  • Hello! I am an AI chatbot trained to assist you with the Roche cobas s 201 system User manual. I’ve already reviewed the document and can help you find the information you need or explain it in simple terms. Just ask your questions, and providing more details will help me assist you more effectively!
cobas s 201 System
Validation Guide
for use with Small Pools
(Pools of 1 and 6)
Configuration C MR1
05922151190-02
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
2
This page intentionally left blank.
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
3
Table of Contents
Preface.......................................................................................................................................................... 7
Revision History .......................................................................................................................................................................... 7
Purpose of this Guide............................................................................................................................................................... 9
How to Use this Guide ...........................................................................................................................................................10
Conventions Used in this Guide .........................................................................................................................................11
References..................................................................................................................................................................................12
Acronyms....................................................................................................................................................................................13
Glossary.......................................................................................................................................................................................14
Section 1: Hamilton Microlab STAR IVD / STARlet IVD Pipettor and Pooling Management
Workstation Validation............................................................................................................................ 15
Traceability Matrix ................................................................................................................................... 16
Traceability Chart.....................................................................................................................................................................16
Installation Qualification (IQ) ................................................................................................................ 17
Overview......................................................................................................................................................................................17
Test Case HIQ-01: Pooling Manager Hardware and Software...............................................................................18
Test Case HIQ-02: Location Verification.........................................................................................................................20
Test Case HIQ-03: Environment Verification.................................................................................................................22
Test Case HIQ-04: Verification of Connections............................................................................................................24
Test Case HIQ-05: Safety and Training Verification...................................................................................................26
Test Case HIQ-06: Consumables ......................................................................................................................................28
Test Case HIQ-07: Documentation Verification...........................................................................................................29
Test Case HIQ-08: Configuration Settings.....................................................................................................................30
Test Case HIQ-09: Verification of Application Security.............................................................................................32
Operational Qualification (OQ).............................................................................................................. 35
Overview......................................................................................................................................................................................35
Dependencies............................................................................................................................................................................37
Test Case HOQ-01: Primary Pools of 6.............................................................................................................................38
Test Case HOQ-02: Primary Pools of 1.............................................................................................................................47
Test Case HOQ-03: Manual Barcode Entry for Donor Tubes...................................................................................57
Test Case HOQ-04: Simultaneous Pools of 6.................................................................................................................62
Test Case HOQ-05: Manual Barcode Entry for a Library Plate................................................................................72
Test Case HOQ-06: Correcting Loading Errors .............................................................................................................76
Test Case HOQ-07: Recovery from Control Pipetting Errors....................................................................................82
Test Case HOQ-08: Recovery from Sample Pipetting Errors (Library Plates Enabled)...................................87
Test Case HOQ-09: Recovery from Sample Pipetting Errors (Library Plates Disabled)..................................94
Test Case HOQ-10: Rejecting Pools and Issuing Change Requests .................................................................. 100
Test Case HOQ-11: Rejecting Library Plate Wells..................................................................................................... 106
Test Case HOQ-12: Resolution Pooling (Library Plates Enabled)........................................................................ 111
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
4
Test Case HOQ-13: Resolution Pooling (Library Plates Disabled) ...................................................................... 118
Test Case HOQ-14: Repeat Pooling (Library Plates Enabled)............................................................................... 124
Test Case HOQ-15: Repeat Pooling (Library Plates Disabled) ............................................................................. 131
Test Case HOQ-16: Correcting S-tube Barcode Reading Errors ......................................................................... 137
Test Case HOQ-17: Aborting a Pooling Run ............................................................................................................... 146
Performance Qualification (PQ)..........................................................................................................151
Overview................................................................................................................................................................................... 151
Section 2: COBAS
®
AmpliPrep Instrument, Optional Docking Station, and AMPLILINK
Software Validation................................................................................................................................155
Traceability Matrix .................................................................................................................................156
Installation Qualification (IQ) ..............................................................................................................157
Overview................................................................................................................................................................................... 157
Materials................................................................................................................................................................................... 157
General Prerequisites .......................................................................................................................................................... 157
Dependencies......................................................................................................................................................................... 157
Test Case APIQ-01: Work Area Verification ................................................................................................................ 158
Test Case APIQ-02: Reagent Storage Area Verification.......................................................................................... 160
Test Case APIQ-03: Hardware and Software Verification ...................................................................................... 162
Test Case APIQ-04: Location Verification................................................................................................................... 165
Test Case APIQ-05: Environment Verification ........................................................................................................... 167
Test Case APIQ-06: Connections and Accessories................................................................................................. 169
Test Case APIQ-07: Safety and Training ..................................................................................................................... 171
Test Case APIQ-08: Documentation ............................................................................................................................. 173
Test Case APIQ-09: Consumables................................................................................................................................. 174
Test Case APIQ-10: Configuration Settings ............................................................................................................... 175
Test Case APIQ-11: Security............................................................................................................................................ 178
Operational Qualification (OQ)............................................................................................................183
Overview................................................................................................................................................................................... 183
Materials................................................................................................................................................................................... 183
Documentation ...................................................................................................................................................................... 183
General Prerequisites .......................................................................................................................................................... 184
Dependencies......................................................................................................................................................................... 185
Test Case APOQ-01: Loading Reagents....................................................................................................................... 186
Test Case APOQ-02: Loading Consumables............................................................................................................... 189
Test Case APOQ-03: Routine Sample Preparation Run .......................................................................................... 193
Test Case APOQ-04: Handling Loading Errors........................................................................................................... 202
Test Case APOQ-05: Handling Reagent Barcode Reading Errors....................................................................... 213
Test Case APOQ-06: Error Handling for Switched Controls.................................................................................. 216
Test Case APOQ-07: Error Handling for Reagents on Incorrect Rack............................................................... 222
Test Case APOQ-08: Error Handling for Empty Reagent Cassettes.................................................................... 231
Test Case APOQ-09: Error Handling for Expired Reagents ................................................................................... 234
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
5
Test Case APOQ-10: Examining System and Rack Status Indicators ................................................................ 237
Test Case APOQ-11: Status Indicators, Logs, and Problem Reports.................................................................. 239
Performance Qualification (PQ)..........................................................................................................243
Overview................................................................................................................................................................................... 243
Section 3: COBAS
®
TaqMan
®
Analyzer and AMPLILINK Software Validation..........................245
Traceability Matrix .................................................................................................................................246
Installation Qualification (IQ) ..............................................................................................................247
Overview................................................................................................................................................................................... 247
Materials................................................................................................................................................................................... 247
General Prerequisites .......................................................................................................................................................... 247
Dependencies......................................................................................................................................................................... 247
Test Case TMIQ-01: Work Area Verification ............................................................................................................... 248
Test Case TMIQ-02: Hardware and Software Verification ..................................................................................... 250
Test Case TMIQ-03: Location Verification.................................................................................................................. 253
Test Case TMIQ-04: Environment Verification .......................................................................................................... 255
Test Case TMIQ-05: Connections and Accessories ................................................................................................ 257
Test Case TMIQ-06: Safety and Training..................................................................................................................... 259
Test Case TMIQ-07: Documentation ............................................................................................................................ 261
Test Case TMIQ-08: Consumables................................................................................................................................ 262
Test Case TMIQ-09: Configuration Settings............................................................................................................... 263
Test Case TMIQ-10: Security........................................................................................................................................... 266
Operational Qualification (OQ)............................................................................................................271
Overview................................................................................................................................................................................... 271
Materials................................................................................................................................................................................... 271
Documentation ...................................................................................................................................................................... 271
General Prerequisites .......................................................................................................................................................... 272
Dependencies......................................................................................................................................................................... 273
Test Case TMOQ-01: Analyzing Primary Pools of 6.................................................................................................. 274
Test Case TMOQ-02: Accepting Results on AMPLILINK Software .................................................................... 278
Test Case TMOQ-03: Blocked Samples........................................................................................................................ 282
Test Case TMOQ-04: Analyzing Primary Pools of 1.................................................................................................. 287
Test Case TMOQ-05: Analyzing Resolution Pools..................................................................................................... 292
Test Case TMOQ-06: Analyzing Additional Batches................................................................................................ 296
Test Case TMOQ-07: Status Indicators, Logs, and Problem Reports.................................................................300
Performance Qualification (PQ)..........................................................................................................305
Overview................................................................................................................................................................................... 305
Section 4: Roche PDM Data Manager Software Validation..........................................................307
Traceability Matrix .................................................................................................................................308
Installation Qualification (IQ) ..............................................................................................................309
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
6
Overview................................................................................................................................................................................... 309
Test Case PIQ-01: Data Management Workstation and Software.....................................................................310
Test Case PIQ-02: Data Manager Communication.................................................................................................. 312
Test Case PIQ-03: PDM Server Hardware .................................................................................................................. 314
Test Case PIQ-04: Safety and Training ........................................................................................................................ 316
Test Case PIQ-05: Documentation ................................................................................................................................ 318
Test Case PIQ-06: Configuration Security................................................................................................................... 319
Operational Qualification (OQ)............................................................................................................325
Overview................................................................................................................................................................................... 325
Test Case POQ-01: Reviewing Results from Primary Pools of 6 .......................................................................... 328
Test Case POQ-02: Reviewing Results from Resolution Pools ............................................................................. 336
Test Case POQ-03: Reviewing Results from Primary Pools of 1 .......................................................................... 341
Test Case POQ-04: Reviewing Invalid Results............................................................................................................ 346
Test Case POQ-05: Defining UDECs.............................................................................................................................. 350
Performance Qualification (PQ)..........................................................................................................355
Overview................................................................................................................................................................................... 355
Appendix: The Validation Process ......................................................................................................359
Introduction............................................................................................................................................................................. 361
Defining User Requirements............................................................................................................................................. 362
Selecting the System............................................................................................................................................................ 362
Analyzing Risk........................................................................................................................................................................ 363
Preparing a Validation Plan............................................................................................................................................... 364
Executing the Validation Plan........................................................................................................................................... 365
Specifying Corrective Action............................................................................................................................................. 366
Using a Waiver Report ........................................................................................................................................................ 366
Preparing the Validation Report ...................................................................................................................................... 366
Forms......................................................................................................................................................................................... 367
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
7
Preface
Revision History
Manual Version Date Action
1.0 August 2009 Original document.
2.0 October 2009 Addition of Configuration to Title Page
Roche Diagnostics Ltd.
CH-6343 Rotkreuz
Switzerland
Distributed by
Roche Diagnostics
Indianapolis, IN 46256 USA
(For Technical Assistance call the
Roche Response Center
toll-free 1-800 526 1247)
Roche Diagnostics
H7V 4A2 Laval, Quebec
(For Technical Assistance call:
Pour toute assistance technique,
appeler le: 1-877 273 3433)
Roche Diagnostics (Schweiz) AG
CH-6343 Rotkreuz
Roche Diagnostics
F-38240 Meylan
Roche Diagnostics GmbH
D-68298 Mannheim, Germany
Distributore in Italia:
Roche Diagnostics SpA
Piazza Durante 11
I-20131 Milano
Roche Diagnostics S.L.
E-08006 Barcelona
Distribuidor em Portugal:
Roche Farmacêutica Química, Lda
P-2700 Amadora
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
8
Questions or comments regarding the contents of this guide can be directed to the
address below or to your Roche representative.
Roche Molecular Systems, Inc.
4300 Hacienda Drive
Pleasanton, California 94588-2722
ROCHE, AMPERASE, AMPLILINK, COBAS, COBAS S, AMPLIPREP, TAQSCREEN,
and TAQMAN are trademarks of a member of the Roche Group.
ROCHE RESPONSE CENTER is a servicemark of a member of the Roche Group.
HP and StorageWorks are trademarks of Hewlett-Packard Company in the United States
and/or other countries.
Microlab is a registered trademark of Hamilton Company.
Microsoft and Windows are either registered trademarks or trademarks of Microsoft
Corporation in the United States and/or other countries.
Copyright © 2009. Roche Molecular Systems, Inc. All rights reserved.
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
9
Purpose of this Guide
The cobas s 201 system includes hardware, software, and reagents. This Validation Guide
assists users in establishing Quality Assurance (QA) procedures with regard to all the
cobas s 201 system components that are used when testing blood products.
The purpose of this Validation Guide is to assist users of the cobas s 201 system in creating
a site-specific Validation Plan to qualify their system according to current guidelines. This
Validation Guide should be used only as a guideline, and users may elect to use, modify,
or not use each section as they deem appropriate.
Refer to the cobas s 201 system Operator’s Manuals for a complete description of the
cobas s 201 system. Refer to the assay-specific package inserts for instructions on
collection, storage, and handling of samples.
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
10
How to Use this Guide
Preface
Lists the reference materials that were used to
prepare this Validation Guide and the terms and
acronyms that are used.
Hamilton Microlab
STAR IVD / STARlet
IVD Pipettor and
Pooling Management
Workstation
Provides sample test cases that can be used to
assist in preparing a Hamilton Microlab STAR
IVD / STARlet IVD Pipettor, Roche PDM Pooling
Wizard software, and Roche PDM Pooling
Manager software Validation Plan for your
facility. The user can select those parts of this
section that are applicable for the validation and
modify them as needed.
COBAS
®
AmpliPrep
Instrument, Optional
Docking Station, and
AMPLILINK Software
Provides sample test cases that can be used to
assist in preparing a COBAS
®
AmpliPrep
Instrument and AMPLILINK software Validation
Plan for your facility. The user can select those
parts of this section that are applicable for the
validation and modify them as needed.
COBAS
®
TaqMan
®
Analyzer and
AMPLILINK Software
Provides sample test cases that can be used to
assist in preparing a COBAS
®
TaqMan
®
Analyzer
and AMPLILINK software Validation Plan for
your facility. The user can select those parts of
this section that are applicable for the validation
and modify them as needed.
Roche PDM Data
Manager Software
Provides sample test cases that can be used to
assist in preparing a Roche PDM Data Manager
and Administrator Software Validation Plan for
your facility. The user can select those parts of
this section that are applicable for the validation
and modify them as needed.
Appendix
Provides an overview of the validation process
and describes how the validation plan is
developed. Templates for commonly used
validation forms are also provided.
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
11
Conventions Used in this Guide
Italics
Indicates a reference to another section of this guide or an
external document. For example:
Refer to the cobas s 201 system Hardware and Software
Reference Manual for a complete description of the cobas s
201 system.
Italics are also used for field, screen, and dialog box titles. For
example:
The Begin Logon dialog box is displayed.
Bold
Emphasizes a term or component that is being described or
indicates a keyboard entry or mouse selection. For example:
Press F1 to display online Help.
Select Export from the File menu.
“Quotes” Designates keyboard input that should be typed exactly as
shown. For example:
Enter “GUEST” then click OK.
1 First Item
2 Second Item
Used to number procedures that should be performed
sequentially and to uniquely identify system requirements.
[Brackets] Indicates a cross-reference to an external document. Outside
source material used in the preparation of the Validation Guide
is listed in the References section. A bracketed category name
and sequential number precede each entry. For example the
first referenced FDA document is listed:
[FDA-1] CFR Title 21
> Separates command options in a multi-level command. For
example:
Select Control Panel > Help > About Windows
+ Indicates more than one key should be pressed
simultaneously. For example:
Press the Ctrl+Alt+Del keys.
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
12
References
(as of the date of this publication)
The following documents were used as sources of information for this Validation Guide:
[AABB-1] Facts About Blood And Blood Banking, AABB Web site:
www.aabb.org/All_About_Blood/FAQs/aabb_faqs.htm
[FDA-1] U.S. Government Code of Federal Regulations (CFR) Title 21 - Food and
Drugs.
[FDA-2] General Principles of Software Validation; Final Guidance for Industry
and FDA Staff (January 11, 2002).
[FDA-3] Glossary of Computerized System and Software Development
Terminology, 1995.
[FDA-4] Guideline on General Principles of Process Validation, May, 1987.
[FDA-5] Docket No. 91N-0450, Guideline for Quality Assurance in Blood
Establishments, July 11, 1995.
[FDA-6] CBER publication "Commemorating 100 Years of Biologics Regulation”,
March 2002.
[FDA-7] CBER publication "Guidance for Industry Q7A Good Manufacturing
Practice Guidance for Active Pharmaceutical Ingredients”, August 2001.
[FDA-8] Medical Device Quality System Manual, HHS Publication FDA 97-4179,
December 1996. Available online at www.fda.gov/cdrh/qsr/
[FDA-9] Development of West Nile virus Testing and Donor screening as a Model
for Screening Bioterrorist Agents, Hira Nakhasi, Ph.D., Director,
DETTD/OBRR, CBER, FDA
[FMEA-1] An Introduction to Risk/Hazard Analysis for Medical Devices; Daniel
Kamm, P.E., C.Q.A.
[GAMP-1] GAMP Good Practice Guide: Validation of Process Control Systems
(VPCS), October 2003.
[GOV-1] GPO Access Web site: http://www.gpoaccess.gov/cfr/index.html. A service
of the U.S. Government Printing Office.
[CEFIC-1] CEFIC Task Force – Computer Validation Guide, ver. 2, December, 2002.
[IEEE-1] IEEE-STD-829-1998; IEEE Standard for Software Test Documentation,
Institute of Electrical and Electronics Engineers, New York.
[ISBT-1] Guidelines for validation and maintaining the validation state of
automated systems in blood banking, July 2003.
[RMS-1]
Roche Diagnostics Web site: www.roche-diagnostics.com/quality.html (2006).
[RMS-2] cobas s 201 system Operator’s Manual for Use with Small Pools
[RMS-3] cobas s 201 system Hardware and Software Reference Manual
[RMS-4] cobas s 201 system PDM Software Administrator’s Guide
[RMS-5] cobas s 201 system Facilities Guide
[RMS-6] cobas s 201 system Operator’s Manual Supplement for Use of the
Docking Station
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
13
Acronyms
The following acronyms are used in this guide:
CBER Center for Biologics Evaluation and Research
CFR Code of Federal Regulations
DNA Deoxyribonucleic Acid
DR Discrepancy Report
FDA U.S. Food and Drug Administration
FMEA Failure Mode Effects Analysis
GAMP Good Automated Manufacturing Practices
HBV Hepatitis B Virus
HCV Hepatitis C Virus
HIV Human Immunodeficiency Virus
IEEE Institute of Electrical and Electronics Engineers
IQ Installation Qualification
ISBT International Society for Blood Transfusion
LAN Local Area Network
LIS Laboratory Information System
MB Megabyte
MHz Megahertz
MPX Multiplexed set of targets: HCV, HBV, HIV-1 (Group M and O), HIV-2
OQ Operational Qualification
PCR Polymerase Chain Reaction
PQ Performance Qualification
QA Quality Assurance
RMEC Roche-Manufactured External Control
RMS Roche Molecular Systems, Inc.
RNA Ribonucleic Acid
TCx Thermal Cycler A, B, C, or D
UDEC User-Defined External Control
URS User Requirements Specification
V Volts
VAC Volts Alternating Current
WNV West Nile Virus
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
14
Glossary
Code of Federal
Regulations
The Code of Federal Regulations (CFR) is the codification of the general
and permanent rules published in the Federal Register by the executive
departments and agencies of the Federal Government. It is divided into 50
titles that represent broad areas subject to Federal regulation.
1
CFR Title 21, FDA, is referenced in this guide.
Concurrent
Validation
Testing that thoroughly monitors the process, allowing tested product to
be released for use. This method can be used when a previously validated
process is modified.
Installation
Qualification
Documented verification that the system, including all software and
hardware components, has been installed correctly.
Operational
Qualification
Documented verification that the system meets all its requirements under
all anticipated operating conditions including boundary and extreme test
cases.
Performance
Qualification
Documented verification that the system produces acceptable results
under normal operating conditions in the customer environment.
Prospective
Validation
Testing that is performed before the product is used.
Quality
Assurance
The actions, planned and performed, to provide confidence that all
systems and components that influence the quality of the product are
working as expected individually and collectively.
2
Quality Control The operational techniques and procedures used to achieve quality
requirements.
2
Retrospective
Validation
Testing that is performed on representative batches of the product after
the product is released. This method is reserved for a well-established
process that has undergone a change that does not affect product quality.
Validation Establishing documented evidence that a process will consistently
perform effectively according to its intended use.
Validation Plan The written and approved document of an experimental sequence of tests,
that when executed as prescribed, are intended to produce documented
evidence that the equipment or system conforms to user needs and
intended uses, and that the particular requirements implemented can be
consistently fulfilled.
For additional information, refer to the Medical Device Quality Systems Manual.
3
1
[GOV-1]
2
[FDA-3]
3
[FDA-8]
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
15
Section 1: Hamilton Microlab
STAR IVD / STARlet IVD Pipettor and Pooling
Management Workstation Validation
This section can be used to assist in preparing a Hamilton Microlab STAR IVD / STARlet IVD Pipettor, PDM Pooling Wizard
software, and PDM Pooling Manager software Validation Plan for your facility.
The contents of this section and the test cases are intended only to be used as a guide. The user can select those parts of this
section that are applicable for the validation and modify them as needed.
Hamilton Microlab STAR IVD / STARlet IVD Pipettor and Roche PDM Pooling Manager Software Validation
Traceability Matrix Traceability Chart
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
16
Traceability Matrix
Requirements for the Hamilton Microlab STAR IVD / STARlet IVD Pipettor, PDM Pooling Wizard software, and PDM Pooling
Manager software and the test case(s) that validate each requirement are summarized below.
Traceability Chart
Requirements Test Case(s)
Hardware and Software HIQ-01
Location HIQ-02
Environment HIQ-03
Connections and Accessories HIQ-04
Safety and Training HIQ-05
Consumables HIQ-06
Documentation HIQ-07
Configuration Settings HIQ-08
Security HIQ-09
Operation HOQ-01 through HOQ-17
Validity Checking HOQ-06 through HOQ-09, HOQ-14, HQO-15
Results Analysis See POQ-01 through POQ-04
Data Management and Reporting See POQ-01 through POQ-04
Audit Log See POQ-01 through POQ-04
Alarms HOQ-03 through HOQ-09
Throughput and Capacity HOQ-01, HOQ-02, HOQ-04
Hamilton Microlab STAR IVD / STARlet IVD Pipettor and Roche PDM Pooling Manager Software Validation
Installation Qualification (IQ) Overview
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
17
Installation Qualification (IQ)
Overview
The Installation Qualification (IQ) verifies that the Hamilton Microlab STAR IVD / STARlet IVD Pipettor and Pooling Management
Workstation software have been installed correctly.
Materials
No special materials are required to perform the IQ.
General Prerequisites
Each IQ test case lists prerequisites required to begin the test case. Additionally, all IQ testing assumes that:
1 The vendor properly installed the system.
2 The operator is familiar with general computer operation and the Windows operating system.
Dependencies
All test cases in the IQ section are designed so they can be run independently.
Hamilton Microlab STAR IVD / STARlet IVD Pipettor and Roche PDM Pooling Manager Software Validation
Installation Qualification (IQ) Test Case HIQ-01: Pooling Manager Hardware and Software
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
18
Test Case HIQ-01: Pooling Manager Hardware and Software
Purpose
This test case verifies that the correct hardware and software is installed for the Pooling Management Workstation.
Prerequisites
This test case assumes the operator:
Has a login ID and password that allow access to the PDM Pooling Wizard and PDM Pooling Manager.
Record the login ID: _________________________________
Acceptance Criteria
This test case is complete and passes when the Expected Result has been obtained and entered in the Observed Result box, Pass is
circled in the Pass/Fail column for each step, and (if required by your facility) each completed step is initialed and dated.
Procedure
Step Action Expected Result Observed Result Pass/Fail By/Date
1
Power on the Pooling Management
Workstation if it is powered off.
A prompt is displayed to press the
CTRL+ALT+DEL keys simultaneously to
logon.
Pass Fail
________
________
2 Press CTRL+ALT+DEL.
The Welcome to Roche PDM dialog box is
displayed.
Pass Fail
________
________
3
Enter the login ID and password
recorded in the Prerequisites
section and then click OK.
The PDM Pooling Manager is started and
the Query tab is displayed. The software
version is displayed in the title bar.
PDM Pooling Manager software version:
__________________________________
Pass Fail
________
________
Hamilton Microlab STAR IVD / STARlet IVD Pipettor and Roche PDM Pooling Manager Software Validation
Installation Qualification (IQ) Test Case HIQ-01: Pooling Manager Hardware and Software
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
19
Step Action Expected Result Observed Result Pass/Fail By/Date
4 Click the Wizard button located on
right side of the Query tab.
Record the software version for
PDM Pooling Wizard software.
The PDM Pooling Wizard main screen is
displayed. The software version is
displayed in the title bar.
PDM Pooling Wizard software version:
__________________________________
Pass Fail
________
________
5 Click Exit.
The PDM Pooling Wizard application
closes. The PDM Pooling Manager Query
tab is displayed.
Pass Fail
________
________
6 Click the Logoff button located on
right side of the Query tab.
The logoff is successful and a prompt is
displayed to press CTRL, ALT, DEL to
logon.
Pass Fail
________
________
Results
Acceptable ________ Unacceptable _________
Comments ________________________________________________________________________________________________________________
__________________________________________________________________________________________________________________________
__________________________________________________________________________________________________________________________
Performed By _______________________________________ Date _________________
Verified By _______________________________________ Date _________________
Hamilton Microlab STAR IVD / STARlet IVD Pipettor and Roche PDM Pooling Manager Software Validation
Installation Qualification (IQ) Test Case HIQ-02: Location Verification
cobas s system 201 Validation Guide for Small Pools 10/2009, Revision 2.0
20
Test Case HIQ-02: Location Verification
Purpose
This test case verifies proper positioning of the Hamilton Microlab STAR IVD / STARlet IVD Pipettor.
Prerequisites
None.
Acceptance Criteria
This test case is complete and passes when the Expected Result has been obtained and entered in the Observed Result box, Pass is
circled in the Pass/Fail column for each step, and (if required by your facility) each completed step is initialed and dated.
Procedure
Action Expected Result Observed Result Pass/Fail By/Date
1
Examine the front and sides of the
Hamilton Microlab STAR IVD /
STARlet IVD Pipettor.
Personnel have access at the front and
sides of the instrument, for operation,
maintenance, and opening of the
protective covers.
Pass Fail
________
________
2
Examine the rear of the Hamilton
Microlab STAR IVD / STARlet IVD
Pipettor.
There is sufficient space for ventilation. Pass Fail
________
________
3
Examine the Hamilton Microlab
STAR IVD / STARlet IVD Pipettor
location.
The Hamilton Microlab STAR IVD /
STARlet IVD Pipettor is positioned away
from:
heating ducts
air conditioning vents
direct sunlight
intense artificial light
laboratory equipment that emits
electromagnetic RF fields (e.g., X-ray
machine, elevator)
Pass Fail
________
________
Note: There are no special location requirements for the Pooling Management Workstation. The USB cable length limits its proximity to the instrument.
/