Gossen MetraWatt METRAwin 90-F/J Operating instructions

Type
Operating instructions

This manual is also suitable for

The Gossen MetraWatt METRAwin 90-F/J is a Windows program that helps you control various calibrators for electrical quantities with the help of a PC, and for documenting calibration results.

The Gossen MetraWatt METRAwin 90-F/J is a Windows program that helps you control various calibrators for electrical quantities with the help of a PC, and for documenting calibration results.

METRAwin 90
Calibration Software
Operating Instructions
3-349-929-
03
Version 2/06.16
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 2
Copyright
Copyright © 2003 - 2013 GMC-I Messtechnik GmbH. All rights reserved.
These operating instructions, as well as the software described herein, are protected by copyright.
Duplication of this document and/or the software is prohibited without previous express consent.
Violators are liable for compensation of damages.
Trademark
METRAwin
®
90 is a registered trademark of GMC-I Messtechnik GmbH.
Products from other manufacturers are mentioned in these operating instructions for strictly
informational purposes, and their mention does not represent any misuse of the respective
trademarks.
Exclusion of Liability
The texts and illustrations included herein have been compiled with the greatest of care. Despite all
of these efforts, errors can never be entirely prevented. For this reason, no guarantee can be made
for the correctness of the contents of this document. We assume no legal responsibility or liability
of any sort for incorrect entries or their consequences. We appreciate your suggestions for
improvement, as well as any information regarding errors.
We make every effort to keep these instructions up-to-date in accordance with the latest software
version. It is nevertheless possible that differences may exist between the software and the
instructions. Further information may be provided in a
README.TXTfile included on the provided
data storage medium, or in the program directory after the software has been installed.
Subject to change without notice.
Preface
Purpose of the Program
METRAwin 90
calibration software is a Windows program for controlling various
calibrators for electrical quantities
with the help of a PC, and for documenting
calibration results.
Contents of these
Operating Instructions
This document describes the use of the general program functions provided by the
software.
The program
s device-specific functions are described in s
eparate documents which
can be accessed, depending on the currently selected device type, by clicking
Device Functions
in the Help menu.
Further information regarding device
-specific functions is included in the operating
instructions for the respective de
vices.
Target Group
These instructions are intended for users of the software.
The software is designed for use with the Microsoft Windows
®
graphic user
interface. Users must be familiar with basic Windows
®
functions.
Scope of Validity
These instructions
are valid for METRAwin 90 as of version 4.00.
i
Note
The version number can be displayed by clicking
About in the Help menu.
Support
If you have any questions about using this software, please contact:
Your GOSSEN METRAWATT sales partner
The address which can be accessed by clicking About in the Help menu
GMC-I Messtechnik GmbH
Product Support Hotline
Phone : +49 911 8602-0
e-mail: support@gossenmetrawatt.com
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 3
Table of Contents
Preface ................................................................................................................................ 2
Table of Contents............................................................................................................... 3
1 Product Description ...................................................................................................... 4
1.1 Variants and Their Use ............................................................................................................................. 4
METRAwin
®
90-2 ................................................................................................................................... 4
METRAwin
®
90-F ................................................................................................................................... 4
METRAwin
®
90-FJ ................................................................................................................................. 5
1.2 Common Features .................................................................................................................................... 5
1.3 System Requirements .............................................................................................................................. 6
Operating System .................................................................................................................................. 6
Hardware Requirements ........................................................................................................................ 6
Other Required Software ....................................................................................................................... 6
2 Installing, Registering and Uninstalling the Program ................................................ 7
2.1 Installation ................................................................................................................................................. 7
2.2 Registration ............................................................................................................................................... 7
2.3 Uninstalling the Software ......................................................................................................................... 8
3 Operation ....................................................................................................................... 9
3.1 Starting and Exiting the Program ............................................................................................................ 9
3.2 Structure of the User Interface .............................................................................................................. 10
3.3 Overview of Menu Functions ................................................................................................................. 11
3.3.1 File Menu Functions ............................................................................................................................... 12
New ...................................................................................................................................................... 12
Open .................................................................................................................................................... 12
Save ..................................................................................................................................................... 13
Print test report .................................................................................................................................... 14
Printer settings ..................................................................................................................................... 15
Export test data .................................................................................................................................... 15
3.3.2 Help Menu .............................................................................................................................................. 18
Program Functions ............................................................................................................................... 18
Device Functions ................................................................................................................................. 18
Online Support ..................................................................................................................................... 18
Display License Agreement ................................................................................................................. 18
Info ....................................................................................................................................................... 18
3.4 Basic Program Settings ......................................................................................................................... 18
3.4.1 Changing the User Interface Language ................................................................................................. 18
3.4.2 Selecting the Type of Calibrator to be Controlled .................................................................................. 18
3.4.3 Port Settings ........................................................................................................................................... 19
a) For METRAtop 90C / FLUKE D9000, D9100, 5100A, 5500A, 5520A and 5700A
multifunction calibrators ............................................................................................... 19
b) For METRAHit CAL and METRACAL MC process calibrators ..................................... 19
c) For METRAHit 28C and 28C light process calibrators ................................................. 20
d) For METRAHit 18C process calibrator ......................................................................... 20
3.4.4 Showing and Hiding the Multimeter Display ........................................................................................... 21
3.4.5 Showing and Hiding the Command Line ................................................................................................ 21
3.4.6 Showing and Hiding the Toolbar ............................................................................................................ 21
3.4.7 Specifying the Temperature Unit of Measure ......................................................................................... 21
3.4.8 Specifying the Thermocouple Cold Junction .......................................................................................... 22
3.5 DIRECT Operating Mode for Interactive Calibrator Operation ........................................................... 23
3.6 TEST Operating Mode ............................................................................................................................ 24
3.6.1 Creating a Calibration Procedure ........................................................................................................... 24
3.6.2 Executing a Calibration Procedure ......................................................................................................... 26
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 4
1 Product Description
1.1 Variants and Their Use
METRAwin
®
90 calibration software is a multilingual Windows program for controlling various calibrators for
electrical quantities with the help of a PC, and for documenting calibration results.
The program is available in the following variants:
METRAwin
®
90-2
Calibration Software for the Following GOSSEN METRAWATT Process Calibrators:
METRACAL MC, METRAHIT CAL, METRAHIT 28C, METRAHIT 28C light and METRA
Hit 18C
These calibra
tors are used primarily for testing and calibrating electrical measuring
instruments and equipment in the field of process measuring and control technology.
They include the following functions: direct voltage simulator, current source and
sink, as well as
simulation of resistors and temperature sensors (thermocouples and
PTCs). The METRACAL MC and METRAHIT 28C versions also include a TRMS
multimeter (V, mA,
, F, Hz, °C/°F). Thanks to inclusion of these multimeter
functions, or by incorporating a separate
METRAHit (in some cases written
METRAHIT) 18S, 28S, 29S, 30M, X
-TRA, EXTRA or ENERGY precision multimeter
with infrared interface into the calibration system, automated calibration of buffer
amplifiers,
temperature measuring transducers and the like can be performed.
The software executes interactive
(DIRECT operating mode) or sequence-
controlled
(TEST operating mode)
setup of the calibrator from the PC via an RS 232 port (with
the help of the BD232 or SI232
-II interface adapter) or a USB port (with the help of
the USB
-HIT or USB-XTRA interface adapter), automatic analysis of measured
values entered manually or read in from the multimeter via the port, and
documentation and archiving of calibration results in a calibration report.
Calibration procedures for
each respective calibration object can be generated and
tested with the program in an uncomplicated fashion. In the case of the METRAHIT
18C and 28C, procedures can also be transferred to the calibrator
s memory module,
from which they can be executed auto
nomously (without PC). Measured values (V,
mA) acquired by the METRAHIT 28C can be saved to memory automatically and
subsequently read out and displayed at a PC, and inserted into a calibration report.
METRAwin
®
90-F
Calibration Software for FLUKE D9100, 5100B, 5500A, 5520A and 5700A Multifunction
Calibrators
The above mentioned calibrators are used primarily for the calibration of indicating
measuring instruments such as multimeters, laboratory recorders, current clamps,
panel
-mount measuring instruments
etc., and include the various electrical quantities
required to this end. By incorporating a separate METRA
Hit (in some cases written
METRAHIT) 18S, 28S, 29S, 30M, X
-TRA, EXTRA or ENERGY precision multimeter
with infrared interface into the calibration sys
tem, automated calibration of buffer
amplifiers,
isolating transformers, measuring transducers and the like can be performed.
The software executes interactive
(DIREKT operating mode) or sequence-
controlled
(TEST operating mode)
setup of the calibrator from the PC via an IEEE488
interface, automatic analysis of measured values entered manually or read in from
the multimeter via an RS
232 port, and documentation and archiving of calibration
results in a calibration report. The program calculates the test uncertainty ratio (TUR)
for each calibration point based upon the calibrators
accuracy specifications which
are integrated into the software, or overall measuring uncertainty in accordance with
directive DKD
-3 (WECC 19-1990).
A calibration procedure for the
respective calibration object can be generated and
tested in an uncomplicated fashion. Tested calibration procedures for numerous
multimeters from our current and previous product spectrum are included with the
software.
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 5
METRAwin
®
90-FJ
Calibration and Adjusting Software for FLUKE D9100, 5100B, 5500A, 5520A and 5700A
Multifunction Calibrators
Above and beyond this, METRA
Hit 22-29S/M handheld multimeters can be
automatically adjusted via their infrared interface
with the METRAwin
®
90-FJ version
(suitable
calibrators: 5500A and 5520A). Type and function-specific adjusting
procedures required to this end are included with the software.
1.2 Common Features
Convenient, interactive control of the calibrator which is connected to the PC with
direct data entry as an individual value
Uncomplicated and fast generation, testing and execution of calibration
procedures
Simple operation: even semiskilled workers can execute qualified calibration
tasks.
Display of procedural instructions which have been generated automatically or
defined by the user before execution of the respective procedure step
With interconnected multimeter: display and continuous refreshing of the
measured value read in via the port
Outstanding flexibility with calibration signal correction (for
analog measured value
indicators, recorders etc.) by entering a displayed measured value with the
keyboard, or by querying the measured value from the multimeter via the port
ISO-9000 compliant calibration documentation as a standardized or user
configured report with all necessary entries regarding the calibration object and
the calibration system, as well as a tabular list of calibration values and associated
evaluations for each calibration point
Dynamic data transfer to report templates created by the user with Microsoft
®
Excel
or Microsoft
®
Word(e.g. with company logo)
Secure archiving of procedures to an external data storage medium
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 6
1.3 System Requirements
Operating System
METRAwin
®
90 can be run with Microsoft
®
Windows XP, Vista and 7 (32 or 64-bit).
Hardware Requirements
METRAwin
®
90 can be used on PCs which fulfil the minimum requirements of the
respective operating system:
CD ROM drive
Microsoft compatible mouse or other pointing device
For controlling METRAHit 18C, 28C or 28C light process calibrators and for reading in
measured values from METRAHIT 2X multimeters with METRAwin
®
90-2:
A least 1 free serial RS
232 COM port for the connection of 1 to 3 BD232 interface
adapters or SI232-II memory adapters
Or USB port(s) for the connection of 1 to 3 USB-HIT interface adapters
For controlling METRAHit CAL or METRACAL MC process calibrators and for reading in
measured values from METRAHIT Starline multimeters with METRAwin
®
90-2:
USB port(s) for the connection of 1 to 3 USB-XTRA interface adapters
For controlling FLUKE multifunction calibrators with METRAwin
®
90-F/-FJ:
1 installed or external GPIB controller interface (IEEE488) with National
Instruments
compatible protocol and drivers suitable for the Windows system,
for example:
National Instruments PCI-GPIB NI-488.2 or
National Instruments GPIB-USB-B NI-488.2
For printing reports:
Local or network printer which is supported by Windows
Other Required Software
For registering the software:
Microsoft
®
Word 97, 2000, 2003, 2007 or 2010 or another word processing
program for the automatic creation of a fax form in RTF file format (rich text
format)
Or Microsoft
®
Outlook or another e-mail program for the automatic creation of an
e-mail with registration data
For the creation of individualized calibra
tion reports with company logo and/or modified
report layouts:
Microsoft
®
Word 97, 2000, 2003, 2007 or 2010
Or Microsoft
®
Excel 97, 2000, 2003, 2007 or 2010
For viewing and/or printing out the integrated help files in PDF format:
Adobe
®
AcrobatReader available free of charge from
www.adobe.com/acrobat
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 7
2 Installing, Registering and Uninstalling the Program
2.1 Installation
A Setup.exe file is included on the supplied CD ROM. The program is installed in dialog with the user by
executing this file. The installation directory and the program file folder can be selected as desired.
i
Note
METRAwin 90 software is available in a variety of versions for various calibrators. These
software variants can be instal
led to the same directory, as long as only the last digits of the
respective version numbers vary from each other (e.g. 4.04 and 4.01). However, the version
with the highest number must be installed last. The version number can be displayed by
clicking
About in the Help menu.
2.2 Registration
The software can be tested for 30 days without a license key. In order to be able to use the software with its
full scope of functions after this trial period a license must be purchased, the software must be registered
with the licensor and it must be permanently enabled by entering the license key (password).
The assigned password depends on the entered company name and the type of calibrator. Registration must
thus be repeated for each type of calibrator.
Setup Menu:
Re
gistration
Opens the dialog box for entering the registration data:
Enter your address information to the corresponding fields. Be certain to enter
the company name correctly because the registration code is generated from it,
as well as other factors.
Enter the serial number which appears on the included program CD under CD
Serial. If you’ve received the software as a download or by e-mail and dont
have a CD serial number, you
ll have to submit proof of purchase
along with your
registration (copy of the invoice or delivery note as a PDF or image file).
The license key can be requested by fax or e
-mail.
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 8
Click the corresponding button: [Create Fax] or [Create Mail].
Select the device type(s) for which the software will be registered in the dialog
box which then appears and acknowledge with [OK]. The selection option
depends on the purchased software variant.
Send the fax form or e-mail which then appears to the specified address.
Youll receive the password with which the program can be permanently ena
bled
for the specified device type(s) without delay.
Open the dialog box for entering the registration information once again to this
end, enter the password to the corresponding field and click the [Register]
button.
The program is then permanently enabled. You can check this in the
registration
information field. Your company name is displayed in the title bar of the program
window after the program has been restarted.
i
Note 1
Registration is only possible for a single company, but for any desired nu
mber of installations,
with a single software license (branch office license). Detailed information regarding licensing
can be viewed by clicking
Show License Agreement in the Help menu.
i
Note 2
Save the returned e
-mail with the password in case reinstallation should become necessary!
2.3 Uninstalling the Software
If desired, the program can be uninstalled by selecting Programs from the Start menu and then clicking
Gossen-Metrawatt / METRAwin / Uninstall.
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 9
3 Operation
3.1 Starting and Exiting the Program
Starting the Program
In order to start the program:
Double click the METRAwin 90 icon on the Windows desktop.
or
Select the program from the Windows start menu:
Start / Programs / METRAwin 90 / METRAwin 90
The program is started.
Exiting the Program
In order to exit the program:
Select Close in the File menu.
or
Simultaneously press the Alt+F4 keys.
or
Click the Close icon at the upper right-hand corner of the title bar.
The program is exited.
i
Note
When the program is exited the selected language and other software setting parameters
(device type, temperature unit of measure, communication parameters etc.) are saved
automatically. When the program is restarted, the last us
ed settings are still valid.
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 10
3.2 Structure of the User Interface
The
METRAwin 90 program window is subdivided into seven segments:
Figure 3.2: Segments of the Program Window
A) Title Bar
The horizontal ba
r at the top of the program window shows the program name and
registration information, as well as the file name of the currently open procedure. It
also includes the system menu at the left which can be opened via the
program icon,
and buttons (
) for minimizing, maximizing and closing
the program window
at the right.
B) Menu Bar
The menu bar includes the names of the main menus from which the various menu
functions and submenus can be accessed.
C) Toolbar
The smart icons in the toolbar allow for quick acces
s to frequently used commands
and functions. The function of any given icon can be displayed by moving the mouse
pointer to it, and is executed by clicking the icon. After clicking
Toolbar in the Setup
men
u, the toolbar can be hidden or shown.
D) Command Line
Commands transmitted to the calibrator appear in this line for double checking. After
clicking
Command Line in the Setup menu, the command line can be hidden or
shown.
E) Control Bar
Program operating mo
des are selected, and sequential control of the test
procedures is made possible using the buttons in the control bar. The type
designation of the currently controlled calibrator is displayed at the far left, which has
been selected via the
Device Type setting in the Setup menu.
F) Workspace
A calibrator control panel appears in this area in the DIRECT operating mode, or a
table is shown in the TEST operating mode which lists the parameters of the test
procedure steps. Various dialog boxes can be opened here
as well.
G) Context Menu
A context menu appears after clicking the right mouse button with the mouse pointer
in the workspace. Just like the toolbar, it allows for quick access to frequently
required functions.
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 11
3.3 Overview of Menu Functions
File Menu
New
Data associated with any previously loaded test calibration procedure are deleted after
confirming the security prompt.
Open
Opens a calibration procedure file with/without results (*.KLT/*.KLF) in the TEST
operating mode.
Save
Saves a calibration procedure file with/without results (*.KLT/*.KLF) in the TEST
operating mode.
Print test report
Prints out a standard calibration report with a predefined structure.
Printer settings
Selects and configures an installed printer for report printing.
Export test data
Exports the calibration data to an individually laid out Word/Excel report template
(available after running the KLT2DOC.DOT or KLT2XLS.XLM macro).
[Alt+F4]
Exit
Exit the Program
Edit Menu (only available in the TEST/PROCEDURE mode)
[Ctrl+Z]
Undo
Undoes the last editing step. The following can be undone: changes to the parameters
in a procedure line, cut and paste operations, the Repeatoperation.
[Ctrl+X]
Cut
Cuts the selected TEST/PROCEDURE line.
[CTRL+C]
Copy
Copies the selected TEST/PROCEDURE line to the clipboard.
[Ctrl+V]
Paste
The TEST/PROCEDURE line which has been copied to the clipboard is inserted into
the selected line.
Duplicate
Duplicates the selected TEST/PROCEDURE line.
Insert line
Adds a new TEST/PROCEDURE line after the selected line.
Insert procedure
Setup Menu
Language
Changes the program menus, commands and descriptions to a different language.
Registration
Registers the software for permanent use via e-mail or fax.
Device type
Sets the program to work with another device type.
Port settings
Selects the port(s) for communication between the PC and the device and configure
the port parameters.
Multimeter display on
Shows or hides the multimeter display.
Command line
Shows or hides the command line.
Toolbar.
Shows or hides the toolbar.
Cold junction
Selects a thermocouple cold junction (internal/external) for the calibrator.
°C/°F°K/
Selects the temperature unit of measure.
Help Menu
[F1]
Program functions
Opens a PDF file with a description of general program functions.
Device functions
Opens a PDF file with a description of device-specific program functions.
Online support
Opens a website with information regarding technical support via the Internet.
Show license agreement
Opens a TXT file with a description of the license agreement.
About
Displays information regarding program version and contact address.
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 12
3.3.1 File Menu Functions
File:
New
Deletion of the data associated with a calibration procedure in the TEST operating mode
All of the steps of any previously created or loaded test calibration procedure are
deleted after confirming the sec
urity prompt.
File:
Open
Open a calibration procedure file with/without results in the TEST operating mode.
A previously saved calibration procedure file can be selected and loaded to
temporary memory after clicking Open in the File menu while in the TEST
operating mode.
The Open dialog box which then appears prompts you to select or enter the
filename
of the file to be opened. As a default function, the last saved or opened
file is suggested.
Figure
3.3.1-a: Dialog Box for Opening a Calibration Procedure File
Calibrating procedure filenames end with
.KLT (procedure with results) or .KLF
(procedure template without results). Program files which can be loaded to the
METRAHit 18C calibrator
have a filename ending with .KLP.
Files located on local drives, or on network drives
to which a connection has been
established, can be opened. Select the directory path to the desired file in the
Look in dialog box.
By clicking a file in the file list, it’s selected as the measurement data file to be
opened. If a comment text was entered when the selected file was saved, the text
appears in the Comment field. Click [Open] in order to open the selected file.
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 13
Additional options can be selected via a further dialog box:
Test Header Only loads only the teat report header parameters of the selected
file for a previously opened/created calibration procedure.
Test Procedure loads the calibrating procedure only without results.
Test Procedure With Results loads the calibration procedure along with the
included calibration results.
i
Note 1
When a file is opened which contains data from a calibrator other the one currently selected
under
Device Type in the Setup menu, the program switches to the current device type
after acknowledgement of the prompt which appears in this case. If this program variant is not
available, i.e. if the required device
-specific program module has not been installed or
enabled, the following message appears
Device is not installed - <device type designation>.
i
Note 2
It
s not possible to open more than one calibration procedure file at a time.
i
Note 3
Further Information regarding elements in this dialog box can be accessed via the direct help
function: First click
? in the title bar of the window, and then click the element for which help
is required. A popup window with an appropriate explanation appears. This explanation can
also be display
ed by clicking the element with the right mouse button and then clicking Direct
help
in the menu which then appears.
File:
Save
After clicking
Save in the file menu, temporary procedure data are permanently
saved to a data storage medium so that they
re still available later on.
Figure
3.3.1-b: Dialog Box for Saving Procedure Files
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 14
You’re prompted to enter a filename in the Save as ... dialog box which then
appears. The KLF file type is used for procedure files without results, and the
KLT file type for procedure files with results.
A text with up to 130 characters including 2 lines can be entered to the
Comment field as a description of the data included in the file. This comment
appears in the Open dialog box before a measurement data file is opened after
the file has been selected from the list.
Files can be saved to local drives, or to network drives to which a connection has
been established. Select the desired directory path to which the file will be saved
in the Save dialog box.
The file is saved to the selected directory using the desired filename by clicking
the [OK] button.
Saving the file can be aborted by clicking the [Cancel] button or by pressing the
ESCkey.
i
Note 1
If the name of an already existing file is entered, a security prompt appears which asks t
he
user if he want
s to overwrite the respective file.
i
Note 2
Further Information regarding elements in this dialog box can be accessed via the direct help
function: First click
? in the title bar of the window, and then click the element for which help
is required. A popup window with an appropriate explanation appears. This explanation can
also be display
ed by clicking the element with the right mouse button and then clicking Direct
help
in the context menu which then appears.
File:
Print test report
Print out a standard calibration report with a predefined structure.
The Test Protocol Header dialog box appears first with numerous entry fields:
Figure
3.3.1-c: The Test Protocol Header Dialog Box
Enter information regarding the device to be calibrated (DUT) and the calibration
conditions to the entry fields in the left-hand column.
Information regarding the utilized calibration equipment can be entered to the
fields in the right-hand column.
Alternatively, these data can be read in from a previously saved test procedure file
(*.KLT/*.KLF). Click the [Load] button to this end and select the file from which the
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 15
test report header data will be loaded from the Open dialog box which then appears.
The test procedure can be saved to the data storage medium with or without
results by clicking [Save] (see
Save in the File menu).
If the test procedure includes results in at least one line, the desired printer can
be selected and the standard calibration report can be printed out after clicking the
[Print]
button. This report is laid out as shown below:
Figure 3.3.1-d: Example of a Standard Calibration Report
Alternatively, a calibration report with individualized layout and content can be
created with the help of the
Export Test Data menu function.
File:
Printer settings
Select and configure an installed printer for report printing.
This command opens the standard Windows dialog box for printer configuration, in
which the printer to be used for printing out the calibration re
port can be selected
and configured.
File:
Export test data
Export the calibration data to an individually laid out Word/Excel report template
.
If the standard calibration report doesn
t contain the desired calibration parameters
or if you want it to in
clude, for example, a company logo, the calibration and report
header data can be exported to an individually laid out Word or Excel report
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 16
template. Corresponding report templates with integrated macros can be found in
the program directory to this end.
Editing the Microsoft
®
Word report template included with the program
Start Microsoft
®
Word
and open the Word template with macro (KLT2DOC.DOT)
located in the program directory.
The included macro (AutoNew.MAIN) can import the following parameters from the
open test procedure with results
based on the variable names:
Report Header Parameter
s
Variable Name
Parameter
#lab
Issuing laboratory
#calby
Calibrated by
#caldate
Calibration date
#nextdate
Next calibration date (= calibration date + 12 months)
#uutmake
Manufacturer of the device under test
#uuttype
Type designation of the device under test
#uutnumb
Serial number of the device under test
#uutdsc
Object (type of device under test)
#envtemp
Ambient temperature
#envhmdt
Relative humidity
#meatype
Type of measurement
#certnum
Certificate number
#caldev
Type designation of the calibrator
#calno
Serial number of the calibrator
#calcert
Calibrator certificate number
#muldev
Type designation of the multimeter
#mulno
Serial number of the multimeter
#mulcert
Multimeter certificate number
#refdev
Type designation of the reference multimeter
#refno
Serial number of the reference multimeter
#refcert
Reference multimeter certificate number
Procedure Step Parameter
s
Variable Name
Parameter
#fun
(Measuring) function (V DC, RTD...)
#mrange
Measuring range (3.0000 V DC”)
#tpoint
Test point
#outval
Output value and unit of measure (2.50000 V DC, 100 °C Pt100” …)
#param
Additional parameter (1.00 kHzfor V AC, Pt100for temperature)
#adig
Permissible deviation in digits
#expval
Expected value
#fexpval
Expected value with measuring function (ACDC)
#minval
Lower limit value
#fminval
Lower limit value with measuring function
#maxval Upper limit value
#fmaxval Upper limit value with measuring function
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 17
#regval Measured value
#fregval Measured value with measuring function
#rdev Measured difference (absolute value)
#frdev Measured difference with measuring function
#rdig Measured difference in digits
#rprc
Measured difference as percentage of the specification
#result
Assessment: pass/limit/fail/---
#text
Comment regarding the test step
#udig
Measuring uncertainty in digits
#uval
Measuring uncertainty (absolute value)
#urat
Measuring uncertainty ratio
General Parameters
Variable Name
Parameter
#passprc
Pass percentage for limit value
#measured
Number of measured steps
#passed
Number of passed steps
#limit
Number of limit steps (between limit value and failure)
#fail
Number of failed steps
Adapt structure and column content
in accordance with your needs and save the
edited file as a Word 97 or 2003 template (*.DOT), or as Word template with
macros (*.DOTM) with a new name.
Running the Word macro and transferring the calibration data t
o the Word report template
Start Microsoft
®
Word and open the desired Word template with macro.
Start the included AutoNew.MAIN macro by clicking Macros and then View
Macros in the View menu.
Respond to the prompt asking whether or not you want to receive data from the
calibration program with [Yes]:
In the window which then appears, youre prompted to open a test procedure
file. If this file has already been opened, click [Cancel].
Now youre prompted to start calibration (if you havent already done so) and
then to select the export test data menu command.
The data are transferred to the Word file and completion is acknowledged with a
ready message.
Calibration data can be transferred to Excel in a similar way. However, the macro and
the report templates are located in separate files in the p
rogram directory in this case.
Start Microsoft
®
Excel and open the KLT2DOC.XLM Excel macro file.
Two icons now appear underneath the add-ins with the designations KLT:Get
calibrator data and KLT:Close.
Click the KLT:Get calibrator data icon and select the report template file
(KLT2XLSx.XLT) to be opened in the window which then appears.
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 18
3.3.2 Help Menu
Program Functions
Opens a PDF file with a description of general program functions.
This menu command starts Adobe
®
Readerin a new window, and opens a PDF
file
which includes descriptions of general program functions.
Device Functions
Opens a PDF file with a description of device-specific program functions.
This menu command starts Adobe
®
Readerin a new window and opens a PDF
document which includes descriptio
ns of specific program functions and
characteristics with reference to the measuring instrument which is currently
selected under
Device Type in the Setup menu.
Online Support
Opens a website with information regarding technical support via the Internet.
This menu command starts your Internet browser and opens a website with
information on how our technical support team can provide you with support via the
Internet.
Display License
A
greement
Opens a TXT file with a description of the license agreement.
Th
is menu command starts your text editor and displays the license agreement in a
new window.
Info
Displays information regarding program version and contact address.
This menu command opens a window which displays information regarding the
revision level o
f the utilized program version, as well as the contact address for
technical support.
3.4 Basic Program Settings
3.4.1 Changing the User Interface Language
When the program is started for the first time, the language which was selected
during installation is use
d for the user interface. This language can be changed by
selecting
English / Deutsch / ... in the Setup menu. The program always starts up
with the last selected language.
3.4.2 Selecting the Type of Calibrator to be Controlled
Select the type of calibrator t
o be controlled under Device Type in the Setup menu.
The various calibrators which are available for selection depend on which version of
the program you purchased and which device(s) your chose during installation.
An image of the selected calibrator appe
ars as an initial window in the workspace
after
the device type has been changed. Depending on the device type, a PDF file
with specific connection and operating instructions can be opened by clicking Device
Functions
in the Help menu.
METRAHit 18C METRAHit 28C/28C light METRAHit Cal/Multical
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 19
METRAtop 90C/D9000/D9100 FLUKE 5100A/5500A/5520A/5700A
3.4.3 Port Settings
The corresponding dialog box appears after clicking
Port Setup in the Setup menu.
Its content depends on which
type of calibrator is currently selected to be controlled.
Device
-specific port settings are described in a separate document. This document
can be accessed after clicking
Device Functions in the Help menu.
a) For METRAtop 90C / FLUKE D9000, D9100, 5100A, 5500A, 5520A and 5700A multifunction calibrators
Figure
3.4.3-a: Port Settings for FLUKE 5XXXA with METRAHit 1XS and SI232 Adapter
b) For METRAHit CAL and METRACAL MC process calibrators
Figure
3.4.3-b: Port Settings for METRAHIT CAL and USB X-TRA with METRAHit X-TRA and
USB X
-TRA
METRAwin 90 - Operating Instructions
GOSSEN METRAWATT 20
c) For METRAHit 28C and 28C light process calibrators
Figure
3.4.3-c: Port Settings for METRAHIT 28C and BD232 with METRAHit 28S and BD232
d) For METRAHit 18C process calibrator
Figure
3.4.3-d: Port Settings for METRAHit 18C and BD232 with METRAHit 18S SI232
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26

Gossen MetraWatt METRAwin 90-F/J Operating instructions

Type
Operating instructions
This manual is also suitable for

The Gossen MetraWatt METRAwin 90-F/J is a Windows program that helps you control various calibrators for electrical quantities with the help of a PC, and for documenting calibration results.