Rosemount NGA 2000 PMD O2 Analyzer Module SW 3.3-Rev B Owner's manual

Type
Owner's manual
748419-B
P
ARAMAGNETIC
D
ETECTOR
A
NALYZER
M
ODULE
S
OFTWARE V
3.3
NGA2000
R
OSEMOUNT
A
NALYTICAL
The information contained in this document is subject to change without notice.
Rosemount Analytical's system of NGA 2000 Modular Gas Analyzers and Controllers are patented, under U.S. Patent
5.787.015.
Manual Part Number 748419-B
June 2000
Printed in U.S.A.
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
NOTICE
Viton-A® is a registered trademark of E.I. duPont de Nemours & Co.
Paliney No. 7â„¢ is a trademark of J.M. Ney Co.
CONTENTS
748419-B
June 2000
i
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
P
REFACE
Purpose/Safety Summary.................................................................................P1
Glossary ....................................................................................................P4
Specifications...................................................................................................P6
Specifications – Sample...................................................................................P6
Specifications - Physical...................................................................................P7
Customer Service, Technical Assistance and Field Service.............................P8
Returning Parts to the Factory..........................................................................P8
Training ....................................................................................................P8
Documentation.................................................................................................P9
Compliances ....................................................................................................P9
S
ECTION
1. I
NTRODUCTION
1.1 Overview ....................................................................................................1
1.2 Typical Applications....................................................................................1
1.3 Theory of Technology.................................................................................1
1.4 Features ....................................................................................................2
S
ECTION
2. I
NSTALLATION
2.1 Unpacking .................................................................................................5
2.2 Assembly...................................................................................................5
2.3 Location ....................................................................................................6
2.4 Gases ....................................................................................................7
2.4.1 Requirements.................................................................................7
2.4.2 Connections...................................................................................9
2.5 Electrical Connections...............................................................................11
CONTENTS
748419-B
June 2000
ii
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
S
ECTION
3. S
TARTUP AND
O
PERATION
3.1 Overview................................................................................................... 13
3.2 Displays ................................................................................................... 13
3.3 Run Mode Display..................................................................................... 13
3.4 Menu Displays .......................................................................................... 13
3.5 Help Displays............................................................................................ 15
3.6 Startup Procedure..................................................................................... 17
3.7 Binding ................................................................................................... 18
3.8 Calibration................................................................................................. 18
3.9 Background Gas Compensation............................................................... 20
3.10 Barometric Pressure Compensation....................................................... 24
S
ECTION
4. M
AINTENANCE AND
T
ROUBLESHOOTING
4.1 Overview................................................................................................... 25
4.2 Printed Circuit Board Replacement........................................................... 25
4.3 Module Fan Replacement......................................................................... 26
4.4 Thermal Fuse Replacement...................................................................... 27
4.5 Flow Sensor Replacement........................................................................ 28
4.6 Power Fuse Replacement......................................................................... 28
S
ECTION
5. R
EPLACEMENT
P
ARTS
5.1 Replacement Parts ................................................................................... 29
CONTENTS
748419-B
June 2000
iii
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
A
PPENDIX
A. PMD I
DENTIFICATION
M
ATRIX
General Precautions For Handling and Storing High Pressure Gas Cylinders
Warranty
Field Service and Repair Facilities
F
IGURES
Figure 1-1. Spherical Body in Non-Uniform Magnetic Field...............................2
Figure 1-2. Paramagnetic Detector Technology................................................3
Figure 2-1. Analyzer Module Installation into Instrument Platform ....................5
Figure 2-2. PMD Front Panel Connections........................................................9
Figure 2-3. PMD Back Panel Connections........................................................9
Figure 2-4. Interconnection of Typical Gas Manifold to PMD Analyzer Module.10
Figure 2-5. PMD Wiring Diagram.......................................................................11
Figure 2-6 Outline and Mounting Dimensions ...................................................12
Figure 3-1. Run Mode Display...........................................................................15
Figure 3-2. Main Menu Display..........................................................................15
Figure 3-3. Basic Controls Menu.......................................................................16
Figure 3-4. Expert Controls and Setup Menu....................................................16
Figure 3-5. Technical Level Configuration Menu...............................................16
Figure 3-6. Typical Help Screen........................................................................17
Figure 4-1. PMD Module - Major Components..................................................26
Figure 4-2. Module Fan Assembly.....................................................................26
Figure 4-3. Detector Assembly..........................................................................27
T
ABLES
Table 3-1. PMD Analyzer Module Alarms..........................................................14
Table 3-2. Calibration Range for Various Zero Based Operating Ranges.........19
Table 3-3. Calibration Range for Various Suppressed Range Operations........19
Table 3-4. Oxygen Equivalents of Common Gases...........................................23
CONTENTS
748419-B
June 2000
iv
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
NOTES
748419-B
June 2000
P1
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
PURPOSE/SAFETY SUMMARY
The purpose of this manual is to provide the procedures for the installation, operation and
maintenance of this NGA 2000 module.
Some sections may describe equipment not used in your NGA 2000 system configuration.
The user should become thoroughly familiar with the operation of this module before
operating it. Read this instruction manual completely.
To avoid explosion, loss of life, personal injury and damage to this equipment
and on-site property, all personnel authorized to install, operate and service
the this equipment should be thoroughly familiar with and strictly follow the
instructions in this manual. SAVE THESE INSTRUCTIONS.
If this equipment is used in a manner not specified in these instructions,
protective systems may be impaired.
DANGER is used to indicate the presence of a hazard which will cause severe personal
injury, death, or substantial property damage if the warning is ignored
WARNING is used to indicate the presence of a hazard which can cause severe personal
injury, death, or substantial property damage if the warning is ignored.
CAUTION is used to indicate the presence of a hazard which will or can cause minor
personal injury or property damage if the warning is ignored.
NOTE is used to indicate installation, operation, or maintenance information which is
important but not hazard-related.
WARNING: ELECTRICAL SHOCK HAZARD
Do not operate without covers secure. Servicing requires access to live parts
which can cause death or serious injury. Refer servicing to qualified personnel.
For safety and proper performance this instrument must be connected to a
properly grounded three-wire source of power.
748419-B
June 2000
P2
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
WARNING: POSSIBLE EXPLOSION HAZARD
This equipment is not designed for and should not be used in the analysis of
flammable samples. Use of this equipment in this way could result in explosion
and death.
Apply leak test liquid to cell or detectors only as a last resort.
WARNING: POSSIBLE EXPLOSION HAZARD
Verify that all gas connections are made as labeled and are leak free. Improper
gas connections could result in explosion or death. See Section 2.4.3 for Leak
test procedure.
WARNING: PARTS INTEGRITY
Tampering or unauthorized substitution of components may adversely affect
safety of this product. Use only factory documented components for repair
WARNING: OVER-VOLTAGE SPIKING
If this Analyzer Module is used with a non-Rosemount Analytical power supply,
adding Rosemount Analytical PN 90331 Current Protector in series with the 24 V
positive line will prevent over-voltage spiking and resultant fuse blowing when
powering up the instrument.
CAUTION: PRESSURIZED GAS
This module requires periodic calibration with a known standard gas. See
General Precautions for Handling and Storing High Pressure Gas Cylinders in
the rear of this manual.
NOTE
748419-B
June 2000
P3
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
CAUTION: HAND INJURY HAZARD
Do not place hands or fingers in Platform front handles when the front panel is
open. Dropping front panel while hand or fingers are inside either handle can
cause serious injury.
CAUTION: OVERBALANCE HAZARD
This Analyzer Module may tip instrument over if it is pulled out too far and the
Platform is not properly supported.
748419-B
June 2000
P4
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
GLOSSARY
Analyzer Module
Self contained analysis modules that are designed to be installed into the NGA 2000
System. One Analyzer Module can be installed into a Single Enclosure containing the
Platform Module. Two Analyzer Modules can be installed into a Dual Enclosure. The
simplest NGA 2000 System consists of one Analyzer Module.
Backplane
The interconnect circuit board which the Controller Board, Power Supply Board, I/O
Board(s) and Expansion Board(s) are plugged into the Backplane
Control module
The operator interface plus the Controller Board.
Controller Board
The Controller Board in the Platform which runs the software program that operates the
Display, Keypad and Network Manager. The Controller Board plugs into the Backplane.
Distribution Assembly
The Distribution Assembly consists of the Backplane and the card cages in the Platform
Module that contain I/O Module(s) and Expansion Module(s).
Expansion Board
The Expansion Board performs special features not related to I/O functions. The Expansion
Board plugs into the Backplane from the Platform front.
I/O Module
An auxiliary module that provides some sort of interface to the outside world. I/O modules
may include analog outputs, relay contacts, and digital interfaces. In general, they are
mounted in platforms as options.
Operator Interface
The Display and Keyboard of the Platform.
748419-B
June 2000
P5
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
Platform
Any combination of the NGA case, the display and computer board, power supply, and I/O
modules. In general, it could be considered to be anything in the NGA system other than
the analyzer modules.
Power Supply
Any of a variety of components that provide conditioned power to other NGA 2000
components, from the Power Supply Board that plugs into the Backplane in a stand-alone
instrument to several larger ones that can power larger collections of modules and
components.
Primary Variable
The measured species concentration value from an Analyzer Module.
Secondary Variable
The current status data placed on the network by an Analyzer Module. This includes
sample flow, source voltage and other diagnostic information.
Softkeys
The five function keys located below the front panel display. The menu function for each
softkey is displayed directly above it and is controlled by the software.
System
A NGA 2000 System consisting of one (or more) Analyzer Modules, an optional Platform,
one or more optional I/O Boards, an optional Expansion Board and an optional
Supplemental Power Supply.
748419-B
June 2000
P6
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
SPECIFICATIONS
MEASUREMENT SPECIES:
Oxygen
R
ANGES
:
0 to 100% oxygen; four fullscale selections, including
suppressed zero ranges in 1% increments
REPEATABILITY:
±1% of fullscale (at constant temperature)
M
INIMUM
D
ETECTABLE
L
EVEL
:
0.01% oxygen
NOISE:
<1% of fullscale, peak-to-peak; <±1% for suppressed ranges
L
INEARITY
:
±1% of fullscale
R
ESPONSE
T
IME
:
0 to 90% of fullscale in 20 seconds (±2 seconds)
D
RIFT
(Z
ERO AND
S
PAN
):
<±1% of fullscale/24 hours, <±2% of fullscale/week at constant
temperature;
<±2% of fullscale/24 hours, <±4% of fullscale/week of range for
99 to 100% (at constant temperature)
E
FFECT OF
T
EMPERATURE
:
<±1% of fullscale over any 10°C interval for rate of change no
greater than 10°C per hour
E
NVIRONMENT
:
Location - Class B controlled, indoor, non-hazardous
A
MBIENT
T
EMPERATURE
:
0 to 45°C (32 to 113°F)
E
FFECT OF
F
LOW
:
<±1% of range when sample flow rate is changed by 20 ml/min.
P
OWER
R
EQUIREMENTS
:
24 VDC ±5%, 50 W max.; ripple and noise: <100 mV
peak-to-peak; line and load regulations: <±1%
SPECIFICATIONS – SAMPLE
T
EMPERATURE
:
Non-flammable;: 10 to 66°C (50 to 150°F)
F
LOW
R
ATE
:
800 to 1400 ml/min.
E
XHAUST
P
RESSURE
:
-345 to 690 hPa-gauge (-5 to 10 psig)
P
ARTICLES
:
filtered to <2 microns
D
EWPOINT
:
below 43°C (110°F), no entrained liquid
M
ATERIALS IN
C
ONTACT WITH
S
AMPLE
:
Glass, 316 stainless steel, titanium, Paliney No. 7, epoxy
resin, Viton A, platinum, nickel, rhodium and MgF
2
748419-B
June 2000
P7
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
SPECIFICATIONS - PHYSICAL
C
ASE
C
LASSIFICATION
:
General purpose for installation in weather-protected
areas
D
IMENSIONS
:
See Outline and Mounting Dimensions, Figure 2-4
W
EIGHT
:
8 kg (17.6 lbs.
MOUNTING:
Inside a Platform or custom-installed in a panel
M
AXIMUM
L
ENGTH OF
LON
C
ABLE
:
1600 m (1 mile) between Analyzer Module and Platform
See the Preface section of the Platform Components manual for specifications regarding
Platform-related components (e.g., case dimensions) and the Preface of the I/O Module
manual for specifications regarding I/O (e.g., relay outputs).
748419-B
June 2000
P8
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
CUSTOMER SERVICE, TECHNICAL ASSISTANCE AND FIELD SERVICE
For order administration, replacement Parts, application assistance, on-site or factory
repair, service or maintenance contract information, contact:
Rosemount Analytical Inc.
Process Analytical Division
Customer Service Center
1-800-433-6076
RETURNING PARTS TO THE FACTORY
Before returning parts, contact the Customer Service Center and request a Returned
Materials Authorization (RMA) number. Please have the following information when you
call: Model Number, Serial Number, and Purchase Order Number or Sales Order Number.
Prior authorization by the factory must be obtained before returned materials will be
accepted. Unauthorized returns will be returned to the sender, freight collect.
When returning any product or component that has been exposed to a toxic, corrosive or
other hazardous material or used in such a hazardous environment, the user must attach
an appropriate Material Safety Data Sheet (M.S.D.S.) or a written certification that the
material has been decontaminated, disinfected and/or detoxified.
Return to:
Rosemount Analytical Inc.
4125 East La Palma Avenue
Anaheim, California 92807-1802
TRAINING
A comprehensive Factory Training Program of operator and service classes is available.
For a copy of the Current Operator and Service Training Schedule contact the Technical
Services Department at:
Rosemount Analytical Inc.
Phone: 1-718-986-7600
FAX: 1-714-577-8006
748419-B
June 2000
P9
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
NAMUR
N
9
6
97-C219
DOCUMENTATION
The following Paramagnetic Detector Analyzer Module instruction materials are available.
Contact Customer Service or the local representative to order.
748419 Instruction Manual (this document)
COMPLIANCES
This product may carry approvals from several certifying agencies, including Factory Mutual
and the Canadian Standards Association (which is also an OSHA accredited, Nationally
Recognized Testing Laboratory), for use in non-hazardous, indoor locations
Rosemount Analytical Inc. has satisfied all obligations from the European Legislation to
harmonize the product requirements in Europe.
This product complies with the standard level of NAMUR EMC. Recommendation (May
1993).
This product satisfies all obligations of all relevant standards of the EMC framework in
Australia and New Zealand.
®
NRTL /C
748419-B
June 2000
P10
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
PREFACE
NOTES
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
1
INTRODUCTION
1
1.1 OVERVIEW
This manual describes the Paramagnetic Detector (PMD) Analyzer Module of Rosemount
Analytical's NGA 200 Series of gas analysis components.
The PMD Analyzer Module is designed to continuously determine the concentration of
oxygen in a flowing gaseous mixture. The concentration is expressed in ppm or percent
volume O
2
.
The entire Analyzer Module is designed as a slide-in module (if configured in stand-alone
instrument fashion), removable from the front of the Platform, with gas connections made
from the rear. All electronics relative to sample detection and conditioning are included in
this module.
1.2 TYPICAL APPLICATIONS
PMD Analyzer Module applications include:
• process control
• continuous emissions monitoring systems (CEMS)
• industrial gas production
• fermentation process monitoring
1.3 THEORY OF TECHNOLOGY
Oxygen is strongly paramagnetic (i.e., capable of becoming a temporary magnet when
placed in a magnetic field) while most other common gases are weakly diamagnetic (i.e.,
tend to be non-magnetic). See Figure 1-1.
The Magnetic susceptibility of the flowing gas sample is sensed in the detector/magnet
assembly. As shown in Figure 1-2, a dumbbell shaped, nitrogen-filled, hollow gas test body
is suspended on a platinum/nickel alloy ribbon in a non-uniform magnetic field.
Because of a "magnetic buoyancy" effect, the spheres of the test body are subjected to
displacement forces, resulting in a displacement torque proportional to the magnetic
susceptibility of the gas surrounding the test body.
Measurement is accomplished by a null-balance system, whereby the displacement torque
is opposed by an equal and opposite restorative torque. The restoring current is
automatically maintained at the correct level by an electro-optical feedback system. A beam
748419-B
June 2000
2
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
1
INTRODUCTION
Shaded
Pole
Piece
Sphere
(Magnetic Susceptibility = k
o
)
F
k
Sample Gas
(Magnetic Susceptibility = k
)
Note:
As percentage of oxygen in sample gas increases,
displacement force (F
k
) increases.
of light from the source LED is reflected off the square mirror attached to the test body onto
a bi-cell (dual photodiode).
The current required to keep the test body to the null position is a linear function of the total
magnetic susceptibility of the sample gas.
See Figures 4-1, 4-2, and 4-3 for component configuration.
F
IGURE
1-1. S
PHERICAL
B
ODY IN
N
ON
-U
NIFORM
M
AGNETIC
F
IELD
1.4 FEATURES
Among the features incorporated into the PMD Analyzer Module is a flow splitter (≈2:1)
that allows for greater sample flow, decreased lag time and faster analyzer response.
The "Time Alignment" feature can be used to delay sending the Primary Variable from
the PMD Analyzer Module for up to 30 seconds in 0.1 second intervals. This feature
allows Primary Variables form more than one PMD Analyzer Module to be "time
aligned" if necessary.
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
748419-B
June 2000
3
INTRODUCTION
1
F
IGURE
1-2. P
ARAMAGNETIC
D
ETECTOR
T
ECHNOLOGY
SIGNAL CONDITIONING
ASSEMBLY
Dual Photocell
Displacement
Tor
q
ue
Restoring
Tor
q
ue
Magnet
Platinum/Nickel Alloy
Suspension Ribbon
Balancing
Weight
Balancing Weight
Nitrogen-Filled Hollow Glass
Test Bod
y
Mirror
Current Loop
TEST BODY DETAIL
Electromagnetic
Axis
Electromagnetic
A
xis
Restoring
Current
Restoring
Current
Displacement
Tor
q
ue
Restoring
Torque
DETECTOR/MAGNET
ASSEMBLY
Test Body
Shaded Pole Pieces (4)
LED Source
Restoring
S
p
an
Zero
SIGNAL OUTPUT
TO NETWORK
NULL BALANCE SYSTEM
748419-B
June 2000
4
Paramagnetic Detector Analyzer Module
Rosemount Analytical NGA 2000
1
INTRODUCTION
NOTES
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94

Rosemount NGA 2000 PMD O2 Analyzer Module SW 3.3-Rev B Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI