Rosemount 185 Colormetric Valve-Pump based Monitor Owner's manual

Type
Owner's manual
Model CFA-3000
Colorimetric Valve-Pump-based Monitor
Instruction Manual
LIQ_MAN_CFA3000/rev. C
September 2014
ESSENTIAL INSTRUCTIONS
READ THIS P
AGE BEFORE PROCEEDING!
Your purchase from Rosemount Analytical, Inc. has
resulted in one of the finest instruments available for
your particular application. These instruments have
been designed, and tested to meet many national
and international standards. Experience indicates
that its performance is directly related to the quality
of the installation and knowledge of the user in oper-
ating and maintaining the instrument. To ensure their
continued operation to the design specifications, per-
sonnel should read this manual thoroughly before
proceeding with installation, commissioning, opera-
tion, and maintenance of this instrument. If this
equipment is used in a manner not specified by the
manufacturer, the protection provided by it against
hazards may be impaired.
Failure to follow the proper instructions may cause
any one of the following situations to occur: Loss
of life; personal injury; property damage; damage
to this instrument; and warranty invalidation.
Ensure that you have received the correct model
and options from your purchase order. Verify that
this manual covers your model and options. If not,
call 1-800-854-8257 or 949-757-8500 to request
correct manual.
For clarification of instructions, contact your
Rosemount representative.
Follow all warnings, cautions, and instructions
marked on and supplied with the product.
Use only qualified personnel to install, operate,
update, program and maintain the product.
Educate your personnel in the proper installation,
operation, and maintenance of the product.
Install equipment as specified in the Installation
section of this manual. Follow appropriate local
and national codes. Only connect the product to
electrical and pressure sources specified in this
manual.
Use only factory documented components for
repair. Tampering or unauthorized substitution of
parts and procedures can affect the performance
and cause unsafe operation of your process.
All equipment doors must be closed and protective
covers must be in place unless qualified personnel
are performing maintenance.
If this equipment is used in a manner not specified
by the manufacturer, the protection provided by it
against hazards may be impaired.
WARNINGS
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
Installation of cable connections and servicing of this
product require access to shock hazard voltage levels.
Main power and relay contacts wired to separate power
source must be disconnected before servicing.
Do not operate or energize instrument with case open!
Signal wiring connected in this box must be rated at least
240 V.
Non-metallic cable strain reliefs do not provide grounding
between conduit connections! Use grounding type bush-
ings and jumper wires.
Unused cable conduit entries must be securely sealed by
non-flammable closures to provide enclosure integrity in
compliance with personal safety and environmental protec-
tion requirements. Unused conduit openings must be
sealed with NEMA 4X or IP65 conduit plugs to maintain
the ingress protection rating (NEMA 4X).
Electrical installation must be in accordance with the
National Electrical Code (ANSI/NFPA-70) and/or any
other applicable national or local codes.
Operate only with front and rear panels fastened and in
place over terminal area.
Safety and performance require that this instrument be
connected and properly grounded through a three-wire
power source.
Proper relay use and configuration is the responsibility of
the user.
CAUTION
This product generates, uses, and can radiate radio frequency
energy and thus can cause radio communication interference.
Improper installation, or operation, may increase such interfer-
ence. As temporarily permitted by regulation, this unit has not
been tested for compliance within the limits of Class A comput-
ing devices, pursuant to Subpart J of Part 15, of FCC Rules,
which are designed to provide reasonable protection against
such interference. Operation of this equipment in a residential
area may cause interference, in which case the user at his
own expense, will be required to take whatever measures may
be required to correct the interference.
WARNING
This product is not intended for use in the light
industrial, residential or commercial environments
per the instrument’s certification to EN50081-2.
Emerson Process Management
Rosemount Analytical Inc.
2400 Barranca Parkway
Irvine, CA 92606 USA
Tel: (949) 757-8500
Fax: (949) 474-7250
http://www.raihom.com
© Rosemount Analytical Inc. 2005
CFA-3000 Colorimetric Analyzer
Instruction Manual
2
Table of
Contents
1.0 INTRODUCTION ....................................................................................... 1
1.1 Thoroughly Read This Manual ................................................................... 2
1.2 Analyzer Benefits ....................................................................................... 2
1.3 WARNING -- Electrical Shock Hazard ....................................................... 3
1.4 CFA-3000 Series Monitors - Approvals...................................................... 3
2.0 INSTALLING THE ANALYZER.................................................................4
2.1 Unpack Parts ............................................................................................. 4
2.2 Mount Main Cabinet................................................................................... 6
2.3 Mount Reagent Cabinet ............................................................................. 6
2.4 Install Waste Tube ..................................................................................... 9
2.5 Mount Overflow Sampling Assembly ......................................................... 9
2.6 Prepare Electrical Connections................................................................ 12
2.7 Install Colorimeter Assembly.................................................................... 13
2.8 Install Valve Pump Assembly................................................................... 14
2.9 Prepare Reagent Containers ................................................................... 16
3.0 OPERATING THE ANALYZER............................................................... 17
3.1 Operational Overview .............................................................................. 17
3.2 Check Sample Stream ............................................................................. 18
3.2.1 Check Fluidics.......................................................................................... 19
3.2.2 Check Reagents ...................................................................................... 19
3.2.3 Turn On Power......................................................................................... 19
3.2.4 Test Valve Pump Operation..................................................................... 20
3.2.5 Set Up Stream Parameters...................................................................... 23
3.3 Set Reagent Supply Time ........................................................................ 25
3.4 Set Scaling Outputs ................................................................................. 25
3.4.1 Record GAIN and mV........................................................................... 26
3.4.2 Initiate Auto Calibration............................................................................ 27
3.5 Process a Grab Sample........................................................................... 28
3.6 Manually Sample a Particular Stream...................................................... 28
3.7 Shutdown................................................................................................. 29
3.8 Quick Start ............................................................................................... 29
4.0 MAINTAINING THE ANALYZER ............................................................ 30
4.1 Analyzer Equipment................................................................................. 30
4.2 Colorimeter Assembly.............................................................................. 30
5.0 TROUBLESHOOTING PROCEDURES ..................................................32
5.1 Troubleshooting Chart ............................................................................. 32
5.2 Test Functions ......................................................................................... 34
6.0 REPAIR PROCEDURES.........................................................................35
6.1 Replace Valve Pump Assembly ............................................................... 35
6.2 Replace Circuit Boards ............................................................................ 37
6.3 Replace Fuses ......................................................................................... 37
6.4 Replace Colorimeter Components ........................................................... 39
Ta
ble of Contents
CFA-3000 Colorimetric Analyzer
Instruction Manual
3
6.4.1 Replace Interference Filter....................................................................... 39
6.4.2 Replace Flowcell...................................................................................... 39
6.4.3 Replace 5 Vdc Source Lamp ................................................................... 40
7.0 INSTRUMENT DESCRIPTION................................................................ 41
7.1 Analyzer Front View................................................................................. 41
7.1.1 Main Cabinet............................................................................................ 41
7.1.2 Reagent Cabinet ...................................................................................... 41
7.2 Keypad Panel........................................................................................... 43
7.3 Card Cage................................................................................................ 44
7.3.1 Card Cage and Door................................................................................ 44
7.3.2 CPU Board............................................................................................... 44
7.3.3 Analog Board ........................................................................................... 44
7.3.4 Valve Board ............................................................................................. 44
7.3.5 Power Supply Board ................................................................................ 44
7.3.6 Analog Board DIP Switches..................................................................... 46
7.4 Colorimeter Assembly.............................................................................. 47
7.5 Rear of Main Cabinet ............................................................................... 48
7.6 Valve Pump Assembly ............................................................................. 49
7.7 Interface Ports.......................................................................................... 50
7.8 Relay Output Option ................................................................................ 50
7.9 Multi-Output Board................................................................................... 50
7.10 Overflow Sampling Assemblies................................................................ 51
8.0 ADDITIONAL FEATURES.......................................................................53
8.1 Chart Recorders and Data Loggers ......................................................... 53
8.2 Computer Interface .................................................................................. 56
8.2.1 Data Output Format ................................................................................. 56
8.2.2 Data Configuration ................................................................................... 56
8.2.3 RS232-C Function Commands ................................................................ 57
8.2.4 Initiating RS232-C Communications to Computer.................................... 58
8.2.5 Printer Interface ....................................................................................... 60
8.2.6 Set Printer Internal Clock ......................................................................... 60
Appendix A. CHEMISTRIES....................................................................................... 61
Appendix B. SPECIFICATIONS ..................................................................................62
Appendix C. POWER INTERCONNECT DESCRIPTION............................................64
Appendix D. BYPASS FOR CALIBRATION................................................................ 66
Appendix E. KEYPAD HELP SHEETS.......................................................................67
ORDERING INFORMATION......................................................................................... 73
CFA-3000 Colorimetric Analyzer
Instruction Manual
4
Figure 2-1. Opened Assembled Cabinet................................................................ 10
Figure 2-2. Mounting Main and Reagent Cabinets................................................. 12
Figure 2-3. Attaching Grounding Strap .................................................................. 12
Figure 2-4a. Single-stream Overflow Sampling Assembly ....................................... 14
Figure 2-4b. Multi-stream Overflow Sampling Assembly ......................................... 14
Figure 2-5. Input/Output Panel............................................................................... 16
Figure 2-6. Colorimeter Assembly.......................................................................... 16
Figure 2-7. Valve Pump Assembly......................................................................... 18
Figure 2-8. Heating Bath and Pump Connectors ................................................... 18
Figure 2-9. Sample Bottle Placement in Cabinet (for all methods; colors refer to
reagent straw colors) ........................................................................... 19
Figure 3-1. Simple Block Diagram ......................................................................... 20
Figure 3-2. Numbered Positions on a Single-Stream Keypad................................ 23
Figure 3-3. Valve Pump LED Designations............................................................ 25
Figure 3-4. Logic Flow for Single-stream Parameter Definition.............................. 26
Figure 3-5. Logic Flow for Multi-stream Parameter Definition ................................ 27
Figure 4-1. Colorimeter Assembly.......................................................................... 34
Figure 6-1. Remove Valve Pump Assembly........................................................... 36
Figure 6-2. Card Cage ........................................................................................... 38
Figure 6-3. Colorimeter Assembly.......................................................................... 39
Figure 6-4. Replace 5 V dc Source Lamp .............................................................. 40
Figure 7-1. Analyzer Front View............................................................................. 42
Figure 7-2. Card Cage and Boards ........................................................................ 45
Figure 7-3. Analog Board DIP Switches (with #3 set in OFF position for multi-stream
analyzer).............................................................................................. 46
Figure 7-4. Colorimeter Assembly.......................................................................... 47
Figure 7-5. Electrical Wiring and Components in Rear Cabinet............................. 48
Figure 7-6. Valve Pump Assembly......................................................................... 49
Figure 7-7. Interface Ports ..................................................................................... 50
Figure 7-8. "D" Connector for Relay Output ........................................................... 50
Figure 7-9. Single-Stream Overflow Sampling Assembly ...................................... 51
Figure 7-10. Multi-stream Overflow Sampling Assembly (three-stream option) ....... 52
Figure 8-1. Input/Output Panel............................................................................... 53
Figure 8-2. Typical Trace Recording with Calibration of Single Stream System .... 54
Figure 8-3. Typical Stream Trace Recording with Calibration of Multi-stream System
............................................................................................................. 55
Figure 8-4. Numbered Positions on a Single-Stream Keypad............................... 58
Figure C-1. CFA-3000 Colorimetric Analyzer Block Diagram ................................. 64
Figure D-1. Calibration Bypass Wiring.................................................................... 65
Figure E-1. Keypad Panel for Single Stream Option .............................................. 66
Figure E-2. Number Positions on a Single-Stream Keypad................................... 66
Figure E-3. Keypad Panel for Multi-stream Option................................................. 69
Figures
CFA-3000 Colorimetric Analyzer
Instruction Manual
5
Table 5-1. Troubleshooting Chart ......................................................................... 32
Table 5-2. Test Functions ..................................................................................... 34
Table 7-2. Main Cabinet Parts .............................................................................. 41
Table 7-2. Reagent Cabinet Parts ........................................................................ 41
Table 7-3. Buttons/LED Functions ........................................................................ 43
Table 7-4. Card Cage Parts.................................................................................. 44
Table 7-5. CPU Board Components ..................................................................... 44
Table 7-6. Analog Board Components.................................................................. 44
Table 7-7. Valve Board Components.................................................................... 44
Table 7-8. Power Supply Board Components....................................................... 44
Table 7-9. Analog Board DIP Switch Positions ..................................................... 46
Table 7-10. Colorimeter Assembly Parts ................................................................ 47
Table 7-11. Back Cabinet Parts .............................................................................. 48
Table 7-12. Valve Pump Parts................................................................................ 49
Table 7-13. Interface Ports ..................................................................................... 50
Table 7-14. Relay Output Option Pin Assignments................................................ 50
Table 7-15. Multi-output Pin Assignments on D Connector................................. 50
Table 7-16. Overflow Sampling Assembly Parts..................................................... 51
Table 8-1. Stream Identification Blips (SIBs) ........................................................ 53
Table 8-2. Normal Data Output Format ................................................................ 56
Table 8-3. Data Configuration............................................................................... 56
Table 8-4. RS232-C Pin Configuration ................................................................. 57
Table 8-5. RS232-C Commands........................................................................... 59
Table 8-6. Printer Port Pin Configuration .............................................................. 60
Table E-1. Single-stream Keypad Help Sheet....................................................... 68
Table E-2. Multi-stream Keypad Help Sheet ......................................................... 71
Tables
CFA-3000 Colorimetric Analyzer
Instruction Manual
6
The Rosemount Analytical CFA-3000 Series analyzers are designed for minimum
maintenance and maximum reliability. The touch of a button starts the microprocessor-
controlled program, calibrates the analyzer, keeps it calibrated, and reports results directly
- unattended - for months at a time. Maintenance is simplified by a modular design with
quick disconnect fittings, labeled tubing, and color-coded components.
The CFA-3000 Series includes analyzers for many common chemical parameters,
including: silica, phosphate, hexavalent chromium (Cr+6), iron, copper, and ammonia.
Sample concentration is clearly indicated on the digital display panel, and output signals
are provided to drive recorders, alarms, or other external devices such as printers or
computers. The specific chemistry of your individual analyzer is discussed in Appendix A,
Chemistries.
Your purchase from Rosemount Analytical provides you with one of the finest instruments
available for your particular application. These instruments have been designed and
tested to meet many national and international standards. Experience indicates that its
performance is directly related to the quality of the installation and knowledge of the user
in operating and maintaining the instrument.
1.0 Introduction
CFA-3000 Colorimetric Analyzer
Instruction Manual
1.1 Thoroughly Read This Manual
This instruction manual describes installation, operation, and maintenance of the CFA-
3000 Colorimetric Analyzer. To ensure continued operation to the design specifications,
personnel should read this manual thoroughly before proceeding with installation,
operation, or maintenance of the analyzer.
Ensure that you have received the correct model and options from your purchase
order. Verify that this manual covers your model and options. If not, call: 1-800-854-
8257 or 949-757-8500 to request the correct manual.
For any inquiries about your analyzer, contact your Rosemount Analytical
representative.
Failure to follow the documented instructions may cause any one of the following
situations to occur: damage to this instrument and warranty invalidation.
Follow all warnings, cautions, and instructions supplied with the product.
Only qualified personnel should install, operate, update, program, and maintain the
product.
All personnel must learn proper installation, operation, and maintenance of the
product.
Install, operate, and maintain equipment exactly as specified in this manual.
Follow appropriate local and national codes.
Only connect the analyzer to electrical sources specified in this manual.
Use only components supplied by Rosemount Analytical or recommended by them
for replacement. Tampering or unauthorized substitution of parts and procedures
may affect the performance of the analyzer as well as void the warranty.
All protective covers must be in place and cabinet panels must be closed at all
times after installation or maintenance.
1.2 Analyzer Benefits
The CFA-3000 Colorimetric Analyzer offers the following benefits:
Maintenance-free pump
Patented blank baseline for silica
7
CFA-3000 Colorimetric Analyzer
Instruction Manual
8
Installation of cable connections and servicing of this product require access to shock
hazard voltage levels.
Main power and relay contacts wired to separate power sources must be disconnected
before servicing.
Non-metallic cable strain relief does not provide grounding between conduit connections!
Use grounding type bushings and jumper wires.
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-
70) and/or any other applicable national or local codes.
Operate only with:
1. Main console rear cover fastened.
2. Left side power terminal cover closed.
3. Keypad assembly latched closed.
Safety and performance require that this instrument be connected and properly grounded
through a three-wire power source.
Proper configuration and use is the responsibility of the user.
1.3 WARNING --Electrical Shock Hazard
1.4 CFA-3000 Series Monitors Approvals
EN 61010-1
Safety of electrical
equipment
EN 55011 Limits of RF equipment
EN 50082-1:92 Generic immunity from
electrostatic discharge,
RF & Fast Transient/
Burst Immunity
9700516
ETL LISTED CONFORMS
TO UL STD. 3101-1
ETL LISTED CERTIFIED
TOCAN/CSA C22.2
No. 1010.1-92
Mark ID Description
CFA-3000 Colorimetric Analyzer
Instruction Manual
9
2.1 Unpack Parts
When you receive your new CFA-3000 analyzer, carefully unpack the items in the
following list.
Main cabinet
Reagent cabinet
Mounting brackets
Overflow sampler assembly
Flowcell assembly
Colorimeter assembly with interference filter assembly
Waste tube assembly
Accessories pack
Startup reagents
Valve pump assembly (integrated housing includes pump, labeled tubes with
straws)
If any of the items are missing or damaged, contact Rosemount Analytical at:
Toll-free number: (800) 854-8257
Direct number: (949) 757-8500
Fax: (949) 863-9159
Inspect each piece to confirm that there is no damage or loose parts. If anything is missing
or appears broken or damaged, contact Rosemount Analytical at the toll-free phone
number above.
Fill out the warranty card located on the back of this manual and mail it
immediately.
2.0 Installing The Analyzer
CFA-3000 Colorimetric Analyzer
Instruction Manual
10
After analyzer has been unpacked and all parts accounted for, install as described in
following sections. Refer to Figure 2-1 to view a completely installed analyzer.
Figure 2-1. Opened Assembled Cabinet
CFA-3000 Colorimetric Analyzer
Instruction Manual
11
General Recommendations:
This cabinet is not for outdoor installation.
Environmental conditions of enclosed facility:
Temperature:
min. 50°F to max. 113°F
(min 10°C max. 45°C)
±10° within regulated range.
Humidity: non-condensing
Mount cabinet on a wall or other suitable vertical surface, at least 42 (107 cm)
wide and 43-1/2 (111 cm) off the floor.
Mounting surface must be able to support 250 pounds (113.6 kg).
Allow 1-1/2 feet (.5 m) of clearance at left side of analyzer for access to
input/output panel, and allow for swing out for rear access.
Allow 12 (.3 m) at right side for access to overflow tube assembly (if single
stream).
Mount cabinet level, or with a slight backward tilt for proper leak detector
operation.
2.2 Mount Main Cabinet
To mount main cabinet, proceed as follows ( see Figure 2-2):
1. Securely attach supplied "L" brackets to wall using suitable hardware. Brackets
must be anchored properly to support 250 pounds (113.6 kg).
2. Attach main cabinet to "L" brackets using supplied four bolts, split ring washers,
and hex nuts.
2.3 Mount Reagent Cabinet
To mount reagent cabinet, proceed as follows:
1. Loosen the four bolts in bottom of main cabinet.
2. Lift reagent cabinet and attach to main cabinet by placing heads of bolts into
keyhole slots in top of reagent cabinet.
3. Slide reagent cabinet back.
4. Feed grounding strap on bottom of main cabinet through notched hole in left rear
corner and attach it to stud on back of reagent cabinet (see Figure 2-3).
5. Tighten screws to attach reagent cabinet to main cabinet.
6. For multi-stream units, plug valve bank cable into positive fit receptacle at left side
of cabinet (refer to Figure 7-1, Multi-stream Valve Panel Connector).
CFA-3000 Colorimetric Analyzer
Instruction Manual
12
-
2. Mounting Main
and Reagent Cabinets
-
3. Attaching
Grounding Strap
CFA-3000 Colorimetric Analyzer
Instruction Manual
13
The Sample intake tube should extend into
The overflow tube approximately ½ 1
(1.25 2.54 cm).
2.4 Install Waste Tube
To install waste tube, proceed as follows (refer to Figure 2-1):
1. Guide waste tube through main cabinet into reagent cabinet and out through
bottom of reagent cabinet.
2. Secure plastic waste tube in clamp in lower right side of main cabinet.
3. Insert Colorimeter drain tube into waste tube.
2.5 Mount Overflow Sampling Assembly
To mount overflow sampling assembly proceed as follows (for sample configurations, see
Figures 2-4a and 2-4b):
1. Mount overflow sampler assembly on outer, right side of reagent cabinet using the
provided four machine screws. For multi-stream units, mount overflow assembly on
left side of cabinet. (If using IF-100 In-Line Filter, refer to instruction sheet
accompanying filter for complete installation instructions.)
2. Place ends of air vent tubing and overflow waste tubing into waste pipe.
3. Connect waste pipe to drain.
4. Supply sample stream to regulating or intake valve with 1/4? tubing (plastic or
steel) as follows: Insert tubing into tube fitting. Make sure tubing rests firmly on
shoulder of fitting and nut is finger-tight. Mark nut at 6 oclock position. While
holding fitting body steady with a wrench, tighten nut 1-1/4 turns (go to 9 oclock
position).
5. For multi-stream assembly, connect sample intake line to bulkhead fitting.
6. For single-stream assembly, attach sample intake line to valve pump.
7. Make sure that sample liquid overflows into overflow chamber.
CFA-3000 Colorimetric Analyzer
Instruction Manual
14
Figure 2-4a. Single-stream Overflow Sampling Assembly
Figure 2-4b. Multi-stream
Overflow Sampling
Assembly
three-stream option)
CFA-3000 Colorimetric Analyzer
Instruction Manual
15
Use cable glands to hold cables securely in holes above input/output panel.
2.6 Prepare Electrical Connections
Electrical connections are made to and from the cabinet at the input/output panel, located on
the left side of the analyzer. See Figure 2-5.
Terminals are provided for ground, power supply, high/low alarms, output signals,
calibration and status relays
.Output signals - a voltage output of 0-5 Vdc and a floating, ungrounded current
output of 4-20 mA are available for remote readout or control of a process.
Also available are contact closures for the following:
High alarm Leak Alarm
Low Alarm Lamp Alarm
Calibration
To make electrical connections, proceed as follows:
1. Remove solid caps from the holes above the input/output panel by puncturing with an
awl and prying them out. (Note: remove caps only from those holes that will be used.)
2. Replace with permanent connectors.
3. Connect alarms and external computer interfaces (if any).
4. Feed 3-wire power cable into analyzer through hole above the input/output panel.
5. Connect wires to appropriate contacts on Terminal Strip TB3.(See next page)
WARNING:
Connect output signals, if any, before connecting the power supply.
CAUTION:
For proper operation and for safety, electrical connections to this equipment
must be made in accordance with local or national electrical code as applicable.
This equipment must be grounded. A qualified electrician should wire this
equipment to an electrical circuit.
CFA-3000 Colorimetric Analyzer
Instruction Manual
2.7 Colorimeter Assembly Installation
(assembly comes factory-installed)
To install colorimeter assembly, proceed as follows:
1. Insert flowcell assembly into slot on top of colorimeter (interference filter and lamp
pre-installed as part of the colorimeter assembly (see Figure 2-6).
2. Hand-tighten thumbscrews on top of flowcell assembly. Then, tighten screws one-
half turn with screwdriver.
3. Slide colorimeter into connector in inside back of cabinet (refer to Figure 2-1).
Housing will latch in place, automatically.
16
Figure 2-6. Colorimeter Assembly
Pls Note: lamp may be halogen or LED; filter or aperture may be used
Figure 2-5. Input/Output Panel
CFA-3000 Colorimetric Analyzer
Instruction Manual
17
2.8 Install Valve Pump Assembly
The pump assembly has a vacuum pressure system that:
o Eliminates internal pump tubing
o Uses significantly less reagent
o Accurately moves fluids through the analyzer.
Perform the following procedure to install valve pump assembly:
1. Tilt pump assembly towards you.
2. .Align the hinge pins and sockets, and slide assembly tray as far as possible to
right (see Figure 2-7). Carefully lay tray down into cabinet
3. Plug in the two connectors to the back of the cabinet (6-position connector for
heating bath; 9-position connector for pump) (see Figure 2-8).
4. Connect 16-pin connector from pump assembly to small connector on CPU board
(yellow ejector) above connector for keypad panel.
5. Attach colorimeter inlet tubing on left side of pump to flowcell inlet (see Figure 2-
1).
6. Continue with procedure depending on whether system is single-stream or multi-
stream
For single-stream, continue:
6. Thread tube labeled Sample" through hole 0n bottom of shelf to the right of pump
assembly, and out through hole on left side of cabinet.
7. Connect Sample tube to top of overflow sample on left side of cabinet.
8. Guide all other tubes that are attached to valve pump assembly through reagent
cabinet access hole.
For multi-stream, continue:
6. Thread Sample tube from pump assembly through reagent access hole on the
right side of the valve pump assembly, and connect to Common (C) on #1 multi-
stream solenoid valve on left side of reagent cabinet.
Guide all other tubes that are attached to valve pump assembly through reagent cabinet
access hole.
CFA-3000 Colorimetric Analyzer
Instruction Manual
18
Figure 2-7. Valve Pump Assembly (comes factory-installed)
Figure 2-8. Pump Connector
(Pls Note: Heating Bath is not usually used on colorimetric Valve-Pump systems,
but may be required for ammonia systems.
CFA-3000 Colorimetric Analyzer
Instruction Manual
2.9 Prepare Reagent Containers
Prepare reagent containers as follows:
1. Remove screw caps from reagent bottles.
2. Using an awl, make one hole in foil for the reagent straw. DO NOT let awl touch
reagent.
3. Place Standard and Reagent bottles into reagent cabinet, as shown in Figure 2-9.
4. Insert color-coded reagent straws and labeled Standard straw leading from valve
pump assembly into holes punched in corresponding bottles.
19
Figure 2-9. Five Liter Sample Bottle Placement in Cabinet (for all methods; colors
refer to reagent straw colors)
  • Page 1 1
  • Page 2 2
  • Page 3 3
  • Page 4 4
  • Page 5 5
  • Page 6 6
  • Page 7 7
  • Page 8 8
  • Page 9 9
  • Page 10 10
  • Page 11 11
  • Page 12 12
  • Page 13 13
  • Page 14 14
  • Page 15 15
  • Page 16 16
  • Page 17 17
  • Page 18 18
  • Page 19 19
  • Page 20 20
  • Page 21 21
  • Page 22 22
  • Page 23 23
  • Page 24 24
  • Page 25 25
  • Page 26 26
  • Page 27 27
  • Page 28 28
  • Page 29 29
  • Page 30 30
  • Page 31 31
  • Page 32 32
  • Page 33 33
  • Page 34 34
  • Page 35 35
  • Page 36 36
  • Page 37 37
  • Page 38 38
  • Page 39 39
  • Page 40 40
  • Page 41 41
  • Page 42 42
  • Page 43 43
  • Page 44 44
  • Page 45 45
  • Page 46 46
  • Page 47 47
  • Page 48 48
  • Page 49 49
  • Page 50 50
  • Page 51 51
  • Page 52 52
  • Page 53 53
  • Page 54 54
  • Page 55 55
  • Page 56 56
  • Page 57 57
  • Page 58 58
  • Page 59 59
  • Page 60 60
  • Page 61 61
  • Page 62 62
  • Page 63 63
  • Page 64 64
  • Page 65 65
  • Page 66 66
  • Page 67 67
  • Page 68 68
  • Page 69 69
  • Page 70 70
  • Page 71 71
  • Page 72 72
  • Page 73 73
  • Page 74 74
  • Page 75 75
  • Page 76 76
  • Page 77 77
  • Page 78 78

Rosemount 185 Colormetric Valve-Pump based Monitor Owner's manual

Type
Owner's manual

Ask a question and I''ll find the answer in the document

Finding information in a document is now easier with AI