Midmark Gas Supply Systems Installation guide

Type
Installation guide

Midmark Gas Supply Systems empower healthcare professionals with cutting-edge technology to ensure the safe and efficient delivery of medical gases. These systems are meticulously engineered to meet the unique demands of healthcare environments, providing reliable and uninterrupted gas supply for critical procedures and patient care.

The systems are highly customizable and can be tailored to specific needs, accommodating various gas types and flow rates. Whether you require a single-gas system for a small clinic or a complex multi-gas system for a large hospital, Midmark Gas Supply Systems offer scalable solutions that adapt to your evolving needs.

Midmark Gas Supply Systems empower healthcare professionals with cutting-edge technology to ensure the safe and efficient delivery of medical gases. These systems are meticulously engineered to meet the unique demands of healthcare environments, providing reliable and uninterrupted gas supply for critical procedures and patient care.

The systems are highly customizable and can be tailored to specific needs, accommodating various gas types and flow rates. Whether you require a single-gas system for a small clinic or a complex multi-gas system for a large hospital, Midmark Gas Supply Systems offer scalable solutions that adapt to your evolving needs.

Midmark
60 Vista Drive
Versailles, Ohio 45380
1-800-Midmark (1-800-643-6275)
www.midmark.com
Part No. 10561700 Rev. B
Canadian
FLO-SAFE
Manifold System
Pre-Installation Kit
Installation Kit
#91307612
#91307613
Reference Only
Style F
Reference Only
i
Section 1.0 Manifold Plate Mounting ....................................... 1-2
1.1 Kit Contents
1.2 Installation of Manifold Mounting Plate
1.2.1 Drywall Mounting
1.2.2 Wall Stud Mounting
1.2.3 Crossmember Mounting
Section 2.0 Piping System ........................................................... 3
2.1 Piping System Installation
2.1.1 Piping materials and Brazing Methods
2.1.2 Testing the Piping System
2.1.3 Purging the Piping System
Section 3.0 Electrical Connections............................................. 3
3.1 Supply Circuit Connection
3.2 Low Voltage Signal Cable
Components designed to comply with the CSA 27396.1-06
Important: Installation must be performed by qualified
Technicians in accordance with local codes.
Warning: To avoid possibility of cross-connection
of oxygen and nitrous oxide, FLO-SAFE components
use a diameter indexing system. All oxygen tubing is
½” O.D. and nitrous oxide tubing 3/8" O.D. It may be
necessary to transition to a larger pipe size when
connecting to the distribution pipeline. When doing
so maintain proper gas line identification.
Caution: When installing Zone Valve and isolation
valve, protect the valve body from the heat damage
by wrapping the connecting tails with wet rags
during brazing.
TABLE OF CONTENTS
CONTENTS
SAFETY
SAFETY WARNINGS AND IMPORTANT INFORMATION
Reference Only
1
COMPONENT MOUNTING
SECTION 1.0
1.1 KIT CONTAINS
The Canadian FLO-SAFE (CFS) Manifold System
Installation Kit is available in two different configurations
based on the zone valve mounting arrangement, being
recessed or surface mount. Installation kit with recessed
zone valve, p/n 91307612 consists of the following:
Description Part No. Qty.
Manifold Installation Kit (includes
mounting plate, 100’ of signal and
hardware package) 40116000 1
Service Isolation Valve 30111600 1
Zone Valve Shut-off Box (Recessed) 91305653 1
Switch Box 30149300 1
Switch Box Support (NLA) 93305355 2
Bushing (NLA- No Longer Available)93305401 1
2-20 Cable 92305263 1
Installation Kit, with surface mount zone valve, p/n
91307613 consists of the following:
Description Part No. Qty.
Manifold Installation Kit (includes
mounting plate, 100’ of signal and
hardware package) 40116000 1
Service Isolation Valve 30111600 1
Zone Valve Shut-off Box (Surface) 91305652 1
Switch Box 30149300 1
Switch Box Support (NLA) 93305355 2
Bushing (NLA- No Longer Available)93305401 1
2-20 Cable 92305263 1
1.2 INSTALLATION OF MANIFOLD MOUNTING PLATE
See Figure 1 to select a location for the manifold
mounting plate. The location selected must be within 36
inches of the proposed location for the gas cylinders
because 36 inch lines are provided to connect the FLO-
SAFE Manifold to the cylinders. Mount the bottom edge
of mounting plate 5 feet above the floor.
FIGURE 2. Installation Options for Manifold Mounting Plate and Main Power Installation
NOTE: UP label indicates proper orientation
1.2.1 Drywall Mounting
1. Use the mounting plate as a template. Mark
locations of the four spacers on back of the four
spacers on back plate and the four corner holes.
2. Drill 1" diameter clearance holes through drywall
at each spacer location so mounting plate fits flush
to wall.
3. Drill ¼” holes at corner holes. Insert hollow wall
anchors in corner holes. Attach plate using anchors.
1.2.2 Wall Stud Mounting
1. Use mounting plate as template. Mark the upper
and lower rear spacer holes.
2. Attach mounting plate to stud using wood screws.
1.2.3 Crossmember Mounting
1. Use mounting plate as a template. Mark the left,
right and lower rear spacer holes.
2. Attach mounting plate using wood screws.
1.3 INSTALLATION OF THE ZONE VALVE
1. Mount Shutoff Box Assy as shown in figure 2.
Note: Do not install pressure switch assemblies
or open packages at this time. Place packaged
assemblies in shutoff box.
2. Run (2) four conductor signal cables from zone
valve box to manifold mounting plate.
3. Feed both cables through plastic bushing at
upper right corner of mounting plate.
4. Replace plaster shield in Zone valve box.
C
B
E
D
A
C
B
E
D
A
MX104800i
DO NOT USE
KNOCKOUTS
ON SIDE OF
ELECTRICAL BOX
IEC
CORD
GAP
SIGNAL
CABLE
ACCESS
FOR WIRING ACCESS
LEAVE THIS AREA OPEN
8” MAXIMUM
LOCATE BOTTOM
EDGE 5’ FROM FLOOR
Mounting Plate Options:
For drywall or surface mounting,
use four corner holes.
For pre-drywall crossmember
mounting, use three holes shown.
For pre-drywall stud mounting, use
two holes shown.
Main Power Installation Options:
Electrical box may be accessed from:
Top knockout
or
Knockout on backside of plate
Reference Only
2
FLOW
MX104900i
Figure 2 Installation Options for Manifold Mounting Plate and Main Power Installation
2-20 CABLE
(6’ PROVIDED)
TRIPLE GANG ELECTRICAL
BOX FOR ALTERNATE
PRE-DRYWALL INSTALLATION
(NOT PROVIDED)
MANIFOLD MOUNTING PLATE:
• LOCATE BOTTOM EDGE 5’ FROM FLOOR
• WITHIN 36” OF SUPPLY GAS REGULATOR
SURFACE MOUNTED
ZONE VALVE
RECESSED
ZONE VALVE
1 3/8”
TWO ZONE VALVE SHUTOFF SIGNAL
CABLE - 4 CONDUCTOR EACH
(100’ PROVIDED)
ALARN SIGNAL CABLE - 4 CONDUCTOR
(100’ PROVIDED)
115 VAC MAIN POWER WITH
DEDICATED CIRCUIT BREAKER
CONNECT WITH WIRE
CONNECTORS (SUPPLIED
WITH ALARM PANEL
12 1/8”
R.O.
2 5/16”
WRAP WET RAGS
AROUND PIPING AT EACH
END WHEN BRAZING
ZONE VALVE
SHUTOFF BOX
IN PATIENT AREA
TO OUTLET
SATATION
ISOLATION
VALVE
2 X 4
1/2”
DRYWALL
O
2
N
2
O
Reference Only
3
2.1 PIPING SYSTEM INSTALLATION
Install O
2
and N
2
O copper tubing (shown in figure 2) as
follows:
1. Leave ½” of lateral flexibility in piping at mounting
plate pigtails to allow for installation of the
FLO-SAFE manifold.
2. When installing Zone Valve and isolation valve,
wrap wet rags around piping to prevent heat
damage to the valve body. See figure 2.
Important: The following information is
offered as a guide and conforms to CSA
Z7396.1-06 as it applies to dental facilities
and small, non-hospital based nitrous
oxide/oxygen piping systems.
2.1.1 Piping Materials and Brazing Methods
1. Piping shall conform to ASTM B819, type K or L
copper tubing, cleaned for oxygen service and
capped prior to installation.
2. Copper to copper joints are to be brazed without
the use of flux.
3. Minimum pipe sizes shall be:
Oxygen gas 12.5 mm (1/2" nom trade size)
Nitrous oxide 6.6 mm (1/4" nom trade size)
Vacuum 19 mm (3/4" nom trade size)
4. While brazing, purge piping system with low flow,
oil-free, dry nitrogen to prevent oxide formation.
2.1.2 Testing the Piping System
After brazing, test piping:
1. Pressurize piping system to 150 psig with oil-free,
dry nitrogen to leak test.
2. Allow piping system to remain pressurized at 150
psig for 24 hrs with source shutoff valve closed.
Allowable pressure changes shall be those caused
by ambient temperature changes.
2.1.3 Purging the Piping System
After successful completion of pressure testing, purge
the piping system:
1. Blow piping system clear with oil-free dry nitrogen.
2. Purge piping systems with intermittent high flow oil
free, dry nitrogen to remove any particulate matter.
Flow each outlet fully until purging process
produces no discoloration in a white cloth.
3. Cap piping system. No further brazing shall be
performed.
4. If further brazing is required, leak test and purge
must be repeated.
3.1 Supply Circuit Connection
All electrical connections must be made to local codes by a
licensed electrician. A dedicated 15A, 115VAC, 14 gauge
supply circuit is required. This must be a dedicated circuit due
to the emergency system requirements, rather than the
amperage demands, of the FLO-SAFE System.
To install the supply line:
1. Feed dedicated branch circuit line to Manifold
Mounting Plate electrical box
For surface-mounted power line, use\
knockout on top of box.
For in-wall power line, use rear center
knockout.
Note: Do not use side knockouts on electrical box”
2. Attach bushing of supplied Romex connector to
incoming cable. Insert connector through chosen
knockout and secure with connector locknut.
3. Connect supply hot line to black or brown wire using
wire connector provided.
4. Connect Supply neutral line to white or blue wire
using wire connector or provided.
5. Connect supply ground to ground post inside
electrical box. Attach ground of Manifold power cord
to same post using #6-32 nut provided.
6. Attach cover to box.
7. To facilitate future servicing, label dedicated
Manifold System supply circuit at service panel.
3.2 Low Voltage Signal Cable
Provided is 300’ of 4 conductor signal cable. Two equal
lengths run between the Manifold Mounting Plate and Zone
Valve Shutoff Box. A third length between the Manifold
mounting Plate and the Alarm panel (see figure 2)
Also provided is a 6’ length of 2 conductor cable which
connects the alarm panel to the 90 dB remote auditory alarm.
Connection details found in the FLO-SAFE Manifold & Alarm
System Installation & Operation Manual #10376600.
ELECTRICAL CONNECTIONS
SECTION 3.0
PIPING SYSTEMSECTION 2.0
Reference Only
Reference Only
145 Mid County Drive
Orchard Park, New York 14127
800 MIDMARK [643-6275] (Customer Service)
888 279 1260 (Technical Service)
www.midmark.com
Part No. 10561700 Rev. A
Reference Only
1-800-Midmark
www.midmark.com
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8

Midmark Gas Supply Systems Installation guide

Type
Installation guide

Midmark Gas Supply Systems empower healthcare professionals with cutting-edge technology to ensure the safe and efficient delivery of medical gases. These systems are meticulously engineered to meet the unique demands of healthcare environments, providing reliable and uninterrupted gas supply for critical procedures and patient care.

The systems are highly customizable and can be tailored to specific needs, accommodating various gas types and flow rates. Whether you require a single-gas system for a small clinic or a complex multi-gas system for a large hospital, Midmark Gas Supply Systems offer scalable solutions that adapt to your evolving needs.