Introduction
ThankyouforselectingtheExtechModelRH30.TheRH30monitorsrelativehumidity,air
temperature,dewpointtemperature,andMAXMINreadings.TheRH30alsoincludesan
audible/visualhighhumidityalarm.Thisdeviceisshippedfullytestedandcalibratedand,with
properuse,willprovideyearsofreliableservice.Pleasevisit
theExtechInstrumentswebsite
(www.extech.com)tocheckforthelatestversionofthisUserGuide.ExtechInstrumentsisanISO
9001certifiedcompany.
Safety
GeneralSafety
o Pleasereadallsafetyandinstructionalinformation beforeusingthisproduct.
o Thisproductisintendedforhomeuseonly.
o Unauthorizedrepairs,modifications,orotherchangestotheproductarenotsupported.
o Thisproductisnotintendedforuseinmedicalpractices.
Caution!Riskof
Injury!
o Keepthisproductanditsbatteriesoutofthereachofchildrenandpets.
o Batteries must not be placed in fire, short circuited, taken apart, or discharged. Risk of
explosion!
o Batteries can be fatal if swallowed. Contact medical emergency personnel if batteries are
swallowed.
o Batteries
contain harmful acids.Weakbatteries should bechanged as soonaspossible to
preventdamagecausedbyleakingbatteries.
ProductSafety!
o Donotplacethisproductnearextremetemperatures,vibration,orshock
o Neverholdaprobedirectlyinoroverfire
o Donotimmersethemetersin
anyliquid
2RH30enGB_v1.57/15
1
2
3
4
5
6
7
8
11
9
10
12
AL. ON
AL. OFF
F
C
o
o
65%
55%
60%
Description
MeterDescription
1. RelativeHumidityreading
2. AirTemperaturereading
3. DewPointreading(Td)
4. HumidityAlarmindicator
5. Batterycompartment
6. Resetbutton
7. MINMAXbutton
8. Meterstand
9. AlarmON/OFFswitch(rear)
10. TemperatureUnitsswitch(rear)
11. RHAlarmsettingswitch(rear)
12. Wallmount
accesshole(rear)
TemperatureandHumiditychart(Forgeneralreferenceonly)
Operation
DisplayProtectiveFoil
TheLCDdisplaymayincludeaprotectivefoilcovering.Carefullyremovethisfilmbeforeuse.
Poweringthemeter
Openthebatterycompartmentbyslidingthefrontbatterycompartmentcoverdownward.Ifthe
meter is new and batteries are installed, the battery insulation strip must be removed before
use
so that thebatteries can makeproper circuit contact and power the meter. If nobatteries
areinstalled,pleaseinserttwo1.5VAAbatteriesorientedforproperpolarity.
Theinstrumentshould now bepowered up andrelative humidity, temperature, and dewpoint
temperatureshouldbedisplayedlefttorightontheLCD.
EU consumers are legally bound by the Battery Ordinance to return used batteries to
communitycollectionpointsorwhereverbatteries/accumulatorsaresold. Disposalin
householdtrashorrefuseisprohibited.
Disposal: Follow the valid legal stipulations in respect of the disposal of the device at
the
endofitslifecycle.
3RH30enGB_v1.57/15
Mounting
TheRH30 canbe wall mountedusing the rear wall mount access hole or it can be placed ona
desktoporothersurfaceusingthesuppliedstandbase.
Selecting
o
C/
o
Funitsofmeasure
Usetherear(center)
o
C/
o
Fswitchtoselectthedesiredtemperatureunitofmeasure.
MAXMINandHOLDFunction
TheRH30storesthehighest(MAX)andlowest(MIN)readingssincethemeterwaslastreset.To
viewthemaximum(MAX)reading,presstheMINMAXbutton(rightside,lower).Allreadings
displayedwillnowrepresent
thehighestreadingsencounteredsincelastreset.TheMAX
indicatorwillalsoappearontheleftandmiddledisplaysforconfirmation.PresstheMINMAX
buttonagaintoviewthelowestreadings(MIN)sincethelastreset.Allreadingsdisplayedwill
nowrepresentthelowestreadingsencounteredalongwiththeMINindicators.Toreset(clear)
theMAXMINmemories,pressRESET(rightside,upperbutton)whileMAXorMINisdisplayed.
RelativeHumidity(RH)Alert
The RH Alarm triggers an audible tone and two visual alerts (frowning face symbol illuminates
anduparrowappearsnexttothehumidityreadout) whenthe
RH exceedsthevaluesetbythe
user.Thelimitswitchontherear(right)ofthemetercanbesetto55%,60%,or65%RH.
WhentheRHreturnstothedesiredrange theaudiblealarmwillstopsoundingand thefrowning
facesymbollightwillswitchoff.
Thearrowwillremainhowevertoshowthatthehumiditywas
higher than the set value at least once in the past. Toggle the rear humidity % limit switch to
turnoffthearrow.
PracticalConsiderations
TheBasics
TheRH30thermohygrometermonitorsambientroomconditionsandalertstheuserwhen the
relativehumidity exceeds the user programmedlimit (use therear humidity 55/60/65%switch
tosetthelimit).
Excessivelyhumidroomaircanaffecthealth,promotingtheformationofmoistureandmold.At
the other extreme,
excessively dry room air can also affect health by drying skin, mucous
membranes,and airways. Pets, houseplants, wood flooring, antique furniture, fine art, musical
instruments, etc. are all affected by extremes in ambient conditions. Effective heating and
ventilation can help achieve a comfortable and healthy living environment and can also
cut
costs.
TheBenefitsofVentilation
Forrooms thataretoohumid, ventilationatfirstmayseemabadideainwinterwhenoutdoor
air is also cold, damp, and just as humid. Cold air however can absorb little or no moisture.
When cold air enters a living area it becomes
warmer and can therefore absorb much more
watervapor;inonlyafewminutesthehumiditydecreases.
4RH30enGB_v1.57/15
DewPointTemperature
DewPointisawindowintotherelationshipbetweenTemperatureandHumidity.Forexample,if
air is cooled at a constant (absolute) humidity, the relative humidity will steadily increase to a
maximumof100%. Iftheairiscooledfurther,theexcesswatervaporwillseparateinthe
form
ofwaterdroplets.IftheRH30indicatesanairtemperatureof20
o
C(60
o
F)andarelativehumidity
of65%,thedewpointtemperaturewillbe13.2
o
C(56
o
F).Thismeansthatifwallsorceilingsina
living area are cooler than 13.2
o
C (56
o
F), the air will condense and form droplets on the walls
andceilingsurfaces.Moldanddamppatchescanresult. Inthissameexample,whentherelative
humidityoftheairis40%,thedewpointtemperatureisonly6
o
C(43
o
F).Surfaceswouldhaveto
besignificantlycolderfortheairtoreachitsdewpointtoformwaterdroplets.
TipsonAchievingOptimumAmbientConditions
o Checkthatthedewpointtemperatureofwallsisatleast15
o
C(59
o
F).
o Donotswitchoffhomeheatingcompletelywhenleavingthehouse.
o Ventilateforshortperiodsonlyinallweatherconditions,evenduringrain.
o Airroomsbyopeningwindowsforashorttimeonly.Partiallyopenedwindowsforprolonged
periodsdonotyieldthedesiredresults,wasteenergy,
andcanfurtherpromotethegrowthof
mold.
o If room air is too dry (as it tends to be in winter), humidifier units are highly recommended
sinceairingatlowoutdoortemperatureswilltendtodecreasehumidityfurther.
Specifications
Display MultifunctionLCD
Sensor Precisionhumiditysensorforquickandaccuratereadings
Measurementrange RelativeHumidity:1.0to99.0%RH
AirTemperatureand
DewPointTemperature: 10to50
o
C(14to122
o
F)
Resolution 0.1
o
C(
o
F)and%
HumidityAccuracy ±4%from35%to75%otherwise±5%
Temp.Accuracy ±1.0
o
C(1.8
o
F)
Powersupply 2x1.5VAAbatteries
Dimensions 95x25x95mm(3.7x1.0x3.7")
Weight 90g(3.2oz.)instrumentonly,batteriesnotincludedinweightspec.
Copyright©20132015FLIRSystems,Inc.
Allrightsreservedincludingtherightofreproductioninwholeorinpartinanyform
ISO-9001 Certified
www.extech.com