Module(s) and Expansion Module(s).
748417-A
August 1999
P5
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
SPECIFICATIONS - GENERAL
M
EASUREMENT
S
PECIES
:
Heteroatomic gases such as ammonia (NH
3
), carbon dioxide
(CO
2
), carbon monoxide (CO), carbon monoxide + carbon
dioxide ethylene (C
2
H
4
), hexane (C
6
H
14
), methane (CH
4
), nitric
oxide (NO) and sulfur dioxide (SO
2
)
RANGES:
10 ppm fullscale to 100% fullscale (application-dependent); 4
fullscale selections, including suppressed zero ranges
R
EPEATABILITY
:
±1% of fullscale (at constant temperature)
M
INIMUM
D
ETECTABLE
L
EVEL
:
0.1% CO
2
(at 1 atm. sample pressure; application dependent)
N
OISE
:
<1% of fullscale, peak-to-peak
L
INEARITY
:
±1% of fullscale with 4th order polynomial
R
ESPONSE
T
IME
:
.05 to 30 seconds (selectable) for 0 to 90% of fullscale
D
RIFT
(Z
ERO AND
S
PAN
):
<±1% of fullscale/24 hours at constant temperature
(application dependent);
<±2% of fullscale/week at constant temperature (application
dependent)
E
FFECT OF
T
EMPERATURE
:
<±1% of fullscale over any 10°C interval for rate of change no
greater than 10°C per hour (application dependent)
E
NVIRONMENT
:
Location - Class B controlled, indoor, non-hazardous
A
MBIENT
T
EMPERATURE
:
0 to 45°C (32 to 113°F)
E
FFECT OF
F
LOW
:
<1% of range when sample flow rate is changed by ≤250
ml/min. (No effect if flow rate is between 0 and 500 ml/min.)
P
OWER
R
EQUIREMENTS
:
24 VDC ±5%, 100 W max.; ripple and noise: <100 mV peak-to-
peak; line and load regulations: <±1%
748417-A
August 1999
P6
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
SPECIFICATIONS - SAMPLE
T
EMPERATURE
:
Non-flammable;: 0°C to 55°C (32°F to 138°F)
FLOW RATE:
500 to 1400 ml/min.
P
RESSURE
:
Maximum 690 hPa-gauge (10 psig), higher pressure in
pressurized cell applications
P
ARTICULATES
:
filtered to <2 microns
D
EWPOINT
:
<40°C (104°F), no entrained liquid
M
ATERIALS IN
C
ONTACT
WITH
S
AMPLE
:
Gold plated Pyrex, sapphire, quartz, Irtran, FEP Teflon, Viton-
A, 316 stainless steel
SPECIFICATIONS - PHYSICAL
C
ASE
C
LASSIFICATION
:
General purpose for installation in weather-protected areas
D
IMENSIONS
:
See Outline and Mounting Dimensions, Figure 2-3
W
EIGHT
:
Standard: 11 kg (24.2 lbs.); extended: 12.5 kg (27.5 lbs.)
M
OUNTING
:
Inside a Platform or custom-installed in a panel
M
AXIMUM
L
ENGTH OF
LON C
ABLE
:
1600 m (1 mile) between Analyzer Module and Platform
See the Preface Section of the Platform manual for specifications regarding Platform
related components.
748417-A
August 1999
P7
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE
For order administration, replacement Parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when you
call: Model Number, Serial Number, and Purchase Order Number or Sales Order Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive or
other hazardous material or used in such a hazardous environment, the user must attach
an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the
material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
TRAINING
A comprehensive Factory Training Program of operator and service classes is available.
For a copy of the Current Operator and Service Training Schedule contact the Technical
Services Department at:
Rosemount Analytical Inc.
Phone: 1-714-986-7600
FAX: 1-714-577-8006
748417-A
August 1999
P8
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
DOCUMENTATION
The following Non-Dispersive Infrared Analyzer Module instruction materials are available.
Contact Customer Service or the local representative to order.
748417 Instruction Manual (this document)
COMPLIANCES
This product may carry approvals from several certifying agencies, including Factory Mutual
and the Canadian Standards Association (which is also an OSHA accredited, Nationally
Recognized Testing Laboratory), for use in non-hazardous, indoor locations
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to
harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May
1993).
This product satisfies all obligations of all relevant standards of the EMC framework in
Australia and New Zealand.
NAMUR
N
9
6
APPROVED
FM
®
97-C219
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
748417-A
August 1999
1
INTRODUCTION
1
1.1 OVERVIEW
This manual describes the Non-Dispersive Infrared (NDIR) Analyzer Module of Rosemount
Analytical's NGA 200 Series of gas analysis components.
The NDIR Analyzer Module is designed to continuously determine the concentration of
oxygen in a flowing gaseous mixture. The concentration is expressed in one of three
fashions:
• parts-per-million
• percent of composition
• percent of fullscale
The user can obtain an output that is linear with concentration by initiating a linearizer,
which is based on a fourth-order polynomial. The linearizer is incorporated in the Analyzer
Module's electronic circuitry and is adjustable through interconnection with the network.
The entire Analyzer Module is designed as a slide-in module (if configured in stand-alone
instrument fashion), removable from the front of the Platform, with gas connections made
from the rear. All electronics relative to sample detection and conditioning are included in
this module.
1.2 TYPICAL APPLICATIONS
The NDIR Analyzer Module is designed to cover a wide range of process, stack and
automotive applications. Typical measurements include:
Chemical and Petroleum
• Carbon dioxide: Manufacture of ethylene oxide, phthalic anhydride and ammonia;
nitrogen generation; and producer gas monitoring
• Carbon Monoxide: Stack monitoring
• Methane: Ammonia manufacture
• Acetylene: Manufacture of acetylene, acrylonitrile, and vinyl chloride
• Sulfur Dioxide: Sulfuric acid stack gas
748417-A
August 1999
2
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
1
INTRODUCTION
Food and Agriculture
• Carbon Dioxide and Water Vapor: Blanketing of perishables, fermentation
processes, photosynthesis studies, personnel protection
Aerospace and Oceanography
• Carbon Dioxide, Carbon Monoxide, and Water Vapor: Diving and space
chambers
Metals and Ceramics
• Carbon Dioxide: Monitoring of producer gas, steel converting, manufacture of
cement, soaking pit, heat treating
• Carbon Monoxide: Inert gas generation, producer gas monitoring, rotary kiln
roasting, tin plate annealing, steel converting, aluminum power processing,
porcelain kilns, tunnels
• Water Vapor: Heat treating, hydrogen brazing, nickel and chrome plating
• Sulfur Dioxide: Flash smelting
• Ammonia: Ammonia dissociation
1.3 THEORY OF TECHNOLOGY
Inside of the Analyzer Module, two equal-energy infrared beams are directed through two
parallel optical cells, a flow-through sample cell and a reference cell. The reference cell
may be sealed or may contain a continuously flowing reference gas. (See Figure 1-1.)
The infrared radiation is interrupted by a chopper at a frequency of 5 Hz.
During analysis, a portion of the infrared radiation is absorbed by the component of interest
in the sample. The quantity of infrared radiation that is absorbed is proportional to the
component concentration.
The detector is a "gas microphone" based on the Luft principle. The detector is generally
filled with the same gas being analyzed. The infrared energy is therefore absorbed at the
same wavelengths in the detector as that in the sample cell, making the detector specific for
the analyzed component. The detector converts the difference in energy between sample
and reference cells to a capacitance change. This change, which is proportional to
component concentration, is processed and expressed as the primary variable on the
network.
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
748417-A
August 1999
3
INTRODUCTION
1
Other modules comprising the NGA 2000 unit then use this variable for a variety of
purposes (e.g., expressing the gas concentration on the Front Panel Display or sending it to
external data acquisition devices).
For a general understanding of the electrical interconnections in the NDIR Analyzer Module,
see Figure 2-4.
F
IGURE
1-1. NDIR T
ECHNOLOGY
1.4 SAMPLE REQUIREMENTS
Maximum allowable sample pressure is 690 hPa-gauge (10 psig) for a standard
configuration NDIR that has a flow restrictor which sets the flow at between 0.5 L/min. to 1
L/min. Special high pressure cells (up to 10,350 hPa-gauge, 150 psig) are available.
Sample temperature range is 0°C to 55°C, and maximum dewpoint is 40°C. The sample
must be filtered to exclude particulates larger than 2 microns in size. Consult factory for
special configurations with specifications outside of those listed above.
SAMPLE IN
SAMPLE
OUT
COMPONENT OF INTEREST
NON-INTERFERING COMPOUNDS
INFRARED
SOURCE
DETECTOR
REFERENCE
CELL
STATIONARY
PLATE
CHOPPER
SAMPLE
CELL
DIAPHRAGM,
DISTENDED
DIAPHRAGM,
DARK STATE
748417-A
August 1999
4
Non-Dispersive Infrared Analyzer Module
Rosemount Analytical NGA 2000
1
INTRODUCTION
1.5 PURGE KITS
A purge kit for the motor source or motor source/flowing reference cell accompanies some
NDIR modules. The purpose of these kits is to improve performance and accuracy through
the reduction of ambient CO
2
interference. They do not provide protection from
explosion hazard. The purge gas vents into the case, which has no outlet fitting for these
types of purge gases.
1.6 FEATURES
Among the features available in the NDIR Analyzer Module are:
• Pressure compensation for barometric fluctuations (optional)
• Flow sensing
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106