  • Hello! I have analyzed the provided document regarding moving the 500, 600, and 700 Box Freezers through a doorway. It includes detailed instructions about disconnecting components such as batteries, the chart recorder, door switch and the grille, which must be removed in a specific order before moving the freezer. I'm ready to answer your questions about these steps.
  • What tools are required for moving the freezer?
    What should I do with the batteries before moving the freezer?
    What should be done with the chart recorder if installed?
    What is the last step to move the unit?
GRILLE FILTER
DOOR
REMOVE THE FRONT FILTER GRILLE BY
SLIDING THE DOOR TO THE RIGHT.
STEP 1
CHART
RECORDER
CHART
RECORDER
DOOR
STEP 2a
UNIT WITH CHART RECORDER INSTALLED ONLY.
IF NO CHART RECORDER PROCEED TO STEP 3.
OPEN THE CHART RECORDER DOOR & REMOVE THE CHART
RECORDER BY UNSCREWING THE 4 SCREWS ON THE CHART
RECORDER WITH PHILLIPS HEAD SCREWDRIVER.
CHART RECORDER
BRACKET
STEP 2b
UNIT WITH CHART RECORDER INSTALLED ONLY.
IF NO CHART RECORDER PROCEED TO STEP 3.
REMOVE THE CHART RECORDER BRACKET BY REMOVING
THE 4 SCREWS MOUNTED ON THE DECK FRAME WITH
PHILLIPS HEAD SCREWDRIVER.
DOOR SWITCH
STEP 3
REMOVE THE DOOR SWITCH BRACKET BY UNSCREWING
2 SCREWS FROM THE DECK WITH PHILLIPS HEAD SCREWDRIVER.
IFA BACKUP SYSTEM IS INSTALLED REMOVE THE SECOND
SWITCH ALSO.
BATTERY BRACKET
WITH BATTERY
STEP 4
DISCONNECT THE BATTERY(S). THERE WILL BE
TWO BATTERIES IF YOU HAVE A BACKUP SYSTEM
INSTALLED. REMOVE THE BATTERY BRACKET(S)
BY UNSCREWING 4 SCREWS WITH A 5/16" PHILLIPS
SCREW DRIVER.
GRILLE FRAME
STEP 5
UNSCREW THE GRILLE FRAME SCREWS WITH A
5/16" DRIVER AND SLIDE THE FRAME TO RIGHT &
THEN DOWN.REMOVE THE GRILLE FRAME.
OPEN THE OUTER DOOR 180° ANGLE AND
PUSH THE UNIT FROM THE SIDE THROUGH
THE DOOR.
STEP 6
INSTRUCTIONS FOR MOVING A 500, 600, 700 BOX FREEZER THROUGH A DOOR WAY BETWEEN 35" AND 36" WIDE
TOOL REQUIRED : PHILLIPS SCREWDRIVER, 5/16" DRIVER
ASSEMBLY IS THE REVERSE OF REMOVAL.
ENSURE THE DOOR SWITCH PROTRUDES
FROM THE HOLE IN THE GRILLE FRAME.
A
60806
2011-09-29
SEE SHEET 1
REV.
ECN #
DATE
DESCRIPTION
REVISIONS
THE INFORMATION CONTAINED HEREIN IS
THE PROPRIETARY DATA OF THERMO
FISHER SCIENTIFIC. IT IS NOT TO BE
REPRODUCED OR DISCLOSED IN WHOLE
OR IN PART WITHOUT PRIOR WRITTEN
CONSENT OF THERMO FISHER SCIENTIFIC.
DO NOT SCALE THIS DRAWING.
275 AIKEN ROAD
ASHEVILLE, NC 28804
TITLE
SIZE
SHEET 2 OF 2
DWG. NO.
3rd ANGLE
PROJECTION
S
SolidWorks 2007-10-25
315404H01
INSTRUCTION
TO PUSH UNIT THROUGH DOORWAY
/