Roxar MultiCorr MKII Corrosion Meter and Data Acquisition Unit Owner's manual

Category
Measuring, testing & control
Type
Owner's manual
0
CUSTOMER:
TITLE:
MultiCorr MKII
- PORTABLE CORROSION METER.
USER MANUAL
DOCUMENT NO.
: C001-011 Tot. no. of pages: 125
PROJECT DOC. NO.
:
04 29.04.96 For distribution MH LK MB
03 20.02.96
For use including SandLog Terminal
Functions
LK RJ RGB
02 02.02.96
For use including ML Terminal Function
RGB RJ RGB
01 14.09.95 FOR ACCEPTANCE RGB RJ RGB
00 01.09.95 FOR INTERNAL CHECK RGB RJ
REV DATE REVISION BY CHK APP
MultiCorr MKII - User Manual Page: 2
WARRANTY
CorrOcean warrants that the MultiCorr MKII Corrosion Instrument is free from manufacturing
defects and will operate in accordance with the procedures detailed in this Users Manual. If during
the first year, from the date of purchase, the MultiCorr MKII becomes defective due solely to poor
workmanship or design CorrOcean will, at it's sole discretion, repair or replace the defective unit.
Damage to the MultiCorr MKII due to misuse, physical damage or other abuse shall not be
covered under any warranty repairs. The repair or replacement of batteries is not considered part of
the one year warranty and is specifically excluded from any repair or replacement during the unit
warranty period.
Limitations of Liability
Except for the express warranty statement above, CorrOcean grants no warranties, either expressed
or implied, with regard to this product, including all implied warranties of merchantability and
fitness for a particular purpose, and the stated express warranty is in lieu of all liabilities or
obligations of CorrOcean for damages arising out of or in connection with the delivery, use or
performance of the MultiCorr MKII product. CorrOcean assumes no responsibility for any
economic or other losses incurred by the Purchaser, User or that any Third Party might experience
from the use of the MultiCorr MKII, poor interpretation of readings, analysis of data obtained or
from conclusions made based on use of the MultiCorr MKII. In no event shall CorrOcean be
liable for lost profits or consequential damages even if CorrOcean is informed of the possibility of
such damage. In no event will CorrOcean be liable for any claim or demand of a third party.
Copyright 1995 by CorrOcean as, Trondheim, Norway
IBMï›› is a registered Trade Mark of the IBM Corporation
MultiCorr MKII - User Manual Page: 3
CONTENTS
1 INTRODUCTION
6
1.1 Capability and Functions 6
1.2 Probe Compatibility 7
1.3 Data Transmission To PC 7
1.4 Manual Use and Information 7
2 REFERENCE DOCUMENTS
8
3 HANDLING AND STORAGE
9
3.1 General 9
3.2 Packing 9
3.3 Initial Inspection 9
3.4 Storage 9
3.5 Charging 10
3.6 NiCd Batteries 10
3.7 Connecting to a Personal Computer 10
3.8 Connecting to a Probe 11
4 MULTICORR MKII INSTRUMENT
12
4.1 General 12
4.2 Entry Keys 12
4.3 Data Screen 13
4.4 Probe Identification System - Probes Measured by MultiCorr MKII 14
4.5 Probe Identification System - Probes measured by MultiLog 14
4.6 Probe Identification System - Probes Measured by SandLog 15
5 MENU DESCRIPTION
16
5.1 The Main Menu 16
5.2 Operator Mode 17
5.3 Advanced Mode 17
5.4 Terminal mode 18
5.5 Menu Paths 18
6 PROBE IDENTIFICATION AND METER SETUP
19
6.1 Selecting Probe Type and Assigning a Probe Number 19
6.2 Installing and Editing Probe Tag Identification 20
6.3 Reading Setup 22
7 DEFINING PROBE CHARACTERISTICS AND PARAMETERS
25
7.1 General 25
7.2 Entering Electrical Resistance (ER) Probe Parameters 26
7.3 Entering Linear Polarization Resistance (LPR) Probe Parameters 28
7.4 Entering Galvanic Probe Parameters 29
7.5 Entering Analog Probe Parameters 30
7.6 Entering Polarization Parameters 31
8 PROBE MEASUREMENTS
33
8.1 Taking A Reading In The Operator Mode 33
8.2 Taking A Reading In The Advanced Mode 34
9 PROBE MEASUREMENT READINGS AND DISPLAY
MultiCorr MKII - User Manual Page: 4
INFORMATION
35
9.1 General 35
9.2 Measurement Error Messages 35
9.3 Selecting The Probe to Be Monitored 35
9.4 Raw Data and Final 36
9.5 Monitoring Electrical Resistance (ER) Probes 37
9.6 Monitoring Linear Polarization Resistance (LPR) Probes 38
9.7 Monitoring Galvanic Probes 39
9.8 Monitoring Analog Probes, Reference Electrodes and CP Systems 40
9.9 Polarization Readings 41
10 DISPLAYING MEASUREMENT DATA
43
10.1 Selecting A Probe 43
10.2 Displaying Raw Data 43
10.3 Displaying Readings 44
10.4 Displaying ER Probe Corrosion Rate 44
11 MultiCorr MKII SYSTEM FEATURES
46
11.1 Memory and Status 46
11.2 Setting Date and Time 46
11.3 System Reset and Default Settings 47
11.4 Clearing The Memory 48
11.5 Changing The Pass Code 49
12 TRANSMITTING READINGS TO A COMPUTER
50
12.1 Data Reception Software 50
12.2 Transmission Cable Connections 50
12.3 Specifying Data to be Transmitted 51
12.4 Setting the Transmission Speed 51
12.5 Starting Data Transmission 52
12.6 Data Transmission Troubleshooting 52
13 AUTOMATIC READINGS
53
13.1 General 53
13.2 Selecting Reading Intervals 53
13.3 Selecting A Probe For Automatic Readings 54
13.4 Starting Automatic Readings 54
13.5 Stopping Automatic Readings 54
13.6 Automatic Reading Technical Information 55
14 TERMINAL MODE 56
14.1 Global settings for MultiCorr MKII 56
MultiCorr MKII - User Manual Page: 5
15 TERMINAL MODE - MULTILOG 57
15.1 MultiLog Terminal Main Menu 57
15.2 Logger Manager 58
15.3 Transfer Data 70
16 TERMINAL MODE - SANDLOG 77
16.1 SandLog Terminal Main Menu 77
16.2 Logger Manager 78
16.3 Transfer Data 87
APPENDIX A: MultiCorr MKII Specifications 95
APPENDIX B: ER Probe Technical Information 97
APPENDIX C: LPR Probe Technical Information 105
APPENDIX D: Galvanic Probe Technical Information 111
APPENDIX E: Analog Probe Technical Information 113
APPENDIX F: Potentiostatic and Potentiodynamic Procedures 114
APPENDIX G: MultiCorr MKII Error Messages 115
APPENDIX H: Service and Technical Information Sources 117
APPENDIX I: Hazardous Area Certificate 118
APPENDIX J: MultiCorr MKII Menu Pattern 123
MultiCorr MKII - User Manual Page: 6
1 INTRODUCTION
The MultiCorr MKII is a software based, computerized, hand held corrosion meter and data
acquisition unit suitable for the interrogation of different types of corrosion probes. It is housed in a
rugged rubber coated housing that is splash resistant and intrinsically safe. Each MultiCorr MKII
can be individually configured to the user's specific needs at the time of purchase. Additionally, the
unit can have its software upgraded at any time to add Functions which were not included at the
time the unit was purchased.
Each MultiCorr MKII has the capacity to store up to 3,000 sets of readings. Readings may be stored
in memory along with a unique probe number, an alphanumerical tag designation, along with an
automatically generated reading date and time. Written records are not required, but can be easily
maintained.
1.1 Capability and Functions
The MultiCorr MKII is capable of monitoring a wide variety of corrosion probes, so there is no
need to purchase or use different corrosion monitoring instruments when measuring different types
of corrosion probes. Shown below is a summary of the Functions available in the MultiCorr MKII
Instrument.
Function Capability
ER Probe Read and Monitor Electrical Resistance Probes
LPR Probe Read and Monitor Linear Polarization Probes
Galvanic Probe Read and Monitor Galvanic Probes
Analog Probe Read and Monitor Reference Electrodes, Monitored
Anodes, Temperature and Pressure Sensors
Polarization Conduct several types of polarization measurements.
Potentiostatic, potentiodynamic and cyclic polarization
may be performed.
Automatic Conduct Automatic Readings at Uniform Time Intervals
Terminal Function for MultiLog Setup of MultiLog Stations, data transfer from MultiLog to the
MultiCorr MKII, data transfer from the MultiCorr MKII to PC, etc.
Terminal Function for SandLog Setup of Sandlog Stations, data transfer from SandLog to the
MultiCorr MKII, data transfer from the MultiCorr MKII to PC, etc.
The user can include all of the above Functions in a single instrument or chose any possible
combination of Functions at the time of purchase. For instance, you may configure an instrument to
read only ER Probes and take automatic readings or have an instrument that can read ER, LPR and
Galvanic Probes, take Automatic readings and perform Polarization measurements.
The MultiCorr MKII can grow with your operations since any Function which was not included
when the unit was delivered may be added at a later date.
MultiCorr MKII - User Manual Page: 7
1.2 Probe Compatibility
The MultiCorr MKII is a versatile instrument which accepts corrosion probes of different designs
and from different manufacturers. There are default settings within the MultiCorr MKII for each
type of probe, and the user can easily modify these settings so that any probe can be read regardless
of the manufacturer. Because of variations in the size and type of connectors it may be necessary to
have different cables for different manufacturers. When ordering, please specify the manufacturer
and probe type and CorrOcean will provide the required probe cables to suit the user's needs.
1.3 Data Transmission To PC
The MultiCorr MKII can communicate directly with any IBM compatible Personal Computer (PC),
which has been loaded with the CorrOcean Corrlist Software provided with the MultiCorr MKII. It
can also communicate directly with other CorrOcean Software programs such as CorrTrend and
CorrChart. CorrTrend is an advanced corrosion data management program and CorrChart can
be utilized to automatically plot and present polarization curves. Data transfer to the PC can be
performed using either direct data transfer cables or via a modem.
When used as a terminal, the data are transferred into the CorrOcean software MultiCom or
SandRater (for MultiLog and SandLog data respectively)
1.4 Manual Use and Information
This manual includes information on the use of every feature of the MultiCorr MKII Instrument.
The manual provides details on the use of functions which may not have been included in the
MultiCorr MKII purchased by the user. Any MultiCorr MKII function may be added to your
instrument after it has been purchased. Call the nearest CorrOcean office listed in the back of this
manual for more information on how to upgrade your MultiCorr MKII.
It is recommended that the MultiCorr MKII Instrument be used while reading this manual so that
the various options may be tried as the user is reading the manual.
MultiCorr MKII - User Manual Page: 8
2 REFERENCE DOCUMENTS
The MultiCorr MKII portable corrosion monitoring system and related software are also described
in the following documents, available from CorrOcean.
Descriptive Literature
MultiCorr MKII Corrosion Meter Data Sheet 155
CorrTrend Software Data Sheet 154
MultiCorr MKII/CorrChart Software Data Sheet 156
MultiLog Data Sheet 158
The CorrOcean Sand Monitoring System Data Sheet 180
Operating Manuals
Corrlist Software User Manual Document No.: 89-3-174
CorrTag Software User Manual Document No.: 91-3-242
CorrTrend Software Inst. and User Manual Document No.: (A): 87-3-016,
(M): 87-3-032
CorrChart Software User Manual Document No.: 89-3-193
MultiCom - User Manual Document No.: 93-3-340
MultiLog Stations -
Installation/Operation/Maintenance Manual Document No.: 94-3-341
SandRater - Sand Monitoring Program
Operating Manual Document No.: 0200-101
Installation Procedure
Sand Monitoring System Document No.: 2072-044
CorrOcean Sand Monitoring System
User and Maintenance Manual Document No.: 2072-029
MultiCorr MKII - User Manual Page: 9
3 HANDLING AND STORAGE
3.1 General
The MultiCorr MKII is built into a rugged splash proof case and is designed for operation in an
offshore, industrial and/or hazardous environment. Carefully selected materials have been used to
meet these requirements. For example, the application of a silicone rubber cover on the bottom part
provides thermal insulation, shock absorption and a friction mantle. The silicone rubber can
withstand environmental temperatures up to 550°F (300°C). However, the user must remember this
is an electronic instrument and if it is left on a hot surface and the internal temperature is allowed to
exceed 160°F (70°C), the instrument may be permanently damaged.
The specifications on the MultiCorr MKII are described in APPENDIX A "MultiCorr MKII
Specifications".
3.2 Packing
The instrument comes complete with a battery charger, PC data transmission cable, Corrlist
software and a users manual. All of the components are packed in a plastic transportation case filled
with shock absorbing foam.
The plastic case is meant for shipment and storage only, and is not intended for use in a
hazardous area.
3.3 Initial Inspection
The instrument was carefully inspected both mechanically and electronically prior to shipment. It
should be free of physical damage upon receipt. To confirm this, the instrument should be inspected
for any damage that may have occurred during shipment. If any damage is observed please contact
CorrOcean immediately.
3.4 Storage
The instrument may be stored or shipped in environments within the following limits:
Temperature: -5°F to 160°F (-20°C to 70°C)
Humidity: Up to 95%
The instrument must be protected from extreme temperatures which may cause condensation to
form within the instrument case.
It is recommended that the instrument is stored in a dry, cool place (not colder than -5°F (-20°C)) to
minimize battery discharge. To prevent loss of stored data and keep the internal clock running,
recharging should be carried out at least once every 3 months. This recharging should be more often
if the unit is stored at temperatures beyond 85°F (30°C) or if the battery is old.
MultiCorr MKII - User Manual Page: 10
3.5 Charging
Charging time for empty batteries is approximately 12 hours. The instrument may also be operated
during charging.
Charging should only take place in nonhazardous areas.
LOW BATT is displayed on the instrument's status line when the battery charge drops below a
certain limit. The batteries should be recharged as soon as possible after this. If operation
continues, the power will automatically be switched off when the voltage drops below the secure
operation limit. The automatic switch off function is to insure that there is sufficient battery charge
to save stored data. If the instrument has automatically switched itself off under these conditions,
it will not be possible to make measurements until it has been recharged.
NiCd batteries can occasionally develop a "memory" which limits the available user time. If this
occurs deplete the battery through use and then recharge. This process of draining the battery and
recharging may have to be repeated.
3.6 NiCd Batteries
The MultiCorr MKII NiCd batteries have a design life of 3 to 5 years depending on the use and care.
The battery package was made especially for the MultiCorr MKII instrument and are designed not
only for a specific power usage but also for use in a hazardous environment. Changing the battery
should only be conducted by personnel authorized by CorrOcean, to insure that the hazardous area
requirements are maintained.
Warning: NiCd batteries slowly loose their charge even when not in use. For this reason it is
important to charge the batteries at least every 3 months, even if the MultiCorr MKII is not
being utilized.
If the NiCd batteries are not recharged when the LOW BATT Warning appears it is possible that
stored data may be lost if the battery charge level drains below the minimum level
necessary to maintain the MultiCorr MKII memory.
The battery voltage can be seen by selecting STATUS in the system or acq. setup menues.
Both the positive and the negative battery should be at more than 6.25 V when they are charged.
The voltages are however depending on the temperature.
3.7 Connecting to a Personal Computer
The battery charging cable also contains a cable branch for connecting it to a PC RS-232 port. The
MultiCorr MKII may be connected to a PC for data transfer whether or not the battery charger is
connected to an outlet. It is recommended that the charger be connected when data transmission is
being accomplished, but this is not required.
The standard connector on the PC Data Cable is a 25 pin, female connector. Some computer ports
may require the use of a 9 pin female connector, in which case an adapter may be utilized.
Connection to a PC should only take place in nonhazardous areas.
MultiCorr MKII - User Manual Page: 11
3.8 Connecting to a Probe
The MultiCorr MKII is connected to the probe using a flexible retractable probe cable. The probe
cable and connector assembly will be different for the different probe types, and may also be
different for probes from different manufacturers. Probe cables can be supplied by CorrOcean
according to user's specifications. For most commercial corrosion probes, probe cables can be
supplied on short notice. Please contact CorrOcean for more information.
The probe cable has an 11 pin male Fisher connector on the instrument side. There is a white
orientation dot on the Fisher connector that should be visible on the top of the connector when
inserting the cable into the MultiCorr MKII. Always insert and remove the cable by grasping the
connector housing and never pull directly on the cable itself.
The internal wiring of probe cables may be different even when the connector appears to fit the
probe. Make sure that both the probe and cable are compatible for both probe type (e.g. ER, LPR)
and manufacturer.
NOTE:
The instrument housing is electrically conductive. When measuring on LPR and galvanic probes,
the instrument must be held in the hand, or otherwise electrically isolated from ground.
MultiCorr MKII - User Manual Page: 12
4 MULTICORR MKII INSTRUMENT
4.1 General
The user interface of MultiCorr MKII is implemented with an efficient and easy to learn on screen
menu system. In this manual the definition of a menu is a list of functions which can be selected
with a single numerical key entry. Throughout this manual the keyboard entries are enclosed in
brackets [ ] to distinguish them from other numerical listings. The definition of a window is a
display screen with specified contents. Probe information may also be edited or new information
may be added utilizing CorrOcean's CorrTag Software which is described later in this manual.
Since the MultiCorr MKII can store hundreds of different probes in its memory the operator will
learn how to select a specific probe so that readings are calculated and stored for that specific probe.
In this manual the probe which is ready to accept and store readings at any given time is referred to
as the Active Probe.
4.2 Entry Keys
The key board on the MultiCorr MKII is generally self descriptive but it is helpful to understand the
use of these keys as they are used the MultiCorr MKII.
Key Use Description
[On/Off] On or Off Use to turn the MultiCorr MKII screen on or off. The MultiCorr
MKII has an internal battery saver which will turn off the instrument if no key is pressed
within 2 minutes of the last key entry.
[Del] Delete This key is used to delete the last erroneous character entered.
You may also use the left arrow key to make a deletion.
[Exit] Exit The [Exit] key is used to leave a menu or a window. Terminating
data entry with the [Exit] key leaves the previous value unchanged.
[Ent] Enter When entering parameters such as probe numbers or the surface
area of an LPR Probe electrode, the input must be terminated with the [Ent] key. The
input value will not be accepted until the entered value is within legal limits. When an
illegal entry is made, the display cursor will return to the first position in the input field and
a new value may be entered.
MultiCorr MKII - User Manual Page: 13
Arrows Up/Down Once probe parameters have been entered into the unit the operator can
display the next probe in sequence by pressing the down arrow or returning to a prior probe
with the up arrow.
When the menu is set to display data the operator can move through the reading summary using the
Up and Down arrows.
These arrows are also utilized when selecting the alphanumeric tag numbers. This use is described
more fully later in the manual.
Arrows Left/Right The left arrow may be used to delete an erroneous character entry. These
arrows are also utilized when selecting the alphanumeric tag numbers. This use is
described more fully later in the manual.
4.3 Data Screen
The MultiCorr MKII uses a Liquid Crystal Display (LCD) screen. As a battery saving feature the
screen will automatically go blank if no key has been pressed during any two (2) minute interval.
The screen will also go blank during the periods between automatic readings.
The LCD display utilizes a standard pattern for the display of information.
Line Location Description
First Line Provides the name of the menu. This corresponds to the previous
selection.
Second Line This is a warning information line. When the battery is low a
"LOW BATT" warning will appear. When the instrument is set to automatically take
readings an "AUTO-ON" Warning will appear. Additionally, if the MultiCorr MKII
memory is full a "MEMORY FULL" Warning is displayed.
Lines 3-8 Menu selection alternatives are listed when a menu is displayed.
When the unit is in an operational mode, the last two lines (7-8) display the Active Probe
in memory. Both the probe number and tag number are displayed.
Line 3-8 is also used to display any electrical measuring errors which may be detected by the
MultiCorr MKII. The meaning of these error messages is provided in APPENDIX G -
MultiCorr MKII Error messages.
MultiCorr MKII - User Manual Page: 14
4.4 Probe Identification System - Probes Measured by MultiCorr MKII
The MultiCorr MKII uses a standard method to identify each probe so that the user can readily
identify the probe which is being interrogated. It is important for the user to understand this
identification system prior to reading the manual. The following identification labels have a specific
meaning that is described below.
Probe No Each probe is assigned a unique number that includes both the type of Probe (e.g.
ER, LPR, Galvanic) and a number from 0 to 99. For instance ER 23, LPR 99 or Glv 0.
All ER Probe Numbers will start with "ER" and all LPR Probe Numbers will start with "LPR".
The MultiCorr MKII will distinguish between probes of the same type by the number
following the probe type designation. A total number of 200 different probes may be
defined.
Probe Numbers must be assigned before the MultiCorr MKII will take a reading.
Tag No The Tag Number allows the user to assign a name to each probe that can assist in
identifying the probe. The Tag Number may be up to 10 alphanumeric characters in length.
For instance, the user could tag probe ER 12 with the additional identification of "Riser
A67". When this probe is selected the MultiCorr MKII screen would display the following
as the Active Probe.
Probe: ER 12
Tag No: Riser A67
Tag Numbers are not required to be entered before the MultiCorr MKII takes a reading.
However, it is recommended that Tag Numbers are assigned.
4.5 Probe Identification System - Probes measured by MultiLog
The probes measured by MultiLog are identified by tag no. etc. as defined in the MultiCom
software (it is referred to document 93-3-340 for description of the probe identification system).
However, to reduce complexity and data amount to be transferred, tag numbers etc. are not given in
the probe identification in the MultiLog stations and Terminal. Identification of the probes are
given by MultiLog station no, probe type and number. Probe number starts at 1 for each probe type
for each logger.
When MultiLog data are stored in the MultiCorr MKII memory, "tag number" will be
automatically generated. This number is used only for display of probe data and is not transmitted
when data are sent to PC. However, these tag numbers will be part of the total capacity of 200 tag
numbers, and are cleared by reset of the MultiCorr MKII only (please see section 11.3 for reset
instructions).
MultiCorr MKII - User Manual Page: 15
4.6 Probe Identification System - Probes Measured by SandLog
The probes measured by SandLog are identified by the SandLog address and probe number. A
SandLog is measuring one or two sand probes. A sand probe may have 2, 3 or 4 measuring
elements, and each element is displayed individually. SandLog address and probe number must be
defined before measurements are initiated, please see section 16.2.3 for logger setup. Also see
figure 32 for data display in SandLog.
The corresponding sand probe identification is found when data is transferred to the MultiCorr
MKII. Section 16.3.3 describes display of the stored data, an example is shown in figure 37.
For complete sand monitoring system setup, it is referred to "SandRater Sand Monitoring Program
Operating Manual".
MultiCorr MKII - User Manual Page: 16
5 MENU DESCRIPTION
5.1 The Main Menu
The MultiCorr MKII Main Menu is displayed when the meter is turned on. This menu allows the
user to select two operating modes which are defined as the "Operator Mode" and the
"Advanced Mode".
Operator Mode This is a limited use mode designed to allow only field measurements and
storage. This limited operational mode permits the user to quickly take readings without
the need to utilize the expanded menu selection.
Advanced Mode This menu mode allows the corrosion specialist to set probe parameters,
take automatic readings, set engineering units, transmit data to a PC and perform a wide
variety of specialized operations. Some areas of this menu are protected by a pass code to
insure that inadvertent changes are not made to the system. Field measurements may also
be taken using the mode.
Terminal Mode This menu mode allows setup of CorrOcean MultiLog and/or SandLog
stations, check of data and data retrieval from MultiLog/SandLog and data transfer to
relevant PC software. Some areas of this menu are protected by a pass code to insure that
inadvertent changes are not made to the system.
Shown in Figure 1 is the Main Menu of the MultiCorr MKII. To enter either the
Operator Mode or the Advanced Mode the user simply presses the corre-
sponding selection number located in front of the desired selection. The
MultiCorr MKII will immediately move to the first menu level for that
operating mode.
Figure 1 shows the menu with five (5) types of probe functions and terminal
function installed. Only the functions which are actually installed in the user's
MultiCorr MKII will be displayed.
*** MultiCorr MKII ***
Probes: ER - LPR
Glv - Ana
Pol
1: Operator Mode
2: Advanced Mode
3: Terminal Mode
Figure 1
MultiCorr MKII - User Manual Page: 17
5.2 Operator Mode
There are only 2 possible selections in this mode. The user can either decide to
take a measurement on the Active Probe or change the Active Probe prior to
making a measurement.
The details on how to install probe information, take measurements and store
the results are provided later in this manual.
Shown in Figure 2 is the Operator Mode Main Menu Window. Details on
making measurements are provided in Section 8.0.
5.3 Advanced Mode
The Advanced Mode allows the user to make measurements in the same
manner as the Operator Mode and also allows the use of the advanced
capabilities in the MultiCorr MKII. The most sensitive functions in the
Advanced Mode are protected with a pass code to prevent data loss or
corruption caused by unauthorized operators.
Shown in Figure 3 is the Advanced Mode Main Menu Window.
There are six (6) general capabilities that can be accessed through the use of the
Advanced Mode. The use of these capabilities are described in more detail later
in this manual. For now it is only important to understand the general
capabilities of the MultiCorr MKII and how these are arranged in the menu
structure.
1. Acquisition Setup This menu route allows the user to enter the various parameters for each
type of probe, enter the tag identification and establish data reading variables such as the
type of engineering units.
2. Measure The user can take field measurements exactly the same way as it is accom-
plished in the Operator Mode.
3. Show Data The user can view all the stored readings for any probe including raw data,
calculated results and corrosion rates.
4. System Setup This menu route allows the user to set the date and time, clear the memory
of readings, clear the unit of probe parameters, change the pass code and also check the
remaining memory.
5. Transmit This feature allows the stored data to be transferred to a computer.
* MEASURE *
1: Measure
2: Select Probe
Probe = ER 12
Tag No: Riser A67
Figure 2
* MAIN MENU *
1: Acq. Setup
2: Measure
3: Show Data
4: System
5: Transmit
6: Automatic
Figure 3
MultiCorr MKII - User Manual Page: 18
6. Automatic THIS IS AN OPTIONAL FUNCTION THAT MAY NOT APPEAR ON
ALL UNITS. It allows the user to automatically take and store readings at regular intervals.
This function may be added to any MultiCorr MKII.
5.4 Terminal mode
The choice of terminal mode allows user to use the MultiCorr MKII as a terminal for MultiLog
and/or SandLog stations (depending on which functions installed).
If both terminal functions are installed the choice of terminal mode will bring you into a submenu
where you can choose between SandLog and MultiLog terminal. When used as a Sand Terminal
you cannot operate MultiLog stations and vice versa.
If only one terminal function is installed the above mentioned submenu is skipped, and you are
brought directly into the installed terminal menu.
5.5 Menu Paths
As menu selections are made, the MultiCorr MKII will move to the next appropriate menu. This
means that if the MultiCorr MKII needs additional information (For instance, probe parameters) it
will automatically move to the appropriate menu. When the unit has the necessary information to
proceed with an entry it will skip unnecessary menus. This feature saves the user time and
minimizes confusion.
There are some general rules which will assist your understanding of this manual and which are
applicable to all menus.
1. Throughout this manual the keyboard entries are enclosed in brackets [ ] to distinguish
them from other numerical information listings.
2. The Enter Key [Ent] is used to confirm your entry and to move to the next screen.
3. The Exit Key [Exit] is used to leave the entries unchanged and moves you forward in the
menu path.
MultiCorr MKII - User Manual Page: 19
6 PROBE IDENTIFICATION AND METER SETUP
There are general procedures that apply to all probes and these procedures are discussed in this
section. Additionally, the user will learn how to check the operational status of the MultiCorr MKII
and select the type of engineering units which are to be used for the probe readings.
The intent of this section of the manual is only to familiarize the user with general procedures for
identifying and labeling all types of probes. Section 7 takes the operator on a step by step procedure
for entering the parameters for each specific type of probe.
Before taking any measurements the user must assign a unique number to each probe. This number
will define the type of probe (For instance ER or LPR) and allow the MultiCorr MKII to know
where the readings will be stored. At the same time that probe number is assigned you may also
assign a location name, called a Tag Number, to assist you in recognizing the probe you are
measuring. Note that if a Tag Number is not assigned, it is possible to enter the Operator Mode
and change the probe parameters without entering the Pass Code. To protect the probe parameter
setup, a Tag Number must be assigned for the probe.
All of the steps necessary to complete each operation are given along with the keys that need to be
pressed and the results that will be obtained.
6.1 Selecting Probe Type and Assigning a Probe Number
This procedure allows the user to assign a Probe Number without also assigning an identifying Tag
Number.
Step Key Description of Action
------------------------------------------------
1 [On/Off] Turn on MultiCorr MKII
2 [2] Go to Advanced Mode
3 [1] Go to Acquisition Set Up
4 [1] To Edit Probe
5 [Pass Code number] Enter your personal passcode
6 [Ent] Enter pass code and move to next menu
7 [1] To select a Probe Type
8 [1] through [5] Select the type of probe which will be assigned a number
NOTE:
Since every MultiCorr MKII is configured for each user, the meter you are using may not have 5
options from which to select. For your general information the five (5) possible key selections and
the respective probe functions are as follows:
[1] ER [4] Analog
[2] LPR [5] Polarization types
[3] Galvanic
In addition, the two teminal modes, MultiLog and SandLog, are available functions.
MultiCorr MKII - User Manual Page: 20
9 [Ent] To move to Next Menu Screen
10 Enter Any No. 0 - 99 The user can assign any number from 0 through 99. The numbers
do not have to be consecutive. It is often helpful to assign the numbers in the same
sequence that the operator encounters the probes while taking field readings. HINT: If
additional probes will be added to the reading circuit, skip some numbers so consecutive
number spaces will be available at a future date.
11 [Ent] Enters the number you selected.
From this point on the entries depend on the type of probe being utilized. Section 7 of this manual
provides details on how to enter the parameters for each type of probe. The same steps as shown
above will be repeated, but the background information will be excluded.
6.2 Installing and Editing Probe Tag Identification
The MultiCorr MKII provides two (2) Methods for installing Tag Numbers and probe parameters.
1. Manual installation through the use of key entries on the MultiCorr MKII.
2. Automated downloading of probe files through the use of CorrTag Software.
6.2.1 Manual Probe and Tag Number Entry Using The MultiCorr MKII
This procedure allows the user to assign a Probe Number and also assign an identifying Tag
Number. Generally, this is a preferred manual procedure to just assigning a probe number as
shown above in Section 6.1. However, to assign a Tag Number using the MultiCorr MKII the user
must know the pass code assigned to that particular MultiCorr MKII.
Step Key Description of Action
------------------------------------------------
1 [On/Off] Turn on MultiCorr MKII
2 [2] Go to Advanced Mode
3 [1] Go to Acquisition Set Up
4 [1] Edit probe
5 [Pass Code Number] Enter your personal passcode
6 [Ent] To Enter Pass Code and move to the next Menu
7 [1] To Select Probe
8 [1] through [5] Select the type of probe which will be assigned a number
9 [Ent] Enter the probe type you selected
10 Enter Any No. 0 - 99 The user can assign any number from 0 through 99. The numbers
do not have to be consecutive. It is often helpful to assign the numbers in the same
sequence that the operator encounters the probes while taking field readings. HINT: If
additional probes will be added to the reading circuit, skip some numbers so consecutive
number spaces will be available at a future date.
11 [Ent] Enters the number you selected.
NOTE:
As explained in Section 6.1 above, at this point the available functions in each MultiCorr MKII will
vary.
12 [2] Edit Tag no.
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78
  • Page 79 79
  • Page 80 80
  • Page 81 81
  • Page 82 82
  • Page 83 83
  • Page 84 84
  • Page 85 85
  • Page 86 86
  • Page 87 87
  • Page 88 88
  • Page 89 89
  • Page 90 90
  • Page 91 91
  • Page 92 92
  • Page 93 93
  • Page 94 94
  • Page 95 95
  • Page 96 96
  • Page 97 97
  • Page 98 98
  • Page 99 99
  • Page 100 100
  • Page 101 101
  • Page 102 102
  • Page 103 103
  • Page 104 104
  • Page 105 105
  • Page 106 106
  • Page 107 107
  • Page 108 108
  • Page 109 109
  • Page 110 110
  • Page 111 111
  • Page 112 112
  • Page 113 113
  • Page 114 114
  • Page 115 115
  • Page 116 116
  • Page 117 117
  • Page 118 118
  • Page 119 119
  • Page 120 120
  • Page 121 121
  • Page 122 122

Roxar MultiCorr MKII Corrosion Meter and Data Acquisition Unit Owner's manual

Category
Measuring, testing & control
Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI